Literature DB >> 35812684

Osteotomy combined with the trephine technique for invisible implant fracture: A case report.

Liang-Wen Chen1, Min Wang1, Hai-Bin Xia1, Dong Chen2.   

Abstract

BACKGROUND: Implant fracture is one of the most serious mechanical complications of dental implants. Conventional treatment necessitates visibility of the apical portion of the fractured implant, whereas for deep and invisible implant fractures, the traditional trephine method has been ineffective. Surgical removal of the marginal bone to expose the fracture surface would be a time-consuming and extensively damaging procedure. Here, we propose a novel technique to address invisible implant fractures. CASE
SUMMARY: A 50-year-old woman was referred to our department with the chief complaint that her right mandibular implant tooth had fallen out 3 mo earlier. Cone-beam computed tomography examination showed an implant fracture with a fracture surface 5.1 mm below the crestal ridge. The patient was treated with osteotomy combined with the trephine technique to expose the surgical field and remove the implant. The invisible fractured implant was successfully removed, with minimal trauma. A modified Wafer technique-supported guided bone regeneration treatment was then administered to restore the buccal bone wall and preserve the bone mass. Six months later, fine regenerative bone and a wide alveolar crest in the edentulous area were observed, and a new implant was placed. Four months later, restoration was completed using a cemented ceramic prosthesis. Clinical and radiographic examinations 12 mo after loading fulfilled the success criteria. The patient reported no complaints and was satisfied.
CONCLUSION: Osteotomy combined with the trephine technique can be effectively used to address deep and invisible implant fractures. ©The Author(s) 2022. Published by Baishideng Publishing Group Inc. All rights reserved.

Entities:  

Keywords:  Case report; Dental implant; Explanation trephine; Invisible implant fracture; Osteotomy; Transformed Wafer technique

Year:  2022        PMID: 35812684      PMCID: PMC9210918          DOI: 10.12998/wjcc.v10.i16.5479

Source DB:  PubMed          Journal:  World J Clin Cases        ISSN: 2307-8960            Impact factor:   1.534


Core Tip: This article provides up-to-date knowledge regarding the clinical presentation, incidence, risk factors, diagnosis, and management of implant fractures. Conventional treatment requires that the apical portion of the fractured implant be visible. For deep and invisible implant fractures, the traditional trephine method has been ineffective. We developed a novel method of osteotomy combined with the trephine technique that can effectively address deep and invisible implant fractures. Meanwhile, modified Wafer technique-supported guided bone regeneration treatment was conducted simultaneously to preserve the bone mass in the surgical area.

INTRODUCTION

Clinical presentation

Implant fracture is a kind of destructive mechanical complication, which is generally accompanied by the following clinical characteristics: Loss of retention, marginal bone resorption[1], soft tissue inflammation[2], and occlusal disorder[3]. Consequently, implant fracture can cause significant loss to patients’ economies, time, and emotions. According to research involving large amounts of data and long-term follow-up, the incidence of implant fracture was between 0.4%[4] and 0.92%[5] after over 5 years of loading.

Diagnosis

The diagnosis of implant fractures requires the detection of implant mobility by clinical examination and of the fracture line by X-ray images[1]. Once diagnosed as an implant fracture, conventional treatment options are: (1) Complete removal of the remaining fractured implant using trephines[6]; (2) Removal and replacement of the coronal portion of the fractured implant with a new prosthetic restoration[7]; (3) Leaving the remaining apical part integrated into the bone and placing a fixed bridge instead[8,9]; and (4) Using a computer-aided design/computer-aided manufacturing surgical guide to remove the buried implant[10].

Treatment

However, conventional treatments necessitate visibility of the fractured implant. In case of deep and invisible implant fractures, the position and orientation of the implants are not perceptible. No effective solution has been reported for invisible implant fractures. Surgical removal of the marginal bone to expose the fracture surface could be time-consuming and extensively damaging. Herein, we report a novel method to address invisible implant fractures, as an attempt to form new treatment modalities for such cases. This manuscript was prepared according to the CARE checklist[11].

CASE PRESENTATION

Chief complaints

A 50-year-old non-smoking woman was referred to our clinic in August 2017 with the chief complaint that her mandibular right implant tooth (tooth 46) had fallen out 3 mo earlier.

History of present illness

According to the medical record, the implant (Ankylos®, 4.5 mm × 9.5 mm, Germany) was placed in the tooth 46 position in October 2013 and loaded after 6 mo. The patient complained of implant restoration mobility after 2 years of function, with the symptoms worsening while chewing. In May 2017, the patient experienced a foreign body in her mouth, which was the fractured portion of the implant connected to the prosthesis (Figure 1A).
Figure 1

The condition of the fractured implant when the patient came to the clinic. A: Fractured portion of the implant connected to the prosthesis; B: Panoramic radiograph showed no signs of peri-implantitis; C: CBCT showed that the remaining implant was 6.0 mm away from the inferior alveolar nerve canal, 3.9 mm away from the buccal cortical bone wall, 5.1 mm away from the lingual bone wall and 5.1 mm beneath the crestal ridge.

The condition of the fractured implant when the patient came to the clinic. A: Fractured portion of the implant connected to the prosthesis; B: Panoramic radiograph showed no signs of peri-implantitis; C: CBCT showed that the remaining implant was 6.0 mm away from the inferior alveolar nerve canal, 3.9 mm away from the buccal cortical bone wall, 5.1 mm away from the lingual bone wall and 5.1 mm beneath the crestal ridge.

History of past illness

The patient reported no history of systemic diseases, long-term medication, or family genetic history.

Personal and family history

No smoking and drinking history, and no hereditary family history were reported.

Physical examination

The patient’s blood pressure was 110/70 mmHg, with a pulse rate of 80 beats per minute (bpm).

Laboratory examinations

The routine blood indicators and coagulation profile were within normal range.

Imaging examinations

Cone-beam computed tomography (CBCT) examination showed that the residual portion of the implant was osseointegrated with no signs of peri-implantitis. Moreover, there appeared to be a high-density image above the fracture surface of the implant. The remaining apical portion of the implant was 6.0 mm away from the inferior alveolar nerve canal, 3.9 mm away from the buccal cortical bone wall, 5.1 mm away from the lingual bone wall, and 5.1 mm away from the crestal ridge (Figure 1B and C).

FINAL DIAGNOSIS

The patient was diagnosed with implant fracture.

TREATMENT

Treatment planning

The risks and benefits of each treatment option were discussed in detail with the patient. The first plan was the fabrication of a three-dimensional (3D) surgical guide for the trephine to achieve precise removal of the remaining implant. However, the 3D surgical guide required at least 20 mm of the trephine with the graduated part, whereas the existing trephines measured only 16 mm. The second plan involved fixed partial restoration, leaving the apical portion of the fractured implant integrated into the bone. However, the patient rejected this plan as she was unwilling to risk damage to her healthy teeth. Consequently, a third treatment plan to perform osteotomy combined with the trephine technique to remove the invisible fractured implant was devised, which was approved by the patient. Informed consent was obtained before treatment.

Surgical procedure

The rapidly growing gingiva had closed the gingival outlet of the implant (Figure 2A). After local anesthesia was administered, a crestal full-thickness flap was raised and an implant hole filled with granulation tissue was observed (Figure 2B). Granulation tissue was removed using a dental excavator and a turbine drill. Thereafter, two vertical incisions and one horizontal incision were made using an ultrasonic osteotome on the buccal side of the alveolar bone where the remaining implant was located (Figure 3A). The buccal bone plate was removed using a bone chisel and hammer (Figure 3B) and soaked in saline temporarily. After the buccal bone plate was removed, the broken end of the fractured implant was clearly exposed (Figure 3C).
Figure 2

Occlusal view before remaining implant removal. A: The rapidly growing gingiva had closed the gingival outlet of the implant; B: After a crestal full-thickness flap was raised, the implant hole filled with granulation tissue was observed.

Figure 3

The osteotomy procedure. A: Two vertical incisions and one horizontal incision were made using an ultrasonic osteotome on the buccal plate; B: The buccal bone plate was removed with a bone chisel and a hammer; C: The broken end of the fractured implant was clearly exposed.

Occlusal view before remaining implant removal. A: The rapidly growing gingiva had closed the gingival outlet of the implant; B: After a crestal full-thickness flap was raised, the implant hole filled with granulation tissue was observed. The osteotomy procedure. A: Two vertical incisions and one horizontal incision were made using an ultrasonic osteotome on the buccal plate; B: The buccal bone plate was removed with a bone chisel and a hammer; C: The broken end of the fractured implant was clearly exposed. Next, the remaining implant was completely removed with a graduated trephine (Nobel Biocare®, Figure 4A), with the inner diameter of the trephine slightly larger than the outer diameter of the implant. From the occlusal view, a ring of uniform thickness was created around the remaining implant, which did not cause any unnecessary bone defects of the lateral wall and was achieved with minimal bone removal (Figure 4B). The remaining implant was then removed (Figure 4C). The surface of the implant was covered with a thin layer of osseointegrated alveolar bone (Figure 4D). After the surgery, a modified Wafer technique was performed; the buccal bone plate was repositioned in situ using a titanium screw with slight rotation (Figure 5A) to build the external wall of the osteogenic space. Guided bone regeneration (GBR) was performed. The alveolar bone defect was filled with demineralized bovine bone matrix (DBBM; Geistlich Bio-Oss), and the filled defect was covered with a biological membrane (Geistlich Bio-Gide). Finally, the wound was closed up tightly (Figure 5B).
Figure 4

Removal of the fractured implant with a trephine. A: The remaining implant was completely removed with a graduated trephine; B: A ring of uniform thickness was created around the remaining implant from the occlusal view; C: The remaining segment was moved out; D: The surface of the implant was covered with a thin layer of osseointegrated alveolar bone.

Figure 5

A modified Wafer technique-supported guided bone regeneration treatment was conducted simultaneously to preserve the horizontal alveolar dimension. A: The buccal bone plate was repositioned in situ using a titanium nail with slight rotation; B: The wound was closed up tightly.

Removal of the fractured implant with a trephine. A: The remaining implant was completely removed with a graduated trephine; B: A ring of uniform thickness was created around the remaining implant from the occlusal view; C: The remaining segment was moved out; D: The surface of the implant was covered with a thin layer of osseointegrated alveolar bone. A modified Wafer technique-supported guided bone regeneration treatment was conducted simultaneously to preserve the horizontal alveolar dimension. A: The buccal bone plate was repositioned in situ using a titanium nail with slight rotation; B: The wound was closed up tightly.

OUTCOME AND FOLLOW-UP

Six months after GBR, fine regenerative bone and wide alveolar crest in the edentulous area were observed. A new implant (Ankylos®, 4.5 mm × 9.5 mm, Germany) was placed in the #30 position. Four months later, restoration was completed using a cemented ceramic prosthesis. Periapical radiographic examination immediately after crown restoration showed a well-osseointegrated implant (Figure 6A). Furthermore, the clinical and radiographic examinations performed 12 mo after loading fulfilled the success criteria outlined by Papaspyridakos et al[13] (Figure 6B)[12,13]. The patient reported no complaints and was very satisfied.
Figure 6

The periapical radiographic examination. A: Immediately after the crown restoration showed a well-osseointegrated new implant; B: After 12 mo of function.

The periapical radiographic examination. A: Immediately after the crown restoration showed a well-osseointegrated new implant; B: After 12 mo of function.

DISCUSSION

The risk factors for implant fractures remain unclear. The possible risk factors include: (1) Implant diameter and length[5,14]—it is considered that the smaller the diameter, the lower the resistance of the implant to fracture; (2) Implant location (i.e., type of bone)[5,15,16], as implants placed in the anterior or premolar region bear less mechanical overloading than those in the posterior region; (3) Direct adjacency to cantilever[17]; (4) Parafunctional habits (teeth clenching and bruxism)[9,18]; and (5) Patient-related factors (smoking and alcohol intake)[19,20]. This case presents a novel method for managing invisible implant fractures. In brief, buccal bone plate osteotomy was conducted at the fractured implant site, followed by complete removal of the implant with a trephine to achieve minimal trauma. Subsequently, the modified Wafer technique was used to preserve the horizontal alveolar dimension. It is worth mentioning that the incision of the buccal bone plate was based on the CBCT results. Accurate osteotomy depth control was achieved by marking the ultrasonic osteotome in advance. In the present case, the surface of the removed implant was covered with a thin layer of osseointegrated alveolar bone. It was assumed that the trephine did not cut into the implant, and no titanium particles remained in the operation area. The use of cortical bone plate supported GBR for bone mass preservation was inspired by the 3D bone reconstruction technique named the Bilaminar cortical tenting grafting technique proposed by Yu et al[21] and the Wafer technique proposed by Merli et al[22] According to the bilaminar cortical tenting grafting technique, an autogenic bone block is bisected into two cortical laminae to construct the buccal and palatal walls of an alveolar ridge defect, and the inter-laminar space is filled with bone graft. In the Wafer technique, an autogenous plate is horizontally harvested and vertically fixed, partially supported by the residual bone wall, and the biomaterial is filled in. Both the techniques provide sufficient bone mass. We used the modified Wafer technique at the implant removal defect site to reconstruct the horizontal alveolar dimension, with the buccal bone plate obtained from the bone defect, thus avoiding a second surgical area[23]. Osteotomy combined with the trephine technique has many advantages in addressing deep and invisible implant fractures. This procedure avoids excessive bone removal or titanium particles, which can affect wound healing. The limitation of this technique is that immediate implantation cannot be achieved owing to insufficient primary stability. Therefore, this technique must be used in conjunction with delayed implantation, which is more time-consuming.

CONCLUSION

This article proposes a novel osteotomy combined with the trephine technique, which effectively addressed an invisible implant fracture 5.1 mm beneath the alveolar crest. In addition, this article provides up-to-date knowledge regarding the clinical presentation, incidence, risk factors, diagnosis, and management of dental implant fractures.
  23 in total

Review 1.  Osseointegrated implant fracture: causes and treatment.

Authors:  Walter Cristiano Gealh; Valéria Mazzo; Francisco Barbi; Edevaldo Tadeu Camarini
Journal:  J Oral Implantol       Date:  2010-06-16       Impact factor: 1.779

2.  A Novel Approach to Bone Reconstruction: The Wafer Technique.

Authors:  Mauro Merli; Marco Moscatelli; Giorgia Mariotti; Alessandro Motroni; Annalisa Mazzoni; Simona Mazzoni; Lorenzo Breschi; Michele Nieri
Journal:  Int J Periodontics Restorative Dent       Date:  2017 May/Jun       Impact factor: 1.840

3.  Implant fracture failure rate and potential associated risk indicators: An up to 12-year retrospective study of implants in 5,124 patients.

Authors:  Dong-Woon Lee; Na-Hong Kim; Young Lee; Yeon-Ah Oh; Jae-Hong Lee; Hyung-Keun You
Journal:  Clin Oral Implants Res       Date:  2019-02-18       Impact factor: 5.977

4.  Incidence and Management of Fractured Dental Implants: Case Reports.

Authors:  Soo-Young Jin; Su-Gwan Kim; Ji-Su Oh; Jae-Seek You; Mi-Ae Jeong
Journal:  Implant Dent       Date:  2017-10       Impact factor: 2.454

5.  Bilamina cortical tenting grafting technique for three-dimensional reconstruction of severely atrophic alveolar ridges in anterior maxillae: A 6-year prospective study.

Authors:  Huajie Yu; Luo Chen; Yibo Zhu; Lixin Qiu
Journal:  J Craniomaxillofac Surg       Date:  2016-04-29       Impact factor: 2.078

6.  Factors influencing the fracture of dental implants.

Authors:  Bruno Ramos Chrcanovic; Jenö Kisch; Tomas Albrektsson; Ann Wennerberg
Journal:  Clin Implant Dent Relat Res       Date:  2017-12-06       Impact factor: 3.932

7.  Bending overload and implant fracture: a retrospective clinical analysis.

Authors:  B Rangert; P H Krogh; B Langer; N Van Roekel
Journal:  Int J Oral Maxillofac Implants       Date:  1995 May-Jun       Impact factor: 2.804

8.  Incidence and pattern of implant fractures: A long-term follow-up multicenter study.

Authors:  Jae-Hong Lee; Yeon-Tae Kim; Seong-Nyum Jeong; Na-Hong Kim; Dong-Woon Lee
Journal:  Clin Implant Dent Relat Res       Date:  2018-05-15       Impact factor: 3.932

Review 9.  A systematic review of the 5-year survival and complication rates of implant-supported single crowns.

Authors:  Ronald E Jung; Bjarni E Pjetursson; Roland Glauser; Anja Zembic; Marcel Zwahlen; Niklaus P Lang
Journal:  Clin Oral Implants Res       Date:  2007-12-07       Impact factor: 5.977

10.  Implant success lies in complications management: A report of two cases.

Authors:  Sapna Rani; Jyoti Devi; Mahesh Verma
Journal:  Contemp Clin Dent       Date:  2016 Apr-Jun
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.