Detection of Platelet Clumps on Peripheral Blood Smears by CellaVision DM96 System and Microscopic Review.

Gene Gulati1, Guldeep Uppal2, Alina Dulau Florea2, Jerald Gong2.   

Abstract

OBJECTIVE: To determine and optimize the sensitivity of the CellaVision DM96 automated image-analysis system in detecting platelet (PLT) clumps on blood smears and to assess the reliability of the traditional laboratory practice of examining only the feather edge of the smear for PLT clumps.
METHODS: We processed 102 blood smears that revealed PLT clumps on microscopic review, using the CellaVision DM96, and reviewed the results for the ability of the analyzer to detect these clumps. We obtained the data regarding relative distribution of PLT clumps on different parts of the blood smear (feather edge, lateral edges, and readable area) from our microscopic-review observations.
RESULTS: The sensitivity of the Cellavision DM96 in detecting PLT clumps was between 40.4% and 82.8%, depending on the number of screens reviewed for this variable. Via microscopic review of the smears, the PLT-clump detection rate increased from 85.3%, obtained by examining only the feather edge, to 99.0%, obtained by examining the feather edge plus the readable area.
CONCLUSION: The sensitivity of the DM96 for detecting PLT clumps can be maximized to 82.8% by reviewing the entire white blood cell screen and the entire PLT screen. Microscopic review of the blood smears yielded a PLT-clump detection rate of 99.0% when we examined the feather edge and the readable area of the smear. Copyright© by the American Society for Clinical Pathology (ASCP).

Entities:  

Keywords:  CellaVision DM96; DM96; blood smears; platelet clumps; platelet estimates; platelet scans

Mesh:

Year:  2014        PMID: 25316670     DOI: 10.1309/LM604RQVKVLRFXOR

Source DB:  PubMed          Journal:  Lab Med        ISSN: 0007-5027


  3 in total

1.  Assessment of the Reliability of the Sysmex XE-5000 Analyzer to Detect Platelet Clumps.

Authors:  Jennifer Hawkins; Gene Gulati; Guldeep Uppal; Jerald Gong
Journal:  Lab Med       Date:  2016-04-25

2.  Evaluation of Scopio Labs X100 Full Field PBS: The first high-resolution full field viewing of peripheral blood specimens combined with artificial intelligence-based morphological analysis.

Authors:  Ben-Zion Katz; Michael D Feldman; Minychel Tessema; Dan Benisty; Grace Stewart Toles; Alicia Andre; Bronka Shtreker; Fatima Maria Paz; Joshua Edwards; Darrin Jengehino; Adam Bagg; Irit Avivi; Olga Pozdnyakova
Journal:  Int J Lab Hematol       Date:  2021-09-21       Impact factor: 3.450

3.  The diagnostic accuracy of Sysmex XN for identification of pseudothrombocytopenia using various thresholds for definition of platelet aggregation.

Authors:  Hanne Elisabeth Lunde; Agnete Nyborg Hjelmtvedt; Erik Koldberg Amundsen
Journal:  Int J Lab Hematol       Date:  2022-06-24       Impact factor: 3.450