The effect of material assignment in nasal cavity on dose calculation for nasopharyngeal carcinoma (NPC) using Acuros XB.

Michael L M Cheung1, Vivian U Y Chow1, Monica W K Kan1,2, Anthony T C Chan1,2.   

Abstract

PURPOSE: To evaluate the effect of material assignment in nasal cavity on dose calculation for the volumetric modulated arc therapy (VMAT) of nasopharyngeal carcinoma (NPC) using Acuros XB (AXB) algorithm.
METHODS: The VMAT plans of 30 patients with NPC were calculated using AXB with material auto-assignment of nasal cavity to lung and reassignment to air respectively. The doses to the planning target volumes (PTVs) overlapping with nasal cavity with material auto-assignment of lung (AXB_Lung) were compared to the values obtained when nasal cavity was reassigned to air (AXB_Air) under the dose-to-medium (Dm ) reporting mode of AXB.
RESULTS: For dose calculated under AXB_Lung, the D98% , D2% , and Dmean of the PTV69.96 _Air Cavity (PTV of prescription dose 69.96 Gy overlapping with nasal cavity) were on average 16.1%, 1.6%, and 8.6% larger than that calculated under AXB_Air, respectively. Up to 19.5% difference in D98% , 3% difference in D2% , and 11.2% difference in Dmean were observed in the worst cases for PTV69.96 . Similar trend was observed for the PTV5940 _Air Cavity, in which the D98% , D2% , and Dmean calculated under AXB_Lung were on average 14.7%, 2.5%, and 10.2% larger than that calculated under AXB_Air, respectively. In the worst cases, the difference observed in D98% , D2% , and Dmean could be up to 17.7%, 4.5%, and 12.7%, respectively.
CONCLUSIONS: Significant dose difference calculated by AXB between the material assignment of lung and air in nasal cavity for NPC cases might imply the possibility of underdosage to the PTVs that overlap with inhomogeneity. Therefore, attention should be put to ensure that accurate material assignment for dose calculation under AXB such that optimal dosage was given for tumor control.
© 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.

Entities:  

Keywords:  Acuros XB algorithm; dose calculation; inhomogeneity; material assignment; nasopharyngeal carcinoma

Mesh:

Year:  2022        PMID: 35699203      PMCID: PMC9359034          DOI: 10.1002/acm2.13698

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.243


INTRODUCTION

Volumetric modulated arc therapy (VMAT) is one of the most common treatment modalities for nasopharyngeal carcinoma (NPC) due to its ability to produce highly conformal dose distributions to the target while minimizing doses to organs‐at‐risk. As the nasopharyngeal region is surrounded by heterogeneous medium such as air cavities, the accuracy of dose calculation for these NPC cases is significantly affected by the ability of algorithms to account for electron transport in air–tissue interface. Anisotropic analytical algorithm (AAA) and collapsed cone convolution (CCC) are widely used convolution/superposition algorithms implemented in commercial treatment planning system. They apply the simplified density scaling of the Monte Carlo (MC)‐derived dose kernels to account for the presence of inhomogeneities so that the secondary electron transport is only modeled macroscopically. Previous studies found that AAA and CCC significantly overpredict the dose near air–tissue interfaces. A more advanced dose calculation algorithm known as Acuros XB (AXB) that can achieve comparable accuracy with MC has been implemented in Eclipse treatment planning system (Varian Medical Systems, Palo Alto, CA, USA). It is a deterministic solver of linear Boltzmann transport equation (LBTE) describing the macroscopic behavior of ionizing particles as they travel through and interact with matter. By solving the LBTE, the electron  fluence is obtained and the dose is generated by using macroscopic electron energy deposition cross sections and the density of materials. In AXB, two dose reporting options are provided, including dose‐to‐water (D w) and dose‐to‐medium (D m). Both options are the same for AXB transport calculation, in which the electron fluence is calculated based on the material properties of crossing media. The main difference arises from post‐processing step, in which the electron fluence is multiplied by a water‐based flux‐to‐dose response function and medium‐based flux‐to‐dose response function for D w and D m, respectively. , , Therefore, the material composition of voxels in computed tomography (CT) images is required for dose calculation. The material is assigned automatically by the conversion of Hounsfield units to mass density using CT calibration curve, followed by looking up the material in the Varian system database. For those voxels with density larger than 3.0 g/cm3, manual material assignment is required in Eclipse to prevent the inaccurate assignment to high‐density materials such as metallic prosthesis. , Several publications have reported that AXB provides an accurate dose calculation in heterogeneous environment. , , , However, no study has been conducted to evaluate the effect of material assignment on dose calculation in the presence of air cavity in nasopharyngeal region using real clinical cases. Therefore, the aim of this study was to assess the dosimetric impact of material assignment to nasal cavity under AXB using real clinical NPC cases.

METHODS

Contouring and prescription of target volumes

A total of 30 NPC VMAT patients with treatment site overlapping nasal cavity were selected for retrospective analysis. All plans were generated using a 6‐MV beam and modulated with a 120 multi‐leaf collimator (MLC) from TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, CA, USA). The target volume of each patient was defined by the oncologist using 3‐mm‐thick axial CT images. The planning target volume (PTV) included the abnormal soft tissue mass with an addition of 5‐mm margin to account for organ movement and patient setup uncertainty. The mean PTV was 313.2 cm3 (ranging from 104.4 to 804.1 cm3) and 921.9 cm3 (ranging from 504.7 to 1590 cm3) for prescription dose 69.96 Gy (PTV6996) and 59.40 Gy (PTV5940), respectively. To evaluate the dosimetric difference of material assignment to nasal cavity, three additional structures were contoured, including the whole nasal cavity (Air Cavity), PTV of prescription dose 69.96 Gy overlapping with nasal cavity (PTV6996_Air Cavity), and PTV of prescription dose 59.40 Gy overlapping nasal cavity (PTV5940_Air Cavity), as shown in Figure 1.
FIGURE 1

Contouring for the dosimetric evaluation of material assignment in nasal cavity. (Left) PTV6996 was in red contour, and its overlapping part with nasal cavity was PTV6996_Air Cavity. (Right) The brown contour was PTV5940, and its overlapping part with nasal cavity was PTV5940_Air Cavity.

Contouring for the dosimetric evaluation of material assignment in nasal cavity. (Left) PTV6996 was in red contour, and its overlapping part with nasal cavity was PTV6996_Air Cavity. (Right) The brown contour was PTV5940, and its overlapping part with nasal cavity was PTV5940_Air Cavity.

Treatment planning

The treatment plans were generated using treatment planning system Eclipse version 13.6 (Varian Medical Systems, Palo Alto, CA, USA). At least 95% of the PTVs received the prescription dose with the optimization criteria followed NRG‐HN001 protocol. Each treatment plan was then calculated using AXB version 13.6.23 (Varian Medical Systems, Palo Alto, CA, USA) with automatic material assignment of nasal cavity to low‐density lung in Eclipse. After that, the Air Cavity was manually assigned to the material of air, and the plans were recalculated with the identical treatment parameters such as monitor unit and MLC setting using AXB. In present study, AXB using D m option with 2.5 mm dose grid resolution was used.

Dosimetric evaluation

The doses PTV6996_Air Cavity and PTV5940_Air Cavity with material auto‐assignment of lung (AXB_Lung) were compared to the values obtained when nasal cavity was reassigned to air (AXB_Air). The D mean, D 2% (minimum dose received by 2% of PTV volume), and D 98% (minimum dose received by 98% of PTV volume) of the Air Cavity, PTV6996_Air Cavity, and PTV5940_Air Cavity were compared. Statistical analyses were performed using SPSS Statistics version 17.0 (SPSS, Inc., Chicago, IL, USA) in this study. Paired t‐test was conducted to investigate if there was significant difference in the dose calculation. The statistical test was two‐sided, and p‐value <0.05 was considered statistically significant.

RESULTS

The comparison between doses received by Air Cavity, PTV6996_Air Cavity, and PTV5940_Air Cavity under AXB_Air and AXB_Lung was shown in Table 1. For dose calculated under AXB_Lung, the D 98%, D 2%, and D mean of the PTV6996_Air Cavity were on average 16.1%, 1.6%, and 8.6% larger than that calculated under AXB_Air, respectively. Up to 19.5% difference in D 98%, 3% difference in D 2%, and 11.2% difference in D mean were observed in the worst cases for PTV6996_Air Cavity.
TABLE 1

Comparison of dose calculated by Acuros XB (AXB) under AXB_Air and AXB_Lung

Dose difference (%)
StructureParameters (Gy)Mean ± SDMinMax p‐Value*
Air Cavity D mean 8.7 ± 1.36.012.8<0.001**
D 2% 3.5 ± 1.67.10.5<0.001**
D 98% 2.4 ± 1.45.50.5<0.001**
PTV6996_Air Cavity D mean 8.6 ± 1.82.211.2<0.001**
D 2% 1.6 ± 0.70.23.0<0.001**
D 98% 16.1 ± 2.74.819.5<0.001**
PTV5940_Air Cavity D mean 10.2 ± 1.46.912.7< 0.001**
D 2% 2.5 ± 1.10.34.5< 0.001**
D 98% 14.7 ± 2.44.017.7< 0.001**

The p‐value was calculated using paired t‐test.

The p‐value has statistical significance of difference (p < 0.05).

Comparison of dose calculated by Acuros XB (AXB) under AXB_Air and AXB_Lung The p‐value was calculated using paired t‐test. The p‐value has statistical significance of difference (p < 0.05). Similar trend was observed for the PTV5940_Air Cavity, in which the D 98%, D 2%, and D mean calculated under AXB_Lung were on average 14.7%, 2.5%, and 10.2% larger than that calculated under AXB_Air, respectively. In the worst cases, the difference observed in D98%, D2%, and D mean could be up to 17.7%, 4.5%, and 12.7%, respectively. Paired t‐test showed that all the dose differences with material assignment of lung and air were statistically significant (p < 0.05).

DISCUSSION

The dosimetric effect owning to inhomogeneities such as air cavity is one of the major concerns for NPC cases, and AXB was proved to be able to achieve comparable accuracy as the MC method in a heterogeneous medium. Although AXB is capable of providing accurate dose calculation, its accuracy depends on the preciseness of material assignment in AXB. , The material composition of each voxel in CT image was assigned automatically from the conversion of mass density from CT value based on the CT calibration curve, followed by looking up the material from material data table in AXB. Fogliata et al. showed that the dose computed inside the air material layer by AXB version 11 presents much better agreement with MC, due to the inclusion of air material assignment that was considered as lung in version 10. , , In AXB version 13.6.23, the automatic assignment of materials includes human material such as “Air,” “Lung,” “Adipose Tissue,” “Muscle, Skeletal,” “Cartilage,” and “Bone.” The density ranges for each material are slightly overlapping as shown in Table 2.
TABLE 2

Look‐up table in Acuros XB (AXB) version 13.6.23 for automatic material assignment

MaterialsDensity range (g/cm3)
Air0.000–0.020
Lung0.011–0.624
Adipose tissue0.554–1.001
Muscle, skeletal0.969–1.093
Cartilage1.056–1.600
Bone1.100–3.000
Look‐up table in Acuros XB (AXB) version 13.6.23 for automatic material assignment In present study, the nasal cavity region was contoured and auto‐assigned with the material of “Lung” for dose calculation in Eclipse. As the nasal cavity is an air‐filled structure, the nasal cavity region was reassigned to “Air” and recalculated using AXB. Our study showed that the calculated D mean and D 98% of PTV6996_Air Cavity and PTV5940_Air Cavity under AXB_Lung were much higher than that under AXB_Air. It implied that the dose reported for PTV6996_Air Cavity under AXB_Lung might be overestimated for an air‐filled nasal cavity, leading to the possibility of underdosage or insufficient coverage to PTV6996. According to AAPM Task Group Report 105, 5% change in dose can result in 10%–20% change in tumor control probability or up to 20%–30% change in normal tissue complication probability if the prescribed dose is within the steepest region of dose–effect curves. , Therefore, attention should be put to ensure the accuracy of material assignment for dose calculation in AXB such that optimal dosage was given for tumor control. Apart from the inherent inhomogeneity (i.e., air cavity) in NP region, imaging artifacts in planning CT and implants inside the patients with density less than 3.0 g/cm3 might also cause the misrepresentation of density information and hence leading to misassignment of materials for AXB dose calculation. Review of material assignment is therefore recommended when using AXB for dose calculation under D m.

CONCLUSION

Although the dose calculation of AXB was superior to other superposition/convolution methods in heterogeneous media, its accuracy is greatly affected by the material assignment. Therefore, a review of automatic material assignment is therefore recommended for accurate dose calculation such that the PTVs are adequately covered with optimal dose for tumor control.

CONFLICT OF INTEREST

The authors declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research reported.

AUTHOR CONTRIBUTION

Michael L. M. Cheung was responsible for idea conceptualization, research design, literature review, data collection, data analysis, and manuscript writing. Vivian U. Y. Chow was responsible for research design, data collection, data analysis, and manuscript editing. Monica W. K. Kan and Anthony T. C. Chan were responsible for research guidance and supervision.
  14 in total

Review 1.  Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning.

Authors:  Indrin J Chetty; Bruce Curran; Joanna E Cygler; John J DeMarco; Gary Ezzell; Bruce A Faddegon; Iwan Kawrakow; Paul J Keall; Helen Liu; C M Charlie Ma; D W O Rogers; Jan Seuntjens; Daryoush Sheikh-Bagheri; Jeffrey V Siebers
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

2.  Dosimetric impact of using the Acuros XB algorithm for intensity modulated radiation therapy and RapidArc planning in nasopharyngeal carcinomas.

Authors:  Monica W K Kan; Lucullus H T Leung; Peter K N Yu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-10-03       Impact factor: 7.038

3.  Verification and dosimetric impact of Acuros XB algorithm on intensity modulated stereotactic radiotherapy for locally persistent nasopharyngeal carcinoma.

Authors:  Monica W K Kan; Lucullus H T Leung; Peter K N Yu
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

4.  Experimental verification of the Acuros XB and AAA dose calculation adjacent to heterogeneous media for IMRT and RapidArc of nasopharygeal carcinoma.

Authors:  Monica W K Kan; Lucullus H T Leung; Ronald W K So; Peter K N Yu
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

5.  Modeling AeroForm tissue expander for postmastectomy radiation therapy.

Authors:  Elaine Dziemianowicz; Stephen J Gardner; Karen Chin Snyder; Ning Wen; Eleanor M Walker; Correen Fraser; Anne Reding; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2019-07-22       Impact factor: 2.102

6.  The effect on tumour control probability of using AXB algorithm in replacement of AAA for SBRT of hepatocellular carcinoma located at lung-liver boundary region.

Authors:  Michael Lok Man Cheung; Monica Wk Kan; Vanessa Ty Yeung; Darren Mc Poon; Michael Km Kam; Louis Ky Lee; Anthony Tc Chan
Journal:  BJR Open       Date:  2021-11-24

7.  The effect of material assignment in nasal cavity on dose calculation for nasopharyngeal carcinoma (NPC) using Acuros XB.

Authors:  Michael L M Cheung; Vivian U Y Chow; Monica W K Kan; Anthony T C Chan
Journal:  J Appl Clin Med Phys       Date:  2022-06-14       Impact factor: 2.243

8.  A review on the use of grid-based Boltzmann equation solvers for dose calculation in external photon beam treatment planning.

Authors:  Monica W K Kan; Peter K N Yu; Lucullus H T Leung
Journal:  Biomed Res Int       Date:  2013-08-27       Impact factor: 3.411

9.  Dosimetric evaluation of the Acuros XB algorithm for a 4 MV photon beam in head and neck intensity-modulated radiation therapy.

Authors:  Kimiko Hirata; Mitsuhiro Nakamura; Michio Yoshimura; Nobutaka Mukumoto; Manabu Nakata; Hitoshi Ito; Haruo Inokuchi; Yukinori Matsuo; Takashi Mizowaki; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

10.  Dosimetric Impact of Acuros XB Dose-to-Water and Dose-to-Medium Reporting Modes on Lung Stereotactic Body Radiation Therapy and Its Dependency on Structure Composition.

Authors:  Carles Muñoz-Montplet; Rafael Fuentes-Raspall; Diego Jurado-Bruggeman; Sebastià Agramunt-Chaler; Albert Onsès-Segarra; Maria Buxó
Journal:  Adv Radiat Oncol       Date:  2021-05-19
View more
  1 in total

1.  The effect of material assignment in nasal cavity on dose calculation for nasopharyngeal carcinoma (NPC) using Acuros XB.

Authors:  Michael L M Cheung; Vivian U Y Chow; Monica W K Kan; Anthony T C Chan
Journal:  J Appl Clin Med Phys       Date:  2022-06-14       Impact factor: 2.243

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.