Introduction: A challenge in the engineering of auto-adjusting prosthetic sockets is to maintain stable operation of the control system while users change their bodily position and activity. The purpose of this study was to test the stability of a socket that automatically adjusted socket size to maintain fit. Socket release during sitting was conducted between bouts of walking. Methods: Adjustable sockets with sensors that monitored distance between the liner and socket were fabricated. Motor-driven panels and a microprocessor-based control system adjusted socket size during walking to maintain a target sensed distance. Limb fluid volume was recorded continuously. During eight sit/walk cycles, the socket panels were released upon sitting and then returned to position for walking, either the size at the end of the prior bout or a size 1.0% larger in volume. Results: In six transtibial prosthesis users, the control system maintained stable operation and did not saturate (move to and remain at the end of the actuator's range) during 98% of the walking bouts. Limb fluid volume changes generally matched the panel position changes executed by the control system. Conclusions: Stable operation of the control system suggests that the auto-adjusting socket is ready for testing in users' at-home settings.
Introduction: A challenge in the engineering of auto-adjusting prosthetic sockets is to maintain stable operation of the control system while users change their bodily position and activity. The purpose of this study was to test the stability of a socket that automatically adjusted socket size to maintain fit. Socket release during sitting was conducted between bouts of walking. Methods: Adjustable sockets with sensors that monitored distance between the liner and socket were fabricated. Motor-driven panels and a microprocessor-based control system adjusted socket size during walking to maintain a target sensed distance. Limb fluid volume was recorded continuously. During eight sit/walk cycles, the socket panels were released upon sitting and then returned to position for walking, either the size at the end of the prior bout or a size 1.0% larger in volume. Results: In six transtibial prosthesis users, the control system maintained stable operation and did not saturate (move to and remain at the end of the actuator's range) during 98% of the walking bouts. Limb fluid volume changes generally matched the panel position changes executed by the control system. Conclusions: Stable operation of the control system suggests that the auto-adjusting socket is ready for testing in users' at-home settings.
Prosthesis users report that socket fit is the single-most important issue related to
use of their prosthesis.[1,2]
A primary source of socket fit problems is a change in residual limb volume. A
socket that automatically adjusts its size in response to residual limb volume
change and maintains fit may benefit people using a lower limb prosthesis. Unlike
traditional methods to accommodate volume fluctuation (e.g., changing sock ply;
manually adjusting socket panels; and hand pumping bladders), an auto-adjusting
socket relieves users of the burden of continually sensing their socket fit and
determining if and how much adjustment is needed. An auto-adjusting socket may
improve satisfaction and limb health and reduce the risk of a fall, particularly for
people with poor limb sensation who have difficulty sensing their socket fit.The first commercial auto-adjusting prosthetic socket was a purely mechanical system
that used a series of bladders and mechanical valves to adjust socket size.
Several research groups extended from this work to pursue electronic
automatic size-adjusting sockets that used a powered actuator to change socket size
and a pressure sensor for feedback control.[4-14] A handful achieved a
closed-loop control system,[5-7,10,13] but only
Pirouzi et al.
tested a system on people with lower limb amputation. Pirouzi et al.
demonstrated that vertical limb displacement at the posterior brimline of the
socket varied linearly with actuator pressure during standing with cyclic weight
bearing, suggesting that actuator pressure could be used to control vertical limb
displacement. However, no walking tests were conducted. In testing on a participant
with transhumeral amputation, Razak et al.’s
control system was unstable and had to be changed to open-loop operation to
avoid saturation, that is, the actuator moving to and staying at the end of its
range. More recently, Gu et al.’s
transhumeral control system, tested on an able-bodied participant, was shown
to maintain consistent limb contact pressures during lifting.We extended from this work to create an automatic movable-panel socket that adjusted
the socket size for people with transtibial amputation based on liner-to-socket
distance data collected using sensors embedded within the surrounding socket wall.
We term the measurement “sensed distance.” We used inductive distance sensing rather
than pressure sensing because it has better sensitivity and resolution, and during
testing on participants with transtibial amputation consistently demonstrated a
linear relationship with socket size.[15-17] The linear relationship
between the actuator variable (panel position) and the sensed variable (distance)
overcame at least part of the control system instability problem experienced by Razak.
In clinical testing on 10 people with transtibial amputation, when socket
size was gradually increased or decreased, the sensor picked up the initial
degradation of fit sooner than a practitioner visually inspecting the participant’s
gait or the participant sensing a need for socket adjustment.Towards the objective of developing a socket capable of automatically adjusting
socket size before the user detects a change in socket fit, we integrated sensors
into a motor-driven adjustable panel socket operated using a microprocessor-based
control system.
The microprocessor was programmed to adjust the radial position of the socket
panels to maintain a target sensed distance during walking. In this previous study,
the target distance was specified by the researcher based on results from
interactive fitting sessions with each participant. The feedback variable used in
the control system, termed the “socket fit metric (SFM),” was the sensed
liner-to-socket distance at a posterior mid-limb location. The control system
adjusted panel position to maintain the SFM at the target distance. The integral of
absolute error (IAE) was used as a metric for characterizing how well the control
system performed. IAE is a standard measure in control system engineering.
It indicates how well a closed-loop control system maintains the feedback
variable (i.e., SFM) at its target value (i.e., the target distance specified by the
researcher). Results from 10 transtibial prosthesis users walking on a treadmill for
bouts of at least 4 min demonstrated that the automatic-adjusting system achieved an
IAE in socket panel position of 0.001 mm–0.005 mm
. The control system maintained the SFM at its target within about 1.1% of the
thickness of a 1-ply cotton sock,
which would be adequate for this application.While the control system performed well and was stable, it did not account for
disturbances to the limb-socket system that would be introduced during clinical use.
In their free-living environments, prosthesis users do not walk continuously.
Instead, their walking is interrupted by bouts of sitting and standing. Bodily
position and activity changes may perturb and destabilize the control system.
Conducting socket release between walking bouts would also be expected to accentuate
instability since it has been shown that releasing socket pressures on a transtibial
residual limb by partially or fully doffing the socket, or by releasing socket
panels increases limb fluid volume.[22-24] An increase in limb fluid
volume would be expected to disturb the sensed distance and challenge the control
system at the outset of the next walking bout.The primary objective of this study was therefore to determine if, in a group of
participants with transtibial amputation, conducting socket release during sitting
and then retightening in preparation for walking destabilized the control system for
a socket that automatically changed size during walking to maintain fit. If the
socket demonstrated good stability in a group of prosthesis users, it would be
considered ready for at-home testing.
Methods
Participants
People were included in this study if they were at least 18 years old, had a
transtibial amputation at least 6 months prior, were using a definitive
prosthesis, and were capable of walking on a treadmill for one 5-min bout and
then multiple 2-min bouts separated by 10-min sits. Their residual limb needed
to be at least 9 cm long from the mid-patellar tendon to the distal end of the
limb (for bioimpedance analysis). Exclusion criteria included presence of skin
breakdown and use of a walking aide (e.g., cane or walker). Participants were
required to have locking pin suspension in their traditional socket. We also
required participants had few or no pads inside their socket so that we could
accurately scan and duplicate the socket. A University of Washington
Institutional Review Board approved all study procedures (IRB #00001779), and
written informed consent was obtained from each participant before study
procedures were initiated. Given the objective to determine if the
auto-adjusting socket was ready for take-home testing, we elected to study a
sample of six individuals. If the socket demonstrated stable performance on all
six participants, then we would consider it appropriate to continue testing in
user take-home environments.
Socket fabrication
Each participant’s traditional socket was scanned so that we could duplicate its
shape for the investigational prosthesis. The investigational prosthesis was
fabricated with three adjustable panels located on load-tolerant areas of the
residual limb (anterior medial, anterior lateral, and posterior midline) (Figure 1). Panel size was
maximized so as to impact socket volume change while avoiding bony prominences
at the anterior distal tibia, fibular head, and tibial crest—areas that may be
sensitive to compression. Sensors that measured the distance between the liner
and socket,[16,25-27] termed socket fit sensors, were positioned within the
socket wall during fabrication at the posterior medial mid-limb, the posterior
lateral mid-limb, and the anterior distal limb (Appendix 1). The stance phase minima from the two posterior
channels were used in the automatic, panel position adjustment algorithm. The
anterior distal channel was used to detect walking, implemented the same way as
in our previous study.
All sockets were made with tether suspension.
Figure 1.
Instrumented investigational prosthesis. Left: Inside view of socket
showing sensors at two posterior mid-limb and one anterior distal
location (white arrows). The tether (yellow arrow) connects to a
short pin (not shown) that provides suspension. The red and green
buttons at the top right are for operation of the powered tether
system. Right: Motors supported to frames mounted to the socket move
the panels radially inward and outward based on socket fit. The
mechanism to control tether length (gray cylinder) is mounted
beneath the socket.
Instrumented investigational prosthesis. Left: Inside view of socket
showing sensors at two posterior mid-limb and one anterior distal
location (white arrows). The tether (yellow arrow) connects to a
short pin (not shown) that provides suspension. The red and green
buttons at the top right are for operation of the powered tether
system. Right: Motors supported to frames mounted to the socket move
the panels radially inward and outward based on socket fit. The
mechanism to control tether length (gray cylinder) is mounted
beneath the socket.To adjust the socket size, we placed direct current (DC) micromotors in frames
that spanned over each panel. Each frame was affixed to the outside of the
socket using custom threaded mounts positioned within the socket wall during
fabrication (Figure 1,
right panel). Each motor included an encoder and gearhead and weighed 26 g
(model 1717006SR 1EH2-4096 15A152:1+MG03, Faulhaber (Micromo), Clearwater,
Florida). The motor unit was of diameter 17.1 mm and length 40.8 mm. The frames
and motors added 865 g to the overall weight on the socket. The motor drove
gearing and a winch assembly that translated the motor’s rotation into radial
displacement of the panel, as described in our prior work.
Unlike cabled-panel sockets, this design allowed the panel to be pulled
radially outward beyond the surrounding socket and relieve panel contact with
the residual limb. Further, a universal joint at the connection of the panel to
the winch minimized stress concentrations at the edge of the panel. Because of
these design features, no cushioning material needed to be placed on the inside
surface of each panel as with a traditional cabled-panel socket. A cable
connection to a PC ran a virtual instrument (VI) (LabVIEW National Instruments)
that adjusted the panels in 1-step increments.
Each step induced a 0.25-mm radial displacement in each of the three
panels. A panel position of 0.00 mm was defined as the position where the panels
were flush with the surrounding socket. When the socket was put in auto mode,
which was activated using the VI, the control scheme described below was
implemented.
Auto-adjustment algorithm
The auto-adjustment algorithm operated during all walking bouts. It started when
continuous walking was detected. It operated using the VI, implementing a custom
program similar to that described in our previous work.
The diagram in Figure
2 illustrates how the auto-adjusting socket operated. Consider a user
who starts with a comfortable socket fit at point “A” in the diagram. The socket
fit metric (SFM) value at this proper fit is termed the “set point.” The user
then gains limb volume during walking and the limb shifts closer to the socket
wall, moving to position “B” on the diagram. The auto-adjusting socket reacts by
increasing socket size to return the user to the SFM set point, traveling along
the blue line to arrive at position “C.” The SFM is now the same as at the
start, but the socket is larger because of the person’s increase in limb volume.
Later, the user sits for an extended period (without socket release), moving to
position “D” on the diagram. The person starts walking, and the auto-adjusting
socket reacts by decreasing socket size to return the user to the SFM set point,
traveling along the green line to arrive at position “E.” The user is now again
at the same SFM as at the start, but the socket size is smaller because of the
decrease in limb volume during sitting. The auto-adjusting socket sampled at
32 Hz. The maximum adjustment rate was 1 change per second.
Figure 2.
Diagram illustrating the design of the control system. The socket fit
metric (SFM) is the mean of the measurements from the two posterior
mid-limb sensors. Slopes of the green, red, and blue lines are the
plant gain. Deviations from the set point reflect changes in socket
fit. An increase in limb fluid volume (“A” to “B”) causes the
controller to increase socket size to return to the set point (“B”
to “C”). A decrease in limb fluid volume (“C” to “D”) causes the
controller to decrease socket size to return to the set point (“D”
to “E”). A goal of socket release/relock is to reduce volume loss
during sitting, that is, retard the change from “C” to “D.”
Diagram illustrating the design of the control system. The socket fit
metric (SFM) is the mean of the measurements from the two posterior
mid-limb sensors. Slopes of the green, red, and blue lines are the
plant gain. Deviations from the set point reflect changes in socket
fit. An increase in limb fluid volume (“A” to “B”) causes the
controller to increase socket size to return to the set point (“B”
to “C”). A decrease in limb fluid volume (“C” to “D”) causes the
controller to decrease socket size to return to the set point (“D”
to “E”). A goal of socket release/relock is to reduce volume loss
during sitting, that is, retard the change from “C” to “D.”To program the auto-adjusting socket for an individual user, we first executed a
test in the lab to characterize the user’s plant gain, the change in SFM induced
by a change in socket volume (slope of the red, blue, and green lines in Figure 2). The
participant walked at a self-selected speed on a treadmill while the researcher
adjusted the socket in 0.25-mm increments across the user’s tolerated socket
size range.
The plant gain is the slope of the least-squares fit to SFM (in counts)
plotted against panel position (in mm). Programmed into the socket’s
microcontroller, the plant gain is used to calculate in real time the change in
panel position the auto-adjusting socket should make when the user’s SFM
deviates from its set point.As described in the testing protocol below, the researcher operated the VI via a
computer interface to adjust socket size during sitting between bouts of
walking. In addition, a motor-driven system mounted beneath the socket, similar
to that described in our prior publication,
was used to draw in and release a tether to the liner.
Testing protocol
Once the investigational prosthesis was fabricated and instrumented, the
participant visited the lab for a fitting and evaluation session. The research
prosthetist evaluated the participant’s gait and adjusted alignment of the foot
and length of the prosthesis if needed. The participant walked on the treadmill
at different panel positions (socket sizes) to ensure the socket was comfortable
and to ensure the instrumentation performed properly.On a separate day, the testing protocol was conducted. After arriving at the lab,
participants sat for at least 10 min with their traditional prosthesis donned to
achieve a homeostatic condition. Participants then doffed their prosthesis, and
their residual limb was instrumented with thin surface electrodes that are part
of a bioimpedance system which was used to monitor limb fluid volume as
described in detail in prior publications.[23,30] The electrodes were
configured to monitor the anterior region and the posterior region of the
residual limb. Data collection from the limb fluid volume monitoring system and
the socket fit sensors was initiated. A plant gain test was conducted, then the
socket was returned to the participant-preferred socket size. The researcher
then used the VI to put the socket in the auto-adjustment mode.A series of eight sitting and walking cycles was conducted (Figure 3). At the beginning of each sit,
the panels were loosened, and the locking pin tether was released 5 cm. The
participant then sat for 10 min in a relaxed position with his or her thighs
horizontal, knees positioned at roughly the same level as the hips, and feet
touching the floor. At the end of the 10-min sit, the researcher drew in the
tether using the motor-driven system mounted beneath the socket. The participant
stood, then the researcher tightened the panels to the preferred socket size
recorded after the plant gain test. The participant stood briefly (5 s) and then
walked on the treadmill at a self-selected walking speed for 2 min, activating
the control system on the auto-adjusting socket. This sit/walk cycle was
repeated, except that at the end of the sit the researcher returned the socket
to its size at the end of the prior walk rather than that recorded at the outset
of the session. The cycles were repeated until the fifth cycle, where the socket
was returned to a size of +1.0% volume larger than that at the end of the prior
walk. The appropriate panel adjustment to achieve a +1.0% change was determined
using a geometric model of the socket shape.
The sixth through eighth cycles were identical to the earlier second
through fourth cycles. After the session, participants were returned to their
traditional socket and left the lab.
Figure 3.
Test protocol. Eight cycles of sit and walk were conducted. The
protocol used for cycles 1–4 and 6–8 is shown in the upper figure.
The protocol used for cycle 5 is shown in the lower figure. During
cycle 5 after the sit, the panels were tightened to the pre-sit
socket volume plus 1.0% socket volume. S = Sit, W = Walk.
Test protocol. Eight cycles of sit and walk were conducted. The
protocol used for cycles 1–4 and 6–8 is shown in the upper figure.
The protocol used for cycle 5 is shown in the lower figure. During
cycle 5 after the sit, the panels were tightened to the pre-sit
socket volume plus 1.0% socket volume. S = Sit, W = Walk.As a subjective measure of the effect of the intervention, participants were
asked to provide a relative socket comfort rating (RSCR) at the end of each
walking bout in cycles 5–8. RSCR has been used previously in prosthetics
research to study relative changes in socket comfort within a session.[32,33] The RSCR
query was phrased, “Compared to the end of the prior walk, is your socket
comfort a lot better, a little better, the same, a little worse, or a lot
worse?”
Data processing and analysis
Data collected across the test session, including the SFM, set point, and panel
position, were downloaded from the auto-adjusting socket and plotted over time
for visual inspection. The walking portion of each cycle was extracted for
further analysis. For each session, the range of panel position was calculated.
For each bout, the absolute error of the auto-adjusting socket control system
(SFM minus set point) was plotted over time and the IAE calculated
as
where SFM
is the measured SFM of the ith temporal
index, SFM
is the SFM set point, and N is the number of data points
in the analysis. Thus, N for an IAE calculated during the first
30 s of a bout included all data points up to 30 s, while N for
an IAE calculated during a whole bout included all points in the bout.Limb fluid volume data from the bioimpedance system were downloaded and converted
to extracellular fluid volume using de Lorenzo’s form of the Cole model.
The data were time-synchronized with the SFM data. The minimum fluid
volume during stance phase of each step was determined for both the anterior and
posterior limb regions and a mean calculated for each bout. The means for a
session were expressed as a percentage change relative to the mean fluid volume
during cycle 4, the cycle before the 1.0% relock socket size increase. This
strategy allowed a consistent reference across participants for the percent
fluid volume change between cycle 4 and subsequent cycles, a variable of
interest in this study.Relative socket comfort rating data were expressed as a change relative to cycle
4, the cycle before the relock socket size increase was executed. “A little
better” and “a little worse” were defined as a plus one-unit change and a minus
one-unit change, respectively. No participants responded with “a lot better” or
“a lot worse,” so no unit change was defined for them.
Results
Six people with transtibial amputation (5 males and 1 female) participated in
this study. All participants had their limb amputation as a result of trauma.
Median age was 44 years (range 36–76), median time since amputation was 14 years
(range 2–40), and median body mass index
(BMI) was 23.6 (range 21.9–26.5) (Table 1). Median residual limb length
was 15.2 cm (range 11.8–18.5) and median mid-limb circumference was 28.2 cm
(range 26.1–32.3). Descriptions of prosthesis componentry are listed in
Appendix 2.
Table 1.
Participant characteristics and plant gains.
Participant
Gender
Age (y)
Height (cm)
Weight (kg)
BMI (kg/m2)
Time since amputation (y)
RL Lengtha (cm)
RL Circumf.b (cm)
Shape
Co-Morb.
Plant gain (counts/mm)
1
M
44
175.0
72.9
24.6
5
12.5
32.3
Cylindrical
HBP
737
2
F
36
160.0
58.2
23.4
13
14.0
26.1
Bulbous
None
2470
3
M
58
188.0
77.3
22.5
34
18.5
28.2
Conical
None
5621
4
M
42
162.5
60.6
23.7
2
11.8
27.3
Cylindrical
Smoker
6332
5
M
43
182.9
71.2
21.9
15
16.3
28.2
Conical
Smoker
7456
6
M
76
180.3
83.2
26.5
40
18.1
28.6
Conical
HBP
8693
afrom the mid-patellar tendon to distal end.
bat 4 cm distal of mid-patellar tendon, 170° knee
flexion.
BMI= body mass index, RL= residual limb, M= male, F= female, HBP=
high blood pressure.
Participant characteristics and plant gains.afrom the mid-patellar tendon to distal end.bat 4 cm distal of mid-patellar tendon, 170° knee
flexion.BMI= body mass index, RL= residual limb, M= male, F= female, HBP=
high blood pressure.Participants’ median plant gain was 5977 counts/mm (range 737–8693) (Table 1). The higher
the plant gain the more sensitive the person’s socket fit was to changes in
socket size.
Control system error
Integral of absolute error increased right after the auto-adjusting socket became
active and then, in general, decreased over time for the rest of the bout (Figure 4). Out of the 48
bouts, 4 of them (bout 6 for participant #1; bouts 6 and 8 for participant #2;
bout 5 for participant #4) did not demonstrate this result and instead showed a
gradual increase in IAE over time that did not stabilize until late in the bout
(Appendix 3, lower panels). The maximum IAE during the first 30 s
of a bout (median 0.008, range 0.002–0.058) was greater than the IAE at the end
of the bout (median 0.006, range 0.001–0.034) by a median of 1.4 times (range
0.5–5.0) (Appendix 4). In 77% of the bouts, IAE was greater during the
first 30 s than later in the bout.
Figure 4.
Example plots illustrating control system performance during the
eight walking bouts. Upper panel: SFM data (black) and control
system set point (orange), both in mm. A consistent scale is not
used across bouts so that the shapes of the curves are clearly
visible. The SFM is closer to the set point at the end of the bout
than at the outset, demonstrating that the control system is
performing well. Lower panel: IAE in mm plotted over the course of
each bout. The shape of the IAE curves over time, a rise and maximum
followed by a decrease to a stable value, is typical for a properly
functioning engineered control system.
Example plots illustrating control system performance during the
eight walking bouts. Upper panel: SFM data (black) and control
system set point (orange), both in mm. A consistent scale is not
used across bouts so that the shapes of the curves are clearly
visible. The SFM is closer to the set point at the end of the bout
than at the outset, demonstrating that the control system is
performing well. Lower panel: IAE in mm plotted over the course of
each bout. The shape of the IAE curves over time, a rise and maximum
followed by a decrease to a stable value, is typical for a properly
functioning engineered control system.Comparing IAE before to after the 1.0% relock socket size increase, three
participants (#1, #2, #4) demonstrated an increase in IAE for cycle 5 compared
with cycle 4, and three participants (#3, #5, #6) demonstrated a decrease for
cycle 5 compared with cycle 4 (Figure 5). For participant #2, IAE for all cycles after the 1.0%
relock socket size increase was greater than that for all cycles before the 1.0%
relock socket size increase. No other participant showed this trend.
Figure 5.
Integral of absolute error at the end of each bout. Dark bars are
from bouts before the 1.0% socket size increase. Light bars are from
bouts after the 1.0% socket size increase. Participants are ordered
from low to high plant gains. There was no consistent trend across
participants of a higher IAE after v. before the 1.0% socket size
increase suggesting that control system performance was not
sensitive to this perturbation.
Integral of absolute error at the end of each bout. Dark bars are
from bouts before the 1.0% socket size increase. Light bars are from
bouts after the 1.0% socket size increase. Participants are ordered
from low to high plant gains. There was no consistent trend across
participants of a higher IAE after v. before the 1.0% socket size
increase suggesting that control system performance was not
sensitive to this perturbation.In one bout, cycle 7 for participant #2, while the auto-adjusting socket was
active, the control system at the outset of the bout held the panels at their
maximum distance and needed to be reset 1 min into cycle 7 using the VI. This
happened because at the end of the preceding cycle 6, the panel position
maximized (Appendix 3). Per the defined protocol, the panels started at
that position at the start of cycle 7. No other cases of control system
saturation occurred during the study.We note from the analysis above that IAE at the end of a bout was not of a
consistent magnitude across all participants (Figure 5). As an exploratory effort, we
investigated if end-of-bout IAE was related to participant characteristics.
Plots of limb length, plant gain, age, height, years since amputation, limb
circumference, and the quotient of limb circumference divided by diameter
demonstrated a weak correlation (Pearson) with participant median end-of-bout
IAE (R < 0.4 for all variables). Body mass index (BMI) was
moderately correlated (R = 0.6). In some bouts for some
participants, the IAE oscillated about the set point (e.g., Appendix 5). This hysteresis may have contributed to the
inconsistent IAE for some participants.During some bouts, there were intermittent connection issues, typically less than
a few seconds, between the VI and the controller, causing the controller to
briefly turn off and restart, resetting the set point. These instances are shown
as discontinuities in figures of controller error (IAE) plotted over time
(Appendix 3). These events were later demonstrated to be a result
of an issue in the VI software, not the control system.
Participant panel positions, limb fluid volume, and pistoning
For 5 of the 6 participants, the range of panel position (maximum–minimum) during
walking across the test session was between 2.26 mm and 6.01 mm (Appendix 6). For the remaining participant (participant #2), the
control system saturated at the maximum panel radial distance (10.00 mm) for
part of one walking bout, contributing to a higher range, 8.75 mm. Panel
position range was not well-correlated with plant gain (R =
0.38).From visual inspection of the data, we note that panel position reduced from
cycles 1 to 4 and from cycles 5 to 8 for three participants (#1, #3, #4) and
increased from cycles 1 to 4 and from cycles 5 to 8 for three participants (#2,
#5, #6) (Figure 6). Per
the specified protocol, the panel position was increased during cycle 5 when the
1.0% relock socket size increase was executed. Panel position data were back
onto their trajectory from earlier cycles 1 to 4 by cycle 6 for participants #4
and #6 and by cycle 7 for participants #2 and #3. Participant #1 did not
demonstrate this behavior and instead maintained a larger panel position (larger
socket size) compared with cycles 1–4. Panel position for participant #5
stabilized to a consistent distance during cycles 5–8 that was larger than that
during cycles 1–4.
Figure 6.
Panel position at the end of each walking bout. Participants #1, #3,
and #4 experienced a socket reduction from the beginning to the end
of the test session. Participants #2, #5, and #6 experienced a
socket enlargement.
Panel position at the end of each walking bout. Participants #1, #3,
and #4 experienced a socket reduction from the beginning to the end
of the test session. Participants #2, #5, and #6 experienced a
socket enlargement.Limb fluid volume change over time followed a similar pattern to the panel
position data, that is, the patterns of change in Figure 7 are similar to those in Figure 6. The shapes of
the plots for anterior and posterior regions were similar to each other for each
participant except for participant #5 who after the intervention experienced a
much greater percent limb fluid volume increase in the anterior region than the
posterior region.
Figure 7.
Percent limb fluid volume during stance phase minima for each bout.
Results from the anterior region (ant) and posterior region (post)
are shown. Participants #1, #3, and #4 lost limb fluid volume over
the session while participants #2, #5, and #6 gained. Differences
are likely a result of participant physiological characteristics.
All participants (#1 to #6) demonstrated an increase in limb fluid
volume from the intervention (between bout 4 and bout 5), suggesting
a common mechanism. However, the changes in both the anterior and
posterior regions for participant #4 and the posterior region for
participant #5 were less than for other participants.
Percent limb fluid volume during stance phase minima for each bout.
Results from the anterior region (ant) and posterior region (post)
are shown. Participants #1, #3, and #4 lost limb fluid volume over
the session while participants #2, #5, and #6 gained. Differences
are likely a result of participant physiological characteristics.
All participants (#1 to #6) demonstrated an increase in limb fluid
volume from the intervention (between bout 4 and bout 5), suggesting
a common mechanism. However, the changes in both the anterior and
posterior regions for participant #4 and the posterior region for
participant #5 were less than for other participants.SFM was controlled during walking bouts thus it did not follow trends similar to
those for panel position or limb fluid volume. As an exploratory effort, we
investigated if there was a trend in the change in peak-to-trough SFM
(pistoning) from before to after the 1.0% relock socket size increase. For the
participants who experienced a loss of limb fluid volume from cycle 1 to cycle 4
(#1, #3, #4), peak-to-trough SFM increased from before to after the 1.0% relock
socket size increase (cycle 4 compared to cycle 5) (Appendix 7). For the participants who experienced a gain of limb
fluid volume from cycle 1 to cycle 4 (#2, #5, #6), peak-to-trough SFM decreased
from before to after the 1.0% relock socket size increase.While enlarging the socket by +1.0% after cycle 4 increased limb fluid volume, 4
of 6 participants reported a worsened socket comfort right after it was executed
(Figure 8). Only
participant #4 reported an improvement in socket comfort, stating that his
socket felt a bit more tightly coupled to his residual limb. Participant #2
reported feeling rubbing on the proximal brim line, participant #3 reported
pistoning, and participant #6 reported looseness at the distal end and back part
of his limb. Despite the decrease in socket comfort while walking after the 1.0%
relock socket size increase, none of the participants decreased their socket
comfort score in cycle 8 compared with cycle 5.
Figure 8.
Change in socket comfort relative to cycle 4 for all participants.
The central horizontal line represents the reference, that is, the
socket comfort during cycle 4 right before the 1.0% relock socket
size increase. The distance between adjacent tick marks on the
y-axis is a relative socket comfort rating
“unit change.” A positive unit change in RSCR is “a little better,”
and a negative unit change in RSCR is “a little worse.” Upon the
1.0% relock socket size increase, four participants indicated a
worsened RSCR, one no change, and one a more favorable RSCR.
Change in socket comfort relative to cycle 4 for all participants.
The central horizontal line represents the reference, that is, the
socket comfort during cycle 4 right before the 1.0% relock socket
size increase. The distance between adjacent tick marks on the
y-axis is a relative socket comfort rating
“unit change.” A positive unit change in RSCR is “a little better,”
and a negative unit change in RSCR is “a little worse.” Upon the
1.0% relock socket size increase, four participants indicated a
worsened RSCR, one no change, and one a more favorable RSCR.As an exploratory effort, we investigated the relationship between percent limb
fluid volume change from before to after the 1.0% relock socket size increase
(cycle 5–cycle 4) and RSCR (Figure 9). The four participants with percent fluid volume changes
greater than 1.0% (# 1,2,3,6) all reported reduced RSCR, while the participant
with a small increase (#4) (0.6%) reported an increased RSCR. The participant
with essentially no percent fluid volume change (#5) (−0.1%) reported no change
in RSCR.
Figure 9.
Relative socket comfort rating (RSCR) v. change in percent fluid
volume. Participants who experienced >1.0 change in percent fluid
volume between cycles 4 and 5 reported a worsened socket comfort
rating while those with a change <1.0% reported no change or a
more favorable socket comfort rating.
Relative socket comfort rating (RSCR) v. change in percent fluid
volume. Participants who experienced >1.0 change in percent fluid
volume between cycles 4 and 5 reported a worsened socket comfort
rating while those with a change <1.0% reported no change or a
more favorable socket comfort rating.
Discussion
In this study, conducting socket release during sitting between bouts of walking was
shown to have the intended effect of changing participants’ limb fluid volume. In
general, the auto-adjusting socket responded well to these perturbations and
maintained stable performance and low error. We believe the results warrant
advancing to testing on prosthesis user participants in their at-home environments,
increasing the number of participants and determining the long-term clinical
outcomes in terms of comfort, prosthesis use, skin health, and other related
outcomes.Part of the reason the control system demonstrated stable performance and low error
was because fluid volume change followed panel position change as shown by the
similarity of curve shapes in Figures 6 and 7. Our prior work and the plant gain tests in the present study showed that,
when socket size adjustments were made, the distance sensed by our custom sensors
changed linearly with socket size.[16,17] Taken together, these results
suggest that the control system operated as intended for all participants—adjusting
panel position based on information from the distance sensors served to adjust limb
fluid volume.It is recognized that clinically it may not be desirable to increase limb fluid
volume over time, as occurred for participants #2, #5, and #6 (Figure 7). However, what is relevant to note
(and was the purpose of the present study) is that the auto-adjusting socket is
capable of being used to control limb fluid volume. The auto-adjusting socket worked
in harmony with the residual limb to take advantage of the limb’s capability for
fluid volume change to accomplish this result. In clinical practice, the
auto-adjustment algorithm would be programmed to maintain a consistent sensed
distance over the day.We expect that the use of distance sensing and the practice of not placing the
sensing elements on the actuators are two key reasons the auto-adjusting socket in
this study performed better than other systems described in the
literature.[4-13] Distance sensing may be more effective than pressure sensing
because the measurement is very sensitive to small changes in socket fit, and
presence of the sensors does not disrupt the regular limb-socket interface.
We did not consider placing the sensing elements on the actuators an
appropriate strategy. The actuators are located at load-tolerant areas of the
residual limb, thus sensing at those locations would not be expected to provide a
clinically meaningful and sensitive measurement of socket fit. In the present study,
the sensors used for auto adjustment were located at a posterior location off of the
midline. It is possible that data from other sensors, for example that monitor
distal limb superior-inferior motion (pistoning) or limb angulation in the sagittal
plane, may be necessary, although results in the present study do not support such a
need. These interpretations are preliminary, however, and need to undergo rigorous
scientific testing on a larger group of participants.
Control system performance
The shape of the IAE curves over time as shown in Figure 4, a rise and maximum followed by
a decrease to a stable value, is typical for a properly functioning engineered
control system. The range of end-of-bout IAE, from 0.001 to 0.034, corresponded
to a median socket volume error of 0.001%–0.033% for the six participants tested
here. For comparison, adding a sock sheath (0.2 mm thickness under stance phase loading
) would change socket volume by an average of 0.97% for participants in
the present study (range 0.80%–1.03%). Thus, the IAE error in this investigation
would be expected to be acceptable in clinical application of the auto-adjusting
socket.The two bouts that showed the greatest increase in IAE over time were both from
the only participant with a bulbous residual limb (participant #2). Unlike the
other five people in the study, this participant found the relock protocol
uncomfortable—drawing in the tether before closing the socket panels—because it
tended to trap her soft tissue distally in the socket. Other participants,
during preliminary investigations prior to this study and during the study
itself, held the opposite opinion, preferring to draw in the tether first and
then close the panels. The high IAE and discomfort stated by participant #2
suggest a need to investigate if a different auto-adjusting control strategy is
necessary for people with much redundant soft tissue in their residual limb.The oscillation of the SFM about the set point observed in some bouts (e.g.,
Appendix 5) may be eliminated by setting a threshold for
execution of a socket size adjustment (known as a “hysteresis band” in control
systems engineering). Because the peak-to-trough range of oscillation observed
in the data collected in this study was typically less than 0.20 mm, we would
expect that a hysteresis band and step size of 0.20 mm would be appropriate. It
is possible that because of the different plant gains across participants, the
threshold size adjustment may need to be tuned to each user, though rigorous
scientific investigation would need to be conducted to test if this is
clinically necessary.As shown in Figure 9, we
observed an interesting relationship between RSCR and percent limb fluid volume
change in the posterior region from before to after the 1.0% socket size
increase (Cycle 5–Cycle 4). The results suggest that matching the socket volume
increase at the start of a new walking bout to the fluid volume increase
experienced by the individual participant between bouts may improve comfort. In
other words, the socket size should be adjusted so that the socket is still
relatively snug on the residual limb and may even restrict its fluid volume
increase. The four participants who experienced increases in percent fluid
volume (cycle 5–cycle 4) of more than 1.0% (#1, #2, #3, and #6) all indicated
that their socket felt too loose during cycle 5. Possibly the high fluid volume
increase they experienced reduced their soft tissue compressive stiffness, made
their limb-socket interface unstable, and caused this sensation. RSCR scores for
the two participants who experienced lower percent fluid volume increase, #4
(−0.1% limb fluid volume change) and #5 (0.6%), were more favorable. Possibly,
participant #4 and #5’s sockets were tight on their residual limb after the 1.0%
socket size increase, helping to ensure a more stable interface. This
interpretation is preliminary, based upon results from a small number of
participants, and would need to be verified in more extensive clinical
studies.We note that by cycle 7 most participants returned to the limb fluid volume
trajectory established before the +1.0% socket size intervention. Thus, while
10-min socket releases helped to reduce limb fluid volume loss, they may not
have been long enough to promote a meaningful limb size increase such that a
socket enlargement in preparation for further walking was warranted. Longer
release durations, however, may show a benefit, though this hypothesis would
need to be tested through rigorous scientific investigation.The increase in peak-to-trough SFM (maximum swing phase SFM minus minimum stance
phase SFM) from before to after the intervention (from cycle 4 to cycle 5) for
participants who experienced a reduction in limb fluid volume over time (#1, #3,
and #4) means that, as expected, these participants’ limb pistoned more in the
socket after the +1.0% relock socket size increase was executed than before it.
The result points to the utility of SFM peak-to-trough data as an additional
socket fit metric for use in future auto-adjusting socket systems.A related prosthetic socket technology, electronic elevated vacuum (EV), applies
a negative pressure between the socket and liner-to-draw residual limb soft
tissues outward. Both EV and the auto-adjusting socket in the present study
attempt to maintain limb volume within a narrow range over time, but they
optimize different metrics to facilitate adjustment, and they use different
actuators (vacuum pressure, panel displacement) to effect change. Results from
the present study indicate that in-socket limb fluid volume response to
automated control of socket size is immediate, while a different investigation
reported in the literature showed that in-socket limb fluid volume response to
automated control of EV is much slower.
As research and development of both auto-adjusting sockets and EV sockets
continue, it will be important to quantify their control system performance
(i.e., capability to maintain a set point based on a socket fit metric), their
capability to manage limb volume, and their benefit to clinical outcomes like
residual limb health.[36-38]A limitation of the study design was that the system was tested on only six
participants. This number of participants was considered acceptable because the
objective of the study was to warrant at-home testing with a large number of
participants.The time between walking bouts in this study was short, approximately 10 min. A
more challenging situation, expected to be encountered during at-home use, is
when there are much longer time periods of sitting, standing, and
weight-shifting between walking bouts, that is, between automatic panel position
changes. In these cases, a more substantial change in the residual limb may
occur and the SFM at the outset of the next walk may be much different than the
set point. If the difference is too great, then the residual limb may not be
able to adjust size quickly, risking control system instability and
necessitating a modification in the control system strategy.The auto-adjusting socket used in this study was heavier than a normal socket.
The instrumentation on the socket added a median of 885 g to the traditional
socket weight (median 589 g). The weight may have affected participant RSCR
scores late in the session because of the accumulated effect of a greater pull
on limb soft tissues compared with users’ traditional sockets. We would not
expect the added weight to have affected control system performance since the
auto-adjusting socket adapts to a change in socket fit. A reduced size frame and
motor and replacement of the control system components with a custom electronics
board could easily be created to reduce the weight since from experience in the
present study the displacement range and resolution needs of the auto-adjusting
socket are now better specified (e.g., Appendix 6). For the revised system, we would expect the weight
difference compared to a traditional prosthesis to be comparable to the
difference between a powered ankle and traditional ankle prosthesis. The LabVIEW
VI instability issue during some bouts in the present study should be resolved
using an on-board microprocessor dedicated to auto-adjustment instead of the
LabVIEW software package.
Conclusion
In this study, the auto-adjusting socket maintained good stability despite
perturbation of panel and tether release/lock during sitting and achieved a low IAE
during subsequent bouts of walking. Therefore, the auto-adjusting socket is ready
for field testing in participant at-home settings.Click here for additional data file.Supplemental Material for Performance of an automatic-adjusting prosthetic socket
during walking with intermittent socket release by Ethan J Weathersby, Andrew C
Vamos, Brian G Larsen, Jake B McLean, Ryan V Carter, Katheryn J Allyn, Daniel
Ballesteros, Horace Wang, Nicholas S deGrasse, Janna L Friedly, Brian J Hafner,
Joseph L Garbini, Marcia A Ciol and Joan E Sanders in Journal of Rehabilitation
and Assistive Technologies Engineering
Authors: M Traballesi; A S Delussu; A Fusco; M Iosa; T Averna; R Pellegrini; S Brunelli Journal: Eur J Phys Rehabil Med Date: 2012-05-28 Impact factor: 2.874
Authors: Yikun Gu; Dapeng Yang; Luke Osborn; Daniel Candrea; Hong Liu; Nitish Thakor Journal: Proc Inst Mech Eng H Date: 2019-06-05 Impact factor: 1.617
Authors: Joan E Sanders; Joseph L Garbini; Jake B McLean; Paul Hinrichs; Travis J Predmore; Jacob T Brzostowski; Christian B Redd; John C Cagle Journal: Med Eng Phys Date: 2019-04-23 Impact factor: 2.242
Authors: Jake B McLean; Christian B Redd; Brian G Larsen; Joseph L Garbini; Jacob T Brzostowski; Brian J Hafner; Joan E Sanders Journal: Clin Biomech (Bristol, Avon) Date: 2019-03-02 Impact factor: 2.063
Authors: Ethan J Weathersby; Clement J Gurrey; Jake B McLean; Benjamin N Sanders; Brian G Larsen; Ryan Carter; Joseph L Garbini; Joan E Sanders Journal: Sensors (Basel) Date: 2019-09-19 Impact factor: 3.576