Kirsten R Nickles1, Alejandro E Relling1, Alvaro Garcia-Guerra2, Francis L Fluharty3, Anthony J Parker1. 1. Department of Animal Sciences, The Ohio State University, Wooster, Ohio, United States of America. 2. Department of Animal Sciences, The Ohio State University, Columbus, Ohio, United States of America. 3. Department of Animal and Dairy Science, University of Georgia, Athens, Georgia, United States of America.
Abstract
Glucose tolerance tests (GTT) are commonly performed in beef cattle to evaluate the glucose-insulin signaling pathway. Blood samples are obtained via a catheter and then transferred back to the laboratory for further analysis. A hand-held glucometer used chute-side can make performing GTT's and quantifying blood glucose concentration much easier and faster for research purposes. The purpose of this study was to evaluate the agreement between a hand-held electronic glucometer (Precision Xtra; Abbott Diabetes Care Inc., Mississauga, ON, Canada) for chute-side use in beef cattle compared with a colorimetric assay in the laboratory (Stanbio Glucose LiquiColor; Stanbio Laboratory, Boerne, TX, USA). A GTT was performed on 13 Simmental × Angus steers during the growing phase. Blood samples were obtained via a jugular catheter. Glucometer readings were taken immediately after blood was sampled from the jugular with no preservative, and laboratory measurements were conducted on plasma preserved with sodium fluoride. A paired t-test (P = 0.40), Pearson's correlation (P < 0.001; r = 0.95), Bland-Altman plot, and Lin's concordance correlation coefficient (LCCC = 0.90) were completed to evaluate the performance of the glucometer relative to the results from the laboratory assay. Based on the results, we conclude that the glucometer is an acceptable method for measuring blood glucose concentration in beef cattle under field conditions.
Glucose tolerance tests (GTT) are commonly performed in beef cattle to evaluate the glucose-insulin signaling pathway. Blood samples are obtained via a catheter and then transferred back to the laboratory for further analysis. A hand-held glucometer used chute-side can make performing GTT's and quantifying blood glucose concentration much easier and faster for research purposes. The purpose of this study was to evaluate the agreement between a hand-held electronic glucometer (Precision Xtra; Abbott Diabetes Care Inc., Mississauga, ON, Canada) for chute-side use in beef cattle compared with a colorimetric assay in the laboratory (Stanbio Glucose LiquiColor; Stanbio Laboratory, Boerne, TX, USA). A GTT was performed on 13 Simmental × Angus steers during the growing phase. Blood samples were obtained via a jugular catheter. Glucometer readings were taken immediately after blood was sampled from the jugular with no preservative, and laboratory measurements were conducted on plasma preserved with sodium fluoride. A paired t-test (P = 0.40), Pearson's correlation (P < 0.001; r = 0.95), Bland-Altman plot, and Lin's concordance correlation coefficient (LCCC = 0.90) were completed to evaluate the performance of the glucometer relative to the results from the laboratory assay. Based on the results, we conclude that the glucometer is an acceptable method for measuring blood glucose concentration in beef cattle under field conditions.
Glucose tolerance tests (GTT) are commonly used in beef cattle research as an
indicator of insulin resistance [1]. This procedure has been modified by several research groups, but
generally consists of an intravenous bolus infusion of 50% glucose followed by
continuous blood sampling to monitor glucose and insulin concentration. The standard
method of measuring glucose in plasma is with a colorimetric assay to quantify
glucose concentration. The Stanbio assay used in our laboratory is based on the
glucose oxidase methodology adapted by Trinder et al. [2]. In this method, glucose is oxidized in the
presence of glucose oxidase. After hydrogen peroxide is formed and reacts with
phenol and 4-aminoantipyrine, a red-violet quinone complex forms. The intensity of
the red-violet color is directly proportional to the glucose concentration (Stanbio
Laboratory, Boerne, TX, USA). Though the laboratory method is the standard, it is
more expensive and time consuming compared with the hand-held glucometer.An alternative method is a hand-held electronic glucose measuring system designed for
use in humans. This system uses electrochemical test strips in which blood is
applied after the test strip is inserted into the glucometer. The blood is then
drawn up the test strip via capillary action. Once it is in the glucometer, it
reacts with glucose oxidase and forms gluconic acid. The gluconic acid then reacts
with the test strip electrodes and creates an electrical current that is
proportional to the concentration of glucose in the blood. This hand-held system has
been validated for use in measuring glucose and β-hydroxybutyrate in dairy cows
[3-6], however, this has not been validated for
measuring glucose chute-side when performing a GTT in beef cattle. It is possible
that beef cattle may have different response curves to a GTT (i.e. peak plasma
glucose concentration, baseline plasma glucose concentration) and thus plasma
glucose concentrations compared with dairy cattle, as it has been previously
reported that plasma glucose and insulin concentrations were different between beef
and dairy cows at the same stage of lactation [7].The objective of the present study was to compare the glucose concentration using
whole blood and the hand-held glucometer compared with the standard laboratory assay
using plasma from samples obtained while completing a GTT on 13 Simmental × Angus
steers during the growing phase. We hypothesized that the two different methods
would show acceptable agreeance and that the hand-held meter would be a suitable
method for quantifying glucose concentration chute-side.
Materials and methods
All procedures were approved by The Ohio State University Institutional Animal Care
and Use Committee (Animal Use Protocol # 2019A00000142).
Animals and procedures
Thirteen Simmental × Angus steer calves were used for this glucose quantification
comparison. Steers were trained in the chute 5 days/week for two weeks to allow
for the steers to become acclimated to standing in the chute and being touched
during the GTT.Steers were fasted for 24 hours before the GTT. The morning of the GTT, steers
were weighed to determine bolus size (0.25 g of glucose/kg BW delivered in a 50%
weight/volume dextrose solution. After being weighed, jugular catheters were
placed in the steers, and then steers were returned to their pen and allowed a
one hour rest period before the GTT began. Blood samples were collected at 5 and
2 minutes before administration of the glucose bolus to determine fasted plasma
glucose concentration. Subsequent blood samples were collected immediately after
glucose bolus infusion (0 minutes), 5, 10, 15, 20, 30, 60, and 120 minutes after
glucose bolus infusion. Before and after each 10 mL blood sample was collected,
the catheter line was flushed with 4–5 mL of sterile heparinized saline (9 g/L
of NaCl). All blood samples were transferred to a tube containing sodium
fluoride and then immediately placed on ice. The sodium fluoride tubes were
transferred back to the laboratory and centrifuged for 25 min at 2500 x g and
4°C. The plasma was then further aliquoted into individual microcentrifuge tubes
to determine plasma glucose concentrations at a later date.
Glucometer method
As blood was being sampled via the jugular catheter for the GTT, blood was
immediately placed on the glucometer test strip to determine whole blood glucose
concentration. The glucometer was used according to the label descriptions and
directions of the manufacturer and is reported to measure blood glucose
concentrations from 20–500 mg/dL (Precision Xtra; Abbott Diabetes Care Inc.,
Mississauga, ON, Canada). According to the user’s manual, if a sample reads
“LO”, the meter has determined that the blood glucose concentration is below 20
mg/dL, and a sample reads “HI”, the meter has determined that the blood glucose
concentration is above 500 mg/dL. One sample that was measured with the
glucometer returned a “HI” result, and the glucose concentration was recorded as
500 mg/dL as recommended by the glucometer’s user manual. All samples measured
with the glucometer were only measured once.
Laboratory trinder method
The laboratory analysis was completed using a colorimetric assay (Stanbio Glucose
LiquiColor (Oxidase) Procedure, Stanbio Laboratory, Boerne, TX, USA). Any sample
that was outside of the linear portion of the standard curve was diluted with a
1:2 dilution. All samples were run in duplicates and the intra- and inter-assay
coefficient of variations were 2.93% and 3.00%, respectively.
Statistical analysis
First, using the raw measurements from the glucometer and the average of the
duplicate laboratory samples, a paired t-test was completed using the TTEST
procedure of SAS (SAS 9.4). Additionally, a Pearson’s Correlation was completed
for the two methods using the CORR procedure of SAS. For both of these models,
data was assessed for normality using the residuals panel in SAS. When plotted,
the differences between the pairs of observations were approximately normally
distributed. Next, the Bland-Altman [8] approach was used to plot the difference
between the two measurements against their mean to determine the two approaches’
agreement. Bland and Altman [8] and Petrie and Watson [9] recommend first performing a paired
t-test to test the null hypothesis that the mean of the differences between the
two methods is zero, and that the differences are evenly scattered above and
below zero. The paired t-test determined if there was evidence of a systematic
difference between the hand-held glucometer and the laboratory assay. The next
step is to perform a Pearson correlation, however, Bland and Altman [8] caution that a Pearson
correlation coefficient only gives indication of how close the observations in
the scatter diagram are to a straight line and do not assess agreement. To
assess agreement, one needs to know how close the points are to the line of
perfect agreement (the 45° line through the origin). The Bland-Altman plot
determines the limits (± 1.96 standard deviations) within which 95% of the
differences are expected to lie when the difference between the two measurements
in a pair are plotted against their mean. If there is no evidence of a
systematic effect, the points should be scattered evenly above and below the
line corresponding to a zero difference. If the variability of the differences
is not constant (i.e. funnel shape), Petrie and Watson [9] recommend transforming the data and
repeating the process. If there is no evidence of a systematic effect in the
Bland-Altman plot of either the raw or transformed data, the next step is to
complete an index of agreement which can either be the intraclass correlation
coefficient (ICC) or the Lin’s concordance correlation coefficient (LCCC). These
two indexes are similar and describe the closeness of the points to the line of
perfect agreement and can be used to assess agreement because both accuracy and
precision are incorporated. A Lin’s concordance correlation coefficient was
calculated to assess agreement between the two methods.
Results
The paired t-test indicated no evidence of a systematic difference between the two
methods of measurement with a test statistic of -0.86, (P = 0.40).
A Pearson correlation coefficient for the two methods of measurement was then
completed on the raw data (Fig
1; P < 0.001; r = 0.95), and indicated precision (i.e.
the random variation describing the tightness of the points about the best-fitting
straight line) between the two methods.
Fig 1
Pearson correlation coefficient (P < 0.001, r = 0.95)
of the raw data (N = 130 samples from 13 steers), with the glucose
concentration as measured by the glucometer on the y-axis (Precision Xtra;
Abbott Diabetes Care, Inc., Mississauga, ON, Canada) plotted against the
plasma preserved in sodium fluoride tubes and analyzed in a laboratory with
a colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure,
Stanbio Laboratory, Boerne, TX, USA) on the x-axis.
A Bland-Altman plot was completed for the raw data between the two measurements
(Fig 2), with the mean
concentration measured by the glucometer plotted against the difference in mean
concentration between the two methods. The raw data demonstrated a funnel shape,
indicating that the variability of the differences was not constant as the mean of
the two measurements increased. Therefore, the raw data was log transformed as
recommended by Petrie and Watson [9]. The transformed Bland-Altman plot is shown in Fig 3. Transforming the data eliminated the
funnel effect, and the points are evenly scattered above and below the line
representing the mean, corresponding to no systematic difference between the two
methods. Additionally, since the scatter of the points is random with no funnel
effect, we can conclude that the size of the discrepancy between the two methods is
not related to the magnitude of the count. In Figs 2 and 3, the dashed lines represent the upper and lower
limits of agreement. We expect 95% of the absolute differences to be less than the
upper and lower limits of agreement, which is the case for the transformed data.
Fig 2
Bland-Altman plot of the raw data (N = 130 samples from 13 steers), with
the mean glucose concentration as measured by the glucometer (Precision
Xtra; Abbott Diabetes Care, Inc., Mississauga, ON, Canada) and the plasma
preserved in sodium fluoride tubes and analyzed in a laboratory with a
colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure, Stanbio
Laboratory, Boerne, TX, USA) on the x-axis, plotted against the difference
in glucose concentrations determined by the glucometer and the laboratory on
the y-axis.
The mean difference is represented by the solid line (mean = 4.6) and the 95%
confidence limits are represented by the dashed lines.
Fig 3
Bland-Altman plot of the log transformed data (N = 130 samples from 13
steers), with the mean glucose concentration as measured by the glucometer
(Precision Xtra; Abbott Diabetes Care, Inc., Mississauga, ON, Canada) and
the plasma preserved in sodium fluoride tubes and analyzed in a laboratory
with a colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure,
Stanbio Laboratory, Boerne, TX, USA) on the x-axis, plotted against the
difference in glucose concentrations determined by the glucometer and the
laboratory on the y-axis. The mean difference is represented by the solid
line (mean = 0.03) and the 95% confidence limits are represented by the
dashed lines.
Bland-Altman plot of the raw data (N = 130 samples from 13 steers), with
the mean glucose concentration as measured by the glucometer (Precision
Xtra; Abbott Diabetes Care, Inc., Mississauga, ON, Canada) and the plasma
preserved in sodium fluoride tubes and analyzed in a laboratory with a
colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure, Stanbio
Laboratory, Boerne, TX, USA) on the x-axis, plotted against the difference
in glucose concentrations determined by the glucometer and the laboratory on
the y-axis.
The mean difference is represented by the solid line (mean = 4.6) and the 95%
confidence limits are represented by the dashed lines.Bland-Altman plot of the log transformed data (N = 130 samples from 13
steers), with the mean glucose concentration as measured by the glucometer
(Precision Xtra; Abbott Diabetes Care, Inc., Mississauga, ON, Canada) and
the plasma preserved in sodium fluoride tubes and analyzed in a laboratory
with a colorimetric assay (Stanbio Glucose LiquiColor (Oxidase) Procedure,
Stanbio Laboratory, Boerne, TX, USA) on the x-axis, plotted against the
difference in glucose concentrations determined by the glucometer and the
laboratory on the y-axis. The mean difference is represented by the solid
line (mean = 0.03) and the 95% confidence limits are represented by the
dashed lines.In addition, since there is no evidence of a systematic effect, we can estimate the
Lin’s concordance correlation coefficient. For this correlation coefficient, a
perfect agreement is achieved when the value is equal to 1, and there is no
agreement when it is equal to 0. For the transformed data, the Lin’s concordance
correlation coefficient is 0.90 (95% confidence interval = 0.87–092). Petrie and
Watson (2013) note that based on a previous article published by McBride et al.
[10], a Lin’s concordance
correlation coefficient of 0.90 ≤ rc ≤ 0.95 is considered moderate
agreement.
Discussion
Taking into consideration the paired t-test, Pearson correlation coefficient,
Bland-Altman plot, and Lin’s concordance correlation coefficient, we accept our
hypothesis that the hand-held Precision Xtra glucometer moderately agrees with the
laboratory method and is acceptable to use for rapid, chute-side measurement of
glucose in beef cattle. Several studies have previously evaluated the agreement of
the Precision Xtra glucometer with laboratory analysis of glucose in dairy cattle,
however, this is the first study to our knowledge that has compared the two methods
in beef cattle during a GTT [3-6].Our results agree with those of previous reports that indicate the Precision Xtra
glucometer has acceptable agreement with laboratory measurement in dairy cattle
[3,5]. Both of these papers demonstrated
Bland-Altman plots and reported that at least 95% of the observations fell within
the 95% confidence intervals, indicating good agreement between the two methods.
Neither paper, however, reported a Lin’s concordance correlation coefficient or an
intraclass correlation coefficient. Therefore, while there is agreement between the
two methods based on the Bland-Altman plots, the assessment of agreement cannot be
definitive without one of the indexes being calculated.Additionally, of the papers that have previously reported unacceptable agreement
between the Precision Xtra and laboratory analysis of glucose concentration only
Lopes et al. [6] reported an
LCCC or ICC while Megahed et al. [4] did not. It is interesting that Lopes et al. [6] showed moderate association between the two
methods with a Pearson correlation coefficient (r = 0.71), indicated that their
reported LCCC of 0.74 demonstrated strong agreement with the reference method, and
reported that the difference between the methods fell within acceptable limits of
agreement (± 1.96 standard deviations) at least 95% of the time, however, continued
to state that according to the American Society for Veterinary Clinical Pathology
guidelines, only 54.6% of the Precision Xtra glucose readings had a total observed
error of ≤ 20% and declared the meter inadequate to measure glucose in dairy cows.
In the present study, we did not measure total observed error, as Bland and Altman
[8] and Petrie and Watson
[9] do not discuss
calculating this total observed error. Additionally, Petrie and Watson [9] follow the guidelines
proposed by McBride [10] that
state that a LCCC less than 0.90 is indicative of poor agreement between two
methods. Following these guidelines, the LCCC of 0.74 reported to be in support of
strong agreement should be observed with caution.Similar to Zakian et al. [5],
we used the glucose oxidase method for the laboratory glucose measurement. Using
this method as the gold standard, we obtained similar results to Zakian et al.
[5] and found the
glucometer to be in moderate agreement to the laboratory measurement. As mentioned
by Zakian et al. [5], one
possibility for this improvement in agreement between the two methods may be the
reference method used, as the other papers that evaluated the Precision Xtra
glucometer compared its measurements to a hexokinase reference method [4,6]. Wittrock et al. [3] reported acceptable agreement between the
two methods, however, did not mention which reference method was used and only
stated that glucose concentrations were determined using a commercial reagent
kit.Of the previous studies that evaluated the agreement of the Precision Xtra glucometer
with the laboratory measurements, only Wittrock et al. [3] similarly performed a GTT. Performing this
method to assess agreement during a GTT is of great interest, as there are many
samples that must be taken during this procedure and the hand-held glucometer could
make performing the procedure quicker and cheaper if glucose concentration could be
analyzed chute side. The present dextrose infusion that was provided to the beef
steers (0.25 g of glucose/kg BW delivered in a 50% weight/volume dextrose solution)
was the same as that used by Wittrock et al. [3] in dairy cows, however, the sampling
timeline was different. Steers were sampled for blood in the current study at 5 and
2 minutes before glucose bolus infusion, and then subsequent samples were taken
immediately after glucose bolus infusion (0 minutes) and then 5, 10, 15, 20, 30, 60,
and 120 minutes after infusion. Wittrock et al. [3] sampled for blood immediately before
dextrose infusion and then at 10 and 80 minutes after infusion. Wittrock et al.
[3] reported that
measurements that were within the physiological range of 2.3 to 5.2 mmol/L were
slightly lower with the glucometer compared with the laboratory value.
Alternatively, the authors found that the high glucose concentrations were generally
overestimated by the glucometer, though the authors were not aware of a
methodological reason for this difference. In the present study, we found that the
glucometer and laboratory measurements were very precise and accurate up to
concentrations ≤ 300 mg/dL. At glucose concentrations greater than 300 mg/dL, we
observed our greatest differences between the glucometer and the laboratory
measurement. Except for one blood sample that was obtained 5 minutes after glucose
infusion, these high concentrations of glucose (> 300 mg/dL) all coincided with
our time 0 blood sample which was sampled immediately after glucose infusion.
Generally, the glucometer overestimated glucose concentration when compared with the
laboratory measurement at these high concentrations. This result is similar to that
reported by Wittrock et al. [3]. However, when the concentrations greater than 300 mg/dL were removed
from the analysis, the LCCC was only improved to 0.92 which still indicates moderate
method agreement according to Petrie and Watson [9]. Therefore, we have only presented the
statistical analyses including all of the data points from the GTT.Since the full range of values included in the data set provided acceptable
Bland-Altman plots and moderate agreement according to the LCCC, we conclude that
the hand-held glucometer is acceptable to use for rapid, chute-side testing of blood
glucose concentration in beef cattle. This glucometer was tested during a GTT, and
while all of the data was included in the analyses presented, we caution its use at
supraphysiological glucose concentrations such as that occurring immediately after
glucose infusion during a GTT. However, it seems that under normal physiologic
conditions the hand-held glucometer agrees with the laboratory glucose oxidase
reference method. Additionally, based on our suppliers and current costs to complete
a GTT for 13 steers, the glucometer method was 57% cheaper on a per sample basis
compared with the laboratory method.(XLSX)Click here for additional data file.23 Mar 2022
PONE-D-21-38258
A comparison between two
glucose measurement methods in beef steers during a glucose tolerance
test
PLOS ONE
Dear Dr. Parker,Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we
feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it
currently stands. Therefore, we invite you to submit a revised version of the
manuscript that addresses the points raised during the review process.Because one of the reviewers raised several crucial questions regarding the
hypothesis tested, methods and statistical analysis, I invite you to revise the
manuscript taken into consideration his comments. Also, please make sure to reply
all his comments with detailed answers to allow myself and the reviewer to better
understand what was proposed in the manuscript. Because we did not have many
comments from one of the reviewers at this time, and was quite difficult to find
available reviewers, I am passing the manuscript for you to review with the comments
made to avoid any further delay. However, after your revision the manuscript shall
be sent to a third reviewer for assessment.Please submit your revised manuscript by May 07 2022 11:59PM. If you will need more
time than this to complete your revisions, please reply to this message or contact
the journal office at plosone@plos.org. When
you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions
Needing Revision' folder to locate your manuscript file.Please include the following items when submitting your revised
manuscript:
A rebuttal letter that responds to each point raised by the academic
editor and reviewer(s). You should upload this letter as a separate file
labeled 'Response to Reviewers'.A marked-up copy of your manuscript that highlights changes made to the
original version. You should upload this as a separate file labeled
'Revised Manuscript with Track Changes'.An unmarked version of your revised paper without tracked changes. You
should upload this as a separate file labeled 'Manuscript'.If you would like to make changes to your financial disclosure, please include your
updated statement in your cover letter. Guidelines for resubmitting your figure
files are available below the reviewer comments at the end of this letter.If applicable, we recommend that you deposit your laboratory protocols in
protocols.io to enhance the reproducibility of your results. Protocols.io assigns
your protocol its own identifier (DOI) so that it can be cited independently in the
future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols.
Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol
articles, which describe protocols hosted on protocols.io. Read more information on
sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.We look forward to receiving your revised manuscript.Kind regards,Marcio Duarte, PhDAcademic EditorPLOS ONEJournal Requirements:When submitting your revision, we need you to address these additional
requirements.1. Please ensure that your manuscript meets PLOS ONE's style requirements, including
those for file naming. The PLOS ONE style templates can be found athttps://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf
andhttps://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf2. In your Data Availability statement, you have not specified where the minimal data
set underlying the results described in your manuscript can be found. PLOS defines a
study's minimal data set as the underlying data used to reach the conclusions drawn
in the manuscript and any additional data required to replicate the reported study
findings in their entirety. All PLOS journals require that the minimal data set be
made fully available. For more information about our data policy, please see
http://journals.plos.org/plosone/s/data-availability.Upon re-submitting your revised manuscript, please upload your study’s minimal
underlying data set as either Supporting Information files or to a stable, public
repository and include the relevant URLs, DOIs, or accession numbers within your
revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.
Any potentially identifying patient information must be fully anonymized.Important: If there are ethical or legal restrictions to sharing your data publicly,
please explain these restrictions in detail. Please see our guidelines for more
information on what we consider unacceptable restrictions to publicly sharing data:
http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.
Note that it is not acceptable for the authors to be the sole named individuals
responsible for ensuring data access.We will update your Data Availability statement to reflect the information you
provide in your cover letter.3. We note that you have stated that you will provide repository information for your
data at acceptance. Should your manuscript be accepted for publication, we will hold
it until you provide the relevant accession numbers or DOIs necessary to access your
data. If you wish to make changes to your Data Availability statement, please
describe these changes in your cover letter and we will update your Data
Availability statement to reflect the information you provide.[Note: HTML markup is below. Please do not edit.]Reviewers' comments:Reviewer's Responses to Questions
Comments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with
data that supports the conclusions. Experiments must have been conducted rigorously,
with appropriate controls, replication, and sample sizes. The conclusions must be
drawn appropriately based on the data presented. Reviewer #1: YesReviewer #2: No********** 2. Has the statistical analysis been performed
appropriately and rigorously? Reviewer #1: YesReviewer #2: Yes********** 3. Have the authors made all data underlying the
findings in their manuscript fully available?The PLOS Data policy requires authors to make all
data underlying the findings described in their manuscript fully available without
restriction, with rare exception (please refer to the Data Availability Statement in
the manuscript PDF file). The data should be provided as part of the manuscript or
its supporting information, or deposited to a public repository. For example, in
addition to summary statistics, the data points behind means, medians and variance
measures should be available. If there are restrictions on publicly sharing
data—e.g. participant privacy or use of data from a third party—those must be
specified. Reviewer #1: YesReviewer #2: Yes********** 4. Is the manuscript presented in an intelligible
fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted
articles must be clear, correct, and unambiguous. Any typographical or grammatical
errors should be corrected at revision, so please note any specific errors
here. Reviewer #1: YesReviewer #2: Yes********** 5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may
also include additional comments for the author, including concerns about dual
publication, research ethics, or publication ethics. (Please upload your review as
an attachment if it exceeds 20,000 characters) Reviewer #1: This manuscript aimed to evaluate the effectiveness of a hand-held
glucometer in beef cattle The manuscript is within the scope of the Plos One and can
be accepted for publication. However, this manuscript should be classified as a
short communication.Reviewer #2: Lines 56-59 - I really didn’t understand why you are pointing out these
limitations and possible errors of the laboratory method as a justification for
using the hand-held glucometer. Because you are using this method as a gold-standard
method for your comparison and recommendation of the alternative method.Lines 60-62 - This sentence sounds very informal to a paper’s introduction. I believe
that should be in a more appropriate place, such as discussion.Lines 68-70 - What is the real necessity to validate the same method for beef cattle?
Why the results should be different for beef cattle when compared with dairy cattle?
We are talking about the method.Line 101 and line 104 – Is the range of reading from 20 to 300 or 20 to 500? There is
confusion here.Lines 105-107 - This is not correct. This sample should be deleted from the database
because you are not certain about the value.Lines 118-120 - Why did you not test for normality?? You need to be certain about
distribution because that could be biasing the results.Lines 120-123 – In a general way, the information concerning statistical analysis and
methods for assessing adequacy were poorly described in this section. Some
explanations about these methods are dispersed over the results and discussion,
which is not the best place for that. I suggest that authors should be providing
these details in the current section.Line 132 - Why did you not include regression to test the adequacy of your new
method?Lines 137-139 - I am really not sure about the confidence to test transformed
variables. It is proved that variables transformation to follow Gaussian
distribution, do not represent a good estimate of the real value when a
back-transformation is applied. I believe that is a similar case. The funnel shape
is also present in the transformed data, but with lower intensity. The authors
should think about that.Lines 142-1444 - It is clear that there is a bias over 300 mg/dL (Figures 1 and 2)
and I believe that should be analyzed and reported in the manuscript.Lines 160-174 - This entire paragraph is closer to a material and methods than a
discussion.Lines 215-218 - Which is the necessity to repeat this text?********** 6. PLOS authors have the option to publish the peer
review history of their article (what does this mean?). If published, this will
include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be
made public.Do you want your identity to be public for this peer review? For
information about this choice, including consent withdrawal, please see our
Privacy Policy. Reviewer #1: Yes: Otavio R. Machado NetoReviewer #2: No[NOTE: If reviewer comments were submitted as an attachment file, they will be
attached to this email and accessible via the submission site. Please log into your
account, locate the manuscript record, and check for the action link "View
Attachments". If this link does not appear, there are no attachment files.]While revising your submission, please upload your figure files to the Preflight
Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps
ensure that figures meet PLOS requirements. To use PACE, you must first register as
a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you
will find detailed instructions on how to use the tool. If you encounter any issues
or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting
Information files do not need this step.3 May 2022I have attached a document titled response to reviewers comments that details the
author's response to both reviewersSubmitted filename: Response
to Reviewers PLOSone Glucose steers.docxClick here for additional data file.6 Jul 2022Short communication: A comparison between two glucose measurement methods in beef
steers during a glucose tolerance testPONE-D-21-38258R1Dear Dr. Parker,We’re pleased to inform you that your manuscript has been judged scientifically
suitable for publication and will be formally accepted for publication once it meets
all outstanding technical requirements.Within one week, you’ll receive an e-mail detailing the required amendments. When
these have been addressed, you’ll receive a formal acceptance letter and your
manuscript will be scheduled for publication.An invoice for payment will follow shortly after the formal acceptance. To ensure an
efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the
'Update My Information' link at the top of the page, and double check that your user
information is up-to-date. If you have any billing related questions, please contact
our Author Billing department directly at authorbilling@plos.org.If your institution or institutions have a press office, please notify them about
your upcoming paper to help maximize its impact. If they’ll be preparing press
materials, please inform our press team as soon as possible -- no later than 48
hours after receiving the formal acceptance. Your manuscript will remain under
strict press embargo until 2 pm Eastern Time on the date of publication. For more
information, please contact onepress@plos.org.Kind regards,Marcio Duarte, PhDAcademic EditorPLOS ONEAdditional Editor Comments (optional):Reviewers' comments:Reviewer's Responses to Questions
Comments to the Author1. If the authors have adequately addressed your comments raised in a previous round
of review and you feel that this manuscript is now acceptable for publication, you
may indicate that here to bypass the “Comments to the Author” section, enter your
conflict of interest statement in the “Confidential to Editor” section, and submit
your "Accept" recommendation. Reviewer #2: All comments have been addressed********** 2. Is the manuscript technically sound, and do the data
support the conclusions?The manuscript must describe a technically sound piece of scientific research with
data that supports the conclusions. Experiments must have been conducted rigorously,
with appropriate controls, replication, and sample sizes. The conclusions must be
drawn appropriately based on the data presented. Reviewer #2: Yes********** 3. Has the statistical analysis been performed
appropriately and rigorously? Reviewer #2: Yes********** 4. Have the authors made all data underlying the
findings in their manuscript fully available?The PLOS Data policy requires authors to make all
data underlying the findings described in their manuscript fully available without
restriction, with rare exception (please refer to the Data Availability Statement in
the manuscript PDF file). The data should be provided as part of the manuscript or
its supporting information, or deposited to a public repository. For example, in
addition to summary statistics, the data points behind means, medians and variance
measures should be available. If there are restrictions on publicly sharing
data—e.g. participant privacy or use of data from a third party—those must be
specified. Reviewer #2: Yes********** 5. Is the manuscript presented in an intelligible
fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted
articles must be clear, correct, and unambiguous. Any typographical or grammatical
errors should be corrected at revision, so please note any specific errors
here. Reviewer #2: Yes********** 6. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may
also include additional comments for the author, including concerns about dual
publication, research ethics, or publication ethics. (Please upload your review as
an attachment if it exceeds 20,000 characters) Reviewer #2: The authors have answered all of my concerns. Because of that, I believe
the manuscript is now ready for publication.********** 7. PLOS authors have the option to publish the peer
review history of their article (what does this mean?). If published, this will
include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be
made public.Do you want your identity to be public for this peer review? For
information about this choice, including consent withdrawal, please see our
Privacy Policy. Reviewer #2: Yes: Alex Lopes da Silva**********11 Jul 2022PONE-D-21-38258R1Short communication: A comparison between two glucose measurement methods in beef
steers during a glucose tolerance testDear Dr. Parker:I'm pleased to inform you that your manuscript has been deemed suitable for
publication in PLOS ONE. Congratulations! Your manuscript is now with our production
department.If your institution or institutions have a press office, please let them know about
your upcoming paper now to help maximize its impact. If they'll be preparing press
materials, please inform our press team within the next 48 hours. Your manuscript
will remain under strict press embargo until 2 pm Eastern Time on the date of
publication. For more information please contact onepress@plos.org.If we can help with anything else, please email us at plosone@plos.org.Thank you for submitting your work to PLOS ONE and supporting open access.Kind regards,PLOS ONE Editorial Office Staffon behalf ofDr. Marcio DuarteAcademic EditorPLOS ONE