| Literature DB >> 35907091 |
R Lopes1,2, T Noailles3, G Padiolleau2, N Bouguennec4, T D Vieira5.
Abstract
PURPOSE: This study evaluates the use of the needle arthroscopy in anatomical reconstruction of the lateral ankle. We hypothesized that the needle arthroscopy would allow anatomical reconstruction to be performed under arthroscopy.Entities:
Keywords: Ankle arthroscopy; Ankle ligaments; Ankle reconstruction; Arthroscopy; Nanoscope; Needle arthroscopy
Year: 2022 PMID: 35907091 PMCID: PMC9339066 DOI: 10.1186/s40634-022-00510-x
Source DB: PubMed Journal: J Exp Orthop ISSN: 2197-1153
Demographic data
| Patient | Sex | Male | Male | Female |
| Age | 22 | 36 | 24 | |
| Side | Right | Right | Left | |
| BMI (kg/m2) | 24 | 21 | 20 | |
| Symptoms | Pure instability | Pure instability | Painful instability | |
| Sports level | Competition | Recreational | None | |
| Intervention | Associated lesions | None | Synovitis Ossification under lateral malleolar | Synovitis |
| Tourniquet time (minutes) | 47 | 50 | 49 |
Fig. 1Standard arthroscope being replaced by the NanoScopeTM. A Standard arthroscope in the anterolateral portal. B NanoScopeTM in the anterolateral portal
Fig. 2NanoScopeTM video system: handpiece provided in a single-use kit
Fig. 3NanoScopeTM 13-inch handheld console
Comparative technical characteristics of the 2 scopes used
| Arthroscope | NanoScope™ | |
|---|---|---|
| Handpiece width | 4.5 cm | 4.1 cm |
| Handpiece length | 10 cm | 25.4 cm |
| Handpiece height | 4.5 cm | 6.4 cm |
| Handpiece weight | 570 g | 150 g |
| Diameter without a cannula | 4 mm | 1.9 mm |
| Diameter with a cannula | 4.6 mm | 2.2 mm |
| Image resolution | 1920 × 1080 pixels | 400 × 400 pixels |
| Frame rates | NA | 30 ips |
| Fields of vision | 100 degrees | 120 degrees |
| Viewing Angle | 30 degrees | 0 degrees |
| Enlightenment | 1.800 lm | ≥ 2 lumens |
| Screen size | 32 inches | 13 inches |
Interest in using the NanoScopeTM at different stages
| Complicated | Complicated | Complicated | ||
| Similar | Similar | Similar | ||
| Impossible | Impossible | Complicated | ||
| Facilitated | Facilitated | Facilitated | ||
| Complicated | Impossible | Impossible | ||
| Similar | Similar | Similar | ||
| Complicated | Complicated | Complicated | ||
Fig. 4Halstead forceps in the lateral talar gutter. A NanoScopeTM view. B Standard arthroscope view
Fig. 5Placement of the cannula in the lateral gutter. A NanoScopeTM view. B Standard arthroscope view
Fig. 6Talar tunnel drilling. A NanoScopeTM view. B Standard arthroscope view
Fig. 7Needle arthroscope ergonomics