Literature DB >> 33717906

Biomechanical study of different internal fixations in Pauwels type III femoral neck fracture - A finite elements analysis.

Anderson Freitas1,2, Joubert Vieira Toledo Júnior1, Adeni Ferreira Dos Santos1, Rubén Jeri Aquino1, Vitor Negreiro Leão1, Weverton Péricles de Alcântara1.   

Abstract

OBJECTIVE: To evaluate biomechanical behavior of different internal fixation methods for the treatment of Pauwels Type III femoral neck fractures.
METHODS: Three internal fixators were developed to treat Pauwels Type III femoral neck fracture using finite elements: dynamic hip screw (DHS); DHS combined with anti-rotation screw; three cannulated screws in an inverted triangular configuration (ASNIS). Under the same conditions, vertical fracture displacement, and maximum and minimum principal, and Von Mises stresses were evaluated.
RESULTS: The vertical displacements evaluated were: 5.43 mm, 5.33 mm and 6.22 mm, rotational displacements values were 1.1 mm, 0.4 mm and 1.3 mm; maximum principle stress values obtained for the upper region of the femoral neck were 3.26 hPa, 2.77 hPa, and 4.5 hPa, minimum principal stress values obtained for the inferior region of the femoral neck were -1.97 hPa, -1.8 hPa and t -3.15 hPa; Von Mises peak stress values were 340.0 MPa, 315.5 MPa and 326.1 Mpa, for DHS, DHS with anti-rotation device, and ASNIS, respectively.
Conclusion: The DHS combined with anti-rotation screw yielded better results in terms of rotational and vertical displacements, traction and compression distributions on fractures, and Von Mises stress, demonstrating mechanical superiority for Pauwels Type III fracture.
© 2020 Delhi Orthopedic Association. All rights reserved.

Entities:  

Keywords:  Biomechanics; Femoral neck fracture; Finite element analysis; Internal fixation

Year:  2020        PMID: 33717906      PMCID: PMC7920117          DOI: 10.1016/j.jcot.2020.06.006

Source DB:  PubMed          Journal:  J Clin Orthop Trauma        ISSN: 0976-5662


  23 in total

1.  Analysis of fracture gap changes, dynamic and static stability of different osteosynthetic procedures in the femoral neck.

Authors:  F A Bonnaire; A T Weber
Journal:  Injury       Date:  2002-12       Impact factor: 2.586

2.  Results of internal fixation of Pauwels type-3 vertical femoral neck fractures.

Authors:  Martyn J Parker
Journal:  J Bone Joint Surg Am       Date:  2009-02       Impact factor: 5.284

3.  Biomechanics of Femoral Neck Fractures and Implications for Fixation.

Authors:  Peter Augat; Emily Bliven; Simon Hackl
Journal:  J Orthop Trauma       Date:  2019-01       Impact factor: 2.512

Review 4.  Biomechanical rationale for implant choices in femoral neck fracture fixation in the non-elderly.

Authors:  Michalis Panteli; Paul Rodham; Peter V Giannoudis
Journal:  Injury       Date:  2015-01-03       Impact factor: 2.586

5.  Fixation of intracapsular fractures of the femoral neck in young patients: risk factors for failure.

Authors:  A D Duckworth; S J Bennet; J Aderinto; J F Keating
Journal:  J Bone Joint Surg Br       Date:  2011-06

Review 6.  Complications following young femoral neck fractures.

Authors:  G P Slobogean; S A Sprague; T Scott; M Bhandari
Journal:  Injury       Date:  2014-10-31       Impact factor: 2.586

7.  Operative treatment of femoral neck fractures in patients between the ages of fifteen and fifty years.

Authors:  George J Haidukewych; Walter S Rothwell; David J Jacofsky; Michael E Torchia; Daniel J Berry
Journal:  J Bone Joint Surg Am       Date:  2004-08       Impact factor: 5.284

8.  Stability of L-shaped and inverted triangle fixation assemblies in treating Pauwels type II femoral neck fracture: a comparative mechanical study.

Authors:  Vincenzo Giordano; Roger Pletsch Paes; Danilo Diniz Alves; Arthur Bonfim Amaral; William D Belangero; Marcos Giordano; Anderson Freitas; Hilton A Koch
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-04-24

9.  Biomechanical Evaluation of Internal Fixation of Pauwels Type III Femoral Neck Fractures: A Systematic Review of Various Fixation Methods.

Authors:  Yong-Han Cha; Jun-Il Yoo; Seok-Young Hwang; Kap-Jung Kim; Ha-Yong Kim; Won-Sik Choy; Sun-Chul Hwang
Journal:  Clin Orthop Surg       Date:  2019-02-18

10.  Analysis on the mechanical resistance of fixation of femoral neck fractures in synthetic bone, using the dynamic hip system and an anti-rotation screw.

Authors:  Anderson Freitas; Gustavo Melo Torres; André Cezar de Andrade de Mello E Souza; Rafael Almeida Maciel; Diogo Ranier de Macedo Souto; George Neri de Barros Ferreira
Journal:  Rev Bras Ortop       Date:  2014-10-27
View more
  6 in total

Review 1.  Does the femoral neck system provide better outcomes compared to cannulated screws fixation for the management of femoral neck fracture in young adults? A systematic review of literature and meta-analysis.

Authors:  Rajesh Kumar Rajnish; Amit Srivastava; Pratik M Rathod; Rehan Ul Haq; Sameer Aggarwal; Prasoon Kumar; Ish Kumar Dhammi; Ankit Dadra
Journal:  J Orthop       Date:  2022-05-11

2.  Clinical Outcome and Biomechanical Analysis of Dynamic Hip Screw Combined with Derotation Screw in Treating Displaced Femoral Neck Fractures Based on Different Reduction Qualities in Young Patients (≤65 Years of Age).

Authors:  Jian Zhu; Yonglong Li; Yingze Zhang; Xiaodong Cheng
Journal:  Biomed Res Int       Date:  2022-01-05       Impact factor: 3.411

3.  Comparative finite element analysis of four different internal fixation implants for Pauwels type III femoral neck fractures in various fracture angles in the sagittal plane.

Authors:  Alaettin Özer; Kerim Öner; Ahmet Emin Okutan; Muhammet Salih Ayas
Journal:  Jt Dis Relat Surg       Date:  2022-07-06

4.  Evaluation of a Locking Autocompression Screw Model in Pauwels Type-3 Femoral Neck Fracture: In Vitro Analysis.

Authors:  Vincenzo Giordano; Anderson Freitas; Robinson Esteves Pires; Leonardo Rigobello Battaglion; Mariana de Oliveira Lobo; William Dias Belangero
Journal:  Bioengineering (Basel)       Date:  2022-09-12

5.  Treatment of Pauwels type III femoral neck fracture with medial femoral neck support screw: a biomechanical and clinical study.

Authors:  Zhichao Gao; Mei Wang; Baojie Shen; Xiaodong Chu; Di Ruan
Journal:  Sci Rep       Date:  2021-11-01       Impact factor: 4.379

6.  Comparing Perioperative Outcome Measures of the Dynamic Hip Screw and the Femoral Neck System.

Authors:  Marcel Niemann; Karl F Braun; Sufian S Ahmad; Ulrich Stöckle; Sven Märdian; Frank Graef
Journal:  Medicina (Kaunas)       Date:  2022-02-26       Impact factor: 2.430

  6 in total