Conflicting effects of haemolysis on plasma sodium and chloride are due to different haemolysis study protocols: A case for standardisation.

Tejas Kalaria1, Henry Gill1, Hayley Sharrod-Cole1, Clare Ford1, Rousseau Gama1,2.   

Abstract

BACKGROUND: Haemolysis has been reported as having a positive, negative or no effect on plasma sodium (PNa) and chloride (PCl). We investigated the haemoltytic effect of different haemolysis protocols on PNa and PCl using modelling and laboratory experiments.
METHODS: In a modelling experiment, percentage change and recovery due to dilution in routinely (in vitro) haemolysed samples were compared against shear stress haemolysis and samples spiked with haemolysate from whole blood freeze-thaw, packed cells freeze-thaw and osmotic shock protocols. The results were compared against a control base pool. Additionally, for the osmotic shock method, results were compared against saline- and deionised water (DIW)-spiked controls. In a laboratory experiment, percentage change and recovery were similarly compared using haemolysate from whole blood freeze-thaw and osmotic shock protocols. PNa, PCl and H-index were measured on the Abbott Architect and haemoglobin on the Sysmex XN-9000.
RESULTS: In the modelling experiment, the percentage decrease in PNa and PCl was similar in in vitro haemolysis, shear stress haemolysis, whole blood freeze-thaw haemolysis and packed cells freeze-thaw haemolysis and this was lower compared to the osmotic shock method. In the laboratory experiment, the change in PNa compared to the base pool was less (p < 0.001) per unit increase in H-index in the freeze-thaw method (-0.33 mmol, 95% CI -0.35 to -0.31) compared to the osmotic shock method (-0.65 mmol, 95% CI -0.66 to -0.64). PCl did not change with haemolysis in the freeze-thaw method and changed by -0.21 ± 0.01 mmol per unit increase in the H-index in the osmotic shock method. Recovery of PNa and PCl increased with increasing H-index in both methods.
CONCLUSION: The osmotic shock protocol is inappropriate for haemolysis studies because of dilution with DIW used for cell lysis. Recovery calculations may incorrectly compensate for genuine dilution caused by haemolysis.

Entities:  

Keywords:  Sodium; chloride; effect; haemolysis; influence; interference

Mesh:

Substances:

Year:  2021        PMID: 34747198     DOI: 10.1177/00045632211040691

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  1 in total

1.  Performance evaluation of the XT MicroSlide assay pairs on the Vitros XT 7600 compared to VITROS single microslide assays on Vitros 5600.

Authors:  Lily Olayinka; Estella Tam; Sridevi Devaraj
Journal:  Pract Lab Med       Date:  2022-05-23
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.