| Literature DB >> 35474646 |
Maya Amichay1, Ortal Shimon2, Eitan Raveh3.
Abstract
Background: Closed system drug-transfer devices (CSTD) allow the reconstitution of hazardous drugs into infusion bags, while preserving the sterility of the product and preventing the escape of liquids and aerosols into the environment. Air-cleaning technology CSTD is based on an activated carbon filter and a membrane which enable maintaining the drug sterile by filtration of air entering the vial during pressure equalization. Objective: The study aimed to investigate if an air-cleaning CSTD can prevent liquid viral contamination by human coronavirus OC43 (HCoV-OC43).Entities:
Keywords: COVID-19; Drug preparation; Viral contamination
Year: 2021 PMID: 35474646 PMCID: PMC9013188 DOI: 10.18549/PharmPract.2021.4.2576
Source DB: PubMed Journal: Pharm Pract (Granada) ISSN: 1885-642X
Figure 1Chemfort CSTD with Toxi-Guard System. Vial adaptor, which firmly connects to any standard vial (left), equipped with the Toxi-Guard system with a 100% activated carbon drug adsorbing matrix (middle) and a 0.2 micron hydrophobic and oleophobic membrane (right), serving together as an effective sterile, particulate, and toxic drug vapor barrier.
Figure 2Tested CSTDs with challenged areas. CSTD were either intact vial adaptor (A) or vial adaptor lacking both the sterilizing membrane and activated carbon system (B). The challenged areas sprayed directly with HCoV-OC43 stock solution are indicated by red arrows
Figure 3Using Chemfort CSTD inside glove box. This procedure mimics actual drug reconstitution and preparation done by pharmacists for IV administration in clinics and hospitals
Figure 4HCoV-OC43 standard curve. A HCoV-OC43 stock titer used for generating the standard curve samples was: 5x106 plaque forming units (PFU)/ml. Six 10-fold dilutions were prepared to obtain the different standard curve samples. The cycle threshold (CT) value was determined by qPCR as described in the Methods section. A negative control containing no virus sample was also used. The obtained standard curve equation is y = -3.429x + 38.889. The range for detected extracted viral RNA was 5x105 to 5PFU/ml
qPCR CT results of air samples filters
| Filter sample | CT value (PFU/ml) | ||
|---|---|---|---|
| Replicate | |||
| 1 | 2 | 3 | |
| Negative control[ | 39.96 (0.5) | ND | ND |
| Positive control[ | 21.47 (120000) | 21.35 (130000) | 21.41 (120000) |
| Endpoint[ | 23.29 (35000) | 23.25 (36000) | 23.28 (36000) |
| Positive control II | 22.89 (46000) | 22.72 (52000) | 22.70 (53000) |
| Endpoint II | 22.64 (55000) | 22.62 (55000) | 22.60 (56000) |
Negative control: filter of air sampled from box aerosolized with sterile medium
Positive control: filter of air sampled from box aerosolized with HCoV-OC43 stock solution
Endpoint: filter of air sampled from box aerosolized with HCoV-OC43 stock after the test of CSTDs was complete
CT=cycle threshold, ND=not detected, PFU=plaque forming units
Detection of HCoV-OC43RNA in samples obtained using Chemfort vial adaptors (with or without Toxi- Guard) under various conditions
| Group[ | Sample description | Biological repeat | CT value (PFU/ml) | ||
|---|---|---|---|---|---|
| Replicate number | |||||
| 1 | 2 | 3 | |||
| Negative control[ | Disinfected[ | 1 | ND | ND | ND |
| 1 | Disinfected[ | 1 | ND | ND | ND |
| 2 | ND | ND | ND | ||
| 3 | ND | ND | ND | ||
| 2 | Challenged[ | 1 | ND | ND | ND |
| 2 | ND | ND | ND | ||
| 3 | ND | ND | ND | ||
| 3 | Challenged and disinfected[ | 1 | ND | 38.86 (1.0) | ND |
| 2 | 35.90 (7.5) | 35.68 (8.6) | 35.80 (8.0) | ||
| 3 | ND | ND | 38.19 (1.6) | ||
| 4 | Challenged vial adaptor | I | ND | ND | 38.10 (1.7) |
| 2 | ND | ND | ND | ||
| 3 | 39.03 (0.9) | 36.12 (6.4) | 37.16 (3.2) | ||
Groups 1-4 were used in an HCoV-OC43 aerosolized box, while the Negative Control group was used in a sterile medium aerosolized box.
“Disinfected”: the septa areas only were disinfected with isopropanol pads before every connection of Chemfort device according to the product’s instructions for use.
“Challenged”: the septa and filter areas were directly sprayed with 1.5 ml of HCoV-OC43 stock solution.
“Toxi-Guard-less”: vial adaptor lacking the protective plastic cover as well as the sterilizing membrane and the activated carbonfilter (i.e., the Toxi-Guard)
CT=cycle threshold, ND=not detected, PFU=plaque forming units
| Fluorophore: SYBR green | ||
|---|---|---|
| Temperature | Time | |
| 95°C | 3min | |
| 95°C | 25sec | X45 cycles |
| 57°C | 30sec | |
| 72°C | 20sec | |