| Literature DB >> 35635553 |
Jeanette Henkelmann1, Ralf Henkelmann2, Timm Denecke3, Dirk Zajonz2,4, Andreas Roth2, Osama Sabri5, Sandra Purz5.
Abstract
PURPOSE: This study investigated the diagnostic value of simultaneous 18F-fluordeoxyglucose positron emission tomography/magnetic resonance imaging (PET/MRI) in suspected periprosthetic joint infection (PJI) of the hip and knee.Entities:
Keywords: Hip and knee arthroplasty; Magnetic resonance imaging; Periprosthetic infections; Positron emission tomography
Mesh:
Substances:
Year: 2022 PMID: 35635553 PMCID: PMC9372014 DOI: 10.1007/s00264-022-05445-7
Source DB: PubMed Journal: Int Orthop ISSN: 0341-2695 Impact factor: 3.479
Scanning parameters. TE, echo time; TR, repetition time; FOV, field of view
| Scanning parameters | Coronal TIRM | Coronal T1W sequence | Axial T1W sequence | Axial T2W sequence | Coronal T1 post-contrast |
|---|---|---|---|---|---|
| TR (ms) | 8000 | 750 | 800 | 4000 | 750 |
| TR (ms) | 22 | 8.1 | 8.1 | 73 | 8.1 |
| Slice thickness (mm) | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 |
| FOV (mm × mm) | 400 × 400 | 400 × 400 | 400 × 400 | 400 × 400 | 400 × 400 |
| Bandwidth (Hz/px) | 504 | 797 | 797 | 507 | 797 |
| Flip angle (°) | 146 | 135 | 136 | 141 | 135 |
| Acquisition time (s) | 3:30 | 3:23 | 3:29 | 2:42 | 3:23 |
Patient characteristics. ASA, Physical Status Classification of the American Society of Anesthesiologists; CCI, Charlson Comorbidity Index
| Patient no | Sex | Age | CCI | ASA | Prosthesis localisation | Prosthesis no | |
|---|---|---|---|---|---|---|---|
| 1 | F | 80 | 7 | 3 | Hip | Right | 1 |
| 2 | M | 77 | 8 | 3 | Hip bilateral | Right | 2 |
| Left | 3 | ||||||
| 3 | M | 55 | 4 | 2 | Hip | Left | 4 |
| 4 | M | 53 | 2 | 1 | Hip | Right | 5 |
| 5 | F | 70 | 7 | 3 | Hip | Left | 6 |
| 6 | F | 74 | 4 | 2 | Knee | Right | 7 |
| 7 | F | 75 | 5 | 2 | Knee | Left | 8 |
| 8 | M | 73 | 4 | 2 | Knee bilateral | Right | 9 |
| Left | 10 | ||||||
| 9 | F | 54 | 4 | 2 | Hip | Right | 11 |
| 10 | M | 58 | 4 | 3 | Hip bilateral | Right | 12 |
| Left | 13 |
Fig. 1A X-ray of a 55-year-old man with bilateral hip prostheses and left-sided cerclage. B Coronal T1-weighted magnetic resonance imaging (MRI) shows slight metal artefacts in both prostheses and left-sided fluid collections around the neck of the prosthesis. Fused.18F-FDG-PET/MRI (C) and the maximum intensity projection image (D) show significant elevated tracer uptake around the prostheses neck, shaft, and soft tissue. In contrast, the left non-infected hip prosthesis reveals a normal tracer pattern. The findings were confirmed by reference standard
Fig. 2A X-ray of a 54-year-old man with left-sided hip prosthesis. B Coronal T1-weighted magnetic resonance imaging (MRI) shows distinct subfacial fluid collections extending around the prostheses neck. Fused.18F-FDG-PET/MRI (C + D) shows an elevated tracer uptake around the prosthesis neck, shaft, and soft tissue. The increased uptake in the periprosthetic bone marrow of the femur (arrow) was confirmed as inflammatory involvement by prosthesis revision and sonication
Diagnostic accuracy for the diagnosis of a periprosthetic joint infection by positron emission tomography/magnetic resonance imaging separated for the predefined anatomical areas. *The analysis of the anatomical regions includes the results of the surgically obtained samples (10 each of the periprosthetic bone margin and soft tissues and 7 samples of the bone marrow)
| Diagnostic accuracy of PET/MRI for the diagnosis of PJI | Anatomical region* | Overall hip or knee arthroplasty | ||
|---|---|---|---|---|
| Periprosthetic bone margin | Bone marrow | Soft tissue | ||
| Sensitivity (%) | 57.1 | 75 | 100 | 100 |
| Specificity (%) | 50.0 | 33.3 | 100 | 100 |
| PPV (%) | 80.0 | 60.0 | 100 | 100 |
| NPV (%) | 25.0 | 50.0 | 100 | 100 |
| True positive | 4 | 3 | 9 | 10 |
| True negative | 1 | 1 | 1 | 3 |
| False positive | 1 | 2 | 0 | 0 |
| False negative | 3 | 1 | 0 | 0 |
| Not evaluable | 1 | 0 | 0 | 0 |