| Literature DB >> 35496451 |
Willy Weng1, Pascal Theriault-Lauzier1, David Birnie1, Calum Redpath1, Mehrdad Golian1, Mouhannad M Sadek1, Andres Klein1, F Daniel Ramirez1, Darryl R Davis1, Pablo B Nery1, Girish M Nair1, Simon Hansom1, Martin S Green1, Alper Aydin1.
Abstract
Background: Ipsilateral approach in patients requiring cardiac implantable electronic device (CIED) revision or upgrade may not be feasible, primarily due to vascular occlusion. If a new CIED is implanted on the contralateral side, a common practice is to explant the old CIED to avoid device interaction. Objective: The purpose of this study was to assess a conservative approach of abandoning the old CIED after implanting a new contralateral device.Entities:
Keywords: Abandoned device; Cardiac implantable electronic device; Elective replacement indicator; Interaction; Simultaneous contralateral device
Year: 2022 PMID: 35496451 PMCID: PMC9043401 DOI: 10.1016/j.hroo.2022.02.005
Source DB: PubMed Journal: Heart Rhythm O2 ISSN: 2666-5018
Figure 1CONSORT diagram of patient screening and study cohort. AI = artificial intelligence; CIED = cardiac implantable electronic device; PACS = Picture Archiving and Communications System.
Baseline clinical characteristics and device details of the 40 patients included in study analysis
| Female | 10 (25) |
| Age (y) | 69.7 ± 13.8 |
| Duration between initial implant and new device implant (y) | 4.8 ± 3.0 |
| Follow-up duration after abandonment (y) | 4.0 ± 3.3 |
| Diabetes | 7 (17.5) |
| Heart failure | 32 (80) |
| Chronic kidney disease (eGFR <30 mL/min/1.73 m2) | 7 (17.5) |
| Estimated PADIT score if the abandoned device had been removed | 4.5 ± 2.2 |
| Device reached end of life | 15 (37.5) |
| 10 (25) | |
| Indication for implant to a different site | |
| Occluded veins | 27 (67.5) |
| Unknown/not recorded | 8 (20) |
| Other | 3 (7.5) |
| Patient preference | 2 (5) |
Values are given as n (%) or mean ± SD.
eGFR = estimated glomerular filtration rate; PADIT = Prevention of Arrhythmia Device Infection Trial.
Need for dialysis line; presence multiple leads; need for cancer radiation therapy to the original side.
Characteristics of abandoned and new devices in the 40 study patients
| Abandoned device | New device | |
|---|---|---|
| Site | ||
| Right subclavicular | 8 (20) | 31 (77.5) |
| Left subclavicular | 31 (77.5) | 9 (22.5) |
| Abdominal | 1 (2.5) | 0 |
| Device type | ||
| Pacemaker | 22 (55) | 8 (20) |
| ICD | 18 (45) | 32 (80) |
| Single chamber | 22 (55) | 12 (30) |
| Dual chamber | 17 (42.5) | 9 (22.5) |
| CRT | 1 (2.5) | 19 (47.5) |
| Brand | ||
| Boston Scientific | 5 (12.5) | 3 (7.5) |
| Medtronic | 33 (82.5) | 37 (92.5) |
| MicroPort | 1 (2.5) | 0 |
| Biotronik | 1 (2.5) | 0 |
Values are given as n (%).
CRT = cardiac resynchronization therapy; ICD = implantable cardioverter-defibrillator.
Figure 2A: Chest radiograph of an abandoned device at ERI, on the patient’s right side. An abandoned lead is seen as well. B: Electrocardiogram. Blue arrowheads indicate pacing spikes from the abandoned device, ∼920 ms apart (corresponding to ERI rate of 65 bpm). The third QRS is sensed, leading to delay of pacing until ∼920 ms later. The abandoned device is not capturing the ventricle. Red arrowheads indicate spikes from the new device, which captures well. ERI = elective replacement indicator.
Figure 3Device interrogation of new device with the programmer, with an abandoned device in situ and at ERI. A: When the new device is temporarily set at VVI 30 bpm, pacing spikes (blue arrowheads) are seen from the abandoned device, ∼920 ms after each QRS, corresponding to the ERI rate of 65 bpm. There is no capture. B: During manual threshold testing, ventricular capture from the new device inhibits pacing from the old device. When capture from the new device is lost (red arrowhead), the abandoned device delivers a pacing spike at cycle length ∼920 ms (65 bpm), with ventricular noncapture. These 2 maneuvers demonstrate normal sensing of the abandoned device as well as appropriate programming of the output settings to below threshold to ensure noncapture. This combination helps ensure no overpacing occurs. EGM = electrogram; ERI = elective replacement indicator.
Summary of device behavior at ERI in pacemakers by major manufacturer
| Brand | Behavior at ERI | ||
|---|---|---|---|
| Mode | Rate | Features disabled | |
| Medtronic | VVI | Set rate (65 bpm) or programmed rate decrease (7%–10%) | Hysteresis |
| Vitatron | VVI | Set rate (65 bpm) or programmed rate decrease (10%–20%) or increased R-R interval (100 ms) | EGM/episode/diagnostic data collection |
| Boston Scientific | No change at ERI; VVI at EOL | No change at ERI; 50 bpm at EOL | |
| Guidant | VVI | Set rate (65 bpm) or programmed rate decrease (11%–20%) | Automatic capture |
| St. Jude | No change | Increased R-R interval (100 ms) | Rate sensor |
| Biotronik | Single or dual chamber, depending on programmed mode | Programmed rate decrease (4.5%–11%) | Atrial pacing |
| ELA Medical (MicroPort) | VVI | Set rate (70 bpm) | Rate response |
AF = atrial fibrillation; EGM = electrogram; EOL = end of life; ERI = elective replacement indicator; NIPS = noninvasive programmed stimulation; PVC = premature ventricular contraction.
Medtronic Inc., Minneapolis, Minnesota, United States
Boston Scientific, Marlborough, Massachusetts, United States
Maastricht, The Netherlands
Abbott Laboratories, Abbott Park, Illinois, United States
Lake Oswego, Oregon, United States
Arvada, Colorado, United States