Zhaoning Xu1, Dongqin Sheng1, Kefan Jiao2, Chi Zhang3, Junping Hao4, Dedong Ma4.
Abstract
INTRODUCTION: Abnormal triggering of non-invasive ventilator (NIV) is the main reason for increasing the possibility of patient intolerance and directly affecting the treatment effect.Entities:
Keywords: abnormal triggering; auto-triggering; ineffective-triggering; non-invasive ventilation; patient comfort
Mesh:
Substances:
Year: 2022 PMID: 35642081 PMCID: PMC9366579 DOI: 10.1111/crj.13497
Source DB: PubMed Journal: Clin Respir J ISSN: 1752-6981 Impact factor: 1.761
FIGURE 1The connection of all devices. 1, non‐invasive ventilator; 2, mask; 3, air compressor; 4, oxygen tube; 5, glass rotor flowmeter (LZB‐10WB™); 6, oxygen injection site: proximal to the mask; 7 oxygen injection site: proximal to the ventilator
FIGURE 2Frequency of auto‐triggering and ineffective‐triggering under two oxygen injection sites, five oxygen flows in two ventilators. *P < 0.05, **P < 0.001. Figure 2A,B shows the results of the frequency of auto‐triggering and ineffective‐triggering of Resmed stellar™ 150 under two oxygen injection sites using five oxygen flows, respectively. Figure 2C,D shows the results of the frequency of auto‐triggering and ineffective‐triggering of curative Flexo ST30 under two oxygen injection sites using five oxygen flows, respectively
FIGURE 3Frequency of auto‐triggering and ineffective‐triggering under two oxygen injection sites, five support pressures in two ventilators. *P < 0.05, **P < 0.001. Figure 3A,B shows the results of the frequency of auto‐triggering and ineffective‐triggering of the Resmed stellar™ 150 under two oxygen injection sites using five IPAP and EPAP, respectively. Figure 3C,D shows the results of the frequency of auto‐triggering and ineffective‐triggering of the curative Flexo ST30 under two oxygen injection sites using five IPAP and EPAP, respectively
Baseline characteristics
| Variable | Total | Male | Female |
|---|---|---|---|
| Frequency | 30 | 12 | 18 |
| Age | 26.1 ± 2.9 | 25.8 ± 0.8 | 26.3 ± 3.7 |
| Height/cm | 167.7 ± 7.3 | 176.0 ± 1.6 | 162.2 ± 3.1 |
| Weight/kg | 64.6 ± 15.8 | 82.7 ± 4.0 | 52.7 ± 5.8 |
| BMI/kg * m−2 | 22.7 ± 3.6 | 26.7 ± 1.0 | 20.0 ± 1.7 |
Abbreviation: BMI, body mass index.
Frequency of auto‐triggering and ineffective‐triggering of Resmed stellar™ 150 and curative Flexo ST30 ventilator under 5 different oxygen flows
| Oxygen flow (L/min) | Frequency of AT (R) | Frequency of IT (R) | Frequency of AT (C) | Frequency of IT (C) | ||||
|---|---|---|---|---|---|---|---|---|
| Mask# | Ventilator# | Mask# | Ventilator# | Mask# | Ventilator# | Mask# | Ventilator# | |
| 5 | 2 (1–2)‡,§,¶,※ | 2 (2–2.25)‡,§,¶,※ | 0 (0–0)‡,§,¶,※ | 0 (0–1)‡,§,¶,※ | 2 (2–2)§,¶,※ | 1 (1–1)§,¶,※ | 0 (0–0)§,¶,※ | 0 (0–0)§,¶,※ |
| 10 | 4 (3–4)†,§,¶,※ | 6 (5–7)†,§,¶,※ | 1 (1–1)†,§,¶,※ | 1 (1–2)†,§,¶,※ | 3 (3–3)§,¶,※ | 1 (1–1)§,¶,※ | 0 (0–0)§,¶,※ | 0 (0–0)§,¶,※ |
| 15 | 7.5 (6–8)†,‡,¶,※ | 8.5 (8–13.25)†,‡,¶,※ | 2 (2–2)†,‡,※ | 3 (2.75–4)†,‡,※ | 6 (6–6.25)†,‡,¶,※ | 4 (3–4)†,‡,¶,※ | 1 (1–1)†,‡,※ | 1 (1–1)†,‡,¶,※ |
| 20 | 12.5 (10–13)†,‡,§,※ | 15 (15–17.75)†,‡,§,※ | 2 (2–2.25)†,‡,※ | 4 (4–5)†,‡ | 12 (10.75–12)†,‡,§,※ | 6 (5–6)†,‡,§,※ | 1 (1–1)†,‡,※ | 2 (2–2)†,‡,§ |
| 25 | 17 (15–17)†,‡,§,¶ | 26 (26–31)†,‡,§,¶ | 6 (3.75–6)†,‡,§,¶ | 5 (5–7)†,‡,§ | 20 (18–22)†,‡,§,¶ | 11 (9.5–12)†,‡,§,¶ | 3 (3–3)†,‡,§,¶ | 2 (2–3)†,‡,§ |
Note: mask# indicates “proximal to the mask”; ventilator# indicates “proximal to the ventilator’; AT indicates ‘auto‐triggering’; IT indicates ‘ineffective triggering’; R indicates ‘Resmed Stellar™ 150 ventilator’; C indicates ‘Curative Flexo ST30 ventilator’; data were shown as IQR (interquartile range) [M (P25, P75)]. †, ‡, §, ¶ and ※ indicate that the comparison within the same oxygen injection and same kind of triggering group was statistically significant compared with oxygen flow is 5, 10, 15, 20 and 25 L/min (P < 0.05) respectively.
Frequency of auto‐triggering and ineffective‐triggering of Resmed stellar™ 150 and curative Flexo ST30 ventilator under 5 different support pressures (IPAP, EPAP)
| IPAP, EPAP (mmHg) | Frequency of AT (R) | Frequency of IT (R) | Frequency of AT (C) | Frequency of IT (C) | ||||
|---|---|---|---|---|---|---|---|---|
| Mask# | Ventilator# | Mask# | Ventilator# | Mask# | Ventilator# | Mask# | Ventilator# | |
| 8, 4 | 1 (0–2)§,¶,※ | 2 (2–2.25)§,¶,※ | 0 (0–0)§,¶,※ | 0 (0–1)‡,§,¶,※ | 2 (1–2)§,¶,※ | 1 (0–1)§,¶,※ | 0 (0–0)§,¶,※ | 0 (0–0)§,¶,※ |
| 10, 4 | 2 (1–4)§,¶,※ | 3 (2.75–5)§,¶,※ | 1 (0–1)§,¶,※ | 1(1–2)a,c,d,e | 3 (2–3)§,¶,※ | 1 (1–2)§,¶,※ | 0 (0–0)§,¶,※ | 0(0–0)§,¶,※ |
| 12, 4 | 5 (3–6)†,‡,¶,※ | 7 (5–9.25)†,‡,¶,※ | 2 (1–2)†,‡,¶,※ | 2.5 (2–3)†,‡,¶,※ | 5 (4–6) | 3 (2–3) | 1(1–1)†,‡,※ | 1 (1–1)†,‡,※ |
| 16, 6 | 10 (6–12.25)†,‡,§,※ | 15 (12–16.25)†,‡,§,※ | 2 (2–3)†,‡,§ | 4 (3–5)†,‡,§,※ | 11 (8–12) | 5 (4–6) | 1 (1–1)†,‡,※ | 2 (1–2)†,‡ |
| 20, 8 | 15.5 (14–20)†,‡,§,¶ | 29.5 (22.75–33)†,‡,§,¶ | 5 (3–6)†,‡,§ | 6 (5–9)†,‡,§,¶ | 22 (17.5–23.25) | 12 (9.75–15) | 3 (3–3)†,‡,§,¶ | 2.5 (2–4)†,‡,§ |
Note: mask# indicates ‘proximal to the mask’; ventilator# indicates ‘proximal to the ventilator’; AT indicates ‘auto‐triggering’; IT indicates ‘ineffective triggering’; R indicates ‘Resmed Stellar™ 150 ventilator’; C indicates ‘Curative Flexo ST30 ventilator’; data were shown as IQR (interquartile range) [M (P25, P75)]. †, ‡, §, ¶ and ※ indicate that the comparison within the same oxygen injection and same kind of triggering group was statistically significant compared with IPAP and EPAP is 8 and 4 cm H2O, 10 and 4 cm H2O, 12 and 4 cm H2O, 16 and 6 cm H2O and 20 and 8 cm H2O (P < 0.05), respectively.
Abbreviations: EPAP, expired positive airway pressure; IPAP, independent inspired positive airway pressure.
FIGURE 4Frequency of auto‐triggering and ineffective‐triggering under two oxygen injection sites, three masks in two ventilators. *P < 0.05, **P < 0.001. Figure 4A,B shows the results of the frequency of auto‐triggering and ineffective‐triggering of the Resmed stellar™ 150 under two oxygen injection sites using three kinds of masks, respectively. Figure 4C,D shows the results of the frequency of auto‐triggering and ineffective‐triggering of the curative Flexo ST30 under two oxygen injection sites using three kinds of masks, respectively
Frequency of auto‐triggering and ineffective‐triggering of Resmed stellar™ 150 and curative Flexo ST30ventilator under three different masks
| Mask | Frequency of AT (R) | Frequency of IT (R) | Frequency of AT (C) | Frequency of IT (C) | ||||
|---|---|---|---|---|---|---|---|---|
| Mask# | Ventilator# | Mask# | Ventilator# | Mask# | Ventilator# | Mask# | Ventilator# | |
| A | 2.5 (1–5)§ | 3 (2–6.25)§ | 0 (0–1.25)‡,§ | 1 (0–2)§ | 3 (2–4)‡,§ | 2 (1–3)‡,§ | 1 (0–2)‡,§ | 1 (0–1)‡,§ |
| B | 4 (1–6) | 6 (2.75–8.25) | 1 (1–2)†,§ | 1 (1–3)§ | 2 (1–3)†,§ | 1 (0.75–1)†,§ | 1 (0–1)†,§ | 0 (0–0)†,§ |
| C | 4.5 (2–8.25)† | 7.5 (5–14)† | 2 (2–3)†,‡ | 3 (3–4)†,‡ | 6 (4–7)†,‡ | 4 (3–5)†,‡ | 2 (1–3)†,‡ | 2 (1–3)†,‡ |
Note: mask# indicates ‘proximal to the mask’; ventilator# indicates ‘proximal to the ventilator’; AT indicates ‘auto‐triggering’; IT indicates ‘ineffective triggering’; R indicates ‘Resmed Stellar™ 150 ventilator’; C indicates ‘Curative Flexo ST30 ventilator’; data were shown as IQR (interquartile range) [M (P25, P75)]. ‘A’ indicates ‘Curative: Bestfit™’; ‘B’ indicates ‘Resmed: Mirage Quattro Full Face Mask™’; ‘C’ indicates ‘Zhongshan: ZS‐MZ‐A™’. †, ‡ and § indicate that the comparison within the same oxygen injection and same kind of triggering group was statistically significant compared with mask is A, B and C (P < 0.05), respectively.