Yusuf Ahmed1, Austin Pereira2, Amrit S Rai2, Victoria C Leung2, Aadam Ahmed3, Amandeep S Rai4,5,6,7.
Abstract
PURPOSE: To evaluate fundus examination accuracy of medical students when using an unmodified iPhone X or a direct ophthalmoscope in comparison to a staff ophthalmologist's retinal examination.Entities:
Keywords: Direct ophthalmoscope; Direct ophthalmoscopy; Medical education; Medical trainees; Telemedicine; iPhone X
Year: 2022 PMID: 35871237 PMCID: PMC9308475 DOI: 10.1007/s10792-022-02377-4
Source DB: PubMed Journal: Int Ophthalmol ISSN: 0165-5701 Impact factor: 2.029
Demographics of patients included in study analysis (n = 24)
| Characteristics | |
|---|---|
| Age, | 72.1 (52 to 87) |
| Number of unique eyes, | 38 |
| Male Sex, | 16 (57.9%) |
| Mean IOP, | 13.0 (3.67) |
| BCVA, | 0.11 (0.15) |
| AMD, | 1 |
| Pseudophakic, | 25 |
| Healthy, | 12 |
Data are listed as mean (range), mean (SD) or number (%). IOP intraocular pressure, BCVA best-corrected visual acuity, AMD Age-related macular degeneration
Patient optic nerve head findings as determined by staff ophthalmologist
| Optic Disc parameter | Staff ophthalmologist finding | |
|---|---|---|
| CDR, | 0.33 (0.18) | |
| Colour, | Pink disc | Pale disc |
| 38 | 0 | |
| Contour, | Crisp margins | Blurred margins |
| 38 | 0 | |
| SVPs, | Yes | No |
| 11 | 27 | |
Data are listed as mean (SD) or number (n). CDR cup-to-disc ratio, SVPs spontaneous venous pulsations
Student CDR observation concordance with staff ophthalmologist findings using the direct ophthalmoscope and iPhone X
| Modality for direct ophthalmoscopy | |||
|---|---|---|---|
| Direct ophthalmoscope (n = 113) | iPhone X (n = 117) | ||
| N (%) | N (%) | ||
| 79 (69.9) | 22 (18.8) | < 0.001 | |
| Exact match | 10 (12.7) | 5 (22.7) | 0.308 |
| ≤ 0.1 Disc Diameter | 30 (38.0) | 9 (40.9) | 0.809 |
| ≤ 0.2 Disc Diameter | 49 (62.0) | 16 (72.3) | 0.454 |
| 34 (30.1) | 95 (81.2) | < 0.001 | |
Data are listed as number and percentage of student responses
“ ≤ 0.1 and 0.2 Disc Diameter” is inclusive
Fig. 1Bland–Altman plot for mean difference of cup-to-disc measurements between 78D slit-lamp biomicroscopy staff ophthalmologist findings and a students observations with either iPhone X or direct ophthalmoscope, b student observations with just the iPhone X, and c student observations with just the direct ophthalmoscope
Student optic nerve head exam findings using the direct ophthalmoscope and iPhone X
| Modality for direct ophthalmoscopy | |||
|---|---|---|---|
| Direct ophthalmoscope (n = 113) | iPhone X (n = 117) | ||
| N (%) | N (%) | ||
| Correct | 89 (82.4) | 63 (88.7) | |
| Incorrect | 19 (17.6) | 8 (11.3) | |
| No response | 5 (4.4) | 46 (39.3) | < 0.001 |
| Correct | 72 (74.2) | 28 (63.8) | |
| Incorrect | 25 (25.8) | 13 (36.2) | |
| No response | 16 (14.2) | 76 (65.0) | < 0.001 |
| Correct | 45 (65.2) | 16 (55.2) | |
| Incorrect | 24 (34.8) | 13 (44.8) | |
| No response | 44 (38.9) | 88 (75.2) | < 0.001 |
Data are listed as number and percentage of student responses. CDR cup-to-disc ratio, SVPs spontaneous venous pulsations