PURPOSE: To assess the influence of contemporary contact lens (CL) materials on human coronavirus attachment and the influence of a rub and rinse step to remove these viruses.Entities:
Keywords: Contact lens materials; HCoV-229E; HCoV-OC43; Human coronavirus; Hydrogel; Rigid gas permeable; Silicone hydrogel; Soft contact lenses
Year: 2022 PMID: 35643848 PMCID: PMC9124925 DOI: 10.1016/j.clae.2022.101719
Source DB: PubMed Journal: Cont Lens Anterior Eye ISSN: 1367-0484 Impact factor: 3.946
Contact lenses evaluated in this study.
| Hydrogel | Acuvue 2 | Johnson&Johnson | Etafilcon A | No | 58 |
| Proclear 1 Day | CooperVision | Omafilcon A | No | 62 | |
| Silicone Hydrogel | Acuvue Oasys | Johnson&Johnson | Senofilcon A | Internal wetting agent | 38 |
| Biofinity | CooperVision | Comfilcon A | No | 48 | |
| Clariti 1 day | CooperVision | Somofilcon A | WetLoc technology | 56 | |
| Air Optix Aqua | Alcon | Lotrafilcon B | Plasma coating | 33 | |
| Dailies Total 1 | Alcon | Delefilcon A | Water gradient technology | 33 | |
| PureVision 2 | Bausch + Lomb | Balafilcon A | Plasma oxidation | 36 |
*USAN: United States Adopted Name.
Fig. 1Protocols used to test the impact of rub and rinse steps on coronaviruses removal from CL surfaces. Experiments were performed in triplicate.
Contact lens care products included in this study.
| Bausch & Lomb, Rochester, NY | Polyaminopropyl biguanide 0.00013% and polyquaternium 0.0001% | 1- Place at least 3 drops of Biotrue on each side of CL and rub for 20 s | |
| Alcon, | Polyquad (Polyquaternium-1) 0.001% and Aldox (Myristamidopropyl Dimethylamine) | 1- Wet each side of the CL with OPTI-FREE and rub the lens for 20 s | |
| Alcon, | 3% hydrogen peroxide | 1- Put CL into the special Clear Care case and rinse with Clear Care solution for at least 5 s | |
| Ophtecs, | 0.05% Povidone-iodine | 1- Add a tablet of cleadew into the case and fill with cleadew solution | |
Fig. 2Infectious viral particles/lens recovered from CLs after “no rub or rinse” treatment. Left panel: HCoV-229E. Right panel: HCoV-OC43. Averages and standard deviations were calculated using results of triplicate samples (3CL). As a control, virus suspension was incubated in the same conditions for each experiment, but with no CL. The titre of control was 8.79 × 104 (±2.27 × 104) MPN/mL and 8.72 x104 MPN/mL (±2.50 × 104) for HCoV-229E and HCoV-OC43, respectively. LQ: limit of quantification.
Fig. 3Infectious viral particles/lens recovered from etafilcon A and lotrafilcon B after disinfection with Biotrue and OPTI-FREE Puremoist with no rub and rinse steps. Averages and standard deviations were calculated using results of triplicate samples (3CL). Biotrue and OPTI-FREE Puremoist were able to reduce the number of infectious viral particles to below the limit of quantification only when rub and rinse steps were performed for both etafilcon A and lotrafilcon B. As a control, virus suspension was incubated in the same conditions, but with no CL. The titre of control was 9.27 × 104 (±4.88 × 103) MPN/mL. LQ: limit of quantification.