| Literature DB >> 35751410 |
Hanne Elisabeth Lunde1,2, Agnete Nyborg Hjelmtvedt1,3, Erik Koldberg Amundsen2,3.
Abstract
INTRODUCTION: The aim of this study was to investigate the diagnostic accuracy of the flag PLT-Clumps from the WNR/WDF and the PLT-F channels from Sysmex XN and to study how different cut-offs for investigation for pseudothrombocytopenia (PTCP) and the definition of platelet aggregation affected the diagnostic accuracy.Entities:
Keywords: Sysmex; hematology instrument flag; platelet aggregation; platelets; pseudothrombocytopenia
Mesh:
Year: 2022 PMID: 35751410 PMCID: PMC9545299 DOI: 10.1111/ijlh.13920
Source DB: PubMed Journal: Int J Lab Hematol ISSN: 1751-5521 Impact factor: 3.450
FIGURE 1(A) Inclusion of samples. (B) Graphic presentation of 30 fields in a smear investigated for platelet aggregation. Created in BioRender
Classification of platelet aggregates in smear review
| Size | Average number of platelets per aggregate (based on 30 fields with 40× objective) | ||
|---|---|---|---|
| 0 | <3 aggregates in 30 fields | ||
| 1 | Small | 3–6 platelets | Based on the definitions from Noklus |
| 5–6 platelets | Based on the definitions from GFHC | ||
| 2 | Medium | 7–19 platelets | |
| 3 | Large | ≥ 20 platelets | |
Observations of aggregates in blood smears based on the Noklus‐definition and GFHC‐definitions of platelet aggregation
| Number of samples | NOKLUS = minimum 3 platelets at least 3 times | GFHC = minimum 5 platelets at least 3 times | ||
|---|---|---|---|---|
| Total number of smears reviewed | 419 | 419 | ||
| Without platelet clumping | 251 | 60% | 316 | 75% |
| With platelet clumping | 168 | 40% | 103 | 25% |
| Size 1 | 98 | 58% | 5 | 5% |
| Size 2 | 70 | 42% | 96 | 93% |
| Size 3 | 0 | 0% | 2 | 2% |
| Quantity 1 | 108 | 64% | 70 | 68% |
| Quantity 2 | 29 | 17% | 15 | 15% |
| Quantity 3 | 31 | 18% | 18 | 17% |
Abbreviations: GFHC, Groupe Francophone de'Hematologie Cellulaire; Noklus, Norwegian quality improvement of laboratory investigations.
Diagnostic accuracy of the flag PLT‐Clumps from WNR/WDF and PLT‐F with different inclusion criteria and definitions of platelet aggregation
| Number of samples without presence of aggregates in smear | Number of samples with presence of aggregates in smear | AUC 95% CI | Sensitivity at Q‐value = 100 95% CI | Specificity at Q‐value = 100 95% CI | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| WNR/WDF | PLT‐F | WNR/WDF | PLT‐F | WNR/WDF | PLT‐F | WNR/WDF | PLT‐F | WNR/WDF | PLT‐F | |
| Noklus All | 205 | 125 | 152 | 68 |
0.56 0.50–0.61 |
0.80 0.73–0.87 |
0.20 0.14–0.28 |
0.57 0.45–0.69 |
0.96 0.92–0.998 |
0.92 0.86–0.96 |
| Noklus All & quantity 2 + 3 | 304 | 159 | 53 | 34 |
0.62 0.53–0.72 |
0.88 0.81–0.96 |
0.40 0.37–0.57 |
0.82 0.66–0.93 |
0.94 0.90–0.96 |
0.87 0.81–0.92 |
| Noklus PLT ≤150 | 198 | 121 | 102 | 46 |
0.54 0.47–0.61 |
0.76 0.67–0.85 |
0.15 0.09–0.23 |
0.50 0.35–0.65 |
0.97 0.93–0.99 |
0.95 0.90–0.98 |
|
Noklus PLT ≤150 Quantity 2 + 3 | 279 | 148 | 21 | 19 |
0.59 0.44–0.74 |
0.88 0.78–0.99 |
0.38 0.18–0.62 |
0.79 0.54–0.94 |
0.95 0.92–0.97 |
0.91 0.85–0.95 |
| Noklus PLT ≤100 | 74 | 90 | 30 | 35 |
0.50 0.38–0.61 |
0.73 0.62–0.84 |
0.17 0.06–0.35 |
0.43 0.26–0.61 |
0.96 0.89–0.99 |
0.96 0.89–0.98 |
| Noklus PLT ≤100 Quantity 2 + 3 | 93 | 113 | 11 | 12 |
0.60 0.40–0.80 |
0.86 0.70–1.00 |
0.36 0.11–0.70 |
0.83 0.52–0.98 |
0.96 0.98–0.99 |
0.92 0.85–0.96 |
| GFHC All | 260 | 149 | 97 | 44 |
0.57 0.50–0.64 |
0.84 0.76–0.92 |
0.26 0.18–0.36 |
0.71 0.55–0.83 |
0.94 0.91–0.97 |
0.88 0.82–0.93 |
|
GFHC All & quantity 2 + 3 | 327 | 175 | 30 | 18 |
0.49 0.37–0.62 |
0.93 0.90–0.97 |
0.27 0.12–0.46 |
0.89 0.65–0.97 |
0.90 0.87–0.93 |
0.81 0.75–0.87 |
| GFHC PLT ≤150 | 244 | 141 | 56 | 26 |
0.56 0.47–0.64 |
0.82 0.71–0.93 |
0.21 0.12–0.34 |
0.65 0.44–0.83 |
0.96 0.93–0.98 |
0.92 0.86–0.96 |
|
GFHC PLT ≤150 Quantity 2 + 3 | 284 | 153 | 16 | 14 |
0.57 0.38–0.75 |
0.97 0.94–0.99 |
0.44 0.20–0.70 |
0.93 0.66–1.00 |
0.95 0.91–0.97 |
0.90 0.84–0.94 |
| GFHC PLT ≤100 | 88 | 108 | 16 | 17 |
0.49 0.33–0.65 |
0.79 0.64–0.94 |
0.25 0.07–0.52 |
0.65 0.38–0.86 |
0.96 0.89–0.99 |
0.93 0.86–0.97 |
|
GFHC PLT ≤100 Quantity 2 + 3 | 95 | 116 | 9 | 9 |
0.59 0.35–0.83 |
0.98 0.96–1.00 |
0.44 0.34–0.79 |
1.00 0.66–1.00 |
0.96 0.90–0.99 |
0.91 0.85–0.96 |
Abbreviations: AUC, area under curve; CI, confidence interval; GFHC: Groupe Francophone de'Hematologie Cellulaire; Noklus, Norwegian quality improvement of laboratory investigations; PLT, Platelet Count ×109/L; PLT‐F, Platelet Fluorescence method for platelet counting on Sysmex XN instrument; Q‐value, numeric value for the PLT‐Clumps flag on Sysmex XN instrument; WNR/WDF, measurement channel for leucocyte count on Sysmex XN instrument.
FIGURE 2Percentage difference between platelet count in Na‐Citrate and EDTA‐blood as a function of the observed number of aggregates in 30 fields with a 40× objective (n = 44)