Vanida Poolnapol1, Taweap Sanghangthum1, Mintra Keawsamur2.
Abstract
Context: Interplay effects have become the significant problem in lung cancer radiotherapy. Since these effects yield dose variation within the target and surrounding tissues. Aim: The aim of this study is to investigate the effect of the dosimetric parameters of interplay effects in 6 MV flattening filter-free (FFF) photon beams for lung cancer. Settings and Design: This study performed planning, measurement, and data analysis sections for examining different breathing amplitudes and phases, doses, dose rates, field sizes, and fractionations. Subjects andEntities:
Keywords: Flattening filter free; interplay effects; lung cancer
Year: 2022 PMID: 35548034 PMCID: PMC9084591 DOI: 10.4103/jmp.jmp_63_21
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Robotic platform and three-dimensional conformal radiotherapy planning parameters setting
| Parameters | Conditions |
|---|---|
| Robotic platform | |
| Amplitude on X-axis (cm) | 0, 0.5, 1, and 2 |
| Amplitude on Y-axis (cm) | 0, 0.5, 1, and 2 |
| Amplitude on XY-axis (cm) | 0, 0.5, 1, and 2 |
| 3D-CRT planning | |
| Dose (cGy) | 500, 800, and 1200 |
| Dose rate (MU/min) | 400, 600, 800, 1000, 1200, and 1400 |
| Field size (cm2) | 4×4, 6×6, 8×8, and 10×10 |
| Number of the fraction | 1 (1,000 cGy/1F), 2 (500 cGy/2F), and 5 (200 cGy/5F) |
3D-CRT: Three-dimensional conformal radiotherapy, F: Fraction
Figure 1The setting up of robotic platform, MapCHECK2, and solid water phantom
The gamma passing rates at 1%/1 mm criteria of standard plans irradiation with different amplitudes and unpaired t-test statistical analysis between 1D (X or Y-axis) and 2D (XY-axis) moving patterns
| Amplitude (cm) | Moving platform in X-axis | Moving platform in Y-axis | Moving platform in XY-axis | Unpaired | |
|---|---|---|---|---|---|
|
| |||||
| X versus XY | Y versus XY | ||||
| 0.0 | 100.0 | 100.0 | 100.0 | - | - |
| 0.5 | 86.1 | 88.8 | 72.4 | 0.72 | 0.93 |
| 1.0 | 55.8 | 60.5 | 35.2 | 0.31 | 0.25 |
| 2.0 | 19.2 | 24.5 | 12.3 | 0.27 | 0.10 |
Figure 2The correlation curves of all moving patterns between the gamma passing rates and the different amplitudes: (a) X-axis, (b) Y-axis, and (c) XY-axis
Figure 3The correlation curve between the gamma passing rates and the different field sizes
The gamma passing rates from 1%/1 mm criteria and the linear equations of standard plans irradiation with different phases, doses, dose rates, and number of fractions
| Dosimetric parameters | Gamma passing rate (%) |
|---|---|
| Phase (s) | |
| 3 | 53.1 |
| 4 | 53.1 |
| 5 | 53.1 |
| Dose (cGy) | |
| 500 | 55.1 |
| 800 | 55.1 |
| 1,200 | 55.1 |
| Dose rate (MU/min) | |
| 400 | 36.7 |
| 600 | 34.7 |
| 800 | 34.7 |
| 1000 | 36.7 |
| 1200 | 34.7 |
| 1400 | 34.7 |
| Number of fraction | |
| 1 | 36.7 |
| 2 | 34.7 |
| 5 | 36.7 |
The gamma passing rates for 3%/2 mm criteria and the planning target volume information of eight clinical plans irradiation to the static and dynamic mode of robotic platform motion
| Number of clinical plan | PTV | Gamma passing rate (%) | ||
|---|---|---|---|---|
|
|
| |||
| Total volume (cm3) | Static platform | Dynamic platform | ||
| 1 | 83.0 | 1 | 100.0 | 85.9 |
| 2 | 71.3 | 1 | 100.0 | 90.8 |
| 3 | 15.0 | 1 | 100.0 | 84.1 |
| 4 | 9.1 | 1 | 100.0 | 84.3 |
| 5 | 31.8 | 2 | 100.0 | 79.7 |
| 6 | 30.2 | 2 | 100.0 | 75.6 |
| 7 | 28.6 | 6 | 99.5 | 71.0 |
| 8 | 9.7 | 2 | 98.5 | 69.7 |
| Mean±SD | 99.8±0.53 | 80.1±7.5 | ||
PTV: Planning target volume, SD: Standard deviation
Figure 4Examples of dose distributions of noncomplex lung stereotactic body radiation therapy plan delivered to MapCHECK2 placing on the robotic platform: (a) Dose distribution during dynamic platform measurement and (b) Dose distribution from the planning step
Figure 5Examples of dose distributions of complex lung stereotactic body radiation therapy plan delivered to MapCHECK2 placing on the robotic platform: (a) Dose distribution during dynamic platform measurement and (b) Dose distribution from the planning step