| Literature DB >> 35747659 |
Maude Joannette Bourguignon1, Sonia Bédard2,3, Frédéric Balg1,3, Karina Lebel4,5, Alexandre Keith Leang1, François Vézina1.
Abstract
Purpose: To determine whether different types of measurement tools can be interchanged without significantly affecting the resulting graft diameter.Entities:
Year: 2022 PMID: 35747659 PMCID: PMC9210359 DOI: 10.1016/j.asmr.2022.01.002
Source DB: PubMed Journal: Arthrosc Sports Med Rehabil ISSN: 2666-061X
Fig 1Graft combinations.
Fig 2Graft drawing. (A) Doubled semitendinosus and gracilis with unfolded quadriceps tendon (4HT+ QT). (B) Tripled semitendinosus and gracilis.
Fig 3Conmed anterior cruciate ligament measurement tools: (A) Slotted block and (B) sizing tubes.
Demographic Information
| Number of specimens | 17 (33 knees) |
| Average age | 76.2 ± 13.6 years |
| Body mass index | 22.6 ± 4 kg/m2 |
| Sex | 52.9% males |
Intrarater and Inter-Rater Reliability
| Slotted sizing tool | Tubes | |||
|---|---|---|---|---|
| ICC | 95% CI | ICC | 95% CI | |
| Intrarater reliability | ||||
| Surgeon 1 | 0.962 | [0.858, 0.990] | 0.931 | [0.738, 0.983] |
| Surgeon 2 | 0.972 | [0.894, 0.993] | 0.988 | [0.956, 0.997] |
| Overall | 0.965 | [0.914, 0.986] | 0.952 | [0.880, 0.981] |
| Inter-rater reliability | 0.980 | [0.970, 0.987] | 0.979 | [0.950, 0.989] |
ICC, intracorrelation coefficient; CI, confidence interval.
Correlation Slotted Versus Tubes
| Grafts combination type | N | Correlation | |
|---|---|---|---|
| Doubled semitendinosus and gracilis (4HT) | 33 | 0.941 | <.001 |
| Tripled semitendinosus and gracilis (6HT) | 33 | 0.965 | <.001 |
| Doubled semitendinosus and gracilis augmented with unfolded quadriceps tendon graft (4HT+QT) | 33 | 0.961 | <.001 |
| Overall | 99 | 0.978 | <.001 |
Statistically significant data.