Size distribution of salbutamol/ipratropium aerosols produced by different nebulizers in the absence and presence of heat and humidification.

Ssu-Han Yang1, Tsung-Ming Yang2, Hui-Ling Lin3, Ying-Huang Tsai4, Tien-Pei Fang5, Gwo-Hwa Wan6.   

Abstract

BACKGROUND: Few studies have evaluated the size distribution of inhaled and exhaled aerosolized drugs, or the effect of heated humidification on particle size and lung deposition. The present study evaluated these aspects of bronchodilator (salbutamol/ipratropium) delivery using a lung model in the absence and presence of heat and humidification.
METHODS: We positioned filters to collect and measure the initial drug, inhaled drug, and exhaled drug. Particle size distribution was evaluated using an 8-stage Marple personal cascade impactor with 0.2-μm polycarbonate filters.
RESULTS: A greater inhaled drug mass was delivered using a vibrating mesh nebulizer (VMN) than by using a small volume nebulizer (SVN), when heated humidifiers were not employed. When heated and humidified medical gas was used, there was no significant difference between the inhaled drug mass delivered by the VMN and that delivered by the SVN. A significantly greater mass of inhaled 1.55-μm drug particles was produced by the VMN than with the SVN, under heated and humidified conditions. However, the mass median aerodynamic diameters (MMADs) of the aerosolized drug produced by the SVN and VMN did not differ significantly under the same conditions.
CONCLUSIONS: The VMN produced more fine particles of salbutamol/ipratropium, and the drug particle size clearly increased in the presence of heat and humidification.
Copyright © 2017 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Bronchodilator; Lung model; Small volume nebulizer; Ventilator; Vibrating mesh nebulizer

Mesh:

Substances:

Year:  2017        PMID: 29054517     DOI: 10.1016/j.pupt.2017.10.009

Source DB:  PubMed          Journal:  Pulm Pharmacol Ther        ISSN: 1094-5539            Impact factor:   3.410


  3 in total

Review 1.  Aerosol delivery via invasive ventilation: a narrative review.

Authors:  Hui-Ling Lin; James B Fink; Huiqing Ge
Journal:  Ann Transl Med       Date:  2021-04

2.  Ventilator dysfunction due to unexpected salbutamol crystallisation while using a nebuliser: a practical advice for intensivists.

Authors:  Jos L M L le Noble; Frank J M Terpoorten; Mary-Anne Funnekotter-van der Snoek; Norbert Foudraine
Journal:  Intensive Care Med Exp       Date:  2022-04-25

3.  Size Distribution of Colistin Delivery by Different Type Nebulizers and Concentrations During Mechanical Ventilation.

Authors:  Ching-Yi Liu; Hsin-Kuo Ko; James B Fink; Gwo-Hwa Wan; Chung-Chi Huang; Yu-Chun Chen; Hui-Ling Lin
Journal:  Pharmaceutics       Date:  2019-09-05       Impact factor: 6.321

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.