Validation of reduced volume VeriFiler™ Express PCR Amplification Kit for buccal swab samples extracted using Prep-n-Go™ Buffer.

Jessica Perry1, Tasnim Munshi1, Thomas Haizel2, Sasitaran Iyavoo1,2.   

Abstract

The efficiency of reduced volume PCR amplification was studied using the VeriFiler™ Express PCR Amplification Kit. Full (25 μL) and reduced (5 μL) volumes were tested in parallel to identify any differences in template DNA sensitivity and other electropherogram parameters. Both volumes produced full DNA profiles down to 0.08 ng/μL DNA concentration at 26 PCR cycles; however, reduced volume produced higher peak heights due to increased signal intensities. Significant difference (p-value ≤ 0.05) in heterozygote peak height ratios was observed between both volumes, where the reduced volume threshold was lowered to 0.6 to accommodate all data points. However, no significant difference (p-value > 0.05) was identified in the stutter ratios between both volumes. The analytical threshold for reduced volume was also determined to be 150 RFU with the presence of template DNA in PCR amplification. When the optimized reduced volume parameters were tested on DNA extracted from buccal swab samples using Prep-n-Go™ Buffer, good quality DNA profiles were produced. Overall, the reduced volume not only showed better results compared to the full volume, but also enable more samples to be processed with a PCR amplification kit, thus reduced the cost.
© 2022 The Authors. Journal of Forensic Sciences published by Wiley Periodicals LLC on behalf of American Academy of Forensic Sciences.

Entities:  

Keywords:  DNA extraction; Prep-n-Go™ Buffer; VeriFiler™ Express kit; buccal swab; reduced volume PCR amplification; short tandem repeats

Mesh:

Substances:

Year:  2022        PMID: 35735089      PMCID: PMC9546378          DOI: 10.1111/1556-4029.15084

Source DB:  PubMed          Journal:  J Forensic Sci        ISSN: 0022-1198            Impact factor:   1.717


Reduction in PCR amplification volume to one‐fifth without compromising the quality of DNA profiles. DNA extraction from buccal swabs and direct PCR amplification eliminated DNA quantification step. Reduction in cost and turn‐around‐time of DNA profile generation from buccal swab samples.

INTRODUCTION

Short tandem repeats are commonly utilized in DNA profiling in both a forensic and relationship testing context. Multiplex kits have been designed to allow for multiple STR loci to be amplified and analyzed at once. It is becoming more common for users of these kits to validate using a lower PCR reaction mix volume to get more PCR reactions out of one kit. The validation of reduced volume PCR has been achieved with many common multiplex STR kits, for example, the PowerPlex® 16 System [1], AmpFLSTR™ Identifiler™ PCR Amplification Kit [2], and GlobalFiler™ PCR Amplification Kit [3]. Overall, reducing the PCR volume reduces cost [4] and has been found to increase the sensitivity and reduce the amount of template DNA required [1]. It has been found that the volume can be reduced by half without impacting the quality of the profile produced [5]. The VeriFiler™ Express PCR Amplification Kit (Applied Biosystems™) has had little to no validation for a reduced PCR volume. Therefore, this study was carried out to validate reduced volume PCR amplification with this kit, which amplifies 22 autosomal loci (D3S1358, vWA, D16S539, CSF1PO, TPOX, D8S1179, D21S11, D18S51, Penta E, D2S441, D19S433, TH01, FGA, D22S1045, D5S818, D13S317, D7S820, D6S1043, D10S1248, D1S1656, D12S391, D2S1338, and Penta D) [6]. This kit is approved by US National DNA Index System (NDIS) and was mainly developed for paternity and kinship testing, where it offers superior genotyping results from single‐source samples with high discrimination power and direct amplification capability [7]. Furthermore, the effectiveness of this reduced volume method was tested on DNA extracted from buccal swabs using Prep‐n‐Go™ buffer; this is a simple and quick DNA extraction buffer designed for direct PCR amplification [8].

MATERIALS AND METHODS

Optimization of reduced volume PCR amplification method

Sample preparation

A serial dilution from a starting DNA concentration of 10 ng/μL was prepared using the 2800M Control DNA (Promega®) giving concentrations of 10, 5, 2.5, 1.25, 0.63, 0.31, 0.16, 0.08, 0.04, 0.02, and 0.01 ng/μL. DNA Control 007, included in the VeriFiler™ Express kit was used as the positive control for both full and reduced reaction volumes, while the deionized water used for the serial dilution was utilized as the negative control.

PCR amplification

The VeriFiler™ Express PCR Amplification Kit (Applied Biosystems™) which includes the master mix and primer set is recommended to be used as follows: 10 μL of master mix, 10 μL of primer set, and 5 μL of DNA solution to make up the 25 μL full reaction volume. The reduced volume reaction mix maintained these ratios, using 2 μL of master mix, 2 μL of primer set, and 1 μL of DNA solution, making up the 5 μL reduced reaction volume. Each serial dilution concentration was processed in triplicate for both full and reduced volumes. PCR amplification was carried out on the GeneAmp® PCR System 9700 thermal cycler (Applied Biosystems™) for 26 cycles with Max ramping mode.

Capillary electrophoresis

Hi‐Di™ Formamide (Applied Biosystems™) and GeneScan™ 600 LIZ™ dye Size Standard v2.0 (Applied Biosystems™) were used in an 8.7–0.3 μL ratio respectively for each sample and then had 1 μL of PCR product (or Verifiler™ Express Allelic Ladder) added to make up the total volume of 10 μL. Samples were run on the 3500xL Genetic Analyzer (Applied Biosystems™) with the 3500 Series Data Collection Software 4, version 4.0.1 (Applied Biosystems™) as per the manufacturer's recommended parameters for VeriFiler™ Express kit.

Data analysis

GeneMapper™ ID‐X v1.6 software (Applied Biosystems™) was used to analyze the raw data to determine the allele calls. Statistical analyses were carried out using the t‐test on Microsoft Excel®.

Analytical threshold study

Twenty‐three replicates of positive and negative controls for the reduced volume PCR amplification method were prepared for the analytical threshold study. The VeriFiler™ Express DNA Control 007 was diluted with deionized water in a 15:15 μL ratio to make the concentration 1 ng/μL to be used as the DNA template for positive controls, while the deionized water employed for the dilution was used for negative controls. For this study, the relative fluorescence units (RFU) on the GM ID‐X software were set to 1 for all dye channels, with all data points in the 60–480 base pair size range captured. For the positive controls, true and stutter peaks were removed, leaving only baseline noise for calculation. The limit of quantification (LOQ) was calculated by adding 10 standard deviations to the average baseline signal [9].

Sensitivity study

Serial dilution triplicates for both reaction volumes were compared to identify the lowest concentration at which full profiles were produced. Full profiles were defined as having all peaks for all alleles above the threshold determined in the analytical threshold study. Average peak height for each locus was also calculated from the serial dilution triplicates for both reaction volumes, to identify any significant difference between them.

Heterozygote peak height ratio study

From the full and reduced volume electropherograms, only heterozygotes where both peaks were in the range of 3000–12,000 RFU were used to calculate the heterozygote peak height ratios, since heterozygotes with no off‐scale allele peaks or allelic dropout would be produced in this range [6]. The calculation was carried out by dividing the height of shorter allele peak by the height of taller allele peak.

Stutter ratio study

For this study, all electropherogram artifacts were removed, leaving only true peaks and stutter peaks. Triplicate profiles with heterozygote peak heights within the 3000–12,000 RFU range were selected along with all homozygotes in those profiles. Stutter ratio was calculated by dividing the stutter peak height by that of the true allele peak height for each locus for all serial dilution concentrations. Then the stutter ratio threshold was calculated by adding three standard deviations to the highest stutter ratio of the locus.

Evaluation of reduced volume PCR amplification method on DNA from buccal swab samples extracted using Prep‐n‐Go™ Buffer

Buccal swab samples

Buccal swab (Puritan®) samples were collected from 104 anonymous volunteers. The volunteers, who provided informed consent, were requested to not consume any food or drink for 30 min prior to sampling. Buccal samples were collected by rubbing the cotton end of the swab against the inside of the cheek for approximately 30 s; the swab was then allowed to air dry at room temperature for approximately 30 s. The swab was then placed swab‐end first into a sealable envelope labeled with the details of the volunteer.

DNA extraction

Prep‐n‐Go™ Buffer (Applied Biosystems™) was used to extract DNA from the buccal swabs. Room temperature protocol [6] was used for the DNA extraction utilizing 400 μl Prep‐n‐Go™ Buffer per sample. The buffer was added to the swabs in 1.5 ml Eppendorf® tubes and left to stand for 20 min at room temperature (20–25°C) to lyse the sample.

DNA quantification

The QuantStudio™ 5 Real‐Time PCR System (Applied Biosystems™) was used to quantify the extracted DNA using the Quantifiler™ HP DNA Quantification Kit (Applied Biosystems™) following manufacturer's recommendation to identify the amount of DNA produced by Prep‐n‐Go™ Buffer extraction on buccal swab samples.

DNA profiling

Extracted DNA samples were directly processed with the optimized reduced volumes method for PCR amplification, capillary electrophoresis, and data analysis on GM‐IDX software, as described in Section 2.1.

RESULTS AND DISCUSSION

Analytical threshold study

Limit of quantification for analytical thresholds can be calculated with only the negative amplification controls [10] where the captured data points will be without any known alleles and stutters. However, presence of template DNA in the PCR amplification can increase the baseline noise at non‐allele product areas [11]. Even though the average baseline noise differences between negative amplification controls and template DNA samples are relatively small [9], inclusion of template DNA samples in the analytical threshold generation is suggested when applied to samples with high levels of DNA [12], such as the buccal swab samples in this study. The highest LOQ for the negative controls was 63.22 RFU in the green dye channel, while the highest LOQ for the positive controls was 147.21 RFU in the purple dye channel. Accordingly, the LOQ threshold for the reduced volume method was set to 150 RFU. This finding demonstrates that the presence of DNA has an impact on the baseline noise level. These results are displayed in Table 1.
TABLE 1

Baseline calculations for each dye set for positive and negative controls.

SampleDye channelHighest baseline noise (RFU)AverageSD10 SDLOQ
PositiveBlue10813.4611.71117.15130.61
Green10820.309.2792.72113.02
Yellow9910.739.8698.57109.29
Red11114.5611.95119.54134.09
Purple11718.6512.86128.57 147.21
NegativeBlue3310.183.1631.5841.76
Green5415.974.7247.25 63.22
Yellow257.052.1321.2928.34
Red3010.122.1921.9132.03
Purple379.173.8438.4147.58

Note: The average, standard deviation (SD), 10 SD and average + 10 SD was calculated to give the limit of quantification (LOQ). Highest LOQ for both controls in bold.

Baseline calculations for each dye set for positive and negative controls. Note: The average, standard deviation (SD), 10 SD and average + 10 SD was calculated to give the limit of quantification (LOQ). Highest LOQ for both controls in bold.

Sensitivity study

Utilizing the 150 RFU analytical threshold, both the full and reduced volume samples produced full profiles down to a DNA concentration of 0.08 ng/μL at 26 cycles. Table 2 shows the minimum DNA concentration required for the correct profile to be generated at each locus across each repeat for both full and reduced volumes. Greater sensitivity down to the lowest DNA concentration in this study (0.01 ng/μL) could be achieved by increasing the cycle number [13]. However, since this reduced volume method was being developed for buccal swab samples extracted using Prep‐n‐Go™ Buffer, which will provide good DNA quantity [14], 26 cycles was maintained.
TABLE 2

The minimum DNA concentrations (ng/μL) required for the correct profile to be generated at each locus for full and reduced volumes with 150 RFU threshold. Overall minimum DNA concentration required to produce a full profile in all repeats for both volumes in bold.

LocusFull volume (25 μL)Reduced volume (5 μL)
Repeat 1Repeat 2Repeat 3MinimumRepeat 1Repeat 2Repeat 3Minimum
D3S13580.040.040.08 0.08 0.040.040.040.04
vWA0.040.040.040.040.040.040.040.04
D16S5390.040.040.040.040.040.040.040.04
CSF1PO0.020.010.020.020.010.010.010.01
TPOX0.040.040.040.040.020.040.040.04
D8S11790.040.040.040.040.020.040.040.04
D21S110.040.040.040.040.020.040.020.04
D18S510.040.020.040.040.040.040.040.04
Penta E0.040.040.040.040.020.040.040.04
D2S4410.040.040.040.040.040.040.040.04
D19S4330.040.040.08 0.08 0.040.040.08 0.08
TH010.040.040.040.040.040.080.08 0.08
FGA0.040.040.040.040.040.040.040.04
D22S10450.020.020.010.020.010.010.020.02
D5S8180.020.040.040.040.010.010.040.04
D13S3170.040.040.040.040.020.040.040.04
D7S8200.040.040.040.040.020.040.040.04
D6S10430.020.040.040.040.020.040.040.04
D10S12480.080.040.04 0.08 0.040.040.040.04
D1S16560.040.040.040.040.040.040.040.04
D12S3910.040.040.040.040.040.020.040.04
D2S13380.020.040.020.040.020.040.040.04
Penta D0.040.040.040.040.040.020.040.04
The minimum DNA concentrations (ng/μL) required for the correct profile to be generated at each locus for full and reduced volumes with 150 RFU threshold. Overall minimum DNA concentration required to produce a full profile in all repeats for both volumes in bold. Peak heights of full and reduced volumes were also compared for each locus using the 150 RFU analytical threshold. Expected allele calls were produced from the 2800M Control DNA [15], with homozygotes at loci CSF1PO, TPOX, D22S1045, and D5S818. Average peak height and standard error for each locus are shown in Table 3. The reduced volume method produced higher average peak heights for all loci from the serial dilution samples used in this study as shown in Figure 1. This observation was also supported by statistical analysis which showed significant difference (p‐value ≤ 0.05) between peak heights of full and reduced volumes. Therefore, even though both reaction volumes have the same sensitivity limit, reduced volume produces electropherograms with higher signal intensities. This can be observed in Figure 2 which shows the electropherograms of full and reduced volumes for the purple dye channel, generated from a serial dilution sample with concentration of 0.31 ng/μL.
TABLE 3

Average peak height (RFU) and standard error for each locus from full (25 μL) and reduced (5 μL) PCR reaction volumes.

LocusAverage peak height (full volume)Standard error (full volume)Average peak height (reduced volume)Standard error (reduced volume)
D3S135884831185.1213,3841694.09
vWA95571445.6814,1261786.10
D16S53910,2031498.5114,2951812.21
CSF1PO15,2912478.6317,8012483.80
TPOX14,0482466.6714,9352446.22
D8S117910,0401552.2512,9991647.44
D21S1182361260.1812,7171634.71
D18S5110,4081380.8013,1761536.54
Penta E90781229.3713,5671638.65
D2S44189731362.6111,5091555.70
D19S43390751416.3011,4331569.35
TH0112,0631651.8912,9471701.15
FGA93731429.9311,8091627.47
D22S104514,8602402.1016,4632283.40
D5S81812,8012454.8215,1522449.44
D13S31787651306.2114,4701781.04
D7S82089701291.0113,6421702.31
D6S104389531341.5714,0161750.83
D10S124884501139.6813,9701690.15
D1S165611,1431621.7415,1571776.43
D12S39194491281.9614,8781778.98
D2S133813,3081704.0014,2141727.73
Penta D11,5111570.6714,9171776.50
FIGURE 1

Comparison of average peak heights between full (25 μL) and reduced (5 μL) PCR reaction volumes.

FIGURE 2

Electropherograms of full (upper pane) and reduced (lower pane) PCR reaction volumes for the purple dye channel, generated from a serial dilution sample with concentration of 0.31 ng/μL.

Average peak height (RFU) and standard error for each locus from full (25 μL) and reduced (5 μL) PCR reaction volumes. Comparison of average peak heights between full (25 μL) and reduced (5 μL) PCR reaction volumes. Electropherograms of full (upper pane) and reduced (lower pane) PCR reaction volumes for the purple dye channel, generated from a serial dilution sample with concentration of 0.31 ng/μL.

Heterozygote peak height ratio study

Overall, the reduced volume method had a lower average heterozygote peak height ratio (0.8870) compared to the full volume method (0.9462). However, when the individual data points were analyzed, reduced volume had the highest ratio (0.9997), compared to the full volume method (0.9982). While all data points for full volume were above the manufacturer's default threshold (0.7) [16], only 2 data points (2.06%) were below this threshold in the reduced volume. However, a significant difference (p‐value ≤ 0.05) was observed between both volume sample sets. The final threshold for reduced volume method was set at 0.6 to cover all heterozygote peak height ratios as shown in Figure 3 with the manufacturer's default threshold.
FIGURE 3

Heterozygote peak height ratio comparison between reduced (5 μL) and full (25 μL) PCR reaction volumes.

Heterozygote peak height ratio comparison between reduced (5 μL) and full (25 μL) PCR reaction volumes.

Stutter ratio study

Stutter alleles are caused by slippage when DNA polymerase pauses during primer extension [17, 18]. It could be one repeat unit smaller than the target STR allele product (minus stutter) or one repeat unit larger (plus stutter) which is less frequent. Plus stutter was observed at locus D22S1045, which is a trinucleotide repeat‐containing locus and this phenomenon is more significant compared to the tetranucleotide repeats [19, 20]. When comparing the full and reduced volumes, there was no significant difference (p‐value > 0.05) identified between the stutter ratio thresholds of both data sets. Results for this study are shown in Table 4, with manufacturer's threshold for comparison.
TABLE 4

Observed stutter calculated for each locus from full (25 μL) and reduced (5 μL) PCR reaction volumes.

LocusObserved stutter for full volumeObserved stutter for reduced volumeManufacturer's threshold
D3S13580.08980.09660.1216
vWA0.13170.14530.1318
D16S5390.13840.14780.1076
CSF1PO0.07400.08840.1054
TPOX0.04100.05640.0520
D8S11790.08780.10540.1085
D21S110.09850.11840.1224
D18S510.14120.15380.1415
Penta E0.05480.07500.0831
D2S4410.05800.06710.0940
D19S4330.06330.06890.1077
TH010.01950.02650.0493
FGA0.11730.11750.1227
D22S1045

0.1038

0.0575 (+3 nt)

0.1151

0.0872 (+3 nt)

0.1800

0.0805 (+3 nt)

D5S8180.07990.09010.1053
D13S3170.10190.11780.1010
D7S8200.08960.10180.0939
D6S10430.12820.14360.1132
D10S12480.12840.14220.1227
D1S16560.06930.07030.1362
D12S3910.18500.19980.1444
D2S13380.14440.17650.1329
Penta D0.02970.03020.0409

Note: Manufacturer's threshold is also included for comparison.

Observed stutter calculated for each locus from full (25 μL) and reduced (5 μL) PCR reaction volumes. 0.1038 0.0575 (+3 nt) 0.1151 0.0872 (+3 nt) 0.1800 0.0805 (+3 nt) Note: Manufacturer's threshold is also included for comparison.

Buccal swab samples with reduced volume PCR amplification method

The Prep‐n‐Go™ Buffer extraction method recommended direct PCR amplification, without quantification [6]. However, DNA quantification on all 104 buccal swab samples was carried out to determine the DNA concentrations obtained using the room temperature protocol of the Prep‐n‐Go™ Buffer. On average, the DNA concentration was 4.27 (1.82 SD) ng/μL with the lowest and highest DNA concentrations being 0.87 ng/μL and 10.02 ng/μL respectively. All buccal swab samples produced full DNA profiles with the finalized reduced volume analysis parameters on the GM‐IDX software. No excessive stutter or heterozygote peak height imbalance indicators were observed on the electropherograms. The lowest heterozygote peak height ratio was 0.6039 which is within the lowered threshold (0.6) for reduced volume method. Any heterozygote peak height ratio falling below this threshold will be flagged by the GM ID‐X software and can be evaluated during the analysis. As size difference between heterozygote alleles increased, a reduction in heterozygote peak balance was observed. No significant change in heterozygote peak height imbalance was observed over the range of DNA concentration of buccal swab samples, but more artifacts were observed as the sample DNA concentration increased, especially pull‐ups (bleed‐through) and split peaks (incomplete adenylation). However, these artifacts had no effect on the true allele calls and were easily identified and corrected.

CONCLUSION

This study demonstrates that reduced volume (5 μL) PCR amplification method is robust and produces reliable results similar to the full volume (25 μL) method. The analytical threshold study results found a threshold of 150 RFU to be appropriate for the reduced volume method. Whilst this is a high threshold, it is acceptable as it takes into consideration the presence of template DNA in PCR amplification. The sensitivity study found the reduced volume method produced full profiles to the same concentration (0.08 ng/μL) as the full volume at 26 cycles. However, when comparing the peak heights of full and reduced volumes, it was shown that the reduced volume had higher signal intensities. The heterozygote peak height ratio study found the full volume method to have better heterozygote peak balance compared to the reduced volume method. However, with the new threshold (0.6), interpretation of the DNA profiles was unaffected. With all these optimized reduced volume analysis parameters, this method is suitable to be applied to samples with optimal DNA concentration. Additionally, the Prep‐n‐Go™ Buffer which was used for DNA extraction from buccal swab samples showed great efficacy in extracting DNA with optimal amounts for this reduced volume method, making the whole system suitable for high‐throughput sample processing workflow with reduced cost.

CONFLICT OF INTEREST

The authors declare no conflicts of interest.
  13 in total

1.  Validation of half-reaction amplification using Promega PowerPlex 16.

Authors:  Nicole H Hoffman; Terry Fenger
Journal:  J Forensic Sci       Date:  2010-04-16       Impact factor: 1.832

2.  Run-specific limits of detection and quantitation for STR-based DNA testing.

Authors:  Jason R Gilder; Travis E Doom; Keith Inman; Dan E Krane
Journal:  J Forensic Sci       Date:  2007-01       Impact factor: 1.832

3.  Characterization of the N+3 stutter product in the trinucleotide repeat locus DYS392.

Authors:  Julio J Mulero; Chien Wei Chang; Lori K Hennessy
Journal:  J Forensic Sci       Date:  2006-09       Impact factor: 1.832

4.  Characterising stutter in forensic STR multiplexes.

Authors:  Clare Brookes; Jo-Anne Bright; SallyAnn Harbison; John Buckleton
Journal:  Forensic Sci Int Genet       Date:  2011-03-08       Impact factor: 4.882

5.  Sequence analysis and characterization of stutter products at the tetranucleotide repeat locus vWA.

Authors:  P S Walsh; N J Fildes; R Reynolds
Journal:  Nucleic Acids Res       Date:  1996-07-15       Impact factor: 16.971

6.  Resizing reaction volumes for the ForenSeq™ DNA Signature Prep kit library preparation.

Authors:  Stefania Turrina; Domenico De Leo
Journal:  Med Sci Law       Date:  2021-01       Impact factor: 1.266

7.  The effects of extra PCR cycles when amplifying skeletal samples with the GlobalFiler® PCR Amplification Kit.

Authors:  Michelle Harrel; David Gangitano; Sheree Hughes-Stamm
Journal:  Int J Legal Med       Date:  2018-05-29       Impact factor: 2.686

8.  Optimization of a reduced volume PCR amplification for PowerPlex® Fusion kit using FTA™ cards and generation of population genetic data for Brunei population.

Authors:  Paul Vun Onn Liew; Laura Natalia Riccardi; Abimbola Olatunde Afolabi; William Goodwin; Sibte Hadi
Journal:  Electrophoresis       Date:  2018-09-02       Impact factor: 3.535

9.  Analytical thresholds and sensitivity: establishing RFU thresholds for forensic DNA analysis.

Authors:  Joli Bregu; Danielle Conklin; Elisse Coronado; Margaret Terrill; Robin W Cotton; Catherine M Grgicak
Journal:  J Forensic Sci       Date:  2012-11-06       Impact factor: 1.832

10.  Internal validation of the GlobalFiler™ Express PCR Amplification Kit for the direct amplification of reference DNA samples on a high-throughput automated workflow.

Authors:  Shahida Flores; Jie Sun; Jonathan King; Bruce Budowle
Journal:  Forensic Sci Int Genet       Date:  2014-01-28       Impact factor: 4.882

View more
  1 in total

1.  Validation of reduced volume VeriFiler™ Express PCR Amplification Kit for buccal swab samples extracted using Prep-n-Go™ Buffer.

Authors:  Jessica Perry; Tasnim Munshi; Thomas Haizel; Sasitaran Iyavoo
Journal:  J Forensic Sci       Date:  2022-06-23       Impact factor: 1.717

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.