| Literature DB >> 32908686 |
Harsh Kumar1, Mithun Thulasidas1.
Abstract
PURPOSE: To compare visual field results obtained using Melbourne Rapid Fields (MRF) iPad-based perimeter software and Humphrey Field Analyzer (HFA) 24-2 Swedish Interactive Threshold Algorithm (SITA) standard program in glaucoma patients.Entities:
Year: 2020 PMID: 32908686 PMCID: PMC7463344 DOI: 10.1155/2020/8384509
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Nine zones with 66 test spots for the Melbourne Rapid Fields (MRF) full test and 54 test spots for the Humphrey Field Analyzer (HFA) 24-2 grid.
Figure 2Fixation changes during Melbourne Rapid Fields (MRF) testing (for 9.7″, 10.5″, and 11″ iPads).
Figure 3The iPad-keyboard-viewing-shield apparatus with a headrest maintaining 33 cm viewing distance for Melbourne Rapid Fields (MRF) testing.
Figure 4Visual field output for Melbourne Rapid Fields (MRF) full test.
Comparison of various test parameters for Melbourne Rapid Fields (MRF) and Humphrey Field Analyzer (HFA) 24-2 Swedish Interactive Threshold Algorithm (SITA) standard program.
| Test parameter | MRF (mean ± SD) | HFA 24-2 SITA standard (mean ± SD) |
|
|---|---|---|---|
| MD (dB) | −8.61 ± 7.21 | −11.71 ± 9.36 | <0.001‡ |
| PSD (dB) | 9.09 ± 3.57 | 7.68 ± 4.19 | 0.005‡ |
| VC (for MRF)/VFI (for HFA) (%) | 70.63 ± 24.56 | 68.43 ± 27.08 | 0.096 |
| Foveal threshold (dB)† | 25.18 ± 7.01 | 34.43 ± 3.17 | <0.001‡ |
| Test time (seconds)† | 342.07 ± 56.70 | 375.11 ± 88.95 | 0.046§ |
| Depressed points with | 21.61 ± 12.86 | 27.61 ± 14.51 | <0.001‡ |
MD = mean deviation, PSD = pattern standard deviation, VC = visual capacity, VFI = visual field index, MRF = Melbourne rapid fields, HFA = Humphrey field analyzer, SITA = Swedish interactive thresholding algorithm, SD = standard deviation. Wilcoxon signed rank test analysis. †Paired t-test analysis. ‡Highly significant with level of significance: P < 0.01. §Significant with level of significance: P < 0.05 at 95.0% confidence limit.
Glaucoma hemifield test/colour coded indicator results obtained with Humphrey Field Analyzer (HFA) 24-2 Swedish Interactive Threshold Algorithm (SITA) standard and Melbourne Rapid Fields (MRF).
| Colour coded indicator/glaucoma hemifield test classification | MRF (n) | HFA 24-2 SITA standard (n) |
|---|---|---|
| Red/outside normal limits | 26 | 28 |
| Amber/borderline | 1 | 0 |
| Green/within normal limits | 1 | 0 |
MRF = Melbourne rapid fields, HFA = Humphrey field analyzer, SITA = Swedish interactive thresholding algorithm.
Figure 5Bland–Altman plots showing agreement in (a) mean deviation (MD), (b) pattern standard deviation (PSD), (c) visual field index (VFI)/visual capacity (VC), and (d) the number of depressed points with P < 5% on PSD probability plot between Humphrey Field Analyzer (HFA) 24-2 Swedish Interactive Threshold Algorithm (SITA) Standard and Melbourne Rapid Fields (MRF).