| Literature DB >> 35844971 |
Yuetian Wang1, Chun Liu2, Huiling Liu3, Haoyong Fu1, Chunde Li1, Lei Yang3,4, Haolin Sun1.
Abstract
Purpose: To evaluate the effect of cement augmentation of cortical bone trajectory (CBT) screws using a novel calcium phosphate-based nanocomposite (CPN). Material andEntities:
Keywords: CBT screws; CPN; PMMA; cement augmentation; osteoporotic spine
Mesh:
Substances:
Year: 2022 PMID: 35844971 PMCID: PMC9278980 DOI: 10.2147/IJN.S365149
Source DB: PubMed Journal: Int J Nanomedicine ISSN: 1176-9114
Figure 1Cortical bone trajectory (CBT) screw placement under the guidance of 3D-printed templates and cement injection.
Information and Characterization of Cadaver Specimens
| Cadaver Specimens | Vertebra | Age (Years) | Mean BMD (g/cm2) | Augmentation Materials Injected on the Left Side | Cement Volume Injected on the Left Side (mL) | Augmentation Materials Injected on the Right Side | Cement Volume Injected on the Right Side (mL) | Biomechanical Test |
|---|---|---|---|---|---|---|---|---|
| 1 | L1 | 81 | 0.792 | Control | 0 | — | — | Torsion test |
| L2 | CPN | 1 | Control | 0 | Torsion test | |||
| L3 | Control | 0 | Control | 0 | Torsion test | |||
| L4 | CPN | 1 | CPN | 1 | Torsion test | |||
| L5 | PMMA | 1 | PMMA | 1 | Torsion test | |||
| 2 | L1 | 63 | 0.822 | Control | 0 | CPN | 1 | Torsion test |
| L2 | PMMA | 1 | PMMA | 1 | Torsion test | |||
| L3 | PMMA | 1 | CPN | 1 | Torsion test | |||
| L4 | PMMA | 1 | Control | 0 | Torsion test | |||
| L5 | PMMA | 1 | CPN | 1 | Torsion test | |||
| 3 | L3 | 78 | 0.741 | PMMA | 1 | CPN | 1 | Torsion test |
| L4 | PMMA | 1 | Control | 0 | Torsion test | |||
| L5 | CPN | 1 | Control | 0 | Torsion test | |||
| L1 | Control | 0 | PMMA | 1 | Pullout test | |||
| L2 | PMMA | 1 | CPN | 1 | Pullout test | |||
| 4 | L1 | 70 | 0.783 | CPN | 1 | Control | 0 | Pullout test |
| L2 | PMMA | 1 | Control | 0 | Pullout test | |||
| L3 | CPN | 1 | PMMA | 1 | Pullout test | |||
| L4 | Control | 0 | CPN | 1 | Pullout test | |||
| L5 | Control | 0 | PMMA | 1 | Pullout test | |||
| 5 | L1 | 67 | 0.802 | PMMA | 1 | CPN | 1 | Pullout test |
| L2 | CPN | 1 | Control | 0 | Pullout test | |||
| L3 | PMMA | 1 | Control | 0 | Pullout test | |||
| L4 | CPN | 1 | PMMA | 1 | Pullout test | |||
| L5 | Control | 0 | CPN | 1 | Pullout test |
Abbreviations: BMD, bone mineral density; CPN, calcium phosphate–based nanocomposite; PMMA, polymethylmethacrylate.
Figure 2Radiological evaluation of cement-augmented cortical bone trajectory (CBT) screws.
Figure 3Biomechanical testing of cemented cortical bone trajectory screws.
Figure 4The results of the setting time and rheological properties.
Figure 5The results of compressive tests.
Figure 6Results of biomechanical tests.
Figure 7Histological analysis of the calcium phosphate–based nanocomposite (CPN) after implantation in a rat cranial defect model.