| Literature DB >> 35836559 |
Thangi Sowjanya1, Krishna Prasad Parvathaneni1, Tbvg Raju1, N Mahendra Varma1, Gowtham Dev Dondapati1, Srivalli Podili1.
Abstract
Introduction and Aim: Postoperative inflammation and endodontic failure occur due to apical extrusion of debris. The purpose of this in vitro study was to assess the amount of apically extruded debris during the root canal preparation using rotary and reciprocating nickel-titanium instrumentation systems. Materials andEntities:
Keywords: Apical extrusion; Flexicon X1; Flexicon X7; Hyflex EDM; continuous rotation motion; reciprocating motion
Year: 2022 PMID: 35836559 PMCID: PMC9274688 DOI: 10.4103/jcd.jcd_631_20
Source DB: PubMed Journal: J Conserv Dent ISSN: 0972-0707
Figure 1(a) Debris collection apparatus (b) glass vials were wrapped with aluminum foil to avoid bias (c) Eppendorf tube with dried debris (d) readings of Eppendorf tubes with dried debris with electronic analytical balance of sensitivity 10 − 5 g
File sequence
| Group | Mode of preparation | Sequence of instrumentation | Torque (NCM)/RPM |
|---|---|---|---|
| Group - A (Hyflex EDM) | Crown down | 0.12/#25 (orifice shaper) → 0.05/#10 (Glidepath file/working length) → 0.06/#25 one file (apical finishing till working length) | 2.5/500 |
| Group - B (flexicon X7) | Crown down | X7: 0.04/#17 (till working length) → 0.06/#17 (till working length) →0.04/#20 (till working length) →0.06/#20 (till working length) →0.04/#25 (apical finishing till working length) → 0.06/#25 (apical finishing till working length) | 2.5/350 |
| Group - C (flexicon X1) | Crown down | X1: 0.04/#17 (till working length) → 0.06/#17 (till working length) →0.06/#20 (till working length) → 0.06/#25 (apical finishing till working length) | 2.5/350 |
EDM: Electric Discharge Machine, NCM: Newton-metre, RPM: Revolutions per minute
The mean values and standard deviation of apically extruded debris between three groups
| Group | Apical extrusion of debris (mg) | ||||
|---|---|---|---|---|---|
|
| |||||
| n | Minimum extrusion | Maximum extrusion | Mean | SD | |
| Group - A (Hyflex EDM) | 25 | −0.00027 | 0.03054 | 0.00741 | 0.00000 |
| Group - B (flexicon X7) | 25 | 0.00215 | 0.02647 | 0.00670 | 0.00001 |
| Group - C (flexicon X1) | 25 | 0.00019 | 0.02647 | 0.00775 | 0.00000 |
SD: Standard deviation, EDM: Electric Discharge Machine
Inter-group comparison of all groups with rotary and reciprocating systems by using one way ANOVA test with a level of significance set at P<0.05
| Comparison between groups | P |
|---|---|
| Group - A (Hyflex EDM) | |
| Group - B (flexicon X7) | 0.606653 |
| Group - C (flexicon X1) | 0.88583 |
| Group - B (flexicon X7) | |
| Group - C (flexicon X1) | 0.720746 |
EDM: Electric Discharge Machine
Bar Diagram 1Mean wise distribution of the apical extrusion of debris for all the three groups. (Group C Highest and B Least debris extrusion among the groups)