Prevention of coronavirus contamination from the environment using an air-cleaning closed system drug-transfer device.

Maya Amichay1, Ortal Shimon2, Eitan Raveh3.   

Abstract

Background: Closed system drug-transfer devices (CSTD) allow the reconstitution of hazardous drugs into infusion bags, while preserving the sterility of the product and preventing the escape of liquids and aerosols into the environment. Air-cleaning technology CSTD is based on an activated carbon filter and a membrane which enable maintaining the drug sterile by filtration of air entering the vial during pressure equalization. Objective: The study aimed to investigate if an air-cleaning CSTD can prevent liquid viral contamination by human coronavirus OC43 (HCoV-OC43).
Methods: Chemfort™ CSTD with (intact) or without (control) a Toxi-Guard system was used to transfer liquids between an IV bag and an empty vial (a total of 5 liquid transfers) inside a sealed glove box contaminated by HCoV-OC43 aerosols. In addition, the vial adaptor was challenged by direct spray of HCoV-OC43 solution on the septum and filter areas. HCoV-OC43 RNA was extracted from samples of the transferred liquid and compared between the devices with or without a Toxi-Guard system.
Results: Use of a CSTD with the Toxi-Guard system resulted in non-detectable cycle threshold (CT) values, indicative of no detectable HCoV-OC43RNA in the transferred liquid, even when the septa and filter areas were directly sprayed with HCoV-OC43 stock solution. In contrast, use of the CSTD with no Toxi-Guard system resulted in a detectable CT value of the transferred liquid. Conclusions: Using Chemfort CSTD with integral Toxi-Guard technology can prevent the introduction of microbial and airborne contaminants into the fluid path, thus potentially protecting patients from infection. Copyright: © Pharmacy Practice.

Entities:  

Keywords:  COVID-19; Drug preparation; Viral contamination

Year:  2021        PMID: 35474646      PMCID: PMC9013188          DOI: 10.18549/PharmPract.2021.4.2576

Source DB:  PubMed          Journal:  Pharm Pract (Granada)        ISSN: 1885-642X


INTRODUCTION

Microbiologically contaminated drug products carry a clinical risk to patients.1-5 Most pharmaceutical facilities have specific procedures in place for controlling contamination by bacteria and fungi. Viral infection can be highly pathogenic, and often there are no effective treatments available. The findings of a recent survey demonstrated that 45% of 20 companies that completed the virus contamination assay reported experiencing at least one virus contamination event between 1985 and 2018.6 Prevention of contaminations is the only way to keep cell cultures for research, development, and the biotech industry free of viruses.7 Closed system transfer devices (CSTD) allow the reconstitution of liquid or pre-dissolved powder drugs into infusion bags, flexible bottles, or syringes, while preserving the sterility of the product, preventing both the entry of contaminants into the system and the escape of the hazardous drug, in whatever form it may exist, into the surrounding environment.8,9 Currently, there are two main technologies for CSTDs: One is based on a physical barrier, where a balloon or a closed chamber holds the air that is exhausted from a vial during drug reconstitution, and the other group of devices uses an air-cleaning system, where the exhausted air leaves the device, but the drug vapor is caught by a filter.10 Chemfort™ CSTD (Simplivia Healthcare Ltd., Kiryat Shmona, Israel;marketed as OnGuard^2 in the USA) is an air-cleaning technology CSTD, using a novel system (Toxi-Guard®) consisting of a hydrophobic filtration membrane with 0.2μm pores and an activated carbon filter. Sterility of the drug in the vial is maintained by filtration of air entering the vial during pressure equalization through a hydrophobic acrylic copolymer membrane with a pore size of 0.2 micron.10,11 The Chemfort CSTD with the Toxi-Guard technology is presented in Figure 1.
Figure 1

Chemfort CSTD with Toxi-Guard System. Vial adaptor, which firmly connects to any standard vial (left), equipped with the Toxi-Guard system with a 100% activated carbon drug adsorbing matrix (middle) and a 0.2 micron hydrophobic and oleophobic membrane (right), serving together as an effective sterile, particulate, and toxic drug vapor barrier.

Chemfort CSTD with Toxi-Guard System. Vial adaptor, which firmly connects to any standard vial (left), equipped with the Toxi-Guard system with a 100% activated carbon drug adsorbing matrix (middle) and a 0.2 micron hydrophobic and oleophobic membrane (right), serving together as an effective sterile, particulate, and toxic drug vapor barrier. Carbon-based materials are known as the most used adsorbents for water and air treatment and are also largely used in virus removal. A recent review analyzed the interaction between viruses and carbon-based systems, among other solutions, regarding efficiency of virus adsorption.12 The authors found that activated carbon adsorbed virus more efficiently than other types of carbon materials. For the specific Toxi-Guard activated carbon system used in our study, laboratory testing performed by the UK’s Health Protection Agency (HPA) using MS-2 coliphage revealed moderate capture level (~10%) of the virus tested. Furthermore, the carbon layer was also shown to have a unique ability to deactivate viruses at a rate of up to 93% without chemical intervention.13 These protective effects of the activated carbon layer enhance the viral filtration efficiency of the 0.2μm membrane. Therefore, both protective layers of the Toxi-Guard system are active against airborne viruses and prevent the risk of virus penetration into the vial. In this study, we examined the ability of Chemfort CSTD with the Toxi-Guard technology to prevent liquid contamination by human coronavirus OC43 (HCoV-OC43) from the outer environment.

METHODS

We compared the efficacy of preventing liquid contamination by an outer environment containing aerosols of human coronavirus of Chemfort CSTD with a vial adaptor in 2 devices: An intact Toxi-Guard vial adaptor versus a non-Toxi-Guard vial adaptor, i.e., a vial adaptor lacking the protective plastic covering and without the activated carbon filter and the sterilizing membrane. Both tested devices are depicted in Figure 2.
Figure 2

Tested CSTDs with challenged areas. CSTD were either intact vial adaptor (A) or vial adaptor lacking both the sterilizing membrane and activated carbon system (B). The challenged areas sprayed directly with HCoV-OC43 stock solution are indicated by red arrows

Tested CSTDs with challenged areas. CSTD were either intact vial adaptor (A) or vial adaptor lacking both the sterilizing membrane and activated carbon system (B). The challenged areas sprayed directly with HCoV-OC43 stock solution are indicated by red arrows The test was performed inside a sealed glove box which was placed in a biological safety laminar flow cabinet. The glove box environment was aerosolized with either sterile growth medium (negative control; NC) or with a known titer of HCoV-OC43 stock solution (5x106 PFU/ml), which was verified twice by a cell plaque assay. The aerosol, comprising 1-5-µm aerosol droplets, was generated by a nebulizer (Air Pro, Medic Spa, China). The CSTDs were handled as described in the device’s instructions for use (IFU) unless otherwise specified, i.e., before connecting the CSTD adaptors to their corresponding items, they were disinfected at the septum areas of all system adaptors with a 70% isopropanol pad (Aplicare Inc., Meriden, CT, USA).Prior to each test, all CSTD adaptors were applied to their corresponding items inside the laminar hood, i.e., the bag adaptor was applied to an intravenous (IV) bag, the vial adaptor was applied to a vial and the syringe adaptor to a syringe. All test articles were inserted into the glove box and the following procedure was performed: a sample of 10 ml of sterile saline was aspirated from the IV bag using a 10-ml syringe equipped with a syringe adaptor and transferred into the glass vial equipped with a vial adaptor. The syringe adaptor was disconnected from the vial adaptor, and the vial was mixed well to mimic proper drug reconstitution. Subsequently, the syringe adaptor was reconnected to the vial adaptor and the sample was transferred from the vial back into the IV bag, via bag adaptor. The syringe adaptor was then disconnected from the bag adaptor, and the IV bag was mixed well. Finally, 10 ml from the IV bag were aspirated into the syringe, which was then disinfected with ethanol before being removed from the glove box. A scheme of the trial procedure is depicted in Figure 3.
Figure 3

Using Chemfort CSTD inside glove box. This procedure mimics actual drug reconstitution and preparation done by pharmacists for IV administration in clinics and hospitals

Using Chemfort CSTD inside glove box. This procedure mimics actual drug reconstitution and preparation done by pharmacists for IV administration in clinics and hospitals A total of 13 CSTDs were used in the study as follows. One CSTD was used according to the IFU in a glove box that was aerosolized by a sterile medium (NC). Twelve other CSTDs were used in a contaminated environment containing aerosolized HCoV-OC43 under the following conditions: (1) Three CSTDs were used with the Toxi-Guard filter according to the IFU, i.e., the septum was disinfected before every connection. (2) Three CSTDs were used with the Toxi-Guard filter but prior to liquid transfers were challenged by spraying HCoV-OC43 stock solution directly onto the outer surfaces of the septum and in the filter area, and then used according to the IFU (i.e., the septum was disinfected before every connection). (3) Three CSTDs without the Toxi-Guard were challenged prior to liquid transfers by spraying HCoV-OC43 stock solution, and then used according to the IFU. (4) Three CSTDs without the Toxi-Guard were challenged prior to liquid transfers by spraying HCoV-OC43 stock solution and were then not disinfected at the septum area before every connection. In addition to sampling the liquid transferred through the CSTD, the air inside the glove box was sampled using an ACD-200 Bobcat Dry Filter Air Sampler (InnovaPrep, Drexel, MO, USA). Air samples were taken (1) at the beginning of the test, after aerosolizing the box with sterile medium (NC), (2) before initiating the liquid sampling using the CSTDs and after aerosolizing the glove box with HCoV-OC43 stock solution, (3) after sampling all CSTDs while the box was aerosolized continuously with virus stock. Two additional air samplings were carried out before (PC II) and after (end-point II) the testing of sampling group 4.

Viral RNA extraction

After sampling, viral RNA was extracted from each sample using the MagCore viral nucleic acid extraction kit (RBC Bioscience, New Taipei City, Taiwan) according to the manufacturer’s instructions. RNA extraction was performed in triplicate samples from each CSTD sampled. Following RNA purification, 30µl from each triplicate of each Chemfort sample set were combined into one tube to generate a total of 12 RNA samples.

Complementary DNA preparation and quantitative PCR

Complementary DNA (cDNA) was synthesized from each pool of extracted RNA sample using the Hy RT PCR Kit (Hylabs, Rehovot, Israel). Briefly, 12.4µl from each RNA sample were used for both the test samples and the no reverse transcriptase (-RT) controls. The cDNA and -RT samples were each diluted 1:2 and then used as templates for quantitative polymerase chain reaction (qPCR), in a mix containing specific primers for HCoV-OC43 (forward:5’-ATTGTCGATCGGGACCCAAG-3’;reverse:5’-TGTGCGCGAAGTAGATCTGG-3’)and platinum SYBR Green qPCR SuperMix-UDG (Invitrogen, Waltham, MA, USA). The specificity of the HCoV-OC43 qPCR specific primers was demonstrated by using the cDNA of influenza A virus H1N1 as a template with the HCoV-OC43 specific primers, which resulted in high cycle threshold (CT; see explanation below)values or no signal at all (38.26, 36.93 and ‘ND’). As positive control, H1N1 cDNA was targeted by its corresponding specific primers (Forward: 5’-CTTCTAACCG AGGTCGAAACGTA-3’; reverse: 5’-GGTGACAGGATTGGTCTTGTCTTTA-3’), which resulted in CTvalues of 24.13, 24.21 and 24.51, respectively. qPCR was performed using the CFX96 instrument (Bio-Rad, CA, USA) as follows: Fluorophore: SYBR green - Plate read following each cycle - Melt curve (65-95°C) + plate read

Determination of cycle threshold values using HCoV-OC43 standard curve

The cycle threshold (CT) is the qPCR cycle during which the fluorescence of the amplified target has crossed the threshold value, i.e., where the PCR product starts to accumulate and becomes detectable in the assay. Therefore, the lower the CT value is, the higher the target nucleic acid concentration is, and accordingly, the higher the CT value is, the lower the target nucleic acid concentration is. A HCoV-OC43 stock solution at a concentration of 5x106 plaque forming units (PFU)/ml was used for generating a standard curve. Six 10-fold dilutions were prepared to obtain the different standard curve samples, ranging from 5 to 5x105 PFU/ml. RNA was extracted from each dilution sample and analyzed by qPCR, as described above, for the samples obtained from the Chemfort devices. The cycle threshold value for each dilution sample was determined and plotted against the PFU/ml value. The HCoV-OC43 standard curve is presented in Figure 4. The range for detected extracted viral RNA was 5x105 to5PFU/ml. NC values of viral RNA were 39 (determined as 5 PFU/ml according to the HCoV-OC43 standard curve) and were defined as the lower limit of detection/lower limit of quantification.
Figure 4

HCoV-OC43 standard curve. A HCoV-OC43 stock titer used for generating the standard curve samples was: 5x106 plaque forming units (PFU)/ml. Six 10-fold dilutions were prepared to obtain the different standard curve samples. The cycle threshold (CT) value was determined by qPCR as described in the Methods section. A negative control containing no virus sample was also used. The obtained standard curve equation is y = -3.429x + 38.889. The range for detected extracted viral RNA was 5x105 to 5PFU/ml

HCoV-OC43 standard curve. A HCoV-OC43 stock titer used for generating the standard curve samples was: 5x106 plaque forming units (PFU)/ml. Six 10-fold dilutions were prepared to obtain the different standard curve samples. The cycle threshold (CT) value was determined by qPCR as described in the Methods section. A negative control containing no virus sample was also used. The obtained standard curve equation is y = -3.429x + 38.889. The range for detected extracted viral RNA was 5x105 to 5PFU/ml

Statistical analysis

CFX Manager 2.1 software package (Bio-Rad, CA, USA) was used for data analysis.

RESULTS

Air sampling

When the glove box was aerosolized with sterile medium (negative control), no viral RNA was detected in the air. However, when the box was aerosolized with the HCoV-OC43 stock solution (positive control), the air samples showed low CT values, indicative of the presence of viral RNA corresponding roughly to 35000-130000 PFU/ml. The air sampling results for negative and positive control for the glove box, before and after the CSTDs were tested, are presented in Table 1.
Table 1

qPCR CT results of air samples filters

Filter sampleCT value (PFU/ml)
Replicate
123
Negative control[1]39.96 (0.5)NDND
Positive control[2]21.47 (120000)21.35 (130000)21.41 (120000)
Endpoint[3]23.29 (35000)23.25 (36000)23.28 (36000)
Positive control II22.89 (46000)22.72 (52000)22.70 (53000)
Endpoint II22.64 (55000)22.62 (55000)22.60 (56000)

Negative control: filter of air sampled from box aerosolized with sterile medium

Positive control: filter of air sampled from box aerosolized with HCoV-OC43 stock solution

Endpoint: filter of air sampled from box aerosolized with HCoV-OC43 stock after the test of CSTDs was complete

CT=cycle threshold, ND=not detected, PFU=plaque forming units

qPCR CT results of air samples filters Negative control: filter of air sampled from box aerosolized with sterile medium Positive control: filter of air sampled from box aerosolized with HCoV-OC43 stock solution Endpoint: filter of air sampled from box aerosolized with HCoV-OC43 stock after the test of CSTDs was complete CT=cycle threshold, ND=not detected, PFU=plaque forming units

Liquid sample transition

The results of HCoV-OC43 RNA detection in samples obtained using Chemfort vial adaptors(with or without Toxi-Guard) under various conditions in different sample groups are presented in Table 2. When liquid samples were transferred using the adaptors containing the Toxi-Guard system and septa were disinfected with isopropanol according to the IFU, there was no evidence of viral RNA traces, even when the vial adaptors were directly sprayed with an HCoV-OC43 stock solution (Table 2, sample groups 1 and 2). In contrast, liquid samples that were transferred by the same procedure using vial adaptors not containing the Toxi-Guard filter that were also challenged by spraying HCoV-OC43 stock solution, were found to be positive for viral RNAranging between 0.9 and 8.6 PFU/ml (Table 2, sample groups 3 and 4). Notably, absence of the Toxi-Guard system, resulted in detection ofviral RNA in the liquid samples regardless of whether the vial adaptor at the septum area was disinfected.
Table 2

Detection of HCoV-OC43RNA in samples obtained using Chemfort vial adaptors (with or without Toxi- Guard) under various conditions

Group[1]Sample descriptionBiological repeatCT value (PFU/ml)
Replicate number
123
Negative control[1]Disinfected[2] Toxi-Guard vial adaptor (negative control)1NDNDND
1Disinfected[2] Toxi-Guard vial adaptor1NDNDND
2NDNDND
3NDNDND
2Challenged[3] and disinfected Toxi-Guard vial adaptor1NDNDND
2NDNDND
3NDNDND
3Challenged and disinfected[2] vial adaptor Toxi-Guard-less[4]1ND38.86 (1.0)ND
235.90 (7.5)35.68 (8.6)35.80 (8.0)
3NDND38.19 (1.6)
4Challenged vial adaptor Toxi-Guard-less[4]INDND38.10 (1.7)
2NDNDND
339.03 (0.9)36.12 (6.4)37.16 (3.2)

Groups 1-4 were used in an HCoV-OC43 aerosolized box, while the Negative Control group was used in a sterile medium aerosolized box.

“Disinfected”: the septa areas only were disinfected with isopropanol pads before every connection of Chemfort device according to the product’s instructions for use.

“Challenged”: the septa and filter areas were directly sprayed with 1.5 ml of HCoV-OC43 stock solution.

“Toxi-Guard-less”: vial adaptor lacking the protective plastic cover as well as the sterilizing membrane and the activated carbonfilter (i.e., the Toxi-Guard)

CT=cycle threshold, ND=not detected, PFU=plaque forming units

Detection of HCoV-OC43RNA in samples obtained using Chemfort vial adaptors (with or without Toxi- Guard) under various conditions Groups 1-4 were used in an HCoV-OC43 aerosolized box, while the Negative Control group was used in a sterile medium aerosolized box. “Disinfected”: the septa areas only were disinfected with isopropanol pads before every connection of Chemfort device according to the product’s instructions for use. “Challenged”: the septa and filter areas were directly sprayed with 1.5 ml of HCoV-OC43 stock solution. “Toxi-Guard-less”: vial adaptor lacking the protective plastic cover as well as the sterilizing membrane and the activated carbonfilter (i.e., the Toxi-Guard) CT=cycle threshold, ND=not detected, PFU=plaque forming units

DISCUSSION

The benefits of CSTD systems in preventing exposure to hazardous drugs, such as antineoplastic agents and antibiotics are well established.14-17 With regards to preventing contamination of the drug preparation, CSTDs have been recently shown to prevent contamination by Gram-positive bacteria, Gram-negative bacteria, yeast, and mold.18 In addition, it was found that vials punctured in ISO5 conditions with a CSTD demonstrated a low frequency of microbial contamination.19 However, viruses are often more difficult to detect than other microbial contaminants.7 As far as we know, our study is the first to demonstrate that CSTD with an air-cleaning technology can prevent ingress of external coronavirus and airborne contaminants into liquids. In this study, CSTDs were used to transfer liquid between an IV bag and an empty vial and back to the bag, which mimics drug reconstitution and preparation for IV administration in clinics and hospitals, as routinely performed by pharmacists (Figure 3). Therefore, the results of this study, demonstrating that CSTDs with air-cleaning technology can prevent internal viral contamination, are highly important for clinical facilities, where patients can be exposed to different hazardous threats. The results of our study demonstrate the efficacy of Chemfort CSTD in preventing viral penetration, as shown by air sample results (Table 1). The generation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) bioaerosols plays an important role in the COVID-19 pandemic. Genetic materials of coronaviruses have been detected in air in several studies.20-26 SARS-CoV-2 was shown to retain infectivity and virion integrity for up to 16 hours in respirable-sized aerosols.27 According to Lee, the minimum size of a respiratory particle that can contain SARS-CoV-2 is approximately 4.7μm.28 The minimum size of the particles can decrease due to the evaporation of water on the particle surfaces. Moreover, high viral loads can decrease the minimum size of respiratory particles containing SARS-CoV-2, thereby increasing the probability of aerosol generation of the viruses. The particle size of the coronavirus used in this study, HCoV-OC43, is 0.8-0.12 µm smaller than the particle size of SARS-CoV-2.29,30 In our study, we investigated the Chemfort CSTD, which is based on air-cleaning technology (Chemfort/OnGuard^2). As mentioned, there are also devices are based on a physical barrier technology, where a balloon or a closed chamber are used to hold the air during drug reconstitution. It may be that our results may not be relevant for physical-barrier CSTDs, which may act differently when exposed to viral contamination during drug reconstitution and preparation.

CONCLUSIONS

In conclusion, Chemfort CSTD with the integral Toxi-Guard system was found effective in preventing viral contamination of liquids transferred by the device. During the COVID-19 epidemic, use of a CSTD with an activated carbon system can be of importance to protect already vulnerable patients from further complications due to viral infection.
Fluorophore: SYBR green
TemperatureTime
95°C3min
95°C25sec  X45 cycles
57°C30sec
72°C20sec
  26 in total

1.  Endemic infusate contamination and related bacteremia.

Authors:  Alejandro E Macias; Samuel Ponce de Leon; Martha Huertas; Ernesto Maravilla; Carmen Romero; Thalpa G Montoya; Juan M Muñoz; Yolanda Lopez-Vidal
Journal:  Am J Infect Control       Date:  2008-02       Impact factor: 2.918

2.  Reduction in surface contamination with antineoplastic drugs in 22 hospital pharmacies in the US following implementation of a closed-system drug transfer device.

Authors:  Paul J M Sessink; Thomas H Connor; James A Jorgenson; Timothy G Tyler
Journal:  J Oncol Pharm Pract       Date:  2010-02-15       Impact factor: 1.809

3.  An Industry Perspective on Compatibility Assessment of Closed System Drug-Transfer Devices for Biologics.

Authors:  Ahmed Besheer; Lori Burton; Richard J Galas; Kedar Gokhale; Pierre Goldbach; Qingyan Hu; Ligi Mathews; Kevin Muthurania; Chakravarthy Narasimhan; Shubhadra N Singh; Elaine S E Stokes; Sarah Weiser; Camellia Zamiri; Shuxia Zhou
Journal:  J Pharm Sci       Date:  2020-10-28       Impact factor: 3.534

4.  Complete genomic sequence of human coronavirus OC43: molecular clock analysis suggests a relatively recent zoonotic coronavirus transmission event.

Authors:  Leen Vijgen; Els Keyaerts; Elien Moës; Inge Thoelen; Elke Wollants; Philippe Lemey; Anne-Mieke Vandamme; Marc Van Ranst
Journal:  J Virol       Date:  2005-02       Impact factor: 5.103

5.  Reduction of Contamination with Antibiotics on Surfaces and in Environmental Air in Three European Hospitals Following Implementation of a Closed-System Drug Transfer Device.

Authors:  Paul J M Sessink; Tibor Nyulasi; Emmie L M Haraldsson; Branislava Rebic
Journal:  Ann Work Expo Health       Date:  2019-04-19       Impact factor: 2.179

Review 6.  Hospital-acquired infections related to contaminated substances.

Authors:  R-P Vonberg; P Gastmeier
Journal:  J Hosp Infect       Date:  2006-12-04       Impact factor: 3.926

7.  Collection of SARS-CoV-2 Virus from the Air of a Clinic Within a University Student Health Care Center and Analyses of the Viral Genomic Sequence.

Authors:  John A Lednicky; Sripriya N Shankar; Maha A Elbadry; Julia C Gibson; Md Mahbubul Alam; Caroline J Stephenson; Arantzazu Eiguren-Fernandez; J Glenn Morris; Carla N Mavian; Marco Salemi; James R Clugston; Chang-Yu Wu
Journal:  Aerosol Air Qual Res       Date:  2020-05-25       Impact factor: 3.063

8.  Multicenter evaluation of a new closed system drug-transfer device in reducing surface contamination by antineoplastic hazardous drugs.

Authors:  Sylvia B Bartel; Timothy G Tyler; Luci A Power
Journal:  Am J Health Syst Pharm       Date:  2018-01-16       Impact factor: 2.637

9.  Aerosol and Surface Distribution of Severe Acute Respiratory Syndrome Coronavirus 2 in Hospital Wards, Wuhan, China, 2020.

Authors:  Zhen-Dong Guo; Zhong-Yi Wang; Shou-Feng Zhang; Xiao Li; Lin Li; Chao Li; Yan Cui; Rui-Bin Fu; Yun-Zhu Dong; Xiang-Yang Chi; Meng-Yao Zhang; Kun Liu; Cheng Cao; Bin Liu; Ke Zhang; Yu-Wei Gao; Bing Lu; Wei Chen
Journal:  Emerg Infect Dis       Date:  2020-06-21       Impact factor: 6.883

10.  Persistence of Severe Acute Respiratory Syndrome Coronavirus 2 in Aerosol Suspensions.

Authors:  Alyssa C Fears; William B Klimstra; Paul Duprex; Amy Hartman; Scott C Weaver; Kenneth S Plante; Divya Mirchandani; Jessica Ann Plante; Patricia V Aguilar; Diana Fernández; Aysegul Nalca; Aysegul Totura; David Dyer; Brian Kearney; Matthew Lackemeyer; J Kyle Bohannon; Reed Johnson; Robert F Garry; Doug S Reed; Chad J Roy
Journal:  Emerg Infect Dis       Date:  2020-06-22       Impact factor: 6.883

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.