Literature DB >> 33592697

Al and Si quantitative analysis in aqueous solutions by LIBS method.

Wojciech Skrzeczanowski1, Maria Długaszek2.   

Abstract

In the paper results of Laser-Induced Breakdown Spectroscopy (LIBS) measurements of Al and Si content in aqueous solutions are presented. Calibration curves exhibit good linearity and satisfying R2 coefficients of determination (over 0.99). The calibration procedures were next used for the determination of Al and Si concentration in biological material transferred into water solutions of unknown Al and Si content. Obtained data were verified by the AAS method (atomic absorption spectrometry). Measurement results showed good conformity. The LIBS spectra were also processed using Factorial Analysis (FA). It was found for Si samples that instead of a classic approach based on the selection of a particular spectral line for the calibration the FA makes possible the use of entire spectra to obtain quantitative information on the investigated material.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Aluminium; Calibration; Laser-induced breakdown spectroscopy; Silicon

Year:  2020        PMID: 33592697     DOI: 10.1016/j.talanta.2020.121916

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057