| Literature DB >> 35631349 |
Pauline Orhon1, Marie-Dominique Desruet1,2, Marie Piquemal1, Nicolas De Leiris2,3, Loïc Djaileb2,3, Jean-Philippe Vuillez2,3, Pierrick Bedouch1,4, Julien Leenhardt1,2.
Abstract
Prostate Specific Membrane Antigen (PSMA) is a highly relevant target in nuclear medicine due to its overexpression in prostate cancer. The 68Ga/177Lu-PSMA-1 combination is a theranostic agent for the detection and treatment of tumors overexpressing the PSMA target. Specifically, 177Lu-PSMA-1 is used in the treatment of castration-resistant prostate cancer that is ineffective or intolerant to the latest generation of chemotherapy and/or hormone therapy. This radiopharmaceutical is manufactured in a radiopharmaceutical synthesizing unit and must pass a quality control where the radiochemical purity (RCP) is assessed prior to release of the batch. RCP evaluation is performed by high-performance liquid chromatography (HPLC) and thin-layer chromatography (TLC). Since there is no monograph for 177Lu-PSMA-1 in the European Pharmacopoeia, we validate the analytical methods according to the EANM recommendations adapted from ICH Q2. Specificity, linearity, accuracy, precision, intermediate precision, limit of quantification (LOQ) and robustness were described for HPLC and TLC in this study. The results obtained demonstrated the robustness and reliability of the HPLC and TLC analytical methods for the evaluation of the RCP of 177Lu-PSMA-1.Entities:
Keywords: 177Lu-PSMA-1; high-performance liquid chromatography (HPLC); prostate cancer; radiochemical purity; radioligand therapy; thin-layer chromatography (TLC); validation of analytical methods
Year: 2022 PMID: 35631349 PMCID: PMC9143309 DOI: 10.3390/ph15050522
Source DB: PubMed Journal: Pharmaceuticals (Basel) ISSN: 1424-8247
Figure 1Retention time of the 175Lu-PSMA I/T standard.
Figure 2Retention time of 177Lu-PSMA-1 complex.
Retention time 117Lu-PSMA, retention time for free 177Lu impurity, peak area, tailing factor, plate count and resolution in HPLC.
| 177Lu-PSMA-1 | Free 177Lu | Peak Area: | Tailing Factor | Plate Count | Resolution | |
|---|---|---|---|---|---|---|
| Mean | 17.32 | 2.32 | 108 895 | 1.25 | 6725.70 | 17.73 |
| Standard | 0.35 | 0.35 | 9270 | 0.08 | 153.60 | 0.44 |
| Variation | 0.05 | 0.38 | 8.51 | 1.18 | 2.28 | 2.50 |
Retardation factor and resolution in TLC.
| 177Lu-PSMA-1 | Free 177Lu | Resolution (Rs) | |
|---|---|---|---|
| Mean | 0.930 | 0.142 | 3.060 |
| Standard | 0.020 | 0.001 | 0.400 |
| Variation | 1.730 | 0.700 | 13.600 |
Accuracy with % recovery in HPLC.
| % 177Lu-PSMA-1 | Recovered RCP | Theoretical RCP | % Recovery |
|---|---|---|---|
| 100 | 99.72 | 99.72 | 100.00 |
| 90 | 90.29 | 89.75 | 100.61 |
| 80 | 80.57 | 79.78 | 101.00 |
| 70 | 70.81 | 69.80 | 101.44 |
| 60 | 60.83 | 59.83 | 101.67 |
Accuracy with % recovery in TLC.
| % 177Lu-PSMA-1 | Recovered RCP | Theoretical RCP | % Recovery |
|---|---|---|---|
| 100 | 99.42 | 99.42 | 100.00 |
| 90 | 89.75 | 89.47 | 100.30 |
| 80 | 80.07 | 79.53 | 100.67 |
| 70 | 69.68 | 69.59 | 100.12 |
| 60 | 58.74 | 59.65 | 98.47 |
Precision/repeatability and intermediate precision in HPLC.
|
|
|
|
|
|
| Repeatability | 98.52 ± 0.15 | 0.15 | 17.53 ± 0.04 | 0.24 |
|
|
|
|
|
|
| Intermediate precision: | 98.79 ± 0.72 | 0.73 | 17.31 ± 1.18 | 0.12 |
Precision/repeatability and intermediate precision in TLC.
|
|
|
|
|
|
| Repeatability | 99.84 ± 0.03 | 0.03 | 0.93 ± 0.01 | 1.26 |
|
|
|
|
|
|
| Intermediate precision: | 99.12 ± 0.89 | 0.90 | 0.92 ± 0.03 | 2.99 |
Figure 3Linearity in HPLC method.
Figure 4Linearity in TLC method.
Summary table of criteria and their results.
| Test | Acceptance Criteria | HPLC Results | TLC Results |
|---|---|---|---|
| % Recovery | 90% < % recovery < 110% | 100.00% to 101.67% | 98.47–100.67% |
| Specificity | Resolution factor (Rs) > 2 | 17.73 | 3.060 |
| Linearity | R2 > 0.99 | 0.9978 | 0.9997 |
| Intermediate Precision | RCP: RSD < 2% | 0.73 | 0.90 |
| Retardation time: RSD < 2% | 0.12 | 2.99 | |
| Repeatability | RCP: RSD < 2% | 0.15 | 0.03 |
| Retardation time: RSD < 2% | 0.24 | 1.26 | |
| Robustness | RSD < 2% | 0.06% (condition A) | 0.51 |
| Quantification limit (LoQ) | Signal-to-noise ratio ≥ 10 | 33.8 kBq (12.7:1) | 61.8 kBq (10.8:1) |
| Range | Reported value | 2.5 to 500 MBq/mL | |