Nic Gillings1, Sergio Todde2, Martin Behe3, Clemens Decristoforo4, Philip Elsinga5, Valentina Ferrari6, Olaug Hjelstuen7, Petra Kolenc Peitl8, Jacek Koziorowski9, Peter Laverman10, Thomas L Mindt11, Meltem Ocak12, Marianne Patt13.
Abstract
BACKGROUND: To fulfil good manufacturing requirements, analytical methods for the analysis of pharmaceuticals for human and vetinary use must be validated. The International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) has published guidance documents on the requirements for such validation activities and these have been adopted by the European Medicines Agency, The U.S. Food and Drug Administration (FDA) and other regulatory bodies. These guidance documents do not, however, fully address all the specific tests required for the analysis of radiopharmaceuticals. This guideline attempts to rectify this shortcoming, by recommending approaches to validate such methods.Entities:
Keywords: Radioanalytical methods; Radiopharmaceuticals; Validation
Year: 2020 PMID: 32052212 PMCID: PMC7016057 DOI: 10.1186/s41181-019-0086-z
Source DB: PubMed Journal: EJNMMI Radiopharm Chem ISSN: 2365-421X
Characteristics to be validated following ICH Q2(R1)
| Type of analytical procedure | ||||
|---|---|---|---|---|
| Identification | Testing for impurities | Assay | ||
| Quantitative | Limit | Dissolution Measurement only | ||
| Characteristic | ||||
| Accuracy | – | + | – | + |
| Precision | ||||
| Repeatability | – | + | – | + |
| Intermediate Precision | – | +a | – | +a |
| Specificityb | + | + | + | + |
| Detection Limit | – | -c | + | – |
| Quantification Limit | – | + | – | – |
| Linearity | – | + | – | + |
| Range | – | + | – | + |
– signifies that this characteristic is not normally evaluated
+ signifies that this characteristic is normally evaluated
ain cases where reproducibility has been tested, intermediate precision is not needed
black of specificity of one analytical procedure could be compensated by other supporting analytical procedure(s)
cmay be needed in some cases
ICH table adapted to radiopharmaceuticals
| Type of analytical procedure | Radioactivity Content (assay) | Radionuclide identity (approx. t½) | Radionuclide identity (spectrometry) | Radiochemical identity (HPLC/TLC) | Radionuclidic purity (limit test) | Radionuclidic purity (spectrometry after decay) | Radiochemical puritya (HPLC/TLC) |
|---|---|---|---|---|---|---|---|
| Characteristics | |||||||
| Accuracy | + | – | + | – | + | + | + |
| Precision (Repeatability) | + | + | – | – | – | (+) | (+) |
| Intermediate Precision | – | – | – | – | – | (+) | (+) |
| Specificity | + | + | + | + | + | + | + |
| Detection Limit | – | – | – | – | + | – | – |
| Quantification Limit | – | – | – | – | – | + | + |
| Linearity | + | + | – | – | – | + | + |
| Range | + | + | – | – | – | + | + |
(+) not always possible (e.g. short half life, see text)
aradioenantiomeric purity measurements should be validated analogously
| Type of analytical procedure | Radiochemical purity (HPLC/TLC/PC) |
|---|---|
| Characteristics | |
| Accuracy | + |
| Precision (Repeatability) | (+) |
| Intermediate Precision | (+) |
| Specificity | + |
| Detection Limit | – |
| Quantification Limit | + |
| Linearity | + |
| Range | + |
+ normally evaluated
(+) - not always possible
| Type of analytical procedure | Radiochemical purity (HPLC/TLC/PC) |
|---|---|
| Characteristics | |
| Accuracy | + |
| Precision (Repeatability) | (+) |
| Intermediate Precision | (+) |
| Specificity | + |
| Detection Limit | – |
| Quantification Limit | + |
| Linearity | + |
| Range | + |
+ normally evaluated
(+) - not always possible
| Accuracy | |||||
|---|---|---|---|---|---|
| Sample | Measured [18F]fluoride content (%) | True [18F]fluoride content (%) | Recovery (%) | ||
| 1 | 2 | Average | |||
| 1 | 1.10 | 1.00 | 1.05 | 1.00 | 105.0 |
| 2 | 2.97 | 3.14 | 3.06 | 2.92 | 104.6 |
| 3 | 4.90 | 5.10 | 5.00 | 4.72 | 105.9 |
| Repeatability | |
|---|---|
| Measurement no. | Measured [18F]fluoride (%) |
| 1 | 5.2 |
| 2 | 5.3 |
| 3 | 5.3 |
| 4 | 5.1 |
| 5 | 5.4 |
| 6 | 5.3 |
| Average (%) | 5.27 |
| Standard deviation | 0.10 |
| CV% | 1.96 |
| Intermediate Precision | |
|---|---|
| Analyst | Measured [18F]fluoride (%) |
| 1 | 5.1 |
| 2 | 5.4 |
| 3 | 5.3 |
| Average (%) | 5.27 |
| Standard deviation | 0.15 |
| CV% | 2.90 |
| Specificity | |
|---|---|
| Resolution of [18F]fluoride and [18F]FET | Rs = 5.7 |
| Limit of Quantification | |
|---|---|
| LOQ | 24 kBq/ml (S/ |
| Linearity | |||
|---|---|---|---|
| Sample | Radioactivity concentration (MBq/ml) | Peak area (decay-corrected) | Correlation coefficient (R) |
| 1 | 12.3 | 27,987 | 0.995 |
| 2 | 61.6 | 124,164 | |
| 3 | 123.2 | 209,463 | |
| 4 | 616.2 | 979,290 | |
| 5 | 1232.5 | 2,407,382 | |
| Validation result summary | ||||
|---|---|---|---|---|
| Test Parameter | Acceptance Criteria | Result | ||
| Accuracy | Recovery 90–110% ([18F]FET spiked with [18F]fluoride) | [18F]fluoride concentration | Recovery | |
| 1% | 105.0% | |||
| 3% | 104.6% | |||
| 5% | 105.9% | |||
| Repeatability | 6 repetitions (5% [18F]fluoride) | RSD ≤5% | RSD = 1.96% | |
| Intermediate Precision | RSD ≤5% (3 analysts) | RSD = 2.90% | ||
| Specificity | Resolution of > 2 between peaks | Resolution of[18F]fluoride and [18F]FET | Rs = 5.7 | |
| Limit of Quantification | S/N ratio ≥ 10 | 24 kBq/ml | ||
| Linearity | 12–1232 kBq/ml (5 concentrations in triplicate) | R > 0.99 | R = 0.995 | |
| Range | Reported value | 4.8–1232 MBq/ml | ||
| Type of analytical procedure | Radionuclidic purity (spectrometry after decay) |
|---|---|
| Characteristics | |
| Accuracy | + |
| Precision (Repeatability) | (+) |
| Intermediate Precision | (+) |
| Specificity | + |
| Detection Limit | – |
| Quantification Limit | + |
| Linearity | + |
| Range | + |
+ normally evaluated
(+) - not always possible
| Product | T1/2 | Nuclear reaction | Threshold (MeV) |
|---|---|---|---|
| Co-55 | 17.5 h | 58Ni(p,α)55Co | 1.36 |
| Co-56 | 77 d | 56Fe(p,n)56Co | 5.44 |
| Co-57 | 272 d | 57Fe(p,n)57Co | 1.65 |
| 60Ni(p,a)57Co | 0.27 | ||
| 58Ni(p, 2p)57Co | 8.31 | ||
| Co-58 | 71 d | 58Fe(p,n)58Co | 3.14 |
| Ni-57 | 35.6 h | 58Ni(p,pn)57Ni | 12.43 |
| Cr-51 | 27.7 d | 52Cr(p,pn)51Cr | 12.27 |
| Mn-52 | 5.6 d | 52Cr(p,n)52Mn | 5.60 |
| Tc-95 | 20 h | 95Mo(p,n)95Tc | 2.50 |
| Tc-96 | 4.3 d | 96Mo(p,n)96Tc | 3.30 |
| Re-181 | 19.9 h | 182W(p, 2n)181Re | 10.65 |
| Mo-93 m | 6.85 h | 93Nb(p,n)93mMo | 3.60 |
| Accuracy | ||
|---|---|---|
| Expected activity: 2545.8 Bq | ||
| Measurement no. | Measured activity (Bq) | Recovery (%) |
| 1 | 2666.2 | 104.7 |
| 2 | 2722.9 | 106.9 |
| 3 | 2636.2 | 103.5 |
| 4 | 2714.8 | 106.6 |
| 5 | 2654.0 | 104.2 |
| 6 | 2654.3 | 104.2 |
| Repeatability | |||
|---|---|---|---|
| Test no. 1 | Starting [18F]FLT activity at T0: 370 Bq | ||
| Measurement no. | Measured activity (Bq) | Decay corrected activity (Bq) | |
| 1 (T0) | 370 | 370 | |
| 2 (T1) | 362 | 372 | |
| 3 (T2) | 353 | 371 | |
| 4 (T3) | 341 | 368 | |
| 5 (T4) | 328 | 363 | |
| 6 (T5) | 326 | 370 | |
| Average | 369.4 | Standard deviation | 3.2 |
| CV% | 0.86% | ||
| Repeatability | |||
|---|---|---|---|
| Test no. 2 | Starting [18F]FLT activity (at T0): 1320 Bq | ||
| Measurement no. | Measured activity (Bq) | Decay corrected activity | |
| 1 (T0) | 1320 | 1320 | |
| 2 (T1) | 1279 | 1311 | |
| 3 (T2) | 1240 | 1305 | |
| 4 (T3) | 1225 | 1322 | |
| 5 (T4) | 1131 | 1307 | |
| 6 (T5) | 1158 | 1314 | |
| Average | 1313 | Standard deviation | 6.8 |
| CV% | 0.52% | ||
| Repeatability | |||
|---|---|---|---|
| Test no. 3 | Starting [18F]FLT activity at T0: 3021 Bq | ||
| Measurement no. | Measured activity (Bq) | Decay corrected activity (Bq) | |
| 1 (T0) | 3021 | 3021 | |
| 2 (T1) | 2933 | 3008 | |
| 3 (T2) | 2853 | 3001 | |
| 4 (T3) | 2793 | 3013 | |
| 5 (T4) | 2710 | 2998 | |
| 6 (T5) | 2659 | 3017 | |
| Average | 3010 | Standard deviation | 9.1 |
| CV% | 0.30% | ||
| Intermediate Precision | |||
|---|---|---|---|
| Measurement no. | Radioactivity (Bq) | ||
| Day 1 | Day 2 | Day 3 | |
| 1 | 895.0 | 858.0 | 870.0 |
| 2 | 867.0 | 868.0 | 858.0 |
| 3 | 867.0 | 879.0 | 868.0 |
| 4 | 868.0 | 866.0 | 869.0 |
| 5 | 871.0 | 880.0 | 879.0 |
| 6 | 875.0 | 865.0 | 857.0 |
| Average | 873.83 | 869.33 | 866.83 |
| Variance | 116.9667 | 73.4667 | 67.7667 |
| Standard deviation | 10.81511 | 8.57127 | 8.23205 |
| CV% | 1.24% | 0.99% | 0.95% |
| Variance analysis | |||
|---|---|---|---|
| Type of variation | Degrees of freedom (DF) | Average squares (AS) | Fcalc |
| Interaction between days | (n-1) | ASB | ASB/ASErr |
| Experimental error | n (r-1) | ASErr | |
| Total | nr-1 | ||
Results: The experimental value, Fcalc. was 0.877, while Fcritical was 3.68.
| Specificity | |||
|---|---|---|---|
| Composition of the multi-nuclide source (in order of increasing energy emissions): | |||
| Radionuclide | Energy (keV) | Resolution (Rs) | |
| 1 | Am-241 | 59.5 | Am-241/Cd-109: 1.82 |
| 2 | Cd-109 | 88 | Cd-109/Co-57: 1.92 |
| 3 | Co-57 | 122 | Sn-113/Co-57: 8.64 |
| 4 | Sn-113 | 392 | Cs-137/Sn-113: 4.47 |
| 5 | Cs-137 | 662 | Y-88(I)/Cs-137: 2.91 |
| 6 | Y-88 (peak I) | 898 | Co-60(I)/Y-88(I): 2.94 |
| 7 | Co-60 (peak I) | 1173 | Co-60(I)/Co-60(II): 1.44 |
| 8 | Co-60 (peak II) | 1837 | Co-60(I)/Y-88(II): 3.62 |
| Linearity | ||
|---|---|---|
| Starting activity of [18F]FLT: 16488 Bq | ||
| Radioactivity (Bq) | Peak area (decay-corrected) | Correlation coefficient (R) |
| 16,487.67 | 4,421,627 | 0.999 |
| 8629.16 | 2,314,149 | |
| 5644.31 | 1,513,678 | |
| 3712.19 | 995,526 | |
| 2241.00 | 654,623 | |
| 1615.31 | 433,190 | |
| 1054.80 | 282,874 | |
| 694.55 | 186,263 | |
| 460.02 | 123,367 | |
| 300.05 | 80,468 | |
| 194.77 | 71,629 | |
| 130.72 | 35,057 | |
| Intermediate Presicion | ||
|---|---|---|
| Mesurement no. | Co-57 radioactivity (Bq) | Co-57 radioactivity (Bq) |
| 1 | 34.41 | 198.54 |
| 2 | 36.89 | 182.17 |
| 3 | 37.60 | 182.81 |
| 4 | 34.66 | 184.41 |
| 5 | 37.47 | 182.45 |
| 6 | 30.98 | 177.46 |
| Average | 35.34 | 184.64 |
| Standard deviation | 2.54 | 7.20 |
| CV% | 7.2 | 3.9 |
| Validation result summary | ||||
|---|---|---|---|---|
| Test Parameter | Acceptance Criteria | Result | ||
| Accuracy | Recovery 90–110% | Measurement | Recovery | |
| 1 | 104.7% | |||
| 2 | 106.9% | |||
| 3 | 103.5% | |||
| 4 | 106.6% | |||
| 5 | 104.2% | |||
| 6 | 104.2% | |||
| Repeatability | 6 repetitions using [18F]FLT, 3 different activities | %CV ≤5% | Radioactivity | %CV |
| 370 Bq | 0.86% | |||
| 1320 Bq | 0.52% | |||
| 3021 Bq | 0.30% | |||
| Intermediate Precision | Fcalc. < Fcritical | Fcalc. = 0.877 | Fcritical = 3.68 | |
| Specificity | Resolution of > 1 between peaks | Peak Pair | Rs | |
| Am-241/Cd-109 | 1.82 | |||
| Cd-109/Co-57 | 1.92 | |||
| Sn-113/Co-57 | 8.64 | |||
| Cs-137/Sn-113 | 4.47 | |||
| Y-88(I)/Cs-137 | 2.91 | |||
| Co-60(I)/Y-88(I) | 2.94 | |||
| Co-60(I)/Co-60(II) | 1.44 | |||
| Co-60(I)/Y-88(II) | 3.62 | |||
| Limit of Quantification | Reported value | 185 Bq | ||
| Linearity | 12 samples of F-18 | R > 0.99 | R = 0.999 | |
| Range | Reported value | 185 Bq - 16 kBq | ||