| Literature DB >> 35852194 |
Alaettin Özer1, Kerim Öner, Ahmet Emin Okutan, Muhammet Salih Ayas.
Abstract
OBJECTIVES: This study aimed to evaluate the performance of four different fixation techniques for Pauwels type III femoral neck fractures considering the fracture morphology in the sagittal plane.Entities:
Mesh:
Year: 2022 PMID: 35852194 PMCID: PMC9361110 DOI: 10.52312/jdrs.2022.676
Source DB: PubMed Journal: Jt Dis Relat Surg ISSN: 2687-4792
Vertical displacements (mm)
| 3CS | DHS+CS | PFLP | 3CS+MBP | |
| 20 anterior | 5.12 | 4.37 | 2.45 | 2.12 |
| Neutral | 4.44 | 2.45 | 2.92 | 1.85 |
| 20 posterior | 4.83 | 3.91 | 2.36 | 1.94 |
| DHS: Dynamic hip screw; CS: Cannulated screws; PFLP: Proximal femoral locking plate; 3CS: Three cannulated screws; MBP: Medial buttress plate. | ||||
Rotational displacements (mm)
| 3CS | DHS+CS | PFLP | 3CS+MBP | |
| 20 anterior | 2.32 | 1.58 | 1.18 | 1.05 |
| Neutral | 2.12 | 1.12 | 1.37 | 0.92 |
| 20 posterior | 2.26 | 1.55 | 1.14 | 1.01 |
| DHS: Dynamic hip screw; CS: Cannulated screws; PFLP: Proximal femoral locking plate; 3CS: Three cannulated screws; MBP: Medial buttress plate. | ||||
Von mises stress of implant (MPa)
| 3CS | DHS+CS | PFLP | 3CS+MBP | |
| 20 anterior | 396.43 | 512.45 | 496.56 | 596.21 |
| Neutral | 361.21 | 453.46 | 412.55 | 567.52 |
| 20 posterior | 321.47 | 405.38 | 425.62 | 528.74 |
| MPa: Megapascal; DHS: Dynamic hip screw; CS: Cannulated screws; PFLP: Proximal femoral locking plate; 3CS: Three cannulated screws; MBP: Medial buttress plate. | ||||
Von mises stress of calcar (MPa)
| 3CS | DHS+CS | DHS+CS | 3CS+MBP | |
| 20 anterior | 46.38 | 29.41 | 19.29 | 16.36 |
| Neutral | 32.25 | 17.45 | 22.13 | 14.11 |
| 20 posterior | 36.43 | 27.84 | 18.34 | 15.82 |
| MPa: Megapascal; DHS: Dynamic hip screw; CS: Cannulated screws; PFLP: Proximal femoral locking plate; 3CS: Three cannulated screws; MBP: Medial buttress plate. | ||||