| Literature DB >> 35754514 |
Chhayly Lim1, Jungyeon Kim2, Jeongseok Kim1, Byeong-Gwon Kang3, Yunyoung Nam4.
Abstract
Wearable health devices and respiratory rates (RRs) have drawn attention to the healthcare domain as it helps healthcare workers monitor patients' health status continuously and in a non-invasive manner. However, to monitor health status outside healthcare professional settings, the reliability of this wearable device needs to be evaluated in complex environments (i.e., public street and transportation). Therefore, this study proposes a method to estimate RR from breathing sounds recorded by a microphone placed inside three types of masks: surgical, a respirator mask (Korean Filter 94), and reusable masks. The Welch periodogram method was used to estimate the power spectral density of the breathing signals to measure the RR. We evaluated the proposed method by collecting data from 10 healthy participants in four different environments: indoor (office) and outdoor (public street, public bus, and subway). The results obtained errors as low as 0% for accuracy and repeatability in most cases. This research demonstrated that the acoustic-based method could be employed as a wearable device to monitor RR continuously, even outside the hospital environment.Entities:
Keywords: Face mask; Microphone; Power spectral density; Respiratory rate; Wearable health device
Year: 2022 PMID: 35754514 PMCID: PMC9206076 DOI: 10.1007/s11227-022-04622-0
Source DB: PubMed Journal: J Supercomput ISSN: 0920-8542 Impact factor: 2.557
Fig. 1Configuration circuits of the developed device
Fig. 2Experiment design
Fig. 3Three types of masks used for the experiment
Fig. 4Overall experimental flowchart procedures
Fig. 5Examples of preprocessed signals from breathing at 12 breaths/min during taking a bus
Fig. 6Example of PSD estimation from a breathing signal envelope at 0.2 Hz (12 breaths/min) that contained both breathing through a nose and mouth signals
The median and IQR of accuracy using three types of masks in four different environments (Nose breathing condition)
| Environment | Beep rate (Hz) | Accuracy error (%) | Types of masks | |||
|---|---|---|---|---|---|---|
| Surgical | KF94 | Reusable | ||||
| Indoor | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| Street | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| Bus | 0.2 | Median | 0 | 0 | 0 | 0.59 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.36 | |
| IQR | 0 | 0 | 0 | |||
| Train | 0.2 | Median | 0 | 0 | 0 | 0.23 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.1 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.32 | |
| IQR | 0 | 0 | 0 | |||
The median and IQR of repeatability using three types of masks in four different environments (Nose breathing condition)
| Environment | Beep rate (Hz) | Repeatability error (%) | Types of masks | |||
|---|---|---|---|---|---|---|
| Surgical | KF94 | Reusable | ||||
| Indoor | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.59 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| Street | 0.2 | Median | 0 | 0 | 0 | 0.36 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| Bus | 0.2 | Median | 0 | 0 | 0 | 0.81 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.36 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.18 | |
| IQR | 0 | 0 | 18.75 | |||
| Train | 0.2 | Median | 0 | 0 | 0 | 0.83 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.31 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.71 | |
| IQR | 0 | 0 | 0 | |||
The median and IQR of accuracy using three types of masks in four different environments (Mouth breathing condition)
| Environment | Beep rate (Hz) | Accuracy error (%) | Types of masks | |||
|---|---|---|---|---|---|---|
| Surgical | KF94 | Reusable | ||||
| Indoor | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| Street | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| Bus | 0.2 | Median | 0 | 0 | 0 | 0.36 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.59 | |
| IQR | 0 | 0 | 0 | |||
| Train | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.36 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.23 | |
| IQR | 0 | 0 | 0 | |||
The median and IQR of repeatability using three types of masks in four different environments (Mouth breathing condition)
| Environment | Beep rate (Hz) | Accuracy error (%) | Types of masks | |||
|---|---|---|---|---|---|---|
| Surgical | KF94 | Reusable | ||||
| Indoor | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.01 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.36 | |
| IQR | 0 | 0 | 0 | |||
| Street | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| Bus | 0.2 | Median | 0 | 0 | 0 | 0.36 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.59 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.33 | |
| IQR | 0 | 0 | 0 | |||
| Train | 0.2 | Median | 0 | 0 | 0 | 0.36 |
| IQR | 18.75 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.59 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.59 | |
| IQR | 0 | 0 | 0 | |||
The median and IQR of accuracy using three types of masks in four different environments (Nose/Mouth breathing condition)
| Environment | Beep rate (Hz) | Accuracy error (%) | Types of masks | |||
|---|---|---|---|---|---|---|
| Surgical | KF94 | Reusable | ||||
| Indoor | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.36 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.59 | |
| IQR | 0 | 0 | 0 | |||
| Street | 0.2 | Median | 0 | 0 | 0 | 0.36 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| Bus | 0.2 | Median | 0 | 0 | 0 | 0.75 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.75 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.96 | |
| IQR | 0 | 0 | 9.37 | |||
| Train | 0.2 | Median | 0 | 0 | 0 | 0.8 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.77 | |
| IQR | 0 | 0 | 0 | |||
The median and IQR of repeatability using three types of masks in four different environments (Nose/Mouth breathing condition)
| Environment | Beep rate (Hz) | Repeatability error (%) | Types of masks | |||
|---|---|---|---|---|---|---|
| Surgical | KF94 | Reusable | ||||
| Indoor | 0.2 | Median | 0 | 0 | 0 | 1.0 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 1.0 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.36 | |
| IQR | 0 | 0 | 0 | |||
| Street | 0.2 | Median | 0 | 0 | 0 | 0.36 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.36 | |
| IQR | 0 | 0 | 0 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.36 | |
| IQR | 0 | 0 | 0 | |||
| Bus | 0.2 | Median | 0 | 0 | 0 | 0.26 |
| IQR | 25 | 0 | 43.75 | |||
| 0.3 | Median | 0 | 0 | 0 | 0.26 | |
| IQR | 16.66 | 0 | 12.49 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.4 | |
| IQR | 9.37 | 0 | 12.5 | |||
| Train | 0.2 | Median | 0 | 0 | 0 | 0.26 |
| IQR | 0 | 0 | 0 | |||
| 0.3 | Median | 0 | 0 | 0 | 0..26 | |
| IQR | 0 | 0 | 12.49 | |||
| 0.4 | Median | 0 | 0 | 0 | 0.40 | |
| IQR | 0 | 0 | 0 | |||