| Literature DB >> 35548321 |
Luyao Li1, Tao Li1, Jin Shen1, Huihui Sun1, Hongyang Duan1, Changping Zhu2, Wei Zhang1, Chen Liang1, Baoying Zhang1, Yan Li1, Liubo Zhang1.
Abstract
Introduction: To explore the feasibility of pulsed ultraviolet (UV) light technology for low-temperature disinfection, a series of experiments were conducted.Entities:
Keywords: Pulsed UV; disinfection; temperature
Year: 2022 PMID: 35548321 PMCID: PMC9081888 DOI: 10.46234/ccdcw2022.080
Source DB: PubMed Journal: China CDC Wkly ISSN: 2096-7071
Figure 1Schematic diagram of the placement of the contaminated vector in the water bath of the heating and cooling circulator during the test.
Killing effect of pulsed UV light on Staphylococcus aureus under different temperature conditions.
|
|
|
|
| |||
|
|
|
|
| |||
| Notes: Negative control sterile growth. Positive control number: 1.00×106–4.20×106 CFU/sheet. The mean logarithmic value of the positive control was 6.32. >6.32 indicates sterile growth in the test group.
| ||||||
| Paper sheets | 20.0 | 1.42 | 2.81 | 6.40 | >6.32 | |
| 0.0 | 1.30 | 2.36 | 6.40 | >6.32 | ||
| −20.0 | 1.27 | 1.86 | 5.80 | 5.29 | ||
| Stainless steel sheet | 20.0 | 2.60 | 5.08 | >6.32 | >6.32 | |
| 0.0 | 2.41 | 3.27 | >6.32 | >6.32 | ||
| −20.0 | 1.51 | 3.38 | >6.32 | >6.32 | ||
| Cloth sheet | 20.0 | >6.32 | >6.32 | >6.32 | >6.32 | |
| 0.0 | >6.32 | >6.32 | >6.32 | >6.32 | ||
| −20.0 | 5.50 | >6.32 | >6.32 | >6.32 | ||
Killing effect of pulsed UV light on Escherichia coli under different temperature conditions.
|
|
|
|
| ||||||
|
|
|
|
|
|
|
| |||
| Notes: Negative control sterile growth. Positive control number: 1.00×106–3.83×106 CFU/sheet. The mean logarithmic value of the positive control was 6.31. >6.31 indicates sterile growth in the test group.
| |||||||||
| Paper sheets | 20.0 | 1.61 | 2.86 | 3.63 | 4.42 | 5.84 | >6.31 | >6.31 | |
| 0.0 | 2.04 | 2.38 | 3.14 | 3.23 | 4.02 | >6.31 | >6.31 | ||
| −20.0 | 1.82 | 2.01 | 2.45 | 3.37 | 3.77 | >6.31 | >6.31 | ||
| Stainless steel sheet | 20.0 | 2.65 | 4.41 | 4.37 | >6.31 | >6.31 | >6.31 | >6.31 | |
| 0.0 | 1.98 | 3.60 | 3.62 | >6.31 | >6.31 | >6.31 | >6.31 | ||
| −20.0 | 1.70 | 2.98 | 3.92 | >6.31 | 4.35 | >6.31 | >6.31 | ||
| Cloth sheet | 20.0 | 2.75 | 4.28 | >6.31 | >6.31 | >6.31 | >6.31 | >6.31 | |
| 0.0 | 1.55 | 4.37 | 6.69 | >6.31 | >6.31 | >6.31 | >6.31 | ||
| −20.0 | 1.37 | 4.88 | 6.21 | >6.31 | >6.31 | >6.31 | >6.31 | ||
Comparison of the effect of different influencing factors on the effectiveness of pulsed UV disinfection.
|
|
|
| |
| Abbreviation: UV=ultraviolet. | |||
| Setting temperature | |||
| 20.0 ℃
| 6.31 (5.27, 6.32)
| 1.792 | 0.408 |
| Irradiation time | |||
| 15 s
| 3.63 (2.41, 6.21)
| 28.667 | 0.000 |
| Bacterial vector | |||
| Paper sheets
| 5.55 (2.54, 6.31)
| 23.292 | 0.000 |
| Test strain | |||
| 6.32 (2.93, 6.32)
| 5.465 | 0.000 | |