Tao Zhan1,2, Xiao-Zhen Feng2, Qi-Qi An2, Shiyong Li1, Mingyue Xue2, Zhencheng Chen1,2, Guo-Cheng Han2, Heinz-Bernhard Kraatz3.
Abstract
Gold electrodes (GE) were modified by the deposition of copper nanoparticles (CuNPs) and cobalt nanoparticles (CoNPs), followed by drop-casting of the ferrocene derivative FcCO-Glu-Cys-Gly-OH (Fc-ECG), resulting in two enzyme-free electrochemical sensors Fc-ECG/CuNPs/GE and Fc-ECG/CuNPs/GE. The ferrocene-peptide conjugate acts as an effective redox mediator for glucose oxidation, while metal nanoparticles acted as non-biological sites for glucose oxidation. Field emission scanning electron microscopy (FESEM), cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) were carried out for characterization, while differential pulse voltammetry (DPV) was used for glucose quantification. Under optimized conditions, DPV shows a linear relationship between glucose concentration and the peak current. Both sensors showed a surprisingly high sensitivity of 217.27 and 378.70 μA mM-1 cm-2, respectively. A comparison to other glucose sensors shows a sensitivity that is 25 times higher. The sensors exhibit good reproducibility, stability, and repeatability. In injection experiments, recovery rates were 87.39-107.65% and 100.00-106.88%, respectively. This journal is © The Royal Society of Chemistry.Entities:
Year: 2022 PMID: 35425584 PMCID: PMC8981370 DOI: 10.1039/d1ra09213h
Source DB: PubMed Journal: RSC Adv ISSN: 2046-2069 Impact factor: 3.361
Scheme 1Preparation of Fc-ECG/CuNPs/GE and Fc-ECG/CoNPs/GE sensors for glucose detection by DPV technique in alkaline solution.
Fig. 1SEM (left) and EDS (right) characterization of the prepared electrodes. (A and B) Screen printing electrode with gold deposited; (C and D) CuNPs/GE; (E and F) Fc-ECG/CuNPs/GE; (G and H) CoNPs/GE; (I and J) Fc-ECG/CoNPs/GE.
Fig. 2(A) DPVs of Fc-ECG/CuNPs/GE in mixed solution of 0.1 M NaOH and different concentrations (0.40–2.30 mM) glucose, inset: calibration plots for various concentrations of glucose; (B) DPVs of Fc-ECG/CuNPs/GE in mixed solution of 0.1 M NaOH and different concentrations (0.70–3.60 mM) glucose, inset: calibration plots for various concentrations of glucose.
Performance comparison of various glucose sensors
| Sensors | Linear ranges (mM) | Sensitivity (μA mM−1 cm−2) | Reference |
|---|---|---|---|
| NPGWE | 0.50–10.00 | 128.80 |
|
| PAA-VS-PANI/GPL-FePc/GOx-CH | 1.00–20.00 | 18.11 |
|
| PAA-rGO/VS-PANI/LuPc2/GOx-MFH | 2.00–12.00 | 15.31 |
|
| Cu/Cu2O nanoporous NPs | 0.01–5.50 | 123.80 |
|
| PdNS-Cu/Cu2O/FTO | 0.0005–2.60 | — |
|
| AuNPs@CuO NWs/Cu2O/CF | 0.0028–2.00 | 1.62 |
|
| NiCo-LDH/CCCH/CuF | 0.001–1.50 | 10.78 |
|
| Fc-ECG/CuNPs/GE | 0.40–2.30 | 217.27 | This work |
| Fc-ECG/CoNPs/GE | 0.70–3.60 | 378.70 | This work |
Fig. 3(A) Selectivity of Fc-ECG/CuNPs/GE (b) and Fc-ECG/CoNPs/GE (a) in the presence of interfering substances by i–t measurements; (B) stability studies of Fc-ECG/CuNPs/GE and Fc-ECG/CoNPs/GE in the presence of 0.4 and 3.6 mM glucose, respectively (the insets are the DPV responses); (C) repeatability studies of Fc-ECG/CuNPs/GE and Fc-ECG/CoNPs/GE by using 4 different preparations of the same electrodes in the presence of glucose (the insets are the DPV responses); (D) reproducibility studies of Fc-ECG/CuNPs/GE and Fc-ECG/CoNPs/GE by measuring 4 times (the insets are the DPV responses).
Detection of glucose in glucose injection samples (n = 3)
| Added glucose (mM) | Measured current (μA) | Concentration founded (mM) | Recovery (%) | RSD (%) | ||||
|---|---|---|---|---|---|---|---|---|
| a | b | a | b | a | b | a | b | |
| 23.54 | 30.42 | |||||||
| 0 | 22.67 | 30.42 | 1.38 ± 0.04 | 1.45 ± 0.01 | — | — | 2.61 | 0.40 |
| 22.38 | 30.26 | |||||||
| 30.52 | 37.35 | |||||||
| 1.70(a)/1.60 (b) | 29.83 | 37.09 | 1.83 ± 0.17 | 1.71 ± 0.10 | 107.65 | 106.88 | 6.44 | 0.58 |
| 2 | 27.13 | 36.91 | ||||||
| 31.61 | 45.01 | |||||||
| 2.00 | 31.26 | 44.18 | 2.00 ± 0.14 | 2.00 ± 0.03 | 100.00 | 100.00 | 2.40 | 0.77 |
| 30.22 | 44.45 | |||||||
| 34.22 | 50.43 | |||||||
| 2.30 | 32.03 | 49.65 | 2.01 ± 0.11 | 2.30 ± 0.15 | 87.39 | 100.00 | 5.11 | 1.38 |
| 31.13 | 49.00 | |||||||
a: Fc-ECG/CuNPs/GE; b: Fc-ECG/CoNPs/GE.