| Literature DB >> 25516367 |
Devin Daems1,2, Alexander L N van Nuijs3, Adrian Covaci3, Ezat Hamidi-Asl1, Guy Van Camp2, Luc J Nagels1.
Abstract
The analytical methods which are often used for the determination of cocaine in complex biological matrices are a prescreening immunoassay and confirmation by chromatography combined with mass spectrometry. We suggest an ultra-high-pressure liquid chromatography combined with a potentiometric detector, as a fast and practical method to detect and quantify cocaine in biological samples. An adsorption/desorption model was used to investigate the usefulness of the potentiometric detector to determine cocaine in complex matrices. Detection limits of 6.3 ng mL(-1) were obtained in plasma and urine, which is below the maximum residue limit (MRL) of 25 ng mL(-1). A set of seven plasma samples and 10 urine samples were classified identically by both methods as exceeding the MRL or being inferior to it. The results obtained with the UPLC/potentiometric detection method were compared with the results obtained with the UPLC/MS method for samples spiked with varying cocaine concentrations. The intraclass correlation coefficient was 0.997 for serum (n =7) and 0.977 for urine (n =8). As liquid chromatography is an established technique, and as potentiometry is very simple and cost-effective in terms of equipment, we believe that this method is potentially easy, inexpensive, fast and reliable.Entities:
Keywords: UPLC; cocaine; plasma; potentiometry; urine
Mesh:
Substances:
Year: 2014 PMID: 25516367 DOI: 10.1002/bmc.3400
Source DB: PubMed Journal: Biomed Chromatogr ISSN: 0269-3879 Impact factor: 1.902