Finger and forehead PPG signal comparison for respiratory rate estimation.

A Hernando1, M D Peláez-Coca, M T Lozano, J Lázaro, E Gil.   

Abstract

OBJECTIVE: An evaluation of the location of the photoplethysmogram (PPG) sensor for respiratory rate estimation is performed. APPROACH: Finger PPG, forehead PPG, and respiratory signal were simultaneously recorded from 35 subjects while breathing spontaneously, and during controlled respiration experiments at a constant rate from 0.1 Hz to 0.6 Hz, in 0.1 Hz steps. Four PPG-derived respiratory (PDR) signals were extracted from each one of the recorded PPG signals: pulse rate variability (PRV), pulse width variability, pulse amplitude variability and the respiratory-induced intensity variability (RIIV). Respiratory rate was estimated from each one of the four PDR signals for both PPG sensor locations. In addition, different combinations of PDR signals, power distribution of the respiratory frequency range and differences of the morphological parameters extracted from both PPG signals have been analysed. MAIN
RESULTS: Results show better performance in terms of successful estimation and relative error when: (i) PPG signal is recorded in the finger; (ii) the respiratory rate is less than 0.4 Hz; (iii) RIIV signal is not considered. Furthermore, lower spectral power around the respiratory rate in the PDR signals recorded from the forehead was observed. SIGNIFICANCE: These results suggest that respiratory rate estimation is better at lower rates (0.4 Hz and below) and that the finger is better than the forehead to estimate respiratory rate.

Entities:  

Mesh:

Year:  2019        PMID: 31422948     DOI: 10.1088/1361-6579/ab3be0

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  3 in total

1.  On the spatial phase distribution of cutaneous low-frequency perfusion oscillations.

Authors:  Stefan Borik; Simon Lyra; Volker Perlitz; Micha Keller; Steffen Leonhardt; Vladimir Blazek
Journal:  Sci Rep       Date:  2022-04-09       Impact factor: 4.379

2.  Estimation of respiratory rate in various environments using microphones embedded in face masks.

Authors:  Chhayly Lim; Jungyeon Kim; Jeongseok Kim; Byeong-Gwon Kang; Yunyoung Nam
Journal:  J Supercomput       Date:  2022-06-18       Impact factor: 2.557

3.  Investigation of Photoplethysmography Behind the Ear for Pulse Oximetry in Hypoxic Conditions with a Novel Device (SPYDR).

Authors:  Brian Bradke; Bradford Everman
Journal:  Biosensors (Basel)       Date:  2020-04-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.