Ventilator dysfunction due to unexpected salbutamol crystallisation while using a nebuliser: a practical advice for intensivists.

Jos L M L le Noble1,2, Frank J M Terpoorten3, Mary-Anne Funnekotter-van der Snoek4, Norbert Foudraine5.   

Abstract

Entities:  

Keywords:  Aerosol; Crystallisation; Nebuliser; Salbutamol; Ventilator malfunction

Year:  2022        PMID: 35467162      PMCID: PMC9038966          DOI: 10.1186/s40635-022-00441-y

Source DB:  PubMed          Journal:  Intensive Care Med Exp        ISSN: 2197-425X


× No keyword cloud information.

To the Editor,

In our ICU, we use a Servo-u® mechanical ventilator (Getinge). During a regular pre-use check, a breathing circuit malfunction was detected along with another malfunction during the expiratory hold manoeuvre. Inspection of the ventilator outlet system revealed unexpected deposition of white crystals of unknown origin, both within the channel and on the outlet valve membrane. A dysfunctional outlet valve membrane may disrupt accurate levels of positive end-expiratory pressure and facilitate backflow of the exhaled air, causing ventilator malfunction and harm to the patient. We hypothesised that aerosol crystallisation during pulmonary drug delivery could be a likely cause. We performed an experimental bench study using a Servo-u® mechanical ventilator connected to a test lung and investigated whether nebulisation of frequently used drugs causes crystal depositions. Standard warming and humidification were performed with an MR950 humidifier (Fisher & Paykel®) [1]. Two types of expiratory filters were used: standard Fisher & Paykel® filters (Part No. RT019) and Getinge® filters (Servo Duo Guard) [1]. Aerosol drug delivery was performed using a mesh nebuliser (Aeroneb® Solo). The tested drugs were 2.0 mL ipratropium (Atrovent®: 500 μg/2 mL), 2.5 mL salbutamol (Ventolin®: 5.0 mg/2.5 mL, Nebules), 2 mL acetylcysteine (Fluimucil®: 100 mg/mL), and 5 ml iloprost (4 μg/mL). Saline solution (5 mL, NaCl 0.9%) served as the control. Tandem mass spectrometry (UPLC-MS/MC) and UPLC PDA-UV spectrophotometry were used to analyse the chemical composition of the crystal deposits. Macroscopically visible crystals were observed only following Ventolin® aerosol delivery using a standard Fisher & Paykel® filter (Fig. 1). No deposition was observed using the Getinge® filter. Further, inconsistencies in expiratory pressures were observed on the ventilator display, indicating outlet valve malfunction. Both UPLC–MS/MC and UPLC PDA-UV analyses of the crystals identified salbutamol.
Fig. 1

a Ventilator outlet system of the Servo-u®. Note the white deposition of salbutamol crystals which occurred following salbutamol nebulisation. The drum on top of the ventilator was removed. b The drum of the ventilator covered with white deposition of salbutamol crystals

a Ventilator outlet system of the Servo-u®. Note the white deposition of salbutamol crystals which occurred following salbutamol nebulisation. The drum on top of the ventilator was removed. b The drum of the ventilator covered with white deposition of salbutamol crystals The main finding of our study is that salbutamol nebulisation can lead to crystallisation in the outlet system of a ventilator (Servo-u®) when a Fisher & Paykel® breathing circuit filter is used. Previously, only one case report has shown intraoperative obstruction of the filter in the expiratory limb of the breathing circuit by nebulised salbutamol in paediatric patients [2]. Ventilated patients are frequently administered high doses of the short-acting β-agonist salbutamol [3, 4]. Heat and humidification may cause supersaturation and precipitation of salbutamol when cooling down in the expiratory limb [1]. Furthermore, differences in expiratory filter characteristics, a large filter pore size, and the size distribution of aerosols could account for salbutamol crystallisation [1, 5]. In our setup, the aerosol administration was not synchronised with inspiratory flow, which may have led to aerosol losses during exhalation, with large amounts of the drug being injected into the expiratory limb and engulfing the outlet valve. Therefore, a breath-synchronised mesh nebuliser has been developed which generates aerosols only during inspiration [3]. Depending on the filter used in the expiratory limb, salbutamol nebulisation may cause crystallisation in the outlet system of the breathing circuit of the ventilator. Therefore, salbutamol prescriptions in the ICU should address nebulisation practices, and aerosols should only be provided during inspiratory flow during mechanical ventilation.
  5 in total

1.  Intraoperative breathing circuit obstruction caused by albuterol nebulization.

Authors:  J S Walton; R Fears; N Burt; B H Dorman
Journal:  Anesth Analg       Date:  1999-09       Impact factor: 5.108

2.  Impact of humidification and nebulization during expiratory limb protection: an experimental bench study.

Authors:  Alexandre Tonnelier; François Lellouche; Pierre Alexandre Bouchard; Erwan L'Her
Journal:  Respir Care       Date:  2012-12-04       Impact factor: 2.258

3.  Size distribution of salbutamol/ipratropium aerosols produced by different nebulizers in the absence and presence of heat and humidification.

Authors:  Ssu-Han Yang; Tsung-Ming Yang; Hui-Ling Lin; Ying-Huang Tsai; Tien-Pei Fang; Gwo-Hwa Wan
Journal:  Pulm Pharmacol Ther       Date:  2017-10-18       Impact factor: 3.410

4.  How Should Aerosols Be Delivered During Invasive Mechanical Ventilation?

Authors:  Rajiv Dhand
Journal:  Respir Care       Date:  2017-09-05       Impact factor: 2.258

5.  The Clinical Practice and Best Aerosol Delivery Location in Intubated and Mechanically Ventilated Patients: A Randomized Clinical Trial.

Authors:  Chuanlin Zhang; Jie Mi; Zeju Zhang; Xueqin Wang; Yunxiao Zhu; Xinyi Luo; Ruiying Gan; Xiaoya Chen; Yujun Zou
Journal:  Biomed Res Int       Date:  2021-04-03       Impact factor: 3.411

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.