Antenna coupling explains unintended thermal injury caused by common operating room monitoring devices.

Nicole T Townsend1, Edward L Jones, Alessandro Paniccia, Joel Vandervelde, Jennifer R McHenry, Thomas N Robinson.   

Abstract

INTRODUCTION: Unintended thermal injury from patient monitoring devices (eg, electrocardiogram pads, neuromonitoring leads) results in third-degree burns. A mechanism for these injuries is not clear. The monopolar "bovie" emits radiofrequency energy that transfers to nearby, nonelectrically active cables or wires without direct contact by capacitive and antenna coupling. The purpose of this study was to determine if, and to what extent, radiofrequency energy couples to common patient monitoring devices.
MATERIALS AND METHODS: In an ex vivo porcine model, monopolar radiofrequency energy was delivered to a handheld "bovie" pencil. Nonelectrically active neuromonitoring and cardiac-monitoring leads were placed in proximity to the monopolar pencil and its cord. Temperature changes of tissue touched by the monitoring lead were measured using a thermal camera immediately after a 5-second activation. The energy-device cords were then separated by 15 cm, the power was reduced from 30 W coag to 15 W coag and different cord angulation was tested. An advanced bipolar device, a plasma-based device, and an ultrasonic device were also tested at standard settings.
RESULTS: The neuromonitoring lead increased tissue temperature at the insertion site by 39 ± 13°C (P<0.001) creating visible char at the skin. The electrocardiogram lead raised tissue temperature by 1.3 ± 0.5°C (P<0.001). Decreasing generator power from 30 W to 15 W and separating the bovie cord from the neuromonitoring cord by 15 cm significantly reduced the temperature change (39 ± 13°C vs. 26±5°C; P<0.001 and 39 ± 13°C vs. 10 ± 5°C; P<0.001, respectively). Lastly, monopolar energy increased tissue temperatures significantly more than argon beam energy (34 ± 15°C), advanced bipolar energy (0.2 ± 0.4°C), and ultrasonic energy (0 ± 0.3°C) (all P<0.001).
CONCLUSIONS: Stray energy couples to commonly used patient monitoring devices resulting in potentially significant thermal injury. The handheld bovie cord transfers energy via antenna coupling to neuromonitoring leads that can raise tissue temperatures over 100°F (39°C) using standard settings. The most effective ways to decrease this energy coupling is to reduce generator power, increase the separation between wires, or utilize lower voltage energy devices such as ultrasonic or bipolar energy.

Entities:  

Mesh:

Year:  2015        PMID: 25793350     DOI: 10.1097/SLE.0000000000000137

Source DB:  PubMed          Journal:  Surg Laparosc Endosc Percutan Tech        ISSN: 1530-4515            Impact factor:   1.719


  6 in total

1.  Stray energy transfer during endoscopy.

Authors:  Edward L Jones; Amin Madani; Douglas M Overbey; Asimina Kiourti; Satheesh Bojja-Venkatakrishnan; Dean J Mikami; Jeffrey W Hazey; Todd R Arcomano; Thomas N Robinson
Journal:  Surg Endosc       Date:  2017-02-15       Impact factor: 4.584

2.  The SAGES Fundamental Use of Surgical Energy program (FUSE): history, development, and purpose.

Authors:  P Fuchshuber; S Schwaitzberg; D Jones; S B Jones; L Feldman; M Munro; T Robinson; G Purcell-Jackson; D Mikami; A Madani; M Brunt; B Dunkin; C Gugliemi; L Groah; R Lim; J Mischna; C R Voyles
Journal:  Surg Endosc       Date:  2017-12-07       Impact factor: 4.584

3.  Monopolar stray energy in robotic surgery.

Authors:  Douglas M Overbey; Heather Carmichael; Krzysztof J Wikiel; Douglas A Hirth; Brandon C Chapman; John T Moore; Carlton C Barnett; Teresa S Jones; Thomas N Robinson; Edward L Jones
Journal:  Surg Endosc       Date:  2020-05-08       Impact factor: 4.584

Review 4.  Electrosurgical unit: Iatrogenic injuries and medico-legal aspect. Italian legal rules, experience and article review.

Authors:  Patrizia Gualniera; Serena Scurria; Daniela Sapienza; Alessio Asmundo
Journal:  Ann Med Surg (Lond)       Date:  2021-01-04

5.  Comparing the safety and effectiveness of dedicated radiofrequency transseptal wires to electrified metal guidewires.

Authors:  Jeremiah Wasserlauf; Bradley P Knight
Journal:  J Cardiovasc Electrophysiol       Date:  2022-01-11       Impact factor: 2.942

6.  Bigger Is Not Always Better: Effects of Electrocautery Setting on Tissue Injury in a Porcine Model.

Authors:  Austin L Shiver; Colton Webber; Taylor Sliker; Patrick Rushford; Aaron Shaw
Journal:  Cureus       Date:  2022-07-14
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.