Ignazio Condello1, Roberto Lorusso2,3, Giuseppe Santarpino1,4,5, Flavio Fiore1, Giuseppe Nasso1, Giuseppe Speziale1.
Abstract
Background: During Cardiopulmonary Bypass (CPB) gaseous micro-emboli (GMEs) decrease the quality of the blood flow and the capillary oxygen delivery, increasing the incidence of postoperative neurocognitive disorders (POCD) following cardiac surgery. In these circumstances, the use of an efficient device, could be crucial for the removal and reduction of micro-embolic activity.Entities:
Keywords: A.L.ONE AF PLUS Oxygenator; GAMPT BCC 300; cardiopulmonary bypass; gaseous micro-emboli; horizon venous reservoir; oxygenation performance; thermal exchange
Year: 2022 PMID: 35629790 PMCID: PMC9146211 DOI: 10.3390/membranes12050465
Source DB: PubMed Journal: Membranes (Basel) ISSN: 2077-0375
Figure 1Study population.
Figure 2Horizon AF PLUS, Eurosets (venous reservoir and oxygenator).
Figure 3A.L.ONE AF PLUS Oxygenator.
Figure 4Probes GAMPT BC300 positioning for micro-embolic activity evaluation on Horizon AF PLUS.
Figure 5Horizon AF PLUS during cardiopulmonary bypass use with GAMPT BC300.
Preoperative profile and operative data.
| Characteristic | Conventional Cardiac Surgery | Minimally Invasive Mitral Valve Repair |
|---|---|---|
| Mean age (y) | 69.9 | 72.5 |
| Male sex | 15 (50) | 6 (60) |
| Mean body surface area (m2) | 1.73 | 1.78 |
| Mean left ventricular ejection fraction (%) | 45 | 50 |
| Median NYHA functional class | 2 | 2 |
| EuroSCORE II (mean) | 2.1 | 2.4 |
| Pre-CPB hematocrit (%) | 34.4 ± 1.2 | 34.4 ± 1.7 |
| Pre-CPB Hb (g/dL) | 10.4 ± 1.1 | 10.8 ± 1.2 |
| No. of chronic obstructive pulmonary disease cases (mean) | 27 | 28 |
| Creatinine (mg/dL) | 1.11 ± 0.4 | 1.09 ± 0.5 |
| Obstructive coronary artery disease (%) | 15 | 0 |
Values are presented as n (%) or mean ± standard deviation. NYHA, New York Heart Association; EuroSCORE, European System for Cardiac Operative Risk Evaluation; CPB, cardiopulmonary bypass; Hb, hemoglobin.
Operative data.
| Parameter | Conventional Cardiac Surgery | Minimally Invasive Mitral Valve Repair | |
|---|---|---|---|
| CPB time (min) | 104 ± 11.1 | 102 ± 9.34 | 0.92 |
| Aortic cross-clamp time (min) | 78 ± 5 | 44 ± 6 | 0.75 |
| Nadir temperature (°C) | 34.9 ± 1.1 | 34.7 ± 2.1 | 0.75 |
| Nadir hemoglobin value (mg/dL) during CPB | 8.73 ± 1.53 | 8.6 ± 1.25 | 0.88 |
| Nadir hematocrit (%) | 26.6 ± 3.4 | 26.3 ± 3.9 | 0.89 |
| Nadir DO2i (mL/min/m2) during CPB | 294 ± 29 | 289 ± 14 | 0.99 |
| O2ERi (%) during CPB | 23 ± 1 | 23 ± 5 | 0.89 |
| Nadir CI (L/min/m2) | 2.5 ± 0.2 | 2.5 ± 0.1 | 0.91 |
| Nadir SvO2 (%) | 81 ± 2 | 80 ± 5 | 0.93 |
Values are presented as mean ± standard deviation. CPB, cardiopulmonary bypass.
Incidence of micro-embolic activity for procedures and duration.
| Unexpected Predisposing Factors for Micro-Embolic Activity | Conventional Cardiac Surgery | Minimally Invasive Mitral Valve Repair | Phenomena Duration (min) |
|---|---|---|---|
| Accidental air embolism from venous line ( | 1 | 3 | 3 ± 1 |
| Low levels in venous reservoir (<250 mL) in association with vacuum-assisted venous drainage (≥40 mmHg) | 8 | 2 | 8 ± 5 |
| Combination of two excessive suctions of aspirators (>2 L/min, >77 RPM) with low levels in venous reservoir (<250 mL) in association with vacuum-assisted venous drainage (≥40 mmHg) | 3 | 1 | 10 ± 2 |
| Absence of predisposing factors for micro-embolic activity | 18 | 4 | 0 |
Values are presented as n (%) or mean ± standard deviation. CPB, cardiopulmonary bypass.
Quantification of micro-embolic activity in the circuit for unexpected predisposing factors for micro-embolic activity on Horizon AF PLUS.
| Unexpected Predisposing Factors for Micro-Embolic Activity on Horizon AF PLUS | Venous Inlet Line of Venous Reservoir | Outlet Line of Venous Reservoir | Arterial Line |
|---|---|---|---|
| Accidental Air Embolism from venous line ( | |||
| Gaseous micro-emboli numbers | 349.299 ± 28 | 129.321 ± 60 | 1.235 ± 73 |
| Diameter >500 μm (%) | 45 | 3.5 | 0 |
| Volume (μL) | 79.9 ± 2 | 1.49 ± 5 | 0.89 ± 1 |
| Low levels in venous reservoir (<250 mL) in association with vacuum-assisted venous drainage (≥40 mmHg) ( | |||
| Gaseous micro-emboli numbers | 199.111 ± 76 | 35.720 ± 40 | 899 ± 47 |
| Diameter > 500 μm (%) | 26 | 1.6 | 0 |
| Volume (μL) | 19.2 ± 2 | 1.19 ± 6 | 0.57 ± 1 |
| Combination of two excessive suctions of aspirators (>2 L/min, >77 RPM) with low levels in venous reservoir (<250 mL) in association with vacuum-assisted venous drainage (≥40 mmHg) ( | |||
| Gaseous micro-emboli numbers | 230.889 ± 100 | 39.119 ± 25 | 1.001 ± 37 |
| Diameter >500 μm (%) | 29 | 1.9 | 0 |
| Volume (μL) | 23.8 ± 2 | 1.25 ± 3 | 0.69 ± 2 |
| Without predisposing factors for micro-embolic activity on Horizon AF PLUS | |||
| Gaseous micro-emboli numbers | 99.000 ± 35 | 18.000 ± 65 | 559 ± 56 |
| Diameter >500 μm (%) | 15 | 1.2 | 0 |
| Volume (μL) | 12.2 ± 3 | 0.97 ± 5 | 0.59 |
Values are presented as mean ± standard deviation.
Difference for gaseous micro-emboli volume (μL) in arterial line at the end of CPBs groups with Horizon AF PLUS.
| Unexpected Predisposing Factors for Micro-Embolic Activity on Horizon AF | Without Predisposing Factors for Micro-Embolic Activity on Horizon AF | ||
|---|---|---|---|
| Gaseous | 0.71 ± 0.2 | 0.59 ± 0.1 | 0.67 |
Values are presented as mean ± standard deviation. CPB, cardiopulmonary bypass.
Evaluation of oxygenation performance in terms of mean values of PaO2 and PaCO2 for the target gas blender of 2.0 ± 0.3 L/min and 55 ± 5% FiO2 during CPBs with Horizon AF PLUS.
| PaO2 mmHg | PaCO2 mmHg | |
|---|---|---|
| Unexpected predisposing factors for micro-embolic activity on Horizon AF PLUS | 260 ± 25 | 38 ± 4 |
| Without predisposing factors for micro-embolic activity on Horizon AF PLUS | 260 ± 23 | 38 ± 6 |
Values are presented as mean ± standard deviation. CPB, cardiopulmonary bypass.
Figure 6Graphic Representation of gaseous micro-emboli in: Numbers (n), Volume (μL), Diameter > 500 μm (%); with and without predisposing factors for micro-embolic activity during Horizon AF PLUS use.
Thermal exchange performance for mild hypothermia during CPBs with Horizon AF PLUS.
| Nasopharyngeal Temperature | From 36 °C to 34 °C | From 34 °C to 36 °C |
|---|---|---|
| Unexpected predisposing factors for micro-embolic activity on Horizon AF PLUS | 5.5 ± 2 | 6.5 ± 1 |
| Without predisposing factors for micro-embolic activity on Horizon AF PLUS | 5.5 ± 1 | 6.5 ± 2 |
Values are presented as mean ± standard deviation. HCD, heater-cooler device; CPB, cardiopulmonary bypass.