Impact of humidification and nebulization during expiratory limb protection: an experimental bench study.

Alexandre Tonnelier1, François Lellouche, Pierre Alexandre Bouchard, Erwan L'Her.   

Abstract

BACKGROUND: Different filtering devices are used during mechanical ventilation to avoid dysfunction of flow and pressure transducers or for airborne microorganisms containment. Water condensates, resulting from the use of humidifiers, but also residual nebulization particles may have a major influence on expiratory limb resistance.
OBJECTIVES: To evaluate the influence of nebulization and active humidification on the resistance of expiratory filters.
METHODS: A respiratory system analog was constructed using a test lung, an ICU ventilator, heated humidifiers, and a piezoelectric nebulizer. Humidifiers were connected to different types of circuits (unheated, mono-heated, new-generation and old-generation bi-heated). Five filter types were evaluated: electrostatic, heat-and-moisture exchanger, standard, specific, and internal heated high-efficiency particulate air [HEPA] filter. Baseline characteristics were obtained from each dry filter. Differential pressure measurements were carried out after 24 hours of continuous in vitro use for each condition, and after 24 hours of use with an old-generation bi-heated circuit without nebulization.
RESULTS: While using unheated circuits, measurements had to be interrupted before 24 hours for all the filtering devices except the internal heated HEPA filter. The heat-and-moisture exchangers occluded before 24 hours with the unheated and mono-heated circuits. The circuit type, nebulization practice, and duration of use did not influence the internal heated HEPA filter resistance.
CONCLUSION: Expiratory limb filtration is likely to induce several major adverse events. Expiratory filter resistance increase is due mainly to the humidification circuit type, rather than to nebulization. If filtration is mandatory while using an unheated circuit, a dedicated filter should be used for ≤ 24 hours, or a heated HEPA for a longer duration.

Entities:  

Keywords:  expiratory limb protection; filters; humidification; mechanical ventilation; nebulization

Mesh:

Year:  2012        PMID: 23232732     DOI: 10.4187/respcare.01785

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  8 in total

Review 1.  Administration of dry powders during respiratory supports.

Authors:  Wei-Ren Ke; Rachel Yoon Kyung Chang; Philip Chi Lip Kwok; Patricia Tang; Lan Chen; Donghao Chen; Hak-Kim Chan
Journal:  Ann Transl Med       Date:  2021-04

2.  In-vitro and in-vivo comparisons of high versus low concentrations of inhaled epoprostenol to adult intubated patients.

Authors:  Jie Li; Ashley E Augustynovich; Payal K Gurnani; James B Fink
Journal:  Respir Res       Date:  2021-08-21

3.  Ventilator dysfunction due to unexpected salbutamol crystallisation while using a nebuliser: a practical advice for intensivists.

Authors:  Jos L M L le Noble; Frank J M Terpoorten; Mary-Anne Funnekotter-van der Snoek; Norbert Foudraine
Journal:  Intensive Care Med Exp       Date:  2022-04-25

4.  Aerosol therapy in intensive and intermediate care units: prospective observation of 2808 critically ill patients.

Authors:  Stephan Ehrmann; Ferran Roche-Campo; Laetitia Bodet-Contentin; Keyvan Razazi; Jonathan Dugernier; Josep Trenado-Alvarez; Alexis Donzeau; François Vermeulen; David Thévoz; Metaxia Papanikolaou; Antoine Edelson; Héctor León Yoshido; Lise Piquilloud; Karim Lakhal; Carlos Lopes; Carlos Vicent; Arnaud Desachy; Gabriela Apiou-Sbirlea; Daniel Isabey; Laurent Brochard
Journal:  Intensive Care Med       Date:  2015-11-24       Impact factor: 17.440

Review 5.  Heat and moisture exchangers (HMEs) and heated humidifiers (HHs) in adult critically ill patients: a systematic review, meta-analysis and meta-regression of randomized controlled trials.

Authors:  Maria Vargas; Davide Chiumello; Yuda Sutherasan; Lorenzo Ball; Antonio M Esquinas; Paolo Pelosi; Giuseppe Servillo
Journal:  Crit Care       Date:  2017-05-29       Impact factor: 9.097

6.  Recommended Approaches to Minimize Aerosol Dispersion of SARS-CoV-2 During Noninvasive Ventilatory Support Can Cause Ventilator Performance Deterioration: A Benchmark Comparative Study.

Authors:  Maxime Patout; Emeline Fresnel; Manuel Lujan; Claudio Rabec; Annalisa Carlucci; Léa Razakamanantsoa; Adrien Kerfourn; Hilario Nunes; Yacine Tandjaoui-Lambiotte; Antoine Cuvelier; Jean-François Muir; Cristina Lalmoda; Bruno Langevin; Javier Sayas; Jesus Gonzalez-Bermejo; Jean-Paul Janssens
Journal:  Chest       Date:  2021-03-02       Impact factor: 9.410

Review 7.  Nebulized antibiotics in mechanically ventilated patients: a challenge for translational research from technology to clinical care.

Authors:  Stephan Ehrmann; Jean Chastre; Patrice Diot; Qin Lu
Journal:  Ann Intensive Care       Date:  2017-08-01       Impact factor: 6.925

8.  Nebulized Heparin in Burn Patients with Inhalation Trauma-Safety and Feasibility.

Authors:  Gerie J Glas; Janneke Horn; Jan M Binnekade; Markus W Hollmann; Jan Muller; Berry Cleffken; Kirsten Colpaert; Barry Dixon; Nicole P Juffermans; Paul Knape; Marcel M Levi; Bert G Loef; David P Mackie; Manu L N G Malbrain; Benedikt Preckel; Auke C Reidinga; K F van der Sluijs; Marcus J Schultz
Journal:  J Clin Med       Date:  2020-03-25       Impact factor: 4.241

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.