Comparative analysis of loop-mediated isothermal amplification combined with microfluidic chip technology and q-PCR in the detection of clinical infectious pathogens.

Enqi Zhang1, Hongling Ou2, Lianling Jia2, Wang Zhang2, Yemei Wang1, Xinru Wang2.   

Abstract

BACKGROUND: Rapid diagnosis of infectious pathogens at an early stage is crucial to stabilize the patient's condition, reduce medical costs, and shorten hospital stays. Currently, some point-of-care tests have their own shortcomings. Therefore, we built a microfluidic chip based on loop-mediated isothermal amplification to can quickly and sensitively detect infectious pathogens.
METHODS: We extracted the DNA of S. aureus, MRSA, Shigella and Klebsiella pneumoniae. Then, the DNA samples were diluted by 10-fold and examined by two methods: LAMP-microfluidic chip and q-PCR, the sensitivity of whom was also compared. In addition, the specificity of the two was also examined by detecting the target bacteria and other microorganisms using the same methods. Finally, we extracted and tested the DNA of clinically infected humoral samples to determine the coincidence rate between the two methods and the bacterial culture method.
RESULTS: For S. aureus, MRSA, Shigella, and Klebsiella pneumoniae, the detection limits of the chip were 2.25 × 103 copies/μl, 5.32 × 103 copies/μl, 2.89 × 103 copies/μl, 6.53 × 102 copies/μl, and the detection limits of q-PCR were 2.25 × 102 copies/μl, 5.32 × 101 copies/μl, 2.89 × 102 copies/μl, 6.53 × 101 copies/μl, respectively. In terms of detection specificity, neither method cross-reacted with other strains. For the detection of infectious humoral samples, the total coincidence rate between the q-PCR and bacterial culture method was 85.7%, 95%, 95%, and 95.5%, and the total coincidence rate between the chip and bacterial culture method was 81%, 95%, 90%, and 86.4%, respectively.
CONCLUSION: LAMP-microfluidic chip provides a simple, sensitive, specific, convenient, and rapid pathogen detection method for clinically infected humoral samples without relying on expensive equipment or technical personnels.
© 2022 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC.

Entities:  

Keywords:  bacteria; chip; detection; loop-mediated isothermal amplification; microfluidics

Mesh:

Substances:

Year:  2022        PMID: 35754145      PMCID: PMC9396168          DOI: 10.1002/jcla.24565

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   3.124


INTRODUCTION

With the rampant application of antibiotics and the increasing cases of invasive operation, critically ill patients are becoming subjects with high risk of various pathogenic infections, who would suffer further worsening of the condition, prolonged hospital stay, incurred with increased treatment costs, even go through life‐threatening effects since they were normally complicated with low immunity and multiple organ failure. , A study based on 49,331 patients with sepsis or septic shock in 149 hospitals in New York State, United States, showed that the use of antibiotics within 3 h increased patient mortality by 1.04% for each hour delayed, and the hospitalization mortality rate for patients who took antibiotics at 3–12 h was 14% higher than patients who took antibiotics within 3 h. Therefore, timely recognition of infectious diseases, rapidly and accurately identify underlying pathogens, and then targeting effective antibiotics are key to successful treatment, not only for better prognosis but also to prevent antibiotic resistance. However, it is often difficult to accurately diagnose infectious diseases in clinical practice due to the wide variety of pathogens that cause infection, and patients may carry more than one pathogen and tend to coexist when the infection occurs. In addition, clinical and laboratory signals of severe infection, such as tachycardia, fever, and changes in white blood cell count, are nonspecific and often appear in other acute diseases. Therefore, the reasons mentioned above have rendered an accurate and rapid diagnosis of infectious diseases a formidable task. Currently, the gold standard for the diagnosis and identification of pathogenic bacteria is still bacterial culture. However, bacterial culture requires a long cycle, high requirements for inspectors, a series of bacterial identification and drug susceptibility equipment, so empiric antibiotic therapy is often required by clinicians and making bacterial resistance became an important issue affecting human society. With the rapid development of molecular biology techniques in the past decades and the emergence of the novel coronavirus 2019 (COVID‐19) pandemic, point‐of‐care testing (POCT) has gradually become the focus of attention, especially POCT based on nucleic acid amplification. In this study, we established a POCT method based on nucleic acid amplification assays that combines LAMP and microfluidic chip technology to exhibit good detection of bacterial DNA in humoral specimens. Loop‐mediated isothermal amplification (LAMP) is a novel nucleic acid amplification technique developed and established by Notomi et al. (2000) as an alternative method to PCR. LAMP is capable of amplifying a limited number of copies of DNA up to a million times in an hour with high specificity, efficiency, and fast under isothermal conditions. LAMP analysis involves four to six primers, including two external primers (F3 and B3), two internal primers (FIP and BIP), and loop primers (LF and/or LB). The reaction uses DNA polymerase with strand displacement function to continuously replicate and amplify DNA at a constant temperature (60–65°C), and fluorescence detection or naked‐eye observation is carried out by fluorescent or colored dye incorporation method. In diagnostics based on nucleic acid amplification, replacing PCR with an LAMP reaction can eliminate the need for large volumes and expensive thermal cycling equipment, and improve the sensitivity and specificity of target gene amplification. At the same time, with the continuous development of nanomaterials, a variety of extremely sensitive biosensors have been produced for detecting markers of pathogenic microorganisms (DNA, RNA, glycoproteins, antibodies, enzymes, etc.). Biosensors for these detection devices are typically based on nanotechnological or microtechnological platforms and rely on the generation of detection signals such as optical, electrical, magnetic or visual, and when biosensors encounter a pathogen or pathogenic byproduct, they can detect a positive signal. Integrating microfluidic systems with nanomaterials and microarray technologies is highly effective for achieving the goal of miniaturization, automation, and portable chips for the analysis of complex microfluidic samples such as cells, nucleic acids and proteins. At present, there are studies that combined LAMP with microfluidic technology for the detection of nucleic acid sequences that showed a good application prospect. In this study, we combined LAMP with microfluidic chip technology, added fluorescent dyes to the constant temperature amplification reaction system, and used the fluorescent dye incorporation method for real‐time fluorescence detection, and the positive sample amplification detection will produce an “S” shaped real‐time fluorescence amplification curve, which completes the amplification and detection of the target gene in one step, and can simultaneously amplify a variety of nucleic acid target genes. We designed TaqMan qPCR primers and fluorescent probes to detection for single pathogens, established a detection system, and compared the differences in detection limits and cross reactions between LAMP‐microfluidic chips and TaqMan qPCR methods. Furthermore, we detected humoral samples of clinical patients' infection sites as evidence, and summarized the advantages, disadvantages, and applicability of microfluidic chips as a new detection method, so as to help the clinical practice to detect infected pathogens in a timely and accurate manner, make up for the shortcomings of traditional detection methods, and able to timely and accurately guide clinical medication.

MATERIALS AND METHODS

Instruments and reagents

This study acquired the following instruments and reagents: a VITEK MS microbial identification mass spectrometer (bioMérieux SA, Marcy L'Etoile, France), a Nano‐400A micro spectrophotometer (Aosheng Instrument, Hangzhou, China), a BacT/ALERT 3D automatic culture system and VITEK‐2 compact ID/AST (bioMérieux SA, Marcy L'Etoile, France), DNA lamp amplification reaction kits and a microfluidic chip detector (Biocare, Tianjin, China), a bacterial genome extraction kit and lysozyme (Tiangen Biotechnology, Beijing, China), a CFX96 Touch™ real‐time PCR detection system (BIO‐RAD, America), and a 2xGoldstar Probe Mixture (CWBIO, Beijing, China).

Experimental strains

All bacterial strains used in the study are collected in Clinical Laboratory of PLA Rocket Characteristic Medical Center with great viability after resuscitation and confirmed by bacterial identification, drug susceptibility system and microbial identification mass spectrometer. Standard strains: Shigella sonnei ATCC25931, Acinetobacter baumannii ATCC BAA‐747, Enterobacter cloacae ATCC700323, Enterococcus faecalis ATCC29212, Pseudomonas aeruginosa ATCC27853, Stenotrophomonas maltophilia ATCC17666, Staphylococcus aureus ATCC29213, Escherichia coli ATCC25922, Klebsiella pneumoniae ATCC700603, Proteus mirabilis ATCC35659. External quality assessment strains: Enterococcus faecium, Carbapenem‐resistant Klebsiella pneumoniae, Carbapenem‐resistant Enterobacterales, Carbapenem‐resistant Acinetobacter baumannii, Salmonella enteritidis, Citrobacter freundii, Serratia marcescens, Enterobacter aerogenes, Klebsiella oxytoca, Staphylococcus haemolyticus, Staphylococcus saprophyticus, S. Paratyphi A, Vibrio parahaemolyticus, Candida tropicalis, Candida albicans, C. parapsilosis. Clinical bacteria strains: Listeria monocytogenes, Shigella flexneri, Chryseobacterium indologenes, Morganella morganii, Methicillin‐resistant Staphylococcus aureus (MRSA), Carbapenem‐resistant Pseudomonas aeruginosa (CRPA), Methicillin‐resistant coagulase negative Staphylococcus (MRCNS).

Clinical samples

We collected blood, urine, alveolar lavage fluid, sputum and drainage fluid and other possible samples from the patients and identified positive samples or negative samples of S. aureus, Klebsiella pneumoniae, MRSA and Shigella by microbiological culture.

DNA extraction of bacterial genomic

DNA extraction from clinical samples and bacterial strain samples was performed following the instructions of the TIANamp Bacteria DNA Kit step by step, respectively. Pretreatment of clinical samples was performed according to the kit instructions, suspend the pretreated sample 200 μl media and centrifuge at 12,000 rpm for 2 min, and discard the supernatant. Then, resuspend the deposit in a 200 μl extracting solution containing 1 mM EDTA, 10 mM Tris–HCl, and a nucleic acid protectant, and transfer to a cracking tube filled with glass beads, heat at 100°C for 5 min after vortex for 5 min. Then, centrifuge with a high‐speed centrifuge at 12,000 rpm for 2 min, transfer the supernatant to a clean centrifuge tube, store at −20°C for later use. For the positive culture samples from lower respiratory tract secretions, 10 g/L trypsin was added to the sample and mixed well with vibration, liquefied at 37°C for 30 min. When there was no solid and stringy state, it was liquefied completely. Then, the DNA extraction was carried out and the concentration and purity of the extracted DNA are detected, and stored at −20°C for later use.

The design, screening, and synthesis of primers and probes

S. aureus, MRSA, Klebsiella pneumoniae, and Shigella primers were all synthesized using the best primers screened from prior studies. LAMP analysis of each strain involves four to six primers, including two external primers (F3 and B3), two internal primers (FIP and BIP), and loop primers (LF and/or LB), where loop primers are used to accelerate the LAMP reaction and shorten the reaction time. At the same time, in order to observe the difference between the detection capability of LAMP‐microfluidic chip and real‐time quantitative PCR (qPCR) methods, the primers of qPCR for the detection of a single pathogen were designed, and a detection system was established to compare the difference between the sensitivity and specificity of the microfluidic chip and qPCR. The primers of qPCR are synthesized by Sangon Biotech and shown in Table 1. The primer sequences of LAMP are not disclosed herein.
TABLE 1

Primer sequence of TaqMan qPCR

SpeciesTarget genePrimersSequence (5′–3′)
Klebsiella pneumoniae fimD ForwardGCAGTCGGTGCGTTTTCTCT
ReverseCGGTAAGTGGTATCGGCAAAG
ProbeATGAGACCGGCACCAACCTGCAG
Staphylococcus aureus femA ForwardAACGGTCAATGCCATGATTTAAT
ReverseTATCTCTGCTGGTTTCTTCTTTATCAAT
ProbeTGCACTGCATAACTTCCGGCAAAATGAC
MRSA mecA ForwardGTGGAAGTTAGATTGGGATCATAGC
ReverseTTGGCCAATTCCACATTGTTT
ProbeTTCCAGGAATGCAGAAAGACCAAAGCATAC
Shigella ipaH ForwardCGTTCCTTGACCGCCTTTC
ReverseGCCATGCAGCGACCTGTT
ProbeCTCTGCACGCAATACCTCCGGATTCC
Primer sequence of TaqMan qPCR

Real‐time quantitative PCR and DNA‐LAMP procedure

The LAMP detection reaction system was composed of 58 μl of isothermal amplification reagent and 14.5 μl of nucleic acids of the sample to be tested. Reaction parameters: 37°C, 3 min; 65°C, 50 min. The microfluidic chip used in this study was designed by SolidWorks 2014 software, and injection molding of medical grade polycarbonate (PC) was implemented. The qPCR reaction consists of 10 μl of 2xGoldstar Probe Mixture (GoldStar Taq DNA Polymerase, PCR Buffer, dNTPs and Mg2+), 2 μl of DNA template, 0.4 μl (1 μM) of upstream primers,0.4 μl (1 μM) of downstream primers, 0.2 μl of fluorescent probes, and 7 μl of sterile deionized water. The cycle was repeated 40 times at 95°C to pre‐denaturation for 10 min, 95°C for 15 s and 60°C for 30 s. The amplification was performed in the BIO‐RAD CFX96TM real‐time system. To detect the presence of contamination, the same volume of PCR‐grade deionized water was used as a negative control, where other conditions are the same.

Data analysis

A positive sample amplification test in an isothermal amplification reaction system produces an “S”‐shaped fluorescence amplification curve. At the same time, the positive control test result is positive, and the negative control test result is negative, indicating that the test result is valid. If the quality control result is wrong, it is determined that the test result is invalid and needs to be re‐examined or find the cause. Positive sample amplification detection of qPCR reaction system produces an “S”‐ shaped fluorescence amplification curve. Ct < 30 was judged positive, Ct > 35 was judged negative, and Ct values between 30 and 35 were judged to be positive after repeated verification. Clinical specimen data were counted according to descriptive research methods, and experimental data were converted into a four‐grid format in Table 2 and calculated according to the formula. Positive coincidence rate = a/(a + c) × 100%; negative coincidence rate = d/(b + d) × 100%; total coincidence rate = (a + d)/(a + b + c + d) × 100%.
TABLE 2

Statistical table for data processing

Reference methodTotal
PositiveNegative
Experimental methodsPositiveaba + b
Negativecdc + d
Totala + cb + da + b + c + d
Statistical table for data processing

RESULTS

Sensitivity

The DNA purity and concentration were measured with a Nano‐400A microspectrophotometer after bacterial culture and extraction. Then, we sequentially dilute the original DNA sample to the designated concentration (labeled as 10–1, 10–2, 10–3, 10–4, 10–5, 10–6, 10–7, 10–8, 10–9,10–10, 10–11, 10–12;the DNA concentration in tube “10–1” to “10–12” is decreased by 10 times in sequence) by 10 times ratio dilution method with sterile deionized water, and after tested using two methods, different bacterial DNA showed different minimum detection limits. The results are shown in Table 3 and Figures 1, 2.
TABLE 3

The lowest detection limit of qPCR and LAMP microfluidic chip

Bacteriaq‐PCR(copies/μl)LAMP (copies/μl)
S. aureus 2.25 × 102 2.25 × 103
MRSA 5.32 × 101 5.32 × 103
Shigella 2.89 × 102 2.89 × 103
Klebsiella pneumonia 6.53 × 101 6.53 × 102

Abbreviation: MRSA, Methicillin‐resistant Staphylococcus aureus.

FIGURE 1

The dilution multiple and sensitivity result of four bacteria detected by q‐PCR

FIGURE 2

The dilution multiple and positive results of four bacteria were detected by microfluidic chip method. The blue curve is the AC quality control line, and the positive signal indicates that the test is effective; TP (min) represents the jump time of the reaction curve signal

The lowest detection limit of qPCR and LAMP microfluidic chip Abbreviation: MRSA, Methicillin‐resistant Staphylococcus aureus. The dilution multiple and sensitivity result of four bacteria detected by q‐PCR The dilution multiple and positive results of four bacteria were detected by microfluidic chip method. The blue curve is the AC quality control line, and the positive signal indicates that the test is effective; TP (min) represents the jump time of the reaction curve signal

Specificity

We extracted DNA from 36 species of bacteria after culture and measured their purity and concentration with a Nano‐400A microspectrophotometer. Then, we used the original concentration of DNA as a sample and used primers of S. aureus, MRSA, Shigella, and Klebsiella pneumoniae to, respectively, detect 36 bacteria, and the specificity of the two methods was shown in Table 4 and Figures 3, 4, 5, 6.
TABLE 4

Comparison of the specificity of qPCR and LAMP microfluidic chip

Primer category S. aureus MRSA Shigella Klebsiella pneumoniae
No.Bacteria speciesq‐PCRLAMPq‐PCRLAMPq‐PCRLAMPq‐PCRLAMP
1 Enterococcus faecalis
2 Enterococcus faecium
3 MRSA
4 S. aureus
5 Staphylococcus epidermidis
6 Pseudomonas aeruginosa
7 Acinetobacter baumannii
8 CRKP
9 CRE
10 Morgan
11 Proteus mirabilis
12 Citrobacter freundii
13 Serratia marcescens
14 Stenotrophomonas maltophilia
15 Escherichia coli
16 Klebsiella pneumoniae
17 Salmonella enteritidis
18 Enterobacter cloacae
19 Chryseobacterium indologenes
20 CRPA
21 CRAB
22 Shigella sonnei
23 Yersinia enterolytica
24 Enterobacter aerogenes
25 Klebsiella oxytoca
26 Staphylococcus haemolyticus
27 Staphylococcus saprophyticus
28 S. Paratyphi A
29 Vibrio parahaemolyticus
30 Listeria monocytogenes
31 Shigella flexneri
32 Candida tropicalis
33 C. glabrata
34 Candida albicans
35 C. parapsilosis
36 Streptococcus pneumonia
37Negative control

Note: +, indicates positive; −, indicates negative; positive results have been marked in red.

Abbreviations: CRKP, carbapenem‐resistant klebsiella pneumoniae; CRE, Carbapenem‐resistant Enterobacterales; CRPA, Carbapenem‐resistant Pseudomonas aeruginosa; CRAB, Carbapenem‐resistant Acinetobacter baumannii; MRSA, Methicillin‐resistant Staphylococcus aureus, the gene for methicillin resistance (mecA).

FIGURE 3

Specificity of detecting S. aureus by qPCR

FIGURE 4

Specificity of detecting MRSA by qPCR

FIGURE 5

Specificity of detecting Shigella by qPCR

FIGURE 6

Specificity of detecting Klebsiella pneumoniae by qPCR

Comparison of the specificity of qPCR and LAMP microfluidic chip Note: +, indicates positive; −, indicates negative; positive results have been marked in red. Abbreviations: CRKP, carbapenem‐resistant klebsiella pneumoniae; CRE, Carbapenem‐resistant Enterobacterales; CRPA, Carbapenem‐resistant Pseudomonas aeruginosa; CRAB, Carbapenem‐resistant Acinetobacter baumannii; MRSA, Methicillin‐resistant Staphylococcus aureus, the gene for methicillin resistance (mecA). Specificity of detecting S. aureus by qPCR Specificity of detecting MRSA by qPCR Specificity of detecting Shigella by qPCR Specificity of detecting Klebsiella pneumoniae by qPCR

Evaluation of the LAMP and q‐PCR assay using clinical specimens

We collected more than 80 humoral samples of patients who had been confirmed by bacterial culture for each of the four kinds of clinically common bacteria to be tested and used bacterial culture as a reference method to assess the diagnostic value of the two detection methods. The results are shown in Tables 5 and 6.
TABLE 5

Coincidence rate of qPCR in humoral samples

Bacteria speciesPositive coincidence rateNegative coincidence rateTotal coincidence rate
S. aureus 81.8%90%85.7%
MRSA 100%90%95%
Shigella 100%90%95%
Klebsiella pneumoniae 100%90%95.5%
TABLE 6

Coincidence rate of LAMP microfluidic chip in humoral samples

Bacteria speciesPositive coincidence rateNegative coincidence rateTotal coincidence rate
S. aureus 72.7%90%81%
MRSA 100%90%95%
Shigella 100%80%90%
Klebsiella pneumoniae 91.2%80%86.4%
Coincidence rate of qPCR in humoral samples Coincidence rate of LAMP microfluidic chip in humoral samples

DISCUSSION

The rapid detection system established by the combination of LAMP technology and microfluidic technology can extract the nucleic acid of the original sample in a short period of time and detect the nucleic acid simply, quickly, specifically and efficiently. This system amplified nucleic acid using DNA polymerase at constant temperature (60–65°C), saving us from complex laboratory equipment. It accomplished the goal of amplifying a limited number of copies of DNA up to 1 million times in 1 h, and allowed simultaneous amplification of nucleic acids in parallel. We compared the sensitivity and specificity of the two methods using q‐PCR as reference and summarized the practicality of the system in clinical practice. The objective was to provide an easy, fast, and accurate way to diagnose bacterial species by detecting humoral specimens and guiding the employment and management of antibiotics in early clinical stages, and to reduce mortality and alleviate the impact incurred on patients. In clinical practice, the top Gram‐positive bacterium in prevalence is S. aureus, accounting for 11% (4833/42,553). In S. aureus, 73% (3302/4515) of the strains are MRSA. The most commonly encountering Gram‐negative bacteria are Klebsiella pneumoniae, accounting for 9.1% (3858/42,553), and Shigella aside from E. coli; therefore, we chose these five bacteria as the subjects of this study. In terms of detection sensitivity, existing studies have shown that both qPCR and LAMP‐microfluidic chips exhibit high detection limits. In this experiment, after detecting the concentration and purity of the DNA sample, we serially diluted 12 times by 10‐fold ratio dilution and labeled “10–1” to “10–12” and detected by two methods, respectively, with PCR grade sterile deionized water as a negative control. The results showed that the two methods show different detection limits for different bacteria, but the lowest detection limits of this system are lower than the q‐PCR method. The detection limits of q‐PCR method were S. aureus 2.25 × 102 copies/μl, Klebsiella pneumocystis 6.53 × 101 copies/μl, MRSA 5.32 × 101 copies/μl, Shigella 2.89 × 102 copies/μl. The detection limits of the microfluidic chip method were S. aureus 2.25 × 103 copies/μl, Klebsiella pneumoniae 6.53 × 102 copies/μl, MRSA 5.32 × 103 copies/μl, Shigella 2.89 × 103 copies/μl. Overall, the sensitivity of q‐PCR is generally 10–102 times higher than LAMP‐microfluidic chip methods. After comparing several detection methods on Alternaria solani, Khan M et al. concluded that qPCR (1.36 × 10−3 ng/μl) has the highest sensitivity and LAMP (1.36 ng/μl) has a sensitivity 103 times lower than qPCR at 63°C. This is similar to our findings. However, there are also some studies that have different conclusions from our study. Choi CW et al. showed that the sensitivity of LAMP was the same as real‐time PCR after comparing the three detection methods simultaneously. In addition, it is worth mentioning that many studies have shown that the sensitivity of LAMP is generally higher than conventional PCR. The amplification results of Yersinia enterocolitica by Ranjbar R et al. showed that minimum copy number of DNA detectable by LAMP and PCR assays was 44 and 440, respectively. Moosavian M et al. detected Lactobacillus pneumophila via its mip gene in respiratory specimens indicating that the detection limit of electrophoresis by PCR was <11.5 pg/μl and the results of naked‐eye inspection by LAMP reaction were <1.15 pg/μl. These results suggested that detection sensitivity of LAMP was tenfold of PCR. Although studies have different views on the sensitivity of LAMP, generally, the detection limits of the two are not much different, and the minimum detection limits are very sensitive in both approaches, which back the detection of humoral specimens with low bacterial concentrations in early clinical infection. In terms of detection specificity, the two methods showed consistent specificity, both showed solely positive signals in the target bacteria, and neither cross reacts with other bacteria. However, most reports currently suggest that LAMP testing provides higher specificity than qPCR, possibly due to the limited scope of strains covered in this study , but it is certain that this is an accurate method with a wide range of detections. A total of 83 humoral samples were collected at the Rocket Army Specialized Medical Center, including blood, urine, sputum, and drainage fluid, etc. In all, 11 humoral specimens of Staphylococcus aureus, 10 humoral specimens of MRSA, 12 humoral specimens of Pneumoniae klebsiella, and a total of 30 samples were negative for the target bacteria. In addition, we also collected 10 cases of simulated stool samples contaminated with Shigella and 10 cases of simulated stool samples contaminated with Salmonella. Both negative and positive samples had been subjected to bacterial identification and drug susceptibility testing in accordance with the operating procedures. In order to evaluate the performance of the detection system, negative and positive samples were extracted with DNA extraction kit instructions. The results are shown that for S. aureus, MRSA, Shigella and klebsiella Pneumoniae, the total coincidence rate between q‐PCR and bacterial culture is 85.7%, 95%, 95%, 95.5% and the total coincidence rate between LAMP‐microfluidic chip and bacterial culture is 81%, 95%, 90%, 86.4%, respectively. Overall, both methods have similar testing performance for humoral samples, whether in terms of positive, negative, or total coincidence rate. For samples that showed positive in bacterial culture and negative in the other two methods, we suspected that DNA loss might be due to improper manipulation during sample handling and bacterial DNA extraction. Due to the limited sample size of this study, further studies should be conducted with more sample types and sample size. Through the comparison of the two methods, it is not difficult to conclude that LAMP microfluidic chip has similar capabilities to qPCR in the clinical detection of infection specimens and possesses the characteristics of high sensitivity and strong detection ability, which provides a reliable basis for rapid clinical diagnosis and precise treatment. However, for ideal diagnostic tests, it should not be limited to the high sensitivity and specificity of the method. According to the requirements of the World Health Organization (WHO) for ideal diagnostic tests in developing countries, the standards for these diagnostic methods should be “Affordable, Sensitive, Specific, User‐friendly (simple to perform in a few steps with minimal training), Robust and Rapid (results available in 30 min), Equipment free, and Deliverable to the end user”. So far, however, only a few rapid diagnostic tests (RDT) have met the criteria proposed by the WHO. Currently, LAMP microfluidic chips exhibit significant advantages over many diagnostic methods based on gene amplification tests and are more likely to meet the WHO criteria for RDT. First of all, the cost of this system is low. One of the main challenges facing most diagnostic methods currently used in developing countries is the enormous cost. The main instruments required for LAMP are heating blocks or water baths, which are readily available in most laboratories, therefore, no additional charges are incurred. A study that estimated the cost of reagents for current diagnostic techniques showed that the cost of testing a sample was the highest at $6.4–7.7 based on PCR technology, compared to just $0.71–2 for LAMP. Second, the detection speed of this method is faster. Conventional PCR reactions take about 1 h 40 min, mainly due to their long gel electrophoresis. q‐PCR can usually be done within 1–2 h of the sample arriving in the laboratory. LAMP microfluidic chip can be completed within 15–60 min, which greatly saves the time spent on testing and can rapidly provide guidance for clinical diagnosis and treatment. , , In addition, the system is also robust. LAMP microfluidic chip has the advantage of amplifying the target DNA from partially processed or unprocessed samples. The LAMP microfluidic chip for detecting microbial nucleic acid in this study mainly includes three steps: sample preparation, amplification and detection. Sample preparation is the most important step in nucleic acid detection, especially for POC diagnosis, as it involves a complex processing procedure that often requires manual preparation, while automated processing requires large instruments in clinical laboratories. However, the method in our study from clinical specimen to sample preparation only needs to be performed in an inexpensive, small, automated microfluidic device that combines nucleic acid amplification and detection. Not only eliminates the need for DNA extraction, avoids possible contamination during the extraction process and the loss of a large amount DNA in the sample but also saves a lot of the time of preprocessing and shortens the entire detection cycle. At the same time, the LAMP microfluidic chip is divided into multiple reaction units, which can parallel detect a variety of pathogens and ensure the independence of each reaction and the accuracy of the results. Finally, the method also has the characteristics of user‐friendliness and simpleness. The system is small and portable, does not require professionals to operate, and is more suitable for under‐resourced areas where often no resources for special training for medical staff. In short, the detection system combined with LAMP technology and microfluidic chip technology provides rapid detection of a variety of pathogens, and the system possesses several merits such as high sensitivity, strong specificity. Furthermore, the coincidence rate of clinical samples is also very high, while having the characteristics of low cost, rapidness, user‐friendliness, etc., which may have considerable practical significance for the prevention and control of major diseases caused by pathogen infection.

CONCLUSION

In conclusion, overall, the LAMP microfluidic chip system is a powerful clinical detection arsenal to tackle the cost and time challenge in clinical infection diagnosis and is as reliable as other existing methods based on nucleic acid amplification. Although several studies enumerated several shortcomings of LAMP microfluidic chip system and suggested further development and optimization, we still believe that this approach can be adapted to future clinical tests and provide an accurate, sensitive, faster, cheaper, and less complex detection tool for clinical bacterial infections.

AUTHOR CONTRIBUTIONS

Enqi Zhang, Hongling Ou, and Xinru Wang involved in conception and design; Lianling Jia and Wang Zhang involved in the provision of study materials or samples; Enqi Zhang, Hongling Ou, and Yemei Wang involved in the experimental technical operation; Lianling Jia, Wang Zhang, and Yemei Wang collected and assembled the data; Enqi Zhang, Hongling Ou, and Xinru Wang analyzed and interpreted the data; All authors wrote the manuscript and provided the final approval of manuscript.

CONFLICT OF INTEREST

All authors have completed the ICMJE Form for Potential Conflicts of Interest and have no conflicts of interest to declare.
  29 in total

1.  Rapid and Sensitive Detection of Salmonella in Chickens Using Loop-Mediated Isothermal Amplification Combined with a Lateral Flow Dipstick.

Authors:  Zhi Liu; Qiu Zhang; Ning-Ning Yang; Ming-Guo Xu; Jin-Feng Xu; Ming-Long Jing; Wen-Xing Wu; Ya-Dong Lu; Feng Shi; Chuang-Fu Chen
Journal:  J Microbiol Biotechnol       Date:  2019-03-28       Impact factor: 2.351

Review 2.  Integration of isothermal amplification methods in microfluidic devices: Recent advances.

Authors:  Maria Chiara Giuffrida; Giuseppe Spoto
Journal:  Biosens Bioelectron       Date:  2016-11-19       Impact factor: 10.618

Review 3.  Point-of-care nucleic acid testing for infectious diseases.

Authors:  Angelika Niemz; Tanya M Ferguson; David S Boyle
Journal:  Trends Biotechnol       Date:  2011-03-04       Impact factor: 19.536

4.  Rapid detection of Salmonella based on loop-mediated isothermal amplification.

Authors:  Hongling Ou; Yan Wang; Jiayi Gao; Jing Bai; Qiaoyun Zhang; Lei Shi; Xinru Wang; Chengbin Wang
Journal:  Ann Palliat Med       Date:  2021-06

5.  Antimicrobial susceptibility testing results from 13 hospitals in Viet Nam: VINARES 2016-2017.

Authors:  Tien Viet Dung Vu; Marc Choisy; Thi Thuy Nga Do; Van Minh Hoang Nguyen; James I Campbell; Thi Hoi Le; Vu Trung Nguyen; Heiman F L Wertheim; Ngoc Thach Pham; Van Kinh Nguyen; H Rogier van Doorn
Journal:  Antimicrob Resist Infect Control       Date:  2021-05-10       Impact factor: 4.887

6.  Detection of Streptococcus equi subsp. equi in guttural pouch lavage samples using a loop-mediated isothermal nucleic acid amplification microfluidic device.

Authors:  Ashley G Boyle; Shelley C Rankin; Kathleen O'Shea; Darko Stefanovski; Jing Peng; Jinzhao Song; Haim H Bau
Journal:  J Vet Intern Med       Date:  2021-03-17       Impact factor: 3.333

7.  Mortality related to hospital-associated infections in a tertiary hospital; repeated cross-sectional studies between 2004-2011.

Authors:  Anne Mette Koch; Roy Miodini Nilsen; Hanne Merete Eriksen; Rebecca Jane Cox; Stig Harthug
Journal:  Antimicrob Resist Infect Control       Date:  2015-12-29       Impact factor: 4.887

Review 8.  Tools for Detection of Schistosomiasis in Resource Limited Settings.

Authors:  Olumide Ajibola; Bashar Haruna Gulumbe; Anthonius Anayochukwu Eze; Emmanuel Obishakin
Journal:  Med Sci (Basel)       Date:  2018-05-23

9.  Point-of-Care Diagnostic Services as an Integral Part of Health Services during the Novel Coronavirus 2019 Era.

Authors:  Tivani P Mashamba-Thompson; Paul K Drain
Journal:  Diagnostics (Basel)       Date:  2020-07-03

10.  Rapid Diagnosis of Infection in the Critically Ill, a Multicenter Study of Molecular Detection in Bloodstream Infections, Pneumonia, and Sterile Site Infections.

Authors:  Jean-Louis Vincent; David Brealey; Nicolas Libert; Nour Elhouda Abidi; Michael O'Dwyer; Kai Zacharowski; Malgorzata Mikaszewska-Sokolewicz; Jacques Schrenzel; François Simon; Mark Wilks; Marcus Picard-Maureau; Donald B Chalfin; David J Ecker; Rangarajan Sampath; Mervyn Singer
Journal:  Crit Care Med       Date:  2015-11       Impact factor: 7.598

View more
  1 in total

1.  Comparative analysis of loop-mediated isothermal amplification combined with microfluidic chip technology and q-PCR in the detection of clinical infectious pathogens.

Authors:  Enqi Zhang; Hongling Ou; Lianling Jia; Wang Zhang; Yemei Wang; Xinru Wang
Journal:  J Clin Lab Anal       Date:  2022-06-26       Impact factor: 3.124

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.