| Literature DB >> 35818511 |
Parul Ichhpujani1, Hennaav Dhillon1.
Abstract
Tablet based perimetry has the potential to be used as a low-cost, portable method for glaucoma screening. With the advent of tablets with large dynamic luminance range and high spatial resolution, perimetry applications can test visual field locations within the central 30 degrees, just like bowl perimetry. Tablet perimetry has garnered interest in this COVID era as it can be promptly done in the waiting room or even from the comfort of home. The current review focuses on evaluating glaucoma detection sensitivities, diagnostic abilities and correlation of each parameter and examination time of the iPad based perimetry applications with those of HFA.Entities:
Keywords: iPad; standard automated perimetry; tablet perimetry; visual field
Year: 2022 PMID: 35818511 PMCID: PMC9270904 DOI: 10.2147/OPTH.S340508
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Basic Parameters of Available Applications (Apps) for Tablet Perimetry
| Application | Test Locations/Degree of Field | Target Size | Sensitivity (%) (Ref No. as Superscript) | Specificity (%) (Ref No. as Superscript) |
|---|---|---|---|---|
| 96; Central 30 degrees | Size V | 77.8 (mild glaucoma) | 52.6 (mild glaucoma) | |
| 67 | 77 | |||
| 66; 34° horizontal X 25° vertically | 40% larger than Goldman Size II | 88.4 | 81 (moderate glaucoma) | |
| 60.9 (moderate to severe glaucoma) | 86.5 (moderate to severe glaucoma) | |||
| 24° horizontally X 20° vertically | 3 mm in diameter | 54–59 (mild glaucoma) | 85–94 | |
| 71–86 (moderate to severe glaucoma) | ||||
| Entire 24° visual field | Size (pixels/px): 2 × 2 (smallest), 4 × 4, 8 × 8, 16 × 16, 32 × 32 (largest) | 91.2 (moderate to severe glaucoma) | – | |
| 24–2 grid; 22 paracentral locations | Size III | 97 | 100 |