Socket size adjustments in people with transtibial amputation: Effects on residual limb fluid volume and limb-socket distance.

Jake B McLean1, Christian B Redd1, Brian G Larsen1, Joseph L Garbini2, Jacob T Brzostowski1, Brian J Hafner3, Joan E Sanders4.   

Abstract

BACKGROUND: Small intermittent adjustments of socket size using adjustable sockets may be a means for people with transtibial amputation to better maintain residual limb fluid volume and limb position while using a prosthesis.
METHODS: Socket size, limb fluid volume, and distance from the limb to the socket, termed "sensed distance," were recorded while participants with transtibial amputation walked on a treadmill wearing a motor-driven, cabled-panel, adjustable socket. Researchers made frequent socket size adjustments using a mobile phone app to identify participants' acceptable socket size range. Limb fluid volume and sensed distance were then monitored as incremental adjustments were made to the socket.
FINDINGS: Prosthesis users in this study (n = 10) accepted socket sizes between -5% and +5% of their neutral socket volume. There was a rapid increase in limb fluid volume and sensed distance upon socket enlargement, and a rapid decrease upon reduction. Subsequently, there were gradual changes in fluid volume and sensed distance. While visually monitoring limb fluid volume data in real time, researchers were able to adjust socket size to maintain consistent limb fluid volume within a -0.7% to +0.9% volume change for 24 min.
INTERPRETATION: Participant residual limbs compensated to socket size adjustment. Using socket-mounted sensors to monitor limb-socket mechanics, an automatic adjustable socket that maintains limb fluid volume may be possible and may improve socket fit in instances where fit deteriorates during use.
Copyright © 2019 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2019        PMID: 30901641     DOI: 10.1016/j.clinbiomech.2019.02.022

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

1.  Performance of an auto-adjusting prosthetic socket during walking with intermittent socket release.

Authors:  Ethan J Weathersby; Andrew C Vamos; Brian G Larsen; Jake B McLean; Ryan V Carter; Katheryn J Allyn; Daniel Ballesteros; Horace Wang; Nicholas S deGrasse; Janna L Friedly; Brian J Hafner; Joseph L Garbini; Marcia A Ciol; Joan E Sanders
Journal:  J Rehabil Assist Technol Eng       Date:  2022-04-28

2.  Thin Magnetically Permeable Targets for Inductive Sensing: Application to Limb Prosthetics.

Authors:  Ethan J Weathersby; Clement J Gurrey; Jake B McLean; Benjamin N Sanders; Brian G Larsen; Ryan Carter; Joseph L Garbini; Joan E Sanders
Journal:  Sensors (Basel)       Date:  2019-09-19       Impact factor: 3.576

3.  Incorporating a Ferrous Polymer Target into Elastomeric Liners for Socket Fit Sensing in Prosthesis Users.

Authors:  Ryan V Carter; Brian G Larsen; Jake B McLean; Joseph L Garbini; Joan E Sanders
Journal:  Sensors (Basel)       Date:  2020-10-01       Impact factor: 3.576