Short communication: validation of a point-of-care glucometer for use in dairy cows.

J A M Wittrock1, T F Duffield, S J LeBlanc.   

Abstract

The purpose of this study was to evaluate the diagnostic performance of a hand-held electronic glucometer (Precision Xtra; Abbott Diabetes Care Inc., Mississauga, ON, Canada) for cow-side use in dairy cattle. This device has been validated for measuring blood concentrations of β-hydroxybutyrate in dairy cows. This study was designed to assess the accuracy of whole-blood glucose measurements from the glucose meter relative to a reference chemical analyzer in a diagnostic laboratory. Duplicate samples were taken from the same cows at the same time, into blood tubes with either the glycolysis-inhibiting preservative sodium fluoride (NaF) or without preservative. Glucometer readings were taken on whole blood with no preservative, and laboratory measurements were conducted on serum preserved with NaF. Blood samples were collected from cows between 3 wk before and 5 wk after calving, including during a glucose tolerance test conducted 1 wk before expected calving. Passing-Bablok and Bland-Altman data analyses were used to evaluate the performance of the glucometer relative to the laboratory results. A strong correlation was observed in 709 samples from 81 cows between the hand-held meter and serum from samples preserved with NaF (R(2)=0.95). Overall, 96% of measurements with the glucometer fell within the 95% confidence limits of analysis in the laboratory, although at higher-than-physiologic glucose concentrations (>5.2mmol/L) the glucometer tended to overestimate. The hand-held glucometer appears suitable for rapid measurement of glucose under field conditions in dairy cattle.
Copyright © 2013 American Dairy Science Association. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23684029     DOI: 10.3168/jds.2012-6533

Source DB:  PubMed          Journal:  J Dairy Sci        ISSN: 0022-0302            Impact factor:   4.034


  6 in total

1.  Accuracy of 5 Point-of-Care Glucometers in C57BL/6J Mice.

Authors:  Linnea A Morley; Thomas H Gomez; Julia L Goldman; Rene Flores; Mary A Robinson
Journal:  J Am Assoc Lab Anim Sci       Date:  2018-01-01       Impact factor: 1.232

2.  Evaluation of an Electrochemical Point-of-Care Meter for Measuring Glucose Concentration in Blood from Periparturient Dairy Cattle.

Authors:  A A Megahed; M W H Hiew; J R Townsend; J B Messick; P D Constable
Journal:  J Vet Intern Med       Date:  2015-10-20       Impact factor: 3.333

3.  Glucose concentration in capillary blood of dairy cows obtained by a minimally invasive lancet technique and determined with three different hand-held devices.

Authors:  B Mair; M Drillich; D Klein-Jöbstl; P Kanz; S Borchardt; L Meyer; I Schwendenwein; M Iwersen
Journal:  BMC Vet Res       Date:  2016-02-24       Impact factor: 2.741

4.  Accuracy of the point-of-care glucose meter for use in calves.

Authors:  Yasunobu Nishi; Yoshiki Murakami; Marina Otsuka; Kenji Tsukano; Tadaharu Ajito; Kazuyuki Suzuki
Journal:  J Vet Med Sci       Date:  2019-07-15       Impact factor: 1.267

5.  Short communication: A comparison between two glucose measurement methods in beef steers during a glucose tolerance test.

Authors:  Kirsten R Nickles; Alejandro E Relling; Alvaro Garcia-Guerra; Francis L Fluharty; Anthony J Parker
Journal:  PLoS One       Date:  2022-07-19       Impact factor: 3.752

6.  Age at First Gestation in Beef Heifers Affects Fetal and Postnatal Growth, Glucose Metabolism and IGF1 Concentration.

Authors:  Sebastian López Valiente; Alejandro M Rodríguez; Nathan M Long; Graciela Quintans; Florencia E Miccoli; Isabel M Lacau-Mengido; Sebastian Maresca
Journal:  Animals (Basel)       Date:  2021-11-27       Impact factor: 2.752

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.