Wojciech Skrzeczanowski1, Maria Długaszek1.
Abstract
Results of laser-induced breakdown spectroscopy measurements of K, Na, Ca, and Mg content in liquid media are discussed in the paper. Calibration results show correct parameters-linearity and R2 coefficients of determination at the levels of 0.94-0.99. Obtained regression equations have been used to determine K, Na, Ca, and Mg concentrations in biological samples with known element content. Measurement results showed acceptable, within the expanded standard uncertainty, conformity with their content in the certified materials. Results have been supported by multivariate factorial analysis, which was especially effective for Ca and Mg samples. For these elements, factorial analysis allows the application of the whole spectra to obtain quantitative data on the tested samples, in contrast to a common method based on the selection of a particular spectral line for the calibration.Entities:
Keywords: calibration; factorial analysis; filter paper; laser-induced breakdown spectroscopy; liquid solution
Year: 2022 PMID: 35629762 PMCID: PMC9146648 DOI: 10.3390/ma15103736
Source DB: PubMed Journal: Materials (Basel) ISSN: 1996-1944 Impact factor: 3.748
Examples of the element content.
| Units | K | Na | Ca | Mg | Refs. | |
|---|---|---|---|---|---|---|
| Sea water | mg/L | 10556 | 380 | 400 | 1262 | [ |
| Surface water | mg/L | 16–63 | 7–24 | 58–178 * | 58–178 * | [ |
| Groundwater | mg/L | 6–27 | 18–110 | 162–280 | 25–74 | [ |
| Intracellular fluids | mmol/L | 156 | 10 | [ | ||
| Interstitial fluids | mmol/L | 3.8 | 144.0 | [ | ||
| Serum | mEq/L | 4 | 142 | 5 | 2 | [ |
| Red cells | mEq/L | 85.78–107.95 | 3.98–19.67 | 0.61–1.97 | 1.6–6.70 | [ |
| 24h urine excretion | mEq/L | 25–100 | 120–220 | 100–250 | 4–16 | [ |
* it is attributed to ref [3].
Examples of the element content.
| Units | K | Na | Ca | Mg | Refs. | |
|---|---|---|---|---|---|---|
| Tap water | mg/100 g | nd 1–20.4 | 0.1–39.1 | nd 1–10 | nd 1–6 | [ |
| Low-mineralized waters | mg/L | 1–56 | 4–15 | 1–110 | [ | |
| High-mineralized waters | mg/L | 900–1419 | 5–176 | 4–60 | [ | |
| Black tea infusion | mg/250 mL | 42.96 | 0.23 | 0.58 | 2.10 | [ |
| Coffee infusion | mg/L | 887.37–1547.70 | 24.74–27.81 | 16.34–25.71 | 77.15–116.30 | [ |
1 not detected.
Figure 1An exemplary view of a filter paper with Na solution after 3 laser shots.
Figure 2The diagram of the measurement stand.
Figure 3Emission spectra of Na I, K I, Ca II, and Mg II in the regions of interest.
Figure 4Double-line approximation of the data for K calibration.
Comparison of the results for K, Na, Ca, and Mg content determined by LIBS method in various CRM samples (μg/mL).
| K | Na | Ca | Mg | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| CRM | LIBS | Δ, % | CRM | LIBS | Δ, % | CRM | LIBS | Δ, % | CRM | LIBS | Δ, % | |
| SRM 1577b Bovine Liver | 92.8 | 104.6 ± 52.0 | 13 | 22.6 | 28.8 ± 6.6 | 27 | 1.1 | - | - | 5.6 | 5.3 | 5 |
| NCT-TL-1 | 165.0 | 157.0 ± 51.0 | 5 | 0.2 | - | - | 56.5 | 45.2 | 20 | 21.8 | 19.8 | 9 |
| NCS ZC91002 | 0.1 | - | - | 2.8 | 3.4 ± 0.9 | 21 | 10.7 | 11.35 | 6 | 1.0 | 1.7 | 70 |
| LOD | 2.8 | 3.4 | 3.1 | 3.3 | ||||||||
Figure 5The Ca LIBS spectra in the FA plane.
Figure 6The Mg LIBS spectra in the FA plane.