Tobias Rainer1, Anna-Sophia Egger2, Ricarda Zeindl1, Martin Tollinger1, Marcel Kwiatkowski2, Thomas Müller1. 1. Institute of Organic Chemistry and Center for Molecular Biosciences (CMBI), Leopold-Franzens University Innsbruck, 6020 Innsbruck, Austria. 2. Institute of Biochemistry and Center for Molecular Biosciences (CMBI), Leopold-Franzens University Innsbruck, 6020 Innsbruck, Austria.
Abstract
Additive manufacturing (3D printing) has greatly revolutionized the way researchers approach certain technical challenges. Despite its outstanding print quality and resolution, stereolithography (SLA) printing is cost-effective and relatively accessible. However, applications involving mass spectrometry (MS) are few due to residual oligomers and additives leaching from SLA-printed devices that interfere with MS analyses. We identified the crosslinking agent urethane dimethacrylate as the main contaminant derived from SLA prints. A stringent washing and post-curing protocol mitigated sample contamination and rendered SLA prints suitable for MS hyphenation. Thereafter, SLA printing was used to produce 360 μm I.D. microcolumn chips with excellent structural properties. By packing the column with polystyrene microspheres and covalently immobilizing pepsin, an exceptionally effective microscale immobilized enzyme reactor (μIMER) was created. Implemented in an online liquid chromatography-MS/MS setup, the protease microcolumn enabled reproducible protein digestion and peptide mapping with 100% sequence coverage obtained for three different recombinant proteins. Additionally, when assessing the μIMER digestion efficiency for complex proteome samples, it delivered a 144-fold faster and significantly more efficient protein digestion compared to 24 h for bulk digestion. The 3D-printed μIMER withstands remarkably high pressures above 130 bar and retains its activity for several weeks. This versatile platform will enable researchers to produce tailored polymer-based enzyme reactors for various applications in analytical chemistry and beyond.
Additive manufacturing (3D printing) has greatly revolutionized the way researchers approach certain technical challenges. Despite its outstanding print quality and resolution, stereolithography (SLA) printing is cost-effective and relatively accessible. However, applications involving mass spectrometry (MS) are few due to residual oligomers and additives leaching from SLA-printed devices that interfere with MS analyses. We identified the crosslinking agent urethane dimethacrylate as the main contaminant derived from SLA prints. A stringent washing and post-curing protocol mitigated sample contamination and rendered SLA prints suitable for MS hyphenation. Thereafter, SLA printing was used to produce 360 μm I.D. microcolumn chips with excellent structural properties. By packing the column with polystyrene microspheres and covalently immobilizing pepsin, an exceptionally effective microscale immobilized enzyme reactor (μIMER) was created. Implemented in an online liquid chromatography-MS/MS setup, the protease microcolumn enabled reproducible protein digestion and peptide mapping with 100% sequence coverage obtained for three different recombinant proteins. Additionally, when assessing the μIMER digestion efficiency for complex proteome samples, it delivered a 144-fold faster and significantly more efficient protein digestion compared to 24 h for bulk digestion. The 3D-printed μIMER withstands remarkably high pressures above 130 bar and retains its activity for several weeks. This versatile platform will enable researchers to produce tailored polymer-based enzyme reactors for various applications in analytical chemistry and beyond.
Apart from rapid prototyping,
3D printing has become a viable method
for manufacturing functional high-performance tools for research and
development. Many exciting applications of additively manufactured
devices in analytical chemistry have been presented in recent years.[1−6] In contrast to the most widely used fused deposition modeling 3D
printing,[2] stereolithography (SLA) can
produce airtight and watertight prints, holding enclosed and complex
fluidic structures on the microscale.[7] Despite
the outstanding performance of SLA 3D printing of photocurable methyl
methacrylate resin for the fabrication of polymethyl methacrylate
microfluidic chips, an open port mass spectrometry (MS) interface[8] as well as a polymer multi-array electrospray
emitter[9] are the only true online electrospray
ionization (ESI)-MS applications where analyte solutions come in direct
contact with the SLA prints. This is not surprising since residual
uncured oligomers and additives tend to leach when in contact with
aqueous and particularly organic solvents. Naturally, such contaminants
are readily detected using a mass spectrometer and greatly interfere
with the detection of low-abundant analytes. It was shown that rinsing
the prints with pure ethanol or UV post-curing of clear resin SLA
prints significantly reduced leaching and therefore enhanced the biocompatibility
of 3D prints.[10] Compounds leaching from
the prints upon contact with aqueous buffers were analyzed by gas
chromatography–MS and hypothesized to be residual monomers
and short-length polymers.[11]Alternatively,
a range of thiol-ene UV-cured resin (TE)-derived
monolithic chips coupled to MS were presented for solid-phase extraction,[12] online protein digestion,[13] or protein hydrogen–deuterium-exchange experiments
(HDX) experiments.[14] While proving highly
effective for their individual tasks, these chips were molded in multi-step
processes and did not support pressures beyond 500 psi (approx. 35
bar). Their pressure tolerance was nonetheless superior to that of
typical polydimethylsiloxane (PDMS) chips. PDMS is the most prominently
used material in microfluidics for modeling open fluidic channels.[15−17] PDMS molds offer high resolution and water tightness but do not
withstand high pressures. Typically, such molded chips are bonded
to a plastic or glass support for sealing the channels. Leakage most
likely occurs at the interconnections or via delamination from the
support, typically at around 100 psi (approx. 7 bar).[18−21] Hence, neither PDMS nor other molded resin chips withstand back
pressures that typically occur in particle-packed microcolumns. Among
other advantages, microparticle supports offer large surface areas
that especially favor applications such as immobilized enzyme reactors
(IMERs). For instance, protease IMERs are powerful tools for fast
and easy sample preparation in bottom-up protein or proteome MS analysis.
Protease IMERs are predominately flow-through devices, which are often
intended for operation in consecutive digestion, trapping, and desalting
steps with subsequent liquid chromatography (LC)-MS/MS analysis. Their
numerous advantages over conventional digestion are well discussed
in the corresponding literature.[22−29] Several open tubular enzyme reactors[30−32] or microfluidic chips
with proteases immobilized on their inner wall[33,34] have been developed. However, solid particle-based IMERs proved
to be superior due to their dramatically larger surface areas and
thus greater enzyme immobilization capacities.[35] A large variety of particle-based[25,36−43] or monolithic[13,44−50] support materials has been utilized for this purpose.Our
goal was to develop a cost-effective and flexible platform
for rapid peptic digestions in the context of protein HDX. In need
of readily customizable, microbore as well as high-pressure-resistant
column housings, we applied high resolution SLA 3D printing. The prints
were rendered fit for application in MS by implementing a comprehensive
washing and curing protocol. For protease immobilization, the benefits
of carboxylated monodisperse polystyrene (PS) particles were exploited
to produce highly effective pepsin μIMERs (here referred to
as IMERs). The 3D-printed IMERs were designed to (i) be operated in
an online LC–MS/MS setup in a plug and play fashion, (ii) withstand
pressures above 100 bar without leaking, (iii) enable temperature
monitoring, and (iv) retain their activity for several weeks. To assess
the performance of the IMERs, we analyzed both recombinantly produced
proteins and complex proteome samples. For all studied proteins, 100%
sequence coverage was achieved, often exceeding 100 unique identifiable
peptides of an average length below 12 amino acid residues. For complex
proteome digestion, the IMER provided dramatically faster and significantly
more efficient digestion compared to bulk digestion. This highlights
the outstanding performance and robustness of the 3D-printed IMER
as the first of its kind stereolithographically produced device for
MS.
Experimental Section
Chemicals and Reagents
See the Supporting Information.
Microcolumn Manufacturing
and SLA 3D Printing
The microcolumn
chip was designed using an Autodesk Inventor 2020 (San Rafael, CA,
U.S.). Inventor files were submitted for 3D printing via the manufacturer’s
software PreForm V3.10.2 (Formlabs, Somerville, MA, U.S.). Printing
was performed using standard Clear V4 resin with
a Form3 low-force stereolithography printing unit
(Formlabs). The layer thickness was set to 25 μm, and supporting
structures were computed by the software PreForm and
adapted manually. Fully open channels were obtained at a 45°
angle tilt of the bore relative to the build platform and in parallel
to the Y-plane. Post printing, the columns were rinsed
for 30 min with tripropylene glycol monomethyl ether in a FormWash unit (Formlabs). Upon removing from the build platform,
the connecting ports and threads were rinsed with fresh isopropyl
alcohol (IPA). After a short drying period, both connecting ports
were equipped with 1/16″ tubing and tightened with flangeless
fittings and ferrules (Upchurch Scientific, Oak Harbor, WA, U.S.).
The column was flushed at 50 μL·min–1 using a syringe pump (Harvard Apparatus, Holliston, MA, U.S.) with
IPA, methanol, or acetone for at least 15 min. After disconnecting
the tubing, the residual solvent was removed from the column using
a microsyringe. The prints were allowed to dry for 1 h at room temperature.
Finally, the prints were photocured for 180 min at 35 °C in a
Form Cure (LED, 405 nm) oven (Formlabs).
Column Packing and Pepsin
Immobilization
A 1/16″
PEEK capillary (130 μm I.D.) was connected to the column outlet
using a flangeless 1/4″-28 PEEK fitting and ferrule. To retain
the packing material, a 2 mm round hydrophilic filter membrane was
clamped between the ferrule and the head of the column. The membrane
(1.2 μm Minisart syringe filters; Sartorius,
Goettingen, Germany) was cut to size using a 2 mm punch. Polybead
carboxylated 3 μm polystyrene (PS) microspheres (2.6% solid
particles in water and traces of the proprietary surfactant) were
used for column packing. 125 μL of the microsphere suspension
(equivalent to 3.2 mg of polystyrene microparticles) was loaded onto
a loop of 1/16″ tubing, which was attached to the inlet of
the microcolumn. The column was packed at a flow rate of 10 μL·min–1 for 18 min. During this process, the back pressure
of the column increased to 80 bar. The pepsin coupling procedure was
optimized for protein concentration, flow rate, and coupling time.
The following protocol turned out to be most effective. 200 μL
of the coupling solution [10 mg·mL–1 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide
(EDAC) in the coupling buffer] was pumped through the packed column
for 20 min at a flow rate of 10 μL·min–1. Subsequently, 200 μL of the freshly prepared pepsin solution
(2.5 mg·mL–1 pepsin in the coupling buffer)
was pumped through the packed column at 4 μL·min–1. The flow through was collected and used for determining the coupling
efficiency by UV/vis spectroscopy. The column was finally flushed
with 0.2% FA for 15 min and stored at 4 °C until use.
LC–MS
Peptide Mapping
A Switchos II microcolumn
switching module (LC Packings, Sunnyvale, CA, U.S.) equipped with
two multi-port valves along with a 130 μm I.D. PEEK tubing was
used for the online LC–MS setup. The loading pump delivered
a flow of 0.2% FA in H2O throughout the online digestion,
trapping, and desalting steps (8 min in total). Sample proteins were
introduced via a 10 μL sample loop. Peptides were trapped on
a SecurityGuard 4.0 mm I.D. × 3.0 mm C18 PreColumn (Phenomenex,
Torrence, CA, U.S.). Separation was performed using a GromSil C18
HPLC column, 2.0 × 60 mm, 3 μm, 100 Å (Grom, Rottenburg,
Germany) by 0.2% FA in H2O (mobile phase A) with 10 min
gradient 5.0–50% mobile phase B (0.2% FA in ACN) plus 2 min
at 50% B. The gradient was delivered using an LC20-AD HPLC system
(Shimadzu, Kyoto, Japan) at 0.5 mL·min–1. This
setup was coupled to a Thermo Scientific QExactive mass spectrometer
(San Jose, CA, U.S), operated in the data-dependent acquisition (DDA)
mode using top5 HCD fragmentation. For detailed MS settings and peptide
identification, see the Supporting Information.
RocC Bulk Digestion
See the Supporting Information.
Testing for Leachables
Direct infusion
MS experiments
were performed to assess the suitability of the 3D-printed chips for
coupling with ESI mass spectrometry. Empty channel chips were coupled
to a LTQ Orbitrap XL mass spectrometer, equipped with a standard ESI
source, and flushed with 0.2% aqueous FA at a flow of 5 μL·min–1. Leachable testing was performed on uncured chips,
and chips which were washed with either isopropanol, acetone, or methanol
for at least 15 min prior to the photocuring step. The efficacy of
the washing step was determined by injections of 5 μL of a 10
μM cytochrome c solution via a sample loop.
Protein Extraction
from Cell Culture and IMER Digestion
HEK293T cells were grown
to full confluency at 37 °C and in
a 5% CO2 atmosphere in DMEM with 4.5 g·L–1 glucose. Proteins were extracted by lysing the cells with 8 M urea
in 100 mM ammonium bicarbonate (pH 8.5). Cells were scraped, resuspended,
and transferred to a tube. Extracted protein amounts were determined
using a colorimetric bichinonic acid assay (Thermo Fisher Scientific,
Dreieich, Germany). Before digestion, proteins were reduced by addition
of 10 mM dithiothreitol (DTT) and alkylated with 20 mM iodoacetamide.
After alkylation, DTT was added to stop overalkylation. Proteome solutions
of 0.125, 0.25, 0.5, and 1 mg·mL–1 protein
in 3 M urea, 100 mM NH4HCO3, and 3% FA were
prepared. Injections of 20 μL were digested online at 10 μL·min–1 for 10 min at room temperature, starting with three
blank injections. The flow through was collected in LoBind Eppendorf
vials. Samples were frozen at −80 °C and lyophilized in
a SpeedVac centrifuge (Thermo Scientific).
Gel Electrophoresis and
Coomassie Staining
Proteome
digests were reconstituted in 20 μL of 100 mM NH4HCO3 buffer, and 5 μL of Laemmli buffer was added.
Sodium dodecyl sulfate-polyacrylamide gel electrophoresis (SDS-PAGE)
was performed by loading the samples on a gel containing 10% acrylamide.
The gels were run at 90 V for 110 min. Afterward, gels were stained
with SimplyBlue SafeStain (Thermo Fisher Scientific, Dreieich, Germany)
for 1 h and washed twice with water, once for 1 h, and once overnight.
UHPLC–MS Proteome Analysis
Dried peptides were
dissolved in 20 μL of 0.1% FA, and 10 μL was injected
onto the trapping column (nanoEase M/Z Symmetry C18 Trap Column, 100
Å, 5 μm, 180 μm × 20 mm, Waters, Manchester,
UK) of a nano-UPLC system (UltiMate 3000 RSLCnano System, Thermo Fisher
Scientific, Bremen, Germany) with a flow rate of 30 μL/min using
5% B (A: 0.1% FA in H2O, B: 80% ACN, 0.1% FA in H2O). Peptides were eluted and separated on a separation column (nanoEase
M/Z Peptide BEH C18 Column, 130 Å, 1.7 μm, 75 μm
× 250 mm, Waters, Manchester, UK) using a gradient from 5 to
27.5% B in 105 min, followed by 27.5–40% B in 10 min at a flow
rate of 300 nl·min-1. The nano-UPLC system was connected
to an orbitrap mass spectrometer (Orbitrap Fusion Lumos, Thermo Fisher
Scientific, San Jose, CA, USA) via a nano-ESI source. The spray was
generated from a steel emitter (Fisher Scientific, Dreieich, Germany)
at a voltage of 1850 V. MS/MS measurements were carried out in the
DDA mode using a normalized HCD collision energy of 30%. Every 3 s,
one MS scan was performed over an m/z range from 375 to 1500, with a resolution of 120,000 at m/z 200 (maximum injection time = 50 ms,
AGC target = 2 × 105). MS/MS spectra were recorded
in the orbitrap with a resolution of 15,000 at m/z 200 (maximum injection time = 54 ms, maximum AGC target
= 5 × 104, intensity threshold: 2.5 × 104, first m/z: 110), a quadrupole
isolation width of 1.6 Da, and an exclusion time of 60 s. For peptide
and protein identification and quantification, LC–MS raw data
were processed with MaxQuant (version 1.6.17.0). For identification,
MS/MS spectra were searched with the Andromeda search engine against
a human swiss-prot database (20,431 entries, www.uniprot.org) and a contaminant
database (298 entries). The searches were performed using the following
parameters: the precursor mass tolerance was set to 20 ppm for a first
peptide search, and the main search was performed with a tolerance
of 4.5 ppm. For MS/MS spectra, a fragment mass tolerance of 20 ppm
was used. Enzyme specificity was set to unspecific, and the following
modifications were considered: carbamidomethylation on cysteine residues
as a fixed modification and oxidation of methionine residues as a
variable modification. Peptides and proteins were identified with
an FDR of 1%. Proteins were kept as correctly identified if at least
one unique peptide was identified. Peptides and proteins were quantified
with the MaxLFQ algorithm, considering only unique peptides and a
minimum ratio count of 1. Bioinformatics data processing (log2-transformation,
normalization), statistical analysis (two-sided Student’s T-test using a permutation-based FDR with an adj. p-value cutoff of 0.05), and data visualization were performed
with Perseus (version 1.6.7.0).
Results and Discussion
IMER Fabrication
and Characterization
The low-force
SLA printer Form3 offers an XY-lateral
resolution of 25 μm, while the Z-resolution,
that is, layer thickness, can be adjusted to 25, 50, or 100 μm.
However, the direct SLA 3D printing of bores below 0.5 mm inner diameter
is very challenging in the case of the targeted diameter to length
ratios (Figure A,B).
In the case of the Form3 printer, narrow open channels
could only be obtained at a 45° angle tilt of the bore relative
to the build platform, while the bore is parallel to the Y-plane. Moreover, the lower limit for the bore diameter was determined
to be 0.5 mm in the CAD drawing, which reproducibly resulted in an
effective column diameter of approx. 360 μm (see the Supporting Information). We hypothesize two factors
to be responsible for this bore narrowing: (i) the laser light penetrating
the transparent resin must have polymerized additional material (Z-overcuring) and (ii) the slight expansion of cured resin
upon post-curing or during the polymerization process. Given the true
column dimensions of 360 μm I.D. × 30 mm, a column volume
of 3.06 μL was determined. Additionally, the formation of a
1/16″ diameter polymer crater surrounding the microbore was
observed (Figure B).
Apparently, the force inflicted by thoroughly tightening the fitting
was sufficient to imprint the PEEK capillary’s shape into the
rigid polymer to create a tight seal. This self-sealing capacity enabled
the operation of the column at high pressures without leaking or pressure
drops.
Figure 1
Photographical image of the transparent SLA-printed cartridge equipped
with a temperature probe (A); microscopy image of the open bore surrounded
by the imprint of the shape and diameter of an 1/16″ PEEK capillary
(B); positive ion mode mass spectra for the determination of leachables
from SLA 3D prints were obtained by direct coupling of the chips to
MS and flushing with 0.2% aqueous FA at a flow rate of 5 μL·min–1: uncured chip (C; left); chips washed with isopropanol,
acetone, or methanol for at least 15 min and cured with 405 nm blue
light at 35 °C for 3 h (C; center left to right); positive ion
mode mass spectra obtained by direct coupling of the chips to MS,
flushing with 0.2% aqueous FA at a flow rate of 5 μL·min–1 and injecting 5 μL of a 10 μM cytochrome
c solution via a sample loop. The intensity of the signal related
to the leachable UDMA at m/z = 471.2685
(marked by an asterisk) is 64,000 a.i. for the uncured chip (D; left)
and 200 a.i. for the cured (D; right). See the Supporting Information for MS settings.
Photographical image of the transparent SLA-printed cartridge equipped
with a temperature probe (A); microscopy image of the open bore surrounded
by the imprint of the shape and diameter of an 1/16″ PEEK capillary
(B); positive ion mode mass spectra for the determination of leachables
from SLA 3D prints were obtained by direct coupling of the chips to
MS and flushing with 0.2% aqueous FA at a flow rate of 5 μL·min–1: uncured chip (C; left); chips washed with isopropanol,
acetone, or methanol for at least 15 min and cured with 405 nm blue
light at 35 °C for 3 h (C; center left to right); positive ion
mode mass spectra obtained by direct coupling of the chips to MS,
flushing with 0.2% aqueous FA at a flow rate of 5 μL·min–1 and injecting 5 μL of a 10 μM cytochrome
c solution via a sample loop. The intensity of the signal related
to the leachable UDMA at m/z = 471.2685
(marked by an asterisk) is 64,000 a.i. for the uncured chip (D; left)
and 200 a.i. for the cured (D; right). See the Supporting Information for MS settings.
Testing for Leachables
Direct infusion MS experiments
were performed to assess the suitability of the 3D-printed chips (Figure A) for coupling with
ESI mass spectrometry. Empty channel chips (Figure B) were coupled to an LTQ Orbitrap XL mass
spectrometer, equipped with a standard ESI source, and flushed with
0.2% aqueous FA at a flow of 5 μL·min–1. Leachable testing was performed on uncured chips, and chips which
were washed with isopropanol, acetone, or methanol (Figure C) for at least 15 min prior
to the photocuring step. The efficacy of the washing step was determined
by injections of 5 μL of a 10 μM cytochrome c solution
via a sample loop. As can be seen in Figure D, the mass spectra obtained from the eluate
of uncured chips showed prominent signals in the range < m/z = 600, which corresponded to contaminants
leaching from the SLA print. It is also evident from Figure D that these compounds will
strongly interfere with any mass spectrometric detection of compounds
in a mass range which is vital for peptide analysis. Further MS/MS
analysis of the most abundant signal at m/z = 471.2685 (z = 1) plus the respective
ammonium, sodium, and potassium adducts revealed characteristic fragmentation
patterns for methacrylate compounds (see Figure S1, Table S1, and Scheme S1). Based on the computed elemental composition and
characteristic fragment ions, the signals were assigned to the compound
urethane dimethacrylate UDMA, a common crosslinking agent in methacrylate
resins. Clearly, the compounds leaching from the SLA prints are mostly
uncured oligomers or crosslinking agents and degradation products
thereof. However, all three post-print protocols drastically decreased
the intensity of leachables to an acceptable level with acetone and
methanol performing the best. In both cases, the ion count of the
UDMA sodium adduct declined from 30,000 pre-curing to below 1000,
while the protonated UDMA species was barely detectable at 200 counts.
Washing with acetone, however, compromised the structural integrity
of the capillary. This was evident by the capillary wall turning from
fully transparent to a cloudy shade. Similar effects of acetone and
acetonitrile on SLA prints have been reported.[51] Hence, methanol-treated chips were chosen for further experiments.
Online/In Situ Enzyme Immobilization
Despite the used
coupling kit and microspheres being intended for coupling and operation
in bulk solution, we chose an online approach after some iterations.
Online coupling is more difficult to perform from a technical perspective
but holds several advantages. Most importantly, lengthy and repetitive
washing, that is, centrifugation steps, can be avoided in flow-through
operation. Furthermore, using the manufacturer’s coupling protocol
was not possible since it involves washing steps of the enzyme-loaded
particles at pH 8 (wash/storage buffer), eventually leading to irreversible
inactivation of pepsin. During the packing process, pressures above
80 bar were observed without leaking (Figure A,B). After column packing, the modified
polystyrene surface was activated with EDAC. Finally, the protease
was immobilized by flushing the activated column with 500 μg
of pepsin dissolved in 200 μL of the coupling buffer (Figure C). The protease
loading was determined using a UV/vis NanoPhotometer Classic (Implen,
Munich, Germany). Online pepsin coupling was successfully employed
as described in the Experimental Section,
yielding a total coupling of 102 ± 3 μg pepsin on 3.2 mg
of the solid support material. This translates to 135% of the manufacturer’s
claim for expected protein loading by offline coupling (300 μg
of IgG on 12.5 mg of particles). The key parameters and characteristics
regarding the column and the IMER are summarized in Supporting Information Table S2.
Figure 2
IMER fabrication and
characterization: photographical image of
the 360 μm diameter column, packed with solid 3 μm PS
particles (A); linear correlation of flow rate and back pressure on
the 30 mm × 360 μm I.D. packed microcolumn (B); pepsin
coupling utilizing EDAC activation (C).
IMER fabrication and
characterization: photographical image of
the 360 μm diameter column, packed with solid 3 μm PS
particles (A); linear correlation of flow rate and back pressure on
the 30 mm × 360 μm I.D. packed microcolumn (B); pepsin
coupling utilizing EDAC activation (C).
Quantitative Assessment of IMER Performance
For online
protein analysis, the IMER was implemented in the LC–MS setup
as shown in Figure A. The IMER was utilized to obtain sequence confirmation by peptide
mapping of three different in-house recombinantly produced proteins.
As the first protein studied, 0.46 μg of samples of the RNA
chaperone FinO-domain RocC (produced as described by Eidelpes et al.[52]) in 0.8% FA and 1 mM NH4HCO3 was digested online at 10 μL·min–1,
trapped on a 4.0 × 3.0 mm C18 column and desalted (8 min in total).
The average pressure during this process was 85 bar. This was followed
by data-dependent LC–MS/MS analysis in a 10 min gradient (for
LC and MS/MS settings, see the Experimental Section). The experiments were performed in three consecutive runs.
Figure 3
Experimental
setup for the online protein analysis using two multi-port
valves for online digestion, desalting, and LC–MS analysis
of proteins (A); quantitative IMER characterization based on the peptic
digestion of in-house recombinant protein RocC: total number of peptide
hits and number of unique identified peptides detected in bulk digestions
(24 h) and IMER online digestions performed on three different days;
the error bars represent the mean plus/minus standard deviation, N = 3 (B); violin plot showing the distribution of intensities
for all quantified peptides. Three replicates are shown for each IMER
or the bulk (C); violin plots showing distribution of CV calculated
for peptide intensities in N = 3 replicates (D).
See Supporting Information Figure S4 for
volcano plots.
Experimental
setup for the online protein analysis using two multi-port
valves for online digestion, desalting, and LC–MS analysis
of proteins (A); quantitative IMER characterization based on the peptic
digestion of in-house recombinant protein RocC: total number of peptide
hits and number of unique identified peptides detected in bulk digestions
(24 h) and IMER online digestions performed on three different days;
the error bars represent the mean plus/minus standard deviation, N = 3 (B); violin plot showing the distribution of intensities
for all quantified peptides. Three replicates are shown for each IMER
or the bulk (C); violin plots showing distribution of CV calculated
for peptide intensities in N = 3 replicates (D).
See Supporting Information Figure S4 for
volcano plots.In spite of the extremely short
digestion time, the IMER was able
to effectively digest the sample proteins and deliver a substantial
number of short-length peptides for sequence analysis and confirmation.
Remarkably, we were able to produce three different IMERs (A, B, and
C) in intervals of at least 1 month, which showed comparable performance
and properties. For all IMERs, a comparable number of peptic RocC
peptides were identified (IMER A: 141 ± 9 peptides, IMER B: 144
± 10 peptides, IMER C: 159 ± 3 peptides, Figure B) and no significant differences
in overall intensities were observed (Figure C). See Supporting Information Figure S2 for the peptide map and S4 for volcano plots. For both IMER and bulk digestion, the quantified
peptides have a coefficient of variation (CV) lower than 25% (Figure D). Each of the IMERs
was able to rival the digestion performance of the 24 h bulk reaction,
highlighting the quality and reproducibility of the presented protocol.
Minor differences can be attributed to varying MS instrument performance,
given the time spans between the experiments. Importantly, all IMERs
showed high run-to-run as well as IMER-to-IMER reproducibility (Figures B–D and S4). All experiments yielded 100% sequence coverage
for RocC with over 100 identified peptides for each experiment. Therefore,
we conclude that the performance of the IMER is highly reproducible.
Complete sequence coverage of the recombinant sample protein was achieved
within an experiment runtime of 20 min per run, which included protein
digestion, desalting, and LC–MS analysis. The IMERs therefore
enable ultra-fast peptide mapping in less than 1.4% of the experiment
time of the bulk digest approach.The above-mentioned online
digestion experiments were repeated
with different IMERs at different lifetimes up to 29 days to assess
the longevity of the surface bound protease. Even after storage for
over 4 weeks at 4 °C, the IMER delivered high and reproducible
performance, resulting in 100% sequence coverage for the studies the
RocC protein, only showing a slight loss of intensity (for details,
see Figure S3).IMER online digestion
and hyphenated LC–MS analysis were
further performed for two additional in-house produced recombinant
proteins: the allergens Mald1.0201 and Act c 8. Here, peptide annotation
was performed using the freely available software MSstudio.[53] For Mal d 1.0201, 0.5 μg injections were
performed at either 5 μL·min–1 or 10
μL·min–1. The lower flow rate naturally
resulted in less back pressure at 45 bar compared to 85–90
bar at 10 μL·min–1. The low flow rate
run yielded 110 peptide hits, while the run at a regular flow rate
resulted in 96 peptide hits with 100% sequence coverage in both cases
(Figure S6). The IMER was further tested
at 15 μL·min–1, which resulted in 135
bar back pressure. Again, no leakage was observed as the pressure
remained constant for several minutes. However, such high flow rates
were not used further so as to ensure greater longevity and prevent
any column bed damage. The manufacturer does not state any pressure
limit for the used polystyrene microspheres as they are not intended
for high-pressure applications. Therefore, 10 μL·min–1 was identified as a suitable operational flow rate.
The observed back pressures at given flow rates are comparable to
those of 3 μm nonporous particle HPLC columns of similar lengths
and diameters.The foodborne allergen Act c 8 was recombinantly
produced and purified
as described by Zeindl and Tollinger.[54] The online digestion and desalting setup allowed for the direct
analysis of the protein in its storing buffer. 2.5 μg of Act
c 8 samples in 3 M guanidine hydrochloride was injected in three consecutive
runs for online digestion at ambient temperature on day 16 of the
IMER lifetime. The addition of the chaotropic agent to the sample
buffer resulted in thorough digestion and 100% sequence coverage with
119.3 ± 9.2 peptide hits per run. A total of 115 of the peptides
were present in at least 2 of 3 runs and therefore labeled as unique
identifiable (Figure S7). We conclude that
the 3D-printed microcolumn IMER can safely be used with high concentrations
of chaotropic agents. See the Supporting Information for additional details. The extremely fast analysis capability enabled
by IMER online digestion is highly favorable when supporting in-house
protein biochemists in the development of protein expression protocols
and downstream processes. The IMER allows us to obtain information
on protein identity and primary sequence integrity in less than 30
min.
Complex Proteome Digestion Efficiency
To examine the
performance of the IMER for global proteome analysis, we investigated
the enzymatic digestion efficiency of the IMER for a complex proteome
sample. Here, different amounts of a proteome extract isolated from
human embryonic kidney 293T cells were loaded onto the pepsin IMER.
The eluents of the IMER were collected and evaporated. The dried eluents
were dissolved in 20 μL of 100 mM ammonium bicarbonate (pH 8.3);
5 μL of 5× Laemmli buffer was added, and the samples were
loaded onto an SDS-PAGE system. As a control, the same amounts of
nondigested proteome extracts were loaded onto the SDS-PAGE system.
The comparison of SDS-PAGE traces of the undigested proteome and IMER-digested
samples showed that IMER efficiently digested proteomes up to 20 μg
(Figure A). The trace
of the IMER-digested proteome amount of 20 μg showed some weak
bands. This indicates that the maximum amount of proteome to be used
for the IMER is about 20 μg.
Figure 4
Quantitative analysis of digestion efficiency
between bulk and
IMER digestion of the complex proteome. (A): SDS-PAGE of the IMER-undigested
proteome extract (u) and digested proteome extract (d). Very left
and very right: protein standard. The different sample amounts loaded
on the SDS-PAGE system are indicated accordingly. (B): Volcano plots
of quantified peptides comparing bulk digestion and IMER digestion.
(C): Bar graphs (median with standard deviation) showing the number
of identified peptides after bulk and IMER digestion. (D): Volcano
plots of quantified proteins comparing bulk digestion and IMER digestion.
(E): Bar graphs (median with standard deviation) showing the number
of identified proteins after bulk and IMER digestion. (C,E): Statistical
analyses were performed using the two-tailed Mann–Whitney test.
(B,D): Adjusted p-value of ≤0.05 was used
as a threshold for statistical significance. Peptides and proteins
showing significant higher amounts in bulk or IMER digestion are colored
in red (bulk) or blue (IMER). Analytes with a fold-change of ≤1.5
are represented as red- or blue-colored semi-transparent dots, and
analytes with fold-change of ≥1.5 are represented as red- or
blue-colored dots. N = 3 independent experiments.
Quantitative analysis of digestion efficiency
between bulk and
IMER digestion of the complex proteome. (A): SDS-PAGE of the IMER-undigested
proteome extract (u) and digested proteome extract (d). Very left
and very right: protein standard. The different sample amounts loaded
on the SDS-PAGE system are indicated accordingly. (B): Volcano plots
of quantified peptides comparing bulk digestion and IMER digestion.
(C): Bar graphs (median with standard deviation) showing the number
of identified peptides after bulk and IMER digestion. (D): Volcano
plots of quantified proteins comparing bulk digestion and IMER digestion.
(E): Bar graphs (median with standard deviation) showing the number
of identified proteins after bulk and IMER digestion. (C,E): Statistical
analyses were performed using the two-tailed Mann–Whitney test.
(B,D): Adjusted p-value of ≤0.05 was used
as a threshold for statistical significance. Peptides and proteins
showing significant higher amounts in bulk or IMER digestion are colored
in red (bulk) or blue (IMER). Analytes with a fold-change of ≤1.5
are represented as red- or blue-colored semi-transparent dots, and
analytes with fold-change of ≥1.5 are represented as red- or
blue-colored dots. N = 3 independent experiments.To compare the performance of proteome digestion
between conventional
bulk digestion and IMER, we digested 5 μg of the proteome extract
with pepsin (protein/protease ratio 100:1) overnight and loaded 5
μg of the proteome extract onto the IMER. Both overnight digested
samples and eluents of the IMER were dried in a SpeedVac centrifuge.
Dried samples were dissolved in 20 and 10 μL thereof were subjected
to LC–MS/MS analysis for label-free quantification. No significant
differences were observed between bulk digestion and online IMER digestion
in terms of the total number of identified peptides (nIMER = 12,141
± 1171; nBULK = 11,760 ± 363; p = 0.700, Figure B) and identified
proteins (nIMER = 1435 ± 60; nBULK = 1499 ± 15; p = 0.100, Figure C). A quantitative comparison at the protein level showed
that a larger number of proteins could be quantified with a significantly
higher amount in general for IMER-digested samples: (nIMER = 221,
nBULK = 156) and at least by a factor of 1.5 (nIMER = 150, nBULK =
114) (Figure D).To compare the performance of bulk and IMER digestion for the analysis
of complex proteomes, the quantitative comparison at the peptide level
is much more relevant because the peptides were the product of pepsin
digestion, and the digestion efficiency can be compared based on the
quantity of peptides generated. The quantitative analysis at the peptide
level showed that 324 peptides were detected in significantly higher
quantities for the bulk digestion, whereas 945 peptides were detected
in significantly higher quantities in samples generated during IMER
digestion (Figure E). It is worth mentioning that 93% of these 945 peptides, that is,
879 peptides, showed at least 1.5-fold higher amount in IMER than
in bulk digestion, where 273 peptides showed at least 1.5-fold higher
amount. These results show that IMER allows a 144-fold faster (10
min for IMER, 24 h for bulk digestion) and significantly more efficient
digestion of complex proteomes in the lower-microgram range compared
to conventional bulk digestion, representing a clear advance in proteome
profiling.
Conclusions
Using high-resolution
additive manufacturing, it was possible to
produce microfluidic chips holding an enclosed 360 μm I.D. microbore
column with an aspect ratio close to 100:1. Leachables derived from
the 3D prints were identified as the crosslinking agent UDMA and degradation
products thereof. The leachable derived sample contamination was drastically
reduced by implementing a comprehensive washing and curing protocol.
Thereafter, SLA prints were suitable for hyphenation with MS and further
utilized to create microscale pepsin-immobilized enzyme reactors.
The IMER proved highly effective for thorough online digestion of
different recombinantly produced proteins, was implemented in an online
setup, and (i) showed outstanding structural integrity, withstanding
astonishingly high pressures above 130 bar without leaking, (ii) achieved
100% sequence coverage for all studied proteins, (iii) displayed excellent
performance even after 4 weeks of storage, and (iv) was successfully
used with 3 M guanidine hydrochloride and urea for improved protein
denaturation. Furthermore, the IMER provided dramatically faster (144-fold)
and significantly more efficient digestion of complex proteomes in
the lower μg range. This IMER therefore represents an interesting
and powerful technology for quantitative proteomics of the smallest
sample amounts.Apart from the pepsin IMER presented here, this
protocol essentially
represents a versatile platform for many more future applications.
The presented protocol, relying on robust coupling chemistry and commercial
products, can be expanded to virtually any protein of choice to create
either different IMERs or other applications where protein immobilization
is favorable, for example, assays for ligand screening. Using the
developed washing and curing protocol, we expect to see many more
exciting applications of SLA prints in the field of analytical chemistry.
Given the unique versatility of 3D printing, some of these might include
multi-bed reactors, composed of multiple columns of varying dimensions
in a single chip or the incorporation of differently loaded particles
in a single column. SLA 3D printing enables the direct manufacturing
of such true 3D geometries. In this context, the use of SLA-printed
components could be expanded to integrated multi-step systems for
sample protein online reduction, deglycosylation, and digestion. These
components will likely exhibit performances on par with milled glass
chips but at much lower costs and faster lead times.As high-resolution
SLA printing is becoming more and more accessible,
researchers are pushing its boundaries toward producing increasingly
smaller and better resolved microfluidic designs. Hence, we anticipate
a striking impact of SLA on the production and prototyping of polymer-based
devices for custom-designed enzyme reactors. Given their highly favorable
self-sealing capacity and excellent pressure resistance, SLA prints
will greatly contribute to the evolving field of high-pressure microfluidics.
Authors: J L Capelo; R Carreira; M Diniz; L Fernandes; M Galesio; C Lodeiro; H M Santos; G Vale Journal: Anal Chim Acta Date: 2009-07-22 Impact factor: 6.558