Performance evaluation of the XT MicroSlide assay pairs on the Vitros XT 7600 compared to VITROS single microslide assays on Vitros 5600.

Lily Olayinka1,2, Estella Tam2, Sridevi Devaraj1,2.   

Abstract

Objectives: Pediatric hospitals are always challenged by specimen volumes and thus any innovation in this realm is very welcome. With the introduction of Microslide assay pairs, we aimed to evaluate the analytical performance of the Vitros XT MicroSlide assay pairs on the Vitros XT 7600 compared to single MicroSlides. Design: Performance characteristics included within-run precision, analytical measurable range, method comparison, and interference verification. We compared six XT MicroSlide pairs on the Vitros XT 7600 with twelve corresponding single slide assays on the Vitros 5600 system.
Results: The XT MicroSlides on Vitros XT 7600 demonstrated excellent precision, equivalent analytical measurable range, and strong method correlation with single slide assays on Vitros 5600 for most of the assays tested. Within-run CVs of the analytes ranged between 0.32% and 2.93% with between-run CV of less than 8.8% and linearity for all analytes was within the manufacturer's specified range. Interference studies showed comparable effects of hemolysis, lipemia, and bilirubin on both instruments. Conclusions: The XT MicroSlides are comparable to the single MicroSlide assays with improved efficiency, turnaround times and lower sample volumes.
© 2022 Published by Elsevier B.V.

Entities:  

Keywords:  Analyte; Assay; Method verification; Performance; Vitros; XT slide

Year:  2022        PMID: 35637639      PMCID: PMC9144011          DOI: 10.1016/j.plabm.2022.e00282

Source DB:  PubMed          Journal:  Pract Lab Med        ISSN: 2352-5517


Introduction

One of the challenges clinical laboratories servicing pediatric centers face is adequate specimen collection from infants and children in addition to performing multiple tests on the limited sample volume [1]. This challenge is significantly heightened in critically ill patients and in patients for whom phlebotomy is challenging. Diagnostic phlembotomy is a known cause of iatrogenic blood loss which can result in the aggravation of hospital-acquired anemia [2]. Both Vitros 5600 and XT 7600 systems integrate dry chemistry (MicroSlide), wet chemistry and immunoturbidimetry with photometric detection (MicroTip), immunoassays with enhanced chemiluminescence (MicroWell), and photometric measurement of sample quality indices (MicroSensor) into a single analyzer. However, one of the unique features of the Vitros XT 7600 includes the introduction of the Vitros XT Microslide pair technology, which combines two film reagents onto the same slide to minimize sample volume, improve efficiency and productivity. Currently available Vitros XT MicroSlide pairs include Alanine Aminotransferase/Aspartate Aminotransferase [ALTV-AST], Triglyceride/Cholesterol [TRIG-CHOL], Albumin/Total Protein [ALB-TP], Urea/Creatinine [UREA-CREA], Glucose/Calcium [GLU-CA], Total Bilirubin/Alkaline Phosphatase [TBIL-ALKP]. These MicroSlides are also available as single slides for use on the Vitros XT 7600 and were tested against the paired slides. For this short communication, only the comparison of the paired slides and single slides on 5600 will be discussed. . In accordance with CLSI guidelines (EP05-A3 and EP-06 A), we evaluated results from XT Microslides for Vitros XT 7600 analyzer compared to the Single MicroSlides assays for Vitros 5600 and report the findings.

Materials and methods

Twelve analytes on the XT MicroSlides: ALTV-AST, TRIG-CHOL, ALB-TP, UREA-CREA, GLU-CA, TBIL-ALKP versus the Vitros 5600 single slide assays were evaluated using the Vitros XT 7600. The methodology, turnaround time and volume required for the single MicroSlides on Vitros 5600 and XT MicroSlide assays on Vitros XT 7600 are listed in Table 1.
Table 1

Sample volume, turnaround time and methodology required for the XT MicroSlides vs. Single slides.

AssaysSingle slides (μL)XT Slides (μL)Turnaround time (TAT)
Methodology
5600XT 7600
Albumin5.54.25 m 3s5 m 3sBromocresol green (BCG)
ALKP115.05 m 10s5 m 10sp-nitrophenyl phosphate substrate
ALTV73.55 m 4s5 m 12sL-alanine - α-ketoglutarate (P-5-P)
AST73.35 m 13s**ALTVL-aspartate - α-ketoglutarate (P-5-P)
BUN/Urea5.54.35 m 4s5 m 3sColorimetric urease reaction
Calcium103.55 m 32s5 m 31sArsenazo III dye
Cholesterol5.53.94 m 45s5 m 3sCholesterol oxidase
Creatinine63.24 m 54s**BUN/UreaCreatine amidinohydrolase
Glucose62.75 m 14s** CalciumGlucose oxidase
Total Bilirubin105.05 m 10s**ALKP4-(N-carboxymethylsulfonyl) benzenediazonium hexafluorophosphate
Total Protein6.54.15 m 3s**AlbuminBiuret reaction
Triglycerides5.52.95 m 4s**CholesterolL-α-glycerophosphate oxidase

AST- Aspartate Aminotransferase; BUN - Blood Urea Nitrogen.

Total TAT for all assays on 5600: 69 min 2 s.

Total TAT for all assays on XT7600: 31 min 2 s.

Total volume for all assays on 5600: 85.5 μL.

Total volume for all assays on XT 7600: 45.6 μL.

**runs simultaneously.

Sample volume, turnaround time and methodology required for the XT MicroSlides vs. Single slides. AST- Aspartate Aminotransferase; BUN - Blood Urea Nitrogen. Total TAT for all assays on 5600: 69 min 2 s. Total TAT for all assays on XT7600: 31 min 2 s. Total volume for all assays on 5600: 85.5 μL. Total volume for all assays on XT 7600: 45.6 μL. **runs simultaneously. Within-run and Between-run precision was assessed by measurement of ten replicates of two-concentration levels of the manufacturer's quality control material (Performance verifier 1 (PV-1) and Performance verifier 2 (PV-2)) for each analyte. Mean concentration and coefficient of variation (%CV) were calculated for each assay. Linearity/Analytical Measurable Range (AMR) was carried out with two replicates of a five-level calibrator (Cal Kit 3–1, Cal Kit 3–2, Cal Kit 3–2, PV-1, PV-2), that span across the manufacturer's reportable range for all 12 analytes. The mean of duplicate measurements of each XT Microslide assays was calculated and compared with the expected value from single slides on the Vitros 5600. Accuracy (method comparison) studies were performed using residual patient serum samples, based on the CLSI EP09-A3 guideline (CLSI guide 2013), except that each sample was measured once due to limited volume. The samples were selected to cover AMR as wide as possible, and twenty-four samples were collected for each assay. Deming regression was used for analysis to calculate an intercept, a slope, and 95% confidence intervals (CI). The percent bias for each assay was compared to the total allowable error (TEa) recommended by Clinical Laboratory Improvement Amendments (CLIA) to assess the clinical significance of the bias [3]. Assay interference was verified by spiking pooled serum samples (12 patients per pool) with different concentrations of hemoglobin (0, 200 mg/dL, 400 mg/dL and 800 mg/dL), bilirubin (0, 15 mg/dL, 30 mg/dL and 60 mg/dL) or intralipid (0, 400 mg/dL, 1000 mg/dL and 3000 mg/dL) using an in-house hemolysate stock and a commercial interference kit for lipemia and icterus (Sun Diagnostics). The concentration corresponded to normal, mild, moderate and significant interference and the hemolysate stock was prepared using the freezing and thawing method [4]. Turnaround time was obtained from the display user interface of each instrument after sample loading and sample volume required for each analyte was obtained from the manufacturer's MicroSlide assay summary. Statistical analyses for method comparison of the single microslides to the paired slides were performed with EP Evaluator (Burlington, VT USA). The method comparison studies were evaluated using Deming regression, and a Bland-Altman plot was generated to assess bias.

Results

The %CVs and 95%CI of within-run and between-run precision were calculated and reported. Within-run %CVs ranged from 0.32% to 2.93%. All assays except total bilirubin met the manufacturer's claimed precision of 1.33%–3.91%. Total bilirubin was observed to exceed the manufacturer's specification at low-level concentration (2.93% vs 2.1%). The between-run precision was 0.7%–8.7%. The AMR of each assay was found to be linear within the manufacturer's specified ranges. The mean slope for all assays ranged from 0.98 to 1.01, with a mean y-intercept ranging from −5.58 to +2.43 and a correlation coefficient from 0.99 to 1.00. All 12 analytes demonstrated a correlation coefficient >0.95. ALT, CA, CREA, CHOL, and GLU met the set criteria of r = > 0.95 and slopes of 0.9–1.1. Five assays (ALB, ALKP, AST, TP, TRIG and UREA) exhibited slopes <0.9. Percent bias for all the assays were within the desirable specification for total allowable error for each analyte (Table 2).
Table 2

Method comparison of XT MicroSlides vs 5600 single slides.

AssaySlope (95%CI)Intercept (95%CI)Correlation coefficient (R)TEaMean %biasWithin-run %CVBetween-run %CV
Albumin0.80(0.70–0.89)0.84(0.44–1.24)0.96110%0.001.111.4
ALKP0.82 (0.75–0.86)25.82(15.09–36.54)0.98430%−4.637.963.0
ALTV0.92(0.85–0.98)2.88(1.26–4.51)0.98620%0.831.171.6
AST0.81(0.78–0.84)7.87(6.61–9.12)0.99720%−0.580.941.5
BUN/Urea1.12(1.03–1.20)−1.39(-2.68 to −0.11)0.9849%/2 mg/dL0.330.731.4
Calcium1.16(1.06–1.26)−1.31(-2.28 to −0.34)0.9801 mg/dL0.220.430.8
Cholesterol0.96(0.90–1.01)6.07(-2.34–14.47)0.99310%−1.421.481.2
Creatinine1.02(0.99–1.04)−0.03(-0.05 to −0.01)0.99815%/0.3 mg/dL−0.020.901.4
Glucose1.01(0.92–1.10)−0.69(-7.45 to 6.08)0.98110%/6 mg/dL0.210.570.7
Total Bilirubin1.30(1.20–1.52)−0.02(-0.32 to −0.10)0.96520%/0.4 mg/dL0.052.938.7
Total Protein0.90(0.78–1.00)0.85(0.08–1.61)0.96210%0.070.751.4
Triglycerides0.86(0.79–0.92)12.08(2.29–21.86)0.98325%−8.421.440.8

CI – Confidence Interval; TEa – total allowable error (CAP); CV – Coefficient of variation; ALKP – Alkaline Phosphatase; ALTV – Alanine Aminotransferase; AST- Aspartate Aminotransferase; BUN Blood Urea Nitrogen.

Method comparison of XT MicroSlides vs 5600 single slides. CI – Confidence Interval; TEa – total allowable error (CAP); CV – Coefficient of variation; ALKP – Alkaline Phosphatase; ALTV – Alanine Aminotransferase; AST- Aspartate Aminotransferase; BUN Blood Urea Nitrogen. With regards to total turnaround time (TAT), the XT7600 had a TAT of 31 min and 2 s compared to 69 min 2 s on the 5600. Similarly, the volume of samples utilized by the XT7600 was 53% less than the volume required by the 5600 (Table 1). All assays were included in the interference studies however, only analytes significantly affected by an interference were reported. The ALB, AST, TBIL and TP assays were affected by mild, moderate and significant hemolysis on the single MicroSlides and XT MicroSlide assays. At mild hemolysis (H-index <250), ALB, AST, TBIL and TP showed a positive bias of 11%, 51%, 1 mg/dL and 11% respectively. Glucose was affected by significant hemolysis with a positive bias of 18%. TP was significantly affected by mild, moderate and significant bilirubin with a positive bias >20%. ALKP was significantly affected by moderate and significant bilirubin with a positive bias >30%. Glucose was affected by intralipid interference on both types of assays with a positive bias of 12–13% (Table 3).
Table 3

Effect of hemolysis, icterus and lipemia on analytes on the Vitros 5600 vs Vitros XT 7600.

AnalyteUnitsVitros 5600
Vitros XT 7600
BaselineMild hemolysisModerate hemolysisSignificant hemolysisBaseline valuesMild hemolysisModerate hemolysisSignificant hemolysis
Hemolysis index values<15209384883<15166328860
ALBg/dL3.5 ± 0.04.0 ± 0.04.5 ± 0.05.6 ± 0.13.5 ± 0.04.0 ± 0.04.5 ± 0.05.6 ± 0.0
ALKPU/l114 ± 1.4102 ± 1.497.5 ± 3.591.5 ± 4.9111.5 ± 1.494.5 ± 3.591.5 ± 0.770.5 ± 7.8
ASTU/l43.5 ± 0.766.0 ± 0.089.0 ± 0.0NR44.5 ± 0.768.5 ± 0.791.5 ± 0.7NR
Glucosemg/dL74.0 ± 0.075.0 ± 0.079.0 ± 0.087.0 ± 0.073.0 ± 0.075.0 ± 0.079.0 ± 0.088.0 ± 0.0
TBILmg/dL0.6 ± 0.01.6 ± 0.02.7 ± 0.0NR0.6 ± 0.01.8 ± 0.12.8 ± 0.0NR
TPg/dL6.1 ± 0.16.9 ± 0.17.7 ± 0.19.3 ± 0.06.2 ± 0.07.0 ± 0.07.8 ± 0.19.5 ± 0.1
BaselineMild icterusModerate icterusSignificant icterusBaselineMild icterusModerate icterusSignificant icterus
Icterus index<2715>25<2516>25
ALKPU/l105.0 ± 0.0127.0 ± 1.4163.5 ± 3.5230.0 ± 3.5106.0 ± 0.0126.0 ± 1.4158.0 ± 2.8209.5 ± 2.8
TPg/dL5.5 ± 0.05.8 ± 0.06.0 ± 0.16.5 ± 0.05.9 ± 0.06.2 ± 0.16.4 ± 0.16.8 ± 0.0
BaselineMild lipemiaModerate lipemiaSignificant lipemiaBaseline valuesMild lipemiaModerate lipemiaSignificant lipemia
Lipemia index values<2036133326<2043127294
Glucosemg/dL64.5 ± 0.765.5 ± 0.767.5 ± 0.773.0 ± 0.760.5 ± 0.761.0 ± 0.062.5 ± 0.767.5 ± 0.7

Data provided as mean ± SD; ALB – Albumin; ALKP – Alkaline Phosphatase; TBIL - Total bilirubin; TP – Total protein; NR - Not reported.

Effect of hemolysis, icterus and lipemia on analytes on the Vitros 5600 vs Vitros XT 7600. Data provided as mean ± SD; ALB – Albumin; ALKP – Alkaline Phosphatase; TBIL - Total bilirubin; TP – Total protein; NR - Not reported.

Discussion

This report describes the method verification of the XT MicroSlide assays: ALTV-AST, TRIG-CHOL, ALB-TP, UREA-CREA, GLU-CA, TBIL-ALKP on the Vitros XT 7600 digital chemistry analyzer in our pediatric population. The XT Microslide assays demonstrated acceptable intra-assay precision and linearity across clinically relevant concentrations of all 12 analytes. Analytical performance evaluation is essential when introducing newly developed assays and instruments to the clinical laboratory to replace existing ones. Ensuring that the new instruments and/or assays provide reliable test values comparable to the instruments being replaced is critical. The XT MicroSlide and single slide assays have multilayered analytical elements coated on a polyester support with similar methodologies for corresponding analytes. Our method comparison studies showed a good correlation between the XT MicroSlides and the single assays. We did observe a small negative percent bias in the XT ALKP and TRIG assays compared to the single slides, −4.83% and −8.42%, respectively. The sample volume required for all analytes on the XT7600 was significantly lower that the volume required for the analytes on the vitro 5600. Insufficient sample volume is a common cause of incomplete laboratory analysis in the pediatric population and phlebotomy has been identified as a major cause of hospital acquired anemia in hospitalized patients. A prospective multicenter study further reported that blood draws accounted for 73% of daily blood loss which led to blood transfusions in hospitalized pediatric patients [[5], [6], [7], [8]]. Therefore, the development of chemistry analyzers like the XT7600 that can accommodate small sample volumes is advantageous for clinical laboratories that serve special populations like pediatrics. Likewise, the introduction of dual slides that allows simultaneous analysis of two analytes for the time allotted for one single slide significantly improved efficiency. There is a lack of comparative data for the XT MicroSlides. However, reports from the manufacturer on the analytical performance of the XT MicroSlide assays using sigma metric methodology demonstrated excellent precision and accuracy when evaluated against the CLIA TEa requirement [9]. Similarly, the comparison of within-lab precision for the XT MicroSlides on the XT 7600 with corresponding single test slides on the Vitros 5600 showed that the XT MicroSlides exhibited comparable or improved precision relative to single test slides [10]. The Vitros 5600 and XT 7600 integrated systems also eliminate carryover using disposable versa tips and disposable cuvettes, although this claim was not verified in this study. ALB, AST, GLU, TBIL, and TP assays were affected by hemolysis. This observation is in congruent with previous findings that hemolysis can increase the serum concentrations of analytes such as AST, LDH and potassium [11]. Although the hemolysis index result for the Vitros XT 7600 was slightly lower than that of the Vitros 5600, the effect of hemolysis on the analytes was similar on both instruments. Glucose was affected by severe lipemia beyond the approved total allowable error recommended by CAP. Of note, the manufacturer does not recommend using grossly lipemic samples with the glucose assay because of lipid interference with assay reactants [12]. There are some limitations in the study which need to be pointed out: i) small number of samples used for comparison (n = 24) due to the pediatric population; ii)dual slides may go against choosing wisely practices; for example cholesterol wouldn't always need to be tested along with TG in all settings. In summary, the XT MicroSlide assays on the XT 7600 demonstrated comparable analytical measurement ranges, precision, and strong method correlation with the Single MicroSlides for all the assays. It is suitable for use in the pediatric setting as it accommodates lower sample volumes and improved turnaround times without any effect on performance.

Credit author statement

Lily Olayinka: Conceptualization, Methodology, Validation, Writing-Original Draft. Estella Tam: Conceptualization, Methodology, Validation. Sridevi Devaraj: Conceptualization, Writing-Original Draft, Supervision.

Declaration of competing interest

None.
  8 in total

1.  Identifying factors to minimize phlebotomy-induced blood loss in the pediatric intensive care unit.

Authors:  Stacey L Valentine; Scot T Bateman
Journal:  Pediatr Crit Care Med       Date:  2012-01       Impact factor: 3.624

2.  Effect of interference from hemolysis, icterus and lipemia on routine pediatric clinical chemistry assays.

Authors:  Shweta Agarwal; German Vargas; Cristina Nordstrom; Estella Tam; Gregory J Buffone; Sridevi Devaraj
Journal:  Clin Chim Acta       Date:  2014-08-13       Impact factor: 3.786

3.  The Effect of Diagnostic Blood Loss on Anemia and Transfusion Among Postoperative Patients With Congenital Heart Disease in a Pediatric Intensive Care Unit.

Authors:  Dan Zhou; Yu-Lan Luo; Shu-Hua Luo; Mei Feng; Meng-Lin Tang
Journal:  J Pediatr Nurs       Date:  2017-11-04       Impact factor: 2.145

4.  Conflicting effects of haemolysis on plasma sodium and chloride are due to different haemolysis study protocols: A case for standardisation.

Authors:  Tejas Kalaria; Henry Gill; Hayley Sharrod-Cole; Clare Ford; Rousseau Gama
Journal:  Ann Clin Biochem       Date:  2021-11-06       Impact factor: 2.057

5.  Blood loss from laboratory testing, anemia, and red blood cell transfusion in the intensive care unit: a retrospective study.

Authors:  Nicholas L Jackson Chornenki; Tyler E James; Rebecca Barty; Yang Liu; Bram Rochwerg; Nancy M Heddle; Deborah M Siegal
Journal:  Transfusion       Date:  2019-12-28       Impact factor: 3.157

6.  A retrospective study investigating the incidence and predisposing factors of hospital-acquired anemia.

Authors:  Peter C Kurniali; Stephanie Curry; Keith W Brennan; Kim Velletri; Mohammed Shaik; Kenneth A Schwartz; Elise McCormack
Journal:  Anemia       Date:  2014-12-21

7.  Iatrogenic blood loss in extreme preterm infants due to frequent laboratory tests and procedures.

Authors:  Clare E Counsilman; Lisanne E Heeger; Ratna Tan; Vincent Bekker; Jaap Jan Zwaginga; Arjan B Te Pas; Enrico Lopriore
Journal:  J Matern Fetal Neonatal Med       Date:  2019-10-06

8.  Avoidable Blood Loss in Critical Care and Patient Blood Management: Scoping Review of Diagnostic Blood Loss.

Authors:  Philipp Helmer; Sebastian Hottenrott; Andreas Steinisch; Daniel Röder; Jörg Schubert; Udo Steigerwald; Suma Choorapoikayil; Patrick Meybohm
Journal:  J Clin Med       Date:  2022-01-10       Impact factor: 4.241

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.