Renato L Gil1, Célia G Amorim1, Maria Cuartero2. 1. LAQV-REQUIMTE, Department of Chemical Sciences, Faculty of Pharmacy, University of Porto, Jorge Viterbo Ferreira, 228, 4050-313 Porto, Portugal. 2. Department of Chemistry, School of Engineering Sciences in Chemistry, Biotechnology and Health, KTH Royal Institute of Technology, Teknikringen 30, SE-100 44 Stockholm, Sweden.
Abstract
An analytical methodology for detecting ammonium ion (NH4 +) in environmental water through potentiometry-ion chromatography (IC) in tandem is presented here. A multielectrode flow cell is implemented as a potentiometric detector after chromatographic separation of cations in the sample. The electrodes are fabricated via miniaturized all-solid-state configuration, using a nonactin-based plasticized polymeric membrane as the sensing element. The overall analytical setup is based on an injection valve, column, traditional conductometric detector, and new potentiometric detector (in that order), permitting the characterization of the analytical performance of the potentiometric detector while validating the results. The limit of detection was found to be ca. 3 × 10-7 M NH4 + concentration after linearization of the potentiometric response, and intra- and interelectrode variations of <10% were observed. Importantly, interference from other cations was suppressed in the tandem potentiometry-IC, and thus, the NH4 + content in fresh- and seawater samples from different locations was successfully analyzed. This analytical technology demonstrated a great potential for the reliable monitoring of NH4 + at micromolar levels, in contrast to the conductivity detector and previously reported NH4 + potentiometric sensors functioning in batch mode or even coupled with IC. Additionally, the suitability of the potentiometric cell for selective multi-ion analysis in the same sample, i.e., Na+, NH4 +, and K+ in water, has been proven.
An analytical methodology for detecting ammonium ion (NH4 +) in environmental water through potentiometry-ion chromatography (IC) in tandem is presented here. A multielectrode flow cell is implemented as a potentiometric detector after chromatographic separation of cations in the sample. The electrodes are fabricated via miniaturized all-solid-state configuration, using a nonactin-based plasticized polymeric membrane as the sensing element. The overall analytical setup is based on an injection valve, column, traditional conductometric detector, and new potentiometric detector (in that order), permitting the characterization of the analytical performance of the potentiometric detector while validating the results. The limit of detection was found to be ca. 3 × 10-7 M NH4 + concentration after linearization of the potentiometric response, and intra- and interelectrode variations of <10% were observed. Importantly, interference from other cations was suppressed in the tandem potentiometry-IC, and thus, the NH4 + content in fresh- and seawater samples from different locations was successfully analyzed. This analytical technology demonstrated a great potential for the reliable monitoring of NH4 + at micromolar levels, in contrast to the conductivity detector and previously reported NH4 + potentiometric sensors functioning in batch mode or even coupled with IC. Additionally, the suitability of the potentiometric cell for selective multi-ion analysis in the same sample, i.e., Na+, NH4 +, and K+ in water, has been proven.
The ammonium
ion (NH4+) is one of the primary compounds involved
in the biogeochemical
cycle of nitrogen (N) in aquatic environments.[1,2] Despite
its distribution in freshwaters being highly variable regionally,
seasonally, and spatially within streams and lakes, the concentration
of total ammonia nitrogen (TAN) in well-oxygenated waters is usually
low (typical values ranging from 7 to 60 μg TAN L–1).[1] When the N-cycle becomes unbalanced,
detrimental consequences manifest for the water life: high NH4+ levels (>2 mg TAN L–1) are
known to translate into undesired eutrophication and poor water quality.[1,2] Excessive NH4+ levels may occur through direct
means (municipal effluent discharges and animal waste) or indirect
sources (nitrogen fixation, air deposition, and runoff from agricultural
lands).[3] Altogether, NH4+ content is considered an important environmental indicator,
and monitoring it is crucial for effective water ecosystem preservation.[2]Analytical methodologies for NH4+ detection
have been developed for many years.[4,5] Spectrophotometric,
conductometric, and potentiometric readouts stand out for the specific
application of water analysis among the available techniques. Potentiometric
ion-selective electrodes (ISEs), based on ion-selective membranes
(ISMs), are particularly interesting because of their capacity for
decentralized measurements.[6] Indeed, a
recent review has thoroughly discussed the analytical features and
applications of NH4+-selective electrodes reported
in the past decade.[7] Considering the ionophore
(or receptor) as the core element of the ISM, effective NH4+ recognition has been accomplished over many years.[7] At present, nonactin is the most utilized ionophore,
despite its significant drawback of forming complexes with other cations
of similar ionic size and charge to NH4+, such
as potassium (K+) and sodium (Na+) ions. Thus,
the presence of excess K+ and Na+ compared to
NH4+ in the sample could hinder accurate determination
with ISEs, rendering the effective detection of NH4+ difficult in either aqueous or other samples. To overcome
this issue of selectivity, several strategies have been proposed in
the literature aside from the search for more selective ionophores.[7]Athavale et al. reported a mathematical
treatment to correct the
recorded ISE potential, which considered the K+ levels
in freshwater together with the NH4+/K+ selectivity coefficient.[8] Briefly, all-solid-state
ISEs were integrated into a submersible device, and the mathematical
treatment was used to measure the NH4+ in lakes.
While this approach provided accurate profiles from the depth where
the NH4+ concentration was greater than 10 μM,
the detection of lower concentrations remained unfeasible. Similarly,
potentiometric electronic tongues for NH4+ detection
have been proposed, which again require the previous determination
of the selectivity coefficients. The Diamond and del Valle groups
have targeted synthetic water samples, as well as river water and
wastewater with this approach, claiming accurate measurements only
at the millimolar level.[9,10] Speciation of N compounds
has also been pursued with electronic tongues.[11] Despite very valuable efforts reported in the literature
until now, demonstrating the utility of a statistical analysis of
potentiometric data (especially in the direction of sample classification),[12−15] the accurate quantification of NH4+ still
remains a challenge in environmental water samples.Some very
interesting analytical concepts have been developed to
reduce or eliminate interference(s) in the sample before applying
potentiometric NH4+ detection.[7] Such strategies have been applied to clinical samples rather
than environmental water, e.g., the method of indirect creatinine
determination reported by Liu et al.[16] All-solid-state
ISEs for NH4+ were modified with the creatinine
deiminase enzyme on top of the ISM and an outer anion-exchange membrane,
which prevents the cations passing from the sample to the core of
the sensing element (the ISM with the enzyme).[16] Hence, the potentiometric response represents only the
NH4+ formed in the quantitative enzymatic creatinine
reaction, without any K+/Na+ interference. Effectively,
separation science may offer suitable opportunities for NH4+ detection in the form of a broad portfolio of techniques
to separate ions in the complex matrix via differences in affinity.[17]This paper presents an analytical methodology
for accurately determining
NH4+ in environmental water samples based on
tandem potentiometry–ion chromatography (IC). A multielectrode
flow cell containing three miniaturized all-solid-state ISEs functionalized
with nonactin as the ionophore is proposed as the detector. The cell
is coupled in-line with an IC system based on a cation-exchange column.
In addition, a traditional conductivity detector is implemented, which
allows the validation of this new methodology. Given the selectivity
limitations generally presented by nonactin-based ISEs, the column
is exploited to provide the unequivocal separation of NH4+ from the main interfering ions (K+ and Na+) in real water samples. In addition, the potentiometric cell
is explored for multication detection in environmental water samples.
Experimental Section
Preparation of the Ion-Selective
Electrodes and Reference Electrode
Miniaturized handmade
glassy carbon electrode bodies were fabricated
as detailed in the Supporting Information.[18] All-solid-state ISEs for NH4+, K+, and Na+ were prepared via
the functionalization of the glassy carbon electrodes with the ion-to-electron
transducer (multiwalled carbon nanotubes, MWCNTs) and ISMs as follows
(Figure a). Membrane
cocktails containing a polymeric matrix, plasticizer, cation exchanger,
and the corresponding ionophore dissolved in tetrahydrofuran (THF)
were prepared to fabricate NH4+-, K+-, and Na+-selective electrodes, as detailed in the Supporting Information. The surface of the glassy
carbon electrode was first modified by drop casting 4 × 5 μL
of 1 mg mL–1 MWCNT solution in ethanol.[16] Each layer was allowed to dry for 10 min before
the next layer was added. Then, the ISM was formed by drop casting
4 × 5 μL of the corresponding membrane cocktail on top
of the MWCNT film. Each layer was allowed to dry for 20 min. Finally,
the electrodes were conditioned at least overnight in a 10–3 mol L–1 solution of the respective cation analyte.
Figure 1
(a) Miniaturized
ISEs based on glassy carbon (GC). The working
electrode (WE) was prepared with MWCNTs and the ISM. The reference
electrode (RE) was prepared with Ag/AgCl ink and the RM on top. (b)
Multielectrode flow cell with three ISEs and the RE, the inlet, and
the outlet. (c) Tandem potentiometry–IC: the sample is injected
through the valve and carried by the mobile phase through the chromatographic
column.
(a) Miniaturized
ISEs based on glassy carbon (GC). The working
electrode (WE) was prepared with MWCNTs and the ISM. The reference
electrode (RE) was prepared with Ag/AgCl ink and the RM on top. (b)
Multielectrode flow cell with three ISEs and the RE, the inlet, and
the outlet. (c) Tandem potentiometry–IC: the sample is injected
through the valve and carried by the mobile phase through the chromatographic
column.The reference electrode was fabricated
by modifying the surface
of the glassy carbon electrode with a silver/silver chloride (Ag/AgCl)
ink (C21310007D3, GWENT Group, UK),[16] followed
by oven-curing at 100 °C for 10 min. The reference membrane (RM,
composition in the Supporting Information) was drop casted on the Ag/AgCl coating (4 × 5 μL); each
layer was dried for 10 min before depositing the next one. The reference
electrode was left to dry overnight and conditioned in 3 M KCl for
at least 48 h. The electrode was stored in a 3 M KCl solution when
not in use.
Preparation of the Multielectrode Potentiometric
Flow Cell
Three miniaturized ISEs were inserted in the multielectrode
flow
cell together with the reference electrode.[18] The flow cell was a cube made of an acrylic block with six drilled
holes: two on opposite sides for the inlet and outlet, one for the
reference electrode, and three others to host the ISEs (Figure b). The electrodes were incorporated
using a plastic screw (flangeless yellow ferrule and a male green
nut, 1/8, IDEX Health & Science). The electrical connection was
made outside the cube with small crocodile clamps connected to each
electrode. For the inlet and outlet, a blue ferrule and a male red
nut (1/16, IDEX Health & Science) were used to connect the PTFE
tubing.
Potentiometry–Ion Chromatography System
The
potentiometry–IC tandem measurements (Figure c) were conducted using an 850 Professional
IC system (Metrohm Nordic) combined with a high-pressure pump and
a six-port high-pressure injection valve. A guard-column (Metrosep
C6 Guard/4.0; 5 × 2.0 mm I.D., 5 μm; Metrohm, Switzerland)
was placed before the analytical cation-exchange column (Metrosep
C6; 150 × 4.6 mm I.D., 4 μm, Metrohm, Switzerland). The
outlet of the cation-exchange column was connected to the Metrohm
conductivity detector. MagicNet chromatography data system software
(Metrohm, Switzerland) was used to control the IC components and for
data processing in the conductivity detector. The multielectrode flow
cell was coupled in-line with the outlet of the conductivity detector
using PEEK tubing (L × O.D. × I.D. = 300 mm × 1/16
in. × 0.25 mm, Metrohm) introduced in the ferrule and nut of
the inlet of the cell.The mobile phase used for the potentiometric
measurements was 2.5 × 10–3 mol L–1, as recommended by the column manufacturer for a proper functioning
and ions’ separation. Notably, different batches of the mobile
phase solution were found to provide slightly different retention
times for analogous ion peaks in the same sample. The injection volume
and flow rate were optimized to be 10 μL and 0.9 mL min–1, respectively.
Results and Discussion
Analytical
Evaluation of Miniaturized Potentiometric Sensors
for Ammonium
The analytical performance of the NH4+-selective electrodes was first evaluated under batch
conditions (background electrolyte: 2.5 × 10–3 mol L–1 nitric acid) against the commercial Ag/AgCl
reference electrode. Figure a shows the dynamic potentiometric response of the NH4+-selective electrode at increasing activities
of NH4+. Fitting the logarithmic NH4+ activity versus the corresponding steady-state potential
to the Nernst equation (inset of Figure a) revealed that the electrode followed a
Nernstian behavior, with a slope of 54.6 mV dec–1, with a linear range of response (LRR) from 3.0 × 10–6 to 3.0 × 10–3 NH4+ activity
and a limit of detection (LOD) of 6.1 × 10–7 NH4+ activity. The response time (t95)[19] was rapid, from 2.3 to 2.9 s within
the LRR. Furthermore, the response was found to be very reproducible
considering subsequent calibrations for the same electrode (slope
of 54.1 ± 0.9 mV dec–1 and intercept of 447.0
± 2.3 mV, n = 3) and equally prepared electrodes
(slope of 54.5 ± 1.8 mV dec–1 and intercept
of 439.9 ± 9.4 mV, n = 3), with variation coefficients
for the calibration parameters of <4%. Analogous calibration parameters,
but with a slight shift in the offset potential, were achieved with
the miniaturized handmade reference electrode (dotted line in Figure a): slope of 55.5
± 0.7 mV dec–1, LRR from 3.0 × 10–6 to 3.0 × 10–3 NH4+ activity and LOD of 2.0 ± 0.5 × 10–7 NH4+ activity (n = 3 electrodes).
Figure 2
(a) Dynamic
response of one NH4+-selective
electrode in batch mode at increasing NH4+ activity
and using the commercial Ag/AgCl reference electrode. Inset: Corresponding
calibration plot and that obtained with the handmade reference electrode.
(b) Response of one NH4+-selective electrode
in the flow cell (0.5 mL min–1). Inset: Corresponding
calibration graph.
(a) Dynamic
response of one NH4+-selective
electrode in batch mode at increasing NH4+ activity
and using the commercial Ag/AgCl reference electrode. Inset: Corresponding
calibration plot and that obtained with the handmade reference electrode.
(b) Response of one NH4+-selective electrode
in the flow cell (0.5 mL min–1). Inset: Corresponding
calibration graph.Three NH4+-selective electrodes were inserted
in the microfluidic cell with the handmade reference electrode (Figure b) for characterization
under flow-mode conditions using a peristaltic pump (0.5 mL min–1, 2.5 × 10–3 mol L–1 nitric acid background). Figure b displays the dynamic response at increasing NH4+ activities and the calibration graph corresponding
to one of the three NH4+-selective electrodes
as an example. The average calibration parameters of the three electrodes
were a slope of 52.5 ± 1.2 mV dec–1, LRR from
3.0 × 10–5 to 1.0 × 10–2 NH4+ activity, an intercept of 410.0 ±
20.3 mV, and a LOD of (7.5 ± 0.2) × 10–6 NH4+ activity. A slightly lower slope was
found, with the LRR starting from an NH4+ activity
1 order of magnitude higher than in the batch mode. The LOD was also
1 order of magnitude higher in the flow mode. Overall, the response
was slightly worse in the microfluidic cell, likely due to the flow-cell
configuration per se (tangential mode) and the hydrodynamic
conditions of the measurements. Other authors found that the tangential
mode hindered achieving approximately 95% of the steady-state signal,
especially in low ion analyte activities. In addition, relatively
low flow rates (such as 0.5 mL min–1) do not propitiate
the removal of infinitesimal concentrations at the membrane surface,
with the result that low analyte concentrations require longer times
to attain the steady-state signal.[20]Potentiometric selectivity coefficients were determined for Mg2+, Ca2+, Li+, Na+, and K+ ions using the well-known separate solution method.[21] Notably, the apparent values were calculated
since this method assumes the slope in the calibration graph of the
interfering cation equal to that of the primary one, which was not
the case. In addition, the interfering cations were tested in order
of their lipophilicity and with NH4+ (the most
preferred cation) in the last place to avoid bias in the values.[21] The potentials provided by 10–3 activity of each interfering cation and for NH4+ were used to calculate the apparent logarithmic selectivity coefficients:
log KNHpot = −4.15 ± 0.22,
−4.30 ± 0.16, −2.50 ± 0.12, −2.20 ±
0.28, and −0.98 ± 0.12 for X = Mg2+, Ca2+, Li+, Na+, and K+ (n = 3), respectively. The K+ ion was found to
give the strongest interference, showing a similar value of the logarithmic
selectivity coefficient to those previously reported for ISEs containing
nonactin as the ionophore (from −0.42 to −1.8).[7]Considering typical K+ concentrations
in freshwater
and seawater (0.1 and 10 mM, respectively) together with the calculated
logarithmic coefficient, the minimum LODs that could be reached for
NH4+ are estimated to be ca. 10–5 and 10–3 M (calculated
as [KNHpot·cK]). As a result, the potentiometric cell
developed in this work is unsuitable for seawater analysis and can
only be applied to freshwater samples with NH4+ exceeding 10–5 M, i.e., heavily
contaminated samples. To overcome this limitation, the tandem potentiometric–IC
is following investigated, wherein the effective separation of NH4+ from any interfering cations present is expected.
Coupling the Potentiometric Cell with the Ion Chromatography
System
The separation capability of the proposed tandem potentiometry–IC
was first evaluated with a mixture containing equal activities (1.0
× 10–3) of Li+, Na+,
K+, and NH4+. The black lines in Figure a and b indicate
the chromatograms observed with both the conductivity detector and
one of the NH4+-selective electrodes in the
potentiometric cell (sample volume of 10 μL, flow rate of 0.9
mL min–1) for the mentioned mixture. Peak identification
was performed from separate solutions of each cation (data not shown).
Figure 3
(a) Conductivity
chromatograms at increasing NH4+ activities
and 1 × 10–3 activity
of the other cations (10 μL volume, 0.9 mL min–1). (b) Potentiometric chromatograms at increasing NH4+ activity and 1 × 10–3 activity of
the rest of the cations (10 μL volume, 0.9 mL min–1). (c) Potentiometric chromatograms with 10 and 20 μL injected
volume of 1 × 10–3 NH4+ activity (0.9 mL min–1). (d) Averaged calibration
graphs (n = 3). (e) Potentiometric chromatograms
with 1 × 10–3 NH4+ activity
at 0.5, 0.7, and 0.9 mL min–1 (10 μL volume).
(f) Averaged calibration graphs (n = 3). Background:
2.5 × 10–3 mol L–1 nitric
acid.
(a) Conductivity
chromatograms at increasing NH4+ activities
and 1 × 10–3 activity
of the other cations (10 μL volume, 0.9 mL min–1). (b) Potentiometric chromatograms at increasing NH4+ activity and 1 × 10–3 activity of
the rest of the cations (10 μL volume, 0.9 mL min–1). (c) Potentiometric chromatograms with 10 and 20 μL injected
volume of 1 × 10–3 NH4+ activity (0.9 mL min–1). (d) Averaged calibration
graphs (n = 3). (e) Potentiometric chromatograms
with 1 × 10–3 NH4+ activity
at 0.5, 0.7, and 0.9 mL min–1 (10 μL volume).
(f) Averaged calibration graphs (n = 3). Background:
2.5 × 10–3 mol L–1 nitric
acid.The order of elution for the cations
was Li+ (5.8 min),
Na+ (8.4 min), NH4+ (9.6 min), and
K+ (14.4 min), with the retention times being ca. 4 s longer for the potentiometric detector than for the conductivity
detector, because of their position in the experimental setup. While
the chromatogram provided by the conductivity detector yields similar
peak areas for the four cations, the chromatogram provided by the
potentiometric cell displayed a higher potential readout for NH4+ than for the other cations (Table S1). Indeed, the peaks for Li+ and Na+ were very weak. Furthermore, when the NH4+ content in the sample was decreased to 1.0 × 10–6 activity, while maintaining the levels for the rest
of the cations (1.0 × 10–3 activity), the peaks
for Li+, Na+, and K+ did not change.
However, the peak height for NH4+ decreased
in both the potentiometric and conductivity chromatograms (Figure a and b and Table S2).Then, the peak for the 1 ×
10–6 NH4+ activity was well-observed
with the potentiometric detector,
but it was difficult to be identified and/or quantified by the conductivity
detector. Notably, in the case of the potentiometric detector, the
returning to the base signal in the NH4+ peak
appears to be slower for increasing activity, which seems to not disturb
the K+ signal, as demonstrated with the overlapping K+ peaks in Figure b. Importantly, this behavior is not expected to influence
the accuracy of the NH4+ analysis even at high
activity/concentrations (around mM), since the signal does return
to the initial baseline at the end of the chromatogram.Next,
the effect of the injected sample volume and flow rate on
the chromatograms was investigated. For this purpose, samples with
increasing activity of NH4+ (from 1 × 10–6 to 1 × 10–3, n = 3) were analyzed using two different sample volumes (10 and 20
μL, flow rate of 0.9 mL min–1) and three flow
rates (0.5, 0.7, and 0.9 mL min–1, sample volume
of 10 μL). Figure c–f compare the chromatograms observed with the same electrode
at 1 × 10–3 NH4+ activity
under different volume and flow rate conditions, together with the
averaged calibration graphs (n = 3) at increasing
NH4+ activity.In the case of the injected
sample volume, and considering a 10–3 NH4+ activity (Figure c), the peak was higher (81.7
mV versus 158.0 mV) and wider (time difference at 5% of the peak maximum
of 2.4 min versus 4.2 min) when the volume was increased from 10 to
20 μL. Also, the retention time was slightly longer (9.8 min
versus 10.3 min). This trend appeared for all tested NH4+ activities (Figure S1). In
principle, an increase in peak height, width, and retention time is
expected with increasing sample volume as the analyte plug injected
into the detector increases in length. Consequently, the separation
efficiency was slightly better with the 10 μL sample (Table S3: number of plates of 2544 versus 1146
for NH4+ activity > 10–5).
However, the average calibration graph (n = 3) with
a 20 μL sample volume displayed a hyper-Nernstian slope (62.1
mV versus 57.3 mV for 20 and 10 μL, respectively, Figure d) as well as a lower LOD (1
× 10–7 versus 6 × 10–6 NH4+ activity, Figure d). The peaks at lower NH4+ activities appear slightly more distinct with the 20 μL
volume than the 10 μL volume (Figure S1).Concerning the flow rate, we found increasing retention
times (9.5,
11.9, and 16.4 min) and widths (1.7, 2.2, and 2.9 min) for the NH4+ peak with decreasing flow rate (Figure e), corresponding to a decrease
in the separation efficiency (number of theoretical plates 1584, 2254,
and 3040 for 0.5, 0.7, and 0.9 mL min–1, respectively, Table S4). Nevertheless, we did not find any
significant difference in the averaged calibration parameters (n = 3) of the different flow rates (Figure f; the corresponding chromatograms are provided
in Figure S2). Given these results, a sample
volume of 10 μL and a flow rate of 0.9 mL min–1 were selected for further experiments as a compromise between separation
efficiency, calibration parameters, and analysis time.
Analytical
Performance of the Tandem Potentiometry–Ion
Chromatography
The sensitivity, LRR, LOD, limit of quantification
(LOQ), and reproducibility were investigated at the optimized experimental
conditions. Potentiometric chromatograms at increasing NH4+ activity (10–7–10–3) are presented in Figure a. The averaged calibration graph for the potentiometric cell
(n = 3 electrodes) revealed a slope of 59.5 ±
1.2 mV dec–1 within the LRR (from 3.0 × 10–5 to 10–3 NH4+ activity) and a LOD of 2.5 × 10–6 (Figure b). Pursuing the
detection of lower NH4+ levels, we investigated
a transformation for the calibration graph via linearization of the
entire logarithmic response.[22] We used
the linear fitting of the (10–1) magnitude (where E refers to the
potential and S to the slope in the LRR when plotting
the potential versus log a) against the NH4+ concentration (Figure c), as previously
reported for other chromatographic–potentiometric tandem systems.[22]
Figure 4
(a) Chromatograms at increasing NH4+ activity.
(b) Corresponding averaged calibration graph (n =
3). (c) Linearization of the averaged calibration graph (2.5 ×
10–3 mol L–1 nitric acid, 10 μL
sample volume, flow rate of 0.9 mL min–1).
(a) Chromatograms at increasing NH4+ activity.
(b) Corresponding averaged calibration graph (n =
3). (c) Linearization of the averaged calibration graph (2.5 ×
10–3 mol L–1 nitric acid, 10 μL
sample volume, flow rate of 0.9 mL min–1).Following this approach, an expanded LRR from 10–6 to 10–3 mol L–1 NH4+ with a sensitivity of 76.6 ± 2.2
L mmol –1 (R2 = 0.9986)
was found. The LOD and
LOQ were calculated as the concentration corresponding to a signal-to-noise
ratio of three and ten times, respectively, yielding values of 3.0
× 10–7 and 1.0 × 10–6 mol L–1 NH4+. Under the
same experimental conditions, the LOD and LOQ with the conductivity
detector were calculated as 1.0 × 10–6 and
3.0 × 10–6 mol L–1 (Figure S3). The reproducibility was evaluated
by the triplicate injection of 3.0 × 10–6,
1.0 × 10–5, and 3.0 × 10–4 mol L–1 NH4+ measured with
three NH4+-selective electrodes and the injection
of 1.0 × 10–5 mol L–1 NH4+ over four consecutive days. Intraelectrode reproducibility
for the peak potentials revealed a variation coefficient < 5%.
Interelectrode reproducibility showed a variation coefficient of <9%.
Finally, variations of ca. 0.5% and 7% were observed
for the calibration parameters of the NH4+-selective
electrodes in the second and fourth days of their continuous usage,
respectively.
Analysis of Natural Water Samples (See Table S5 for Sample Identification)
Recovery studies were
accomplished with a river water sample (R1.0, Table S6) spiked with 1.0 × 10–5 (R1.1)
and 5.0 × 10–5 (R1.2) mol L–1 NH4+ concentrations. Each sample was analyzed
in triplicate using both the potentiometric and conductivity detectors
(Table and Table S6). The recovery percentages were acceptable
for both detectors, showing the appropriate accuracy of the potentiometry–IC
methodology.
Table 1
Quantification of Ammonium Levels
in Different Natural Water Samples Using the Potentiometry–IC
and Conductivity–IC Setupsa
Potentiometry–IC
Conductivity–IC
Sample
μmol L–1
RSD
μmol L–1
RSD
%Diff
R1.0
5.5
2.3
4.2
7.9
31
R1.1
13.0
2.8
14.0
1.2
7
R1.2
56.5
1.8
57.0
0.9
1
R2
2.6
8.4
ND
R3
5.2
5.3
6.7
8.4
22
R4
4.7
3.1
ND
L1
3.6
9.8
ND
L2
4.0
7.1
ND
L3
12.5
3.6
12.8
7.5
2
S1
4.2
8.3
ND
S2
9.9
3.7
8.9
9.3
6
S3
4.0
6.7
ND
S4
23.7
2.6
22.9
7.4
3
ND = nondetectable. %Diff = percentage
of difference between the results provided by potentiometry and conductivity.
RSD = relative standard deviation.
ND = nondetectable. %Diff = percentage
of difference between the results provided by potentiometry and conductivity.
RSD = relative standard deviation.Furthermore, the NH4+ content
in ten water
samples from different locations in Portugal, Sweden, and Spain (Table S5) was estimated. Some representative
chromatograms are shown in Figure . Notably, the NH4+ peak was
visible in all samples, approaching the level of noise (and potentiometric
baseline) if the NH4+ content was close to the
micromolar level. Peaks for K+ and Na+ were
also observed in all the tested samples, presenting higher levels
in the seawater samples. Nevertheless, there was no detected overlap
with the NH4+ peak that could adversely affect
its detection. The quantification of NH4+ was
possible in all samples using the potentiometric detector, whereas
the conductivity detector only detected NH4+ concentrations higher than 5 × 10–6 mol L–1. In those samples where a comparison was feasible,
the results from both detectors agreed better at higher NH4+ concentrations. Overall, the potentiometric cell could
detect and quantify lower NH4+ concentrations
than the conductivity detector. The RSDs provided by both techniques
were similar and always <10%.
Figure 5
Chromatograms observed for two water samples: freshwater (river)
and seawater.
An interesting advantage of
the potentiometric detector developed
here is the possibility of simultaneously combining ISEs that are
selective for different ions in the same cell for multi-ion detection.
Accordingly, NH4+-, Na+-, and K+-selective electrodes were incorporated into the potentiometric
cell as a preliminary proof-of-concept. The chromatograms and the
corresponding calibration graphs obtained by using solutions containing
the three cations at the same (increasing) concentrations are displayed
in Figure . As observed,
all the cations can be selectively detected by the corresponding electrode
with no interference from the other cations. In more detail, the Na+-selective electrode presented a negligible response for NH4+ and a very low response for K+, whereas
the K+-selective electrode displayed a negligible response
for Na+ and some response for NH4+.
Figure 6
Chromatograms at increasing NH4+, Na+, and K+ activity provided by the
(a) ammonium-selective
electrode, (b) sodium-selective electrode, and (c) potassium-selective
electrode. Insets: Linearized calibration graphs (2.5 × 10–3 mol L–1 nitric acid, 10 μL
sample volume, flow rate of 0.9 mL min–1).
Chromatograms observed for two water samples: freshwater (river)
and seawater.Chromatograms at increasing NH4+, Na+, and K+ activity provided by the
(a) ammonium-selective
electrode, (b) sodium-selective electrode, and (c) potassium-selective
electrode. Insets: Linearized calibration graphs (2.5 × 10–3 mol L–1 nitric acid, 10 μL
sample volume, flow rate of 0.9 mL min–1).Regarding the calibration graph, in the case of
Na+ and
K+ electrodes, the LRR of the typical potentiometry plot
(i.e., EMF versus logarithmic activity) was found
to include the levels expected in water samples (0.1–100 mM),
and hence, implementing the linearization approach is not necessary
(see Figure S4 for the calibration graphs
and Table S7 for the corresponding calibration
parameters). However, aiming at only one type of calibration, we used
the linearized version of the calibration graph (as described above)
for each of the three cations (NH4+, Na+, and K+) with the corresponding electrode. The
calibration parameters in the LRR are collected in Table S8. Linear regression lines were obtained for NH4+, Na+, and K+ in the concentration
range from 10–6 to 10–3 with sensitivities
of 61.2 ± 2.5 (R2 = 0.9983), 24.9
± 0.4 (R2 = 0.9979), and 33.1 ±
0.6 (R2 = 0.9973), respectively. The LOD
and LOQ were calculated as the concentration corresponding to a signal-to-noise
ratio of three and ten times, respectively, producing values of 3.0
× 10–7 (LOD) and 1.0 × 10–6 (LOQ) mol L–1 for the three cations.The
river samples R1.0, R1.1, and R1.2 were analyzed with the multi-cation
potentiometric cell, revealing appropriate recovery percentages ranging
from 81 to 102% for the potentiometric detector and from 73 to 140%
for the conductivity one (Table ). Considering the results provided by the conductivity
detector, an excellent agreement was found, with a percentage of difference
always lower than 18%, except for the R1.0 sample, which presented
a 26% difference (Table ). These results confirmed the great potential of the potentiometry–IC
tandem for multi-ion detection.
Table 2
Quantification of
Ammonium, Sodium,
and Potassium Levels in Different Natural Water Samples Using the
Potentiometry–IC and Conductivity–IC Setupsa
Potentiometry–IC
Conductivity–IC
Sample
Cation
μmol L–1
RSD
%Rec
μmol L–1
RSD
%Rec
%Diff
R1.0
NH4+
4.1
2.6
5.2
8.8
26
Na+
140.3
0.4
148.4
1.0
6
K+
13.2
3.6
12.6
2.9
5
R1.1
NH4+
12.2
3.6
82
14.4
4.4
92
18
Na+
149.8
3.1
95
162.4
1.3
140
8
K+
21.3
6.1
81
19.9
5.0
73
7
R1.2
NH4+
47.8
3.6
88
53.8
6.7
97
12
Na+
191.1
4.0
102
206.3
2.5
116
8
K+
57.6
4.1
89
65.6
8.1
106
14
%Rec = percentage of recovery.
%Diff = percentage of difference between the results provided by potentiometry
and conductivity.
%Rec = percentage of recovery.
%Diff = percentage of difference between the results provided by potentiometry
and conductivity.
Analysis of
the Developed Tandem Potentiometry–Ion Chromatography
Method with Respect to the State-of-the-Art
To the best of
our knowledge, the first attempts to coupling potentiometry to IC
for (multi)ion detection used inner-filing solution-type ISEs.[23−26] While the design of the electrode hindered the miniaturization of
the entire system, these works demonstrated the great potential of
potentiometry to detect ions after chromatographic separation, as
it was sometimes superior to the analytical performance of the conductivity
detector.[27,28] It is worth mentioning that three main strategies
have been followed in terms of the membrane composition: (i) a membrane
containing a single ionophore that is selective for only one ion;
(ii) a multi-ionophore membrane that is selective for at least two
ions; and (iii) a membrane with a nonselective profile.[23−26,29]The concept of a multi-ionophore
membrane was translated to the solid-contact electrode type by Lee
et al., who demonstrated the simultaneous analysis of K+, NH4+, Na+, and Ca2+ with a membrane containing four ionophores, one for each cation.[30] However, the separation between NH4+, Na+, and K+ peaks was incomplete,
likely due to the complex cross-selectivity profile normally found
for this type of membrane. For NH4+ in particular,
the observed LOD was close to 10–5 M, which is more
than one order of concentration higher than that presented in this
paper. Notably, NH4+ detection in real samples
was not demonstrated.Five years later, Isildak et al. presented
tubular-shaped membrane
electrodes with nonselective profiles for either cation or anion detection
after chromatographic separation.[27,31] Their cation
detector was able to identify Na+, NH4+, K+, Rb+, Cs+, and Tl+ with well-separated peaks but with what appears to be a continuously
drifting baseline, statement made based on the reported figures. The
electrode was used in conjunction with a double-junction calomel reference
electrode, which was not a solid-contact type because this technology
was not yet common. For NH4+, the LOD was ca.
6 μM, which is again higher than that achieved in this paper
and hence unsuitable for NH4+ detection in any
water sample. The NH4+ concentration in two
water samples (river and seawater) was calculated to be close to 20
μM, while the cation was not detectable in a tap water sample.
Unfortunately, the results in real water samples were not validated
with parallel measurements using a gold standard technique.Shen et al. have reported chromatographic separation coupled to
a solid-contact sensor array (i.e., conducting wires
covered by the membrane) in which each electrode was selective for
one cation: Na+, NH4+, K+, Mg2+, and Ca2+.[32,33] The authors achieved a constant baseline without any noticeable
drift by including a 1 μM concentration of each cation in the
eluent solution (i.e., the background for the potentiometric
measurements). The sensors presented LRRs from 0.05 to 1 mM. The ammonium
content in some synthetic samples and a hydroponic solution was analyzed.
However, the LRR did not allow the detection of NH4+ in a series of commercial and natural water samples.The most recent paper reporting on NH4+ detection
in mixtures dates from 2016 and is based on a solid-contact electrode
(copper wire) modified with a carbon composite containing the membrane
components.[34] In this case, the membrane
presented a nonselective profile, and after chromatographic separation,
the NH4+ and K+ peaks appear with
a high degree of overlap; thus, the LRR ranged between 5 × 10–5 and 10–2 M, which is not suitable
for NH4+ detection in water samples. However,
it is suitable for K+ detection.Overall, the potentiometry–IC
approach developed in the
present paper is proposed within a state-of-the-art context that lacks
a functional solution to detect NH4+ together
with other cations in any water sample. The technology put forward
regarding the potentiometry–chromatography system that can
additionally include a conductivity detector is a unique method to
validate all the measurements and demonstrates the advantages of the
potentiometric readout over conductometry. The performance of the
potentiometric sensors used to detect NH4+ (but
also K+ and Na+) is superior to other sensors
reported at the time of writing. Importantly, the potentiometric cell
presents huge versatility for the detection of any ion.
Conclusions
We demonstrate the necessity of the tandem potentiometry–IC
system together with the linearization of the potentiometric response
to address NH4+ detection at micromolar levels
in any water sample where other cations (such as K+ and
Na+) are present at higher levels. To pursue the decentralization
of such analytical determination in diverse water environments, a
miniaturized potentiometric detector based on a multielectrode flow
cell was fabricated and characterized to provide the best analytical
performances after in-line coupling with chromatographic separation.
Advantageously, the potentiometric detector can be equipped with three
similar ISEs to obtain reproducibility in the NH4+ quantification; or three electrodes, each selective for different
ions, for multi-ion detection within the same sample. Preliminary
results revealed the accurate and simultaneous detection of NH4+, Na+, and K+ in river samples.
Soon, the implementation of the potentiometry–IC analytical
concept in true in situ measurements is foreseen via potentiometric-ion-chromatography-on-a-chip technology.
Simultaneous multi-ion detection is expected to be feasible in any
kind of water (even such a complex matrix as seawater) and with high
temporal and spatial resolution, thanks to further implementation
in water platforms considering microfluidics.
Authors: J Gallardo; S Alegret; R Muñoz; M De-Román; L Leija; P R Hernández; M del Valle Journal: Anal Bioanal Chem Date: 2003-07-19 Impact factor: 4.142
Authors: Devin Daems; Alexander L N van Nuijs; Adrian Covaci; Ezat Hamidi-Asl; Guy Van Camp; Luc J Nagels Journal: Biomed Chromatogr Date: 2014-12-16 Impact factor: 1.902
Authors: Rohini Athavale; Ilga Kokorite; Christian Dinkel; Eric Bakker; Bernhard Wehrli; Gastón A Crespo; Andreas Brand Journal: Anal Chem Date: 2015-12-01 Impact factor: 6.986