Literature DB >> 35386351

Evaluating metallic artefact of biodegradable magnesium-based implants in magnetic resonance imaging.

Jonathan Espiritu1, Mostafa Berangi2,3,4, Christina Yiannakou5, Eduarda Silva6, Roberto Francischello7,8, Andre Kuehne2, Thoralf Niendorf2,3,4, Sören Könneker3, Regine Willumeit-Römer9, Jan-Marten Seitz1.   

Abstract

Magnesium (Mg) implants have shown to cause image artefacts or distortions in magnetic resonance imaging (MRI). Yet, there is a lack of information on how the degradation of Mg-based implants influences the image quality of MRI examinations. In this study, Mg-based implants are analysed in vitro, ex vivo, and in the clinical setting for various magnetic field strengths with the aim to quantify metallic artefact behaviour. In vitro corroded Mg-based screws and a titanium (Ti) equivalent were imaged according to the ASTM F2119. Mg-based and Ti pins were also implanted into rat femurs for different time points and scanned to provide insights on the influence of soft and hard tissue on metallic artefact. Additionally, MRI data of patients with scaphoid fractures treated with CE-approved Mg-based compression screws (MAGNEZIX®) were analysed at various time points post-surgery. The artefact production of the Mg-based material decreased as implant material degraded in all settings. The worst-case imaging scenario was determined to be when the imaging plane was selected to be perpendicular to the implant axis. Moreover, the Mg-based implant outperformed the Ti equivalent in all experiments by producing lower metallic artefact (p < 0.05). This investigation demonstrates that Mg-based implants generate significantly lower metallic distortion in MRI when compared to Ti. Our positive findings suggest and support further research into the application of Mg-based implants including post-operative care facilitated by MRI monitoring of degradation kinetics and bone/tissue healing processes.
© 2022 The Authors.

Entities:  

Keywords:  Biodegradable implants; Magnesium; Magnetic resonance imaging; Medical imaging; Translational medicine

Year:  2022        PMID: 35386351      PMCID: PMC8958470          DOI: 10.1016/j.bioactmat.2021.11.035

Source DB:  PubMed          Journal:  Bioact Mater        ISSN: 2452-199X


Anterior/Posterior American Society of Testing and Materials Main Magnetic Field Digital Imaging and Communications in Medicine Field of View Gradient Echo Head/Foot Left/Right Magnetic Resonance Imaging Orthogonal Parallel Radio Frequency Sagittal Total Artefact Volume Echo Time Titanium Time Point Repetition Time Transverse Turbo Spin Echo Weighted

Introduction

Traditional permanent orthopaedic implants have shown to cause inflammatory responses [1] in addition to further refracture risk [2]. These inappropriate bodily reactions and mechanical failures suggest that other solutions be investigated for temporary bone support, such as magnesium (Mg)- based alloy materials. The ability to safely degrade in vivo as a load-bearing implant material is arguably the most attractive property [[3], [4], [5]]. The similar elastic modulus and densities between Mg alloys and natural bone prevent negative mechanical conditions such as stress-shielding [6]. Recognizing the increasing usage of Mg material in the clinical setting magnetic resonance imaging (MRI) presents a viable diagnostic imaging modality for the examination of the degradation status of biodegradable implants and for the study of the implant/tissue interface. MRI is a mainstay of today's diagnostic imaging due to its ability to provide excellent soft tissue contrast at high spatial resolution [[7], [8], [9]]. Computed tomography (CT) is widely used in clinical practice to image bone fractures and corresponding implants. However, CT provides little information on soft tissue processes. The use of ionising radiation in serial CT examinations adds to the cumulative radiation dose of patients which may be associated with a small but significantly increased risk of malignancy [10]. Since MRI does not administer any type of ionising radiation, long-term harmful effects are considered insignificant [11]. However, diagnostic imaging quality can be compromised by image artefacts attributed to the patient (motion and flow artefacts), signal processing (wrap around and chemical shifts), or to the hardware (static magnetic field or radiofrequency (RF) transmission field inhomogeneities). In MRI metallic implants may cause image distortion artefacts, or susceptibility artefacts. Metals have intrinsic magnetic susceptibilities that produce significant local field disturbances. These local changes in the resonance frequency shift image pixels away from their true positions leading to significant geometric distortions including signal reduction or voids (black areas in the image) and signal enhancements (bright areas in the image). These distortions are created by perturbations of the static magnetic field and are related to the specific material properties of the implant. Although there have been many methods demonstrating a reduction in artefacts [12], complete metallic artefacts cannot be removed. Metallic implants made of materials, like Mg, have shown to cause image artefacts due to the inherent magnetic susceptibility of the material. However, their impact on MRI quality is less compared to other current permanent metallic implants [13]. The evaluation of artefact production in MRI is necessary when choosing a biomaterial. Artefact limitation is vital since compromised or poor image quality may disturb medical diagnosis or anatomical visualisation. The aim of this investigation is to quantify the metallic artefact production of Mg-based implants during the degradation process in various MRI settings with the ultimate goal to support the use of Mg-based implants in the clinical setting due to successful visualisation in MRI. To benchmark the MRI performance of Mg-based implants standard, commercial Ti based implants are used as a reference.

Materials and methods

In vitro investigation

Orthopaedic compression screws 40 mm in length and diameter of 3.2 mm made of WE43-based material manufactured by Syntellix AG (Hannover, Germany) underwent in vitro degradation. The volume of the WE43 screw is 151.51 mm3 with a surface area of 552.54 mm2. Chemical composition of WE43 includes Yttrium (3.7–4.3%), Rare Earth Elements (2.4–4.4%), Zirconium (0.4%), with the remainder composing of Mg [14]. A non-degraded WE43 screw and a titanium equivalent of similar geometry were included for comparative results (Fig. 1).
Fig. 1

40 mm Mg-based orthopaedic compression screw MAGNEZIX® CS ø3.2 (top) and Ti market-equivalent (bottom).

40 mm Mg-based orthopaedic compression screw MAGNEZIX® CS ø3.2 (top) and Ti market-equivalent (bottom). The WE43 samples were degraded via a modified ASTM F3268 Standard [15] with Dulbecco's Phosphate Buffered Saline (DPBS). A total of five screws were degraded in the DPBS with different immersion times varying by one week with a total of five time points (1 week, 2 weeks, 3 weeks, 4 weeks, and 5 weeks). After immersion time was completed, the samples were cleaned with distilled water to remove any excess salts and rinsed with ethanol to prevent further oxidation before drying and storage. For the in vitro MRI measurements, a 1.5 T (GE Signa Artist, GE Healthcare, Waukesha, USA), 3.0 T (Skyra Fit, Siemens, Erlangen, Germany) and 7.0 T (Magnetom, Siemens, Erlangen, Germany) MRI Scanner were used. Further details on the imaging protocol and imaging techniques employed for the in vitro measurements are documented in Table 1.
Table 1

Parameters of sequences used for in vitro MRI investigations.

B0 Field Strength [T]
1.5
3.0
7.0
RF Coil TypeSpinalSpinalBasic Volume
SequenceGRETSEGRETSEGRETSE
TR [ms]10050050011905501190
TE [ms]152013201330
FOV [mm]300300180200185185
Matrix Size [pixel]256256256256256256
Slice Thickness [mm]222222
Parameters of sequences used for in vitro MRI investigations. The following in vitro procedure has been adapted from ASTM F2119 [16] and consists of two parts: (1) the acquisition of images with the test object's long axis perpendicular to the main magnetic field (B0), and (2) the acquisition of images with the test object's long axis parallel to the main magnetic field. Implant orientation relative to the main magnetic field, imaging sequence applied, slice orientation, phase-encoding direction, and material were varied for each combination before each image acquisition. The test object was suspended in a home built phantom acrylic container via a 3D printed holder filled with CuSO4 solution (Merck Millipore, USA) to reduce T1 relaxation effects. Artefact dimensions were determined by applying the ASTM F2119 definition for artefacts via a custom MATLAB (The MathWorks Inc., USA) script. Examples of measurements completed by the script are shown in Fig. 2. The diameter artefact is measured by determining the longest artefact along the diameter of the screw. The length artefact is determined by measuring the longest artefact along the length of the screw. The largest measurements are taken as the ‘worst-case’ artefact.
Fig. 2

Script workflow of diameter (to p) and length (bottom) artefact measurement of image data acquired at 3 T MRI. A control and a sample image are loaded and compared to detect a 30% intensity change. This change is used to create a binary image where a longest distance can then be measured (red indicator line).

Script workflow of diameter (to p) and length (bottom) artefact measurement of image data acquired at 3 T MRI. A control and a sample image are loaded and compared to detect a 30% intensity change. This change is used to create a binary image where a longest distance can then be measured (red indicator line). Calculation of test object distortion was performed to determine the artefact size in relation to the test object dimension. Test object distortion is calculated by:

Ex vivo investigation

Experimental animals

All in vivo experiments were carried out in accordance with the National Ethical Guidelines (Italian Ministry of Health; D.L.vo 26/2014) and the guidelines from Directive 2010/63/EU of the European Parliament. The protocol was approved by the Instituto Superiore di Sanità on behalf of Italian Ministry of Health and Ethical Panel (Prot. n° 299/2020-PR) and the local ethics committee. Additionally, the protocol conformed to the ARRIVE guidelines [17]. A total of three 12-week-old female Wistar rats were used in this experiment. On surgery, animals were anesthetized with an intraperitoneal injection of 5 mg/kg Xylazine (Rompum Elanco, Italy) and 10 mg/kg Zoletil (Virbac, Switzerland). The depth of anaesthesia was controlled by toe pinch response.

Surgical approach

Through a lateral approach, an incision was made on the skin and the muscles are carefully teased away to expose the mid-diaphyseal region of the femur. A drill with 1.55 mm diameter is used to prepare the transcortical implantation bed with the longitudinal axis of this drill perpendicular to the longitudinal axis of the femoral diaphysis. Drilling is performed at low rotational speed and profuse physiological saline (Fresenius Kabi, Italy) irrigation is applied via syringe to minimise frictional heat and thermal necrosis. A pin implant of 8 mm in length and 1.5 mm in diameter made of WE43 or Ti was implanted in this procedure. The cylindrical implant is inserted by gentle tapping, resulting in a uniform press fit. After transcortical placement is ensured, the operating field is cleaned with sterile cotton tips for remaining bone debris and the wound is closed with resorbable sutures (Johnson & Johnson Medical, Italy). Thereafter, the contralateral side is operated in the same way and using the same implant type. The sacrifice of the animals at various time-points is performed with a lethal dose of isoflurane (IsoFlo, Zoetis, Italy); hind paws are explanted cutting through the hip joint and stored at −20 °C until further analysis.

Explant imaging

The explants were scanned to obtain information on the influence of hard and soft tissue on the artefact production, moving closer to mimicking the true physiological environment. Ex vivo MRI measurements of the hind paws were performed at Consiglio Nationale delle Ricerche Area della Ricerca di Pisa (Pisa, Italy) using a 7 T small animal MRI scanner (Bruker Avance III HD 300). Further details and sequences employed for measurements are documented in Table 2.
Table 2

MRI sequences applied during explant imaging at 7 T.

B0 Field Strength [T]7.0
RF CoilQuadrature Bird Cage
SequenceT1 RARE
TR [ms]1500
TE [ms]7
Spatial resolution [mm]0.12
FOV [mm]30
Matrix Scan [pixel]256
Slice Thickness [mm]0.8
MRI sequences applied during explant imaging at 7 T. DICOM images were loaded and analysed with MATLAB. The artefact area of each scan was measured within each sequence set for one explant containing a WE43-based pin one day after surgery, one explant containing a WE43-based pin thirty days after surgery, and one explant containing a Ti-based equivalent. A single time-point was selected for Ti since the material is not expected to degrade, unlike the WE43 material. A total of 18 tracings were completed (six tracings per sample). To calculate the total artefact volume (TAV) for each scan, the artefact areas measured for each slice were summed and multiplied by the slice thickness as described by the Debatin method [18].

Clinical investigation

Clinical trial

Imaging data in humans was provided by the randomised controlled clinical trial SCAMAG (Comparison of scaphoid fracture osteosynthesis by magnesium-based headless Herbert screws with titanium Herbert screws) [19]. The trial is a blinded observer controlled parallel two-group post-market trial with a follow-up of one year per patient. Each involved patient is treated with either a Ti or Mg-based compression screw. The primary aim of SCAMAG is to demonstrate Mg-based compression screws as non-inferior when compared to Ti in the treatment of scaphoid fractures. The study protocol adheres to the SPIRIT [20] and CONSORT [21] statements. The SCAMAG trial was approved by the ethics committee of the Hanover Medical School (MHH) on September 27, 2017 (registration number: 7614) including registration with the German Register for Clinical Trials (DRKS, drks.de) on December 4, 2017 (registration number: DRKS00013368).

Patient imaging

Randomly chosen patients involved in the study treated with Mg-based compression screw (MAGNEZIX® CS, Syntellix AG, Hannover, Germany) participated in MRI examinations. The scans were performed on a 3 T MR scanner (Vario, Siemens, Erlangen, Germany) at the Medical School Hannover (Hannover, Germany). Typical wrist imaging MRI protocols were performed for this study, which are summarised in Table 3.
Table 3

MRI scanning protocols from SCAMAG clinical trial. Scanning sequences used include proton density-weighted turbo spin echo with fat saturation (PDw TSE with FS) and T1/T2-weighted turbo spin echo (T1/T2w TSE).

Field Strength [T]3.0
SequencePDw TSE with fat saturationT1w TSET2w TSET2w TSE
Slice PlaneAxialCoronal 1Coronal 2Sagittal
TR [ms]3220–4290485–5863090–31602430–2920
TE [ms]37323997
Spatial resolution [mm]0.220.200.250.25
FOV [mm]100 × 80100 × 80100 × 80100 × 80
Slice Thickness [mm]2222
MRI scanning protocols from SCAMAG clinical trial. Scanning sequences used include proton density-weighted turbo spin echo with fat saturation (PDw TSE with FS) and T1/T2-weighted turbo spin echo (T1/T2w TSE). DICOM images were loaded and analysed with MATLAB R2019b (The MathWorks Inc., US). A total of four patient follow-up scans were included in this analysis. The first, second, third, and fourth patients were treated with a single compression screw of sizes Ø 2.7 mm with length 26 mm, Ø 3.2 mm with length 26 mm, Ø 3.2 mm with length 20 mm, Ø 3.2 mm with length 24 mm, respectively. The artefact area of each scan was measured within each sequence set for the available time points: 1.5-, 3-, 6-, and 12-months post operation. A total of 64 tracings were completed. The TAV was calculated similarly as in Section 2.2.

Statistics

Statistical analyses were performed using RStudio (RStudio, USA). A p-value of <0.05 was chosen to be statistically significant. To compare the artefacts produced by Ti, Mg, and various time-corroded Mg samples in the in vitro measurements with respect to implant orientation, sequence type, slice orientation, and encoding direction, an analysis of variance (ANOVA) was completed. If a significant p-value was determined, pairwise comparisons were performed using a Tukey's post hoc multiple comparison procedure. For ex vivo investigations, the ANOVA test was applied to determine differences between material implanted into the rat femur. The ANOVA test was also applied in the clinical cases to determine the significant effects of slice orientation and healing time on the TAV.

Results

In vitro results

Titanium vs magnesium

By varying different MRI scanning parameters (seen in Fig. 3) the significant differences were found for the Ti and Mg samples. For all main field magnetic strengths, the Mg-samples produced significantly less artefacts than the Ti counterparts. Turbo spin echo sequences produced lower artefacts versus gradient echo sequence. The averaged diameter- and length-measured artefacts are summarised in Table 4. Averaged distortions or artefact measured based on material and magnetic field strength are documented in Fig. 4. An increase in magnetic field strength results in an increase in artefact size for both materials. The application of the gradient echo sequence and use of Ti material generally produced higher distortions than the turbo spin echo sequence and the use of Mg with the exception of the PAR-TSE-TRA-AP/LR protocol at 3.0 T and 7.0 T of the PAR-TSE-SAG-AP protocol at 7.0 T. The Tukey post-hoc test revealed significant pairwise differences between the artefacts size obtained at 1.5 T and 7 T and at 3 T versus 7 T. No significant difference in artefact size was found between 1.5 T and 3 T. The p-values of parameters changed in this investigation are listed in Table 5.
Fig. 3

Measured artefact lengths of Ti and Mg in various magnetic strengths. Implant position relative to the main magnetic field, sequence applied, slice plane, and phase encoding direction were altered for artefact measurements. (Reading format of y-axis: IMPLANT POSITION-SEQUENCE-SLICE PLANE-ENCODING DIRECTION).

Table 4

Average measured metallic artefact of Ti and Mg in MRI along the diameter and length of the implant.

B0 Field Strength [T]
1.5
3.0
7.0
MaterialTiMgTiMgTiMg
Average Diameter [mm]16.7 ± 3.912.5 ± 3.918.5 ± 2.716.2 ± 5.031.1 ± 5.324.6 ± 2.2
Average Length [mm]50.2 ± 3.946.3 ± 3.251.0 ± 2.447.5 ± 2.168.7 ± 6.960.7 ± 5.9
Fig. 4

Averaged artefact distortions of diameter (left) and length (right) measurements of Ti and Mg in MRI.

Table 5

ANOVA and post hoc significance of varied parameters on metallic artefact produced by Ti vs Mg in MRI. Encoding directions are abbreviated as follows: anterior/posterior (AP), head/foot (HF), and right/left (RL).

ParameterDiameter p-valueLength p-value
B0 Magnetic Strength<0.05<0.05
3 T:1.5 T0.110.73
7 T:1.5 T<0.05<0.05
7 T:3 T<0.05<0.05
Implant Position0.680.57
Imaging technique (GRE:TSE)<0.05<0.05
Phase Encoding Direction0.620.47
HF:AP0.770.67
RL:AP0.660.99
RL:HF0.980.70
Material<0.05<0.05
Measured artefact lengths of Ti and Mg in various magnetic strengths. Implant position relative to the main magnetic field, sequence applied, slice plane, and phase encoding direction were altered for artefact measurements. (Reading format of y-axis: IMPLANT POSITION-SEQUENCE-SLICE PLANE-ENCODING DIRECTION). Average measured metallic artefact of Ti and Mg in MRI along the diameter and length of the implant. Averaged artefact distortions of diameter (left) and length (right) measurements of Ti and Mg in MRI. ANOVA and post hoc significance of varied parameters on metallic artefact produced by Ti vs Mg in MRI. Encoding directions are abbreviated as follows: anterior/posterior (AP), head/foot (HF), and right/left (RL).

The influence of Mg in vitro corrosion time on artefact production

Examining the effect of the Mg implant corrosion time on the artefact size revealed statistically significant results (Table 6). For all magnetic field strengths a general decrease of image distortion was found for increasing in vitro corrosion times of the implant (Fig. 5).The spin echo sequence produced lower artefacts than the gradient echo sequence (Fig. 6).
Table 6

ANOVA and post hoc significance of varied parameters on metallic artefact produced by corroded Mg samples for different time points (tp). Encoding directions are abbreviated as follows: anterior/posterior (AP), head/foot (HF), and right/left (RL).

ParameterDiameter p-valueLength p-value
B0 Magnetic Strength<0.05<0.05
3 T:1.5 T<0.050.40
7 T:1.5 T<0.05<0.05
7 T:3 T<0.05<0.05
Implant Position0.600.13
Imaging Technique (GRE:TSE)<0.05<0.05
Encoding Direction0.440.12
HF:AP0.510.54
RL:AP0.610.82
RL:HF0.990.32
Corrosion Time<0.05<0.05
1tp:0tp0.240.98
2tp:0tp0.210.61
3tp:0tp<0.050.11
4tp:0tp<0.050.05
5tp:0tp<0.05<0.05
2tp:1tp0.980.95
3tp:1tp0.980.43
4tp:1tp0.470.26
5tp:1tp0.08<0.05
3tp:2tp0.990.91
4tp:2tp0.510.79
5tp:2tp0.090.24
4tp:3tp0.890.99
5tp:3tp0.350.82
5tp:4tp0.940.94
Fig. 5

Averaged diameter (left) and length (right) artefact distortions of Mg-based samples as a function of in vitro corrosion time.

Fig. 6

Averaged diameter (left) and length (right) measured artefacts of Mg-based samples as a function of in vitro corrosion time and scanning sequences.

ANOVA and post hoc significance of varied parameters on metallic artefact produced by corroded Mg samples for different time points (tp). Encoding directions are abbreviated as follows: anterior/posterior (AP), head/foot (HF), and right/left (RL). Averaged diameter (left) and length (right) artefact distortions of Mg-based samples as a function of in vitro corrosion time. Averaged diameter (left) and length (right) measured artefacts of Mg-based samples as a function of in vitro corrosion time and scanning sequences.

Ex vivo results

Fig. 7 illustrates MR images obtained for rat explants with Mg and Ti pins using the small bore animal 7 T MRI scanner. Ti produced higher distortion of approximately 3 times more TAV than the Mg-based samples (p < 0.05) when scanned with the Turbo RARE sequence. TAV averaging highlighted in Fig. 8 indicates a very minor decrease in TAV from one day to thirty days of the Mg samples post-surgery, although not significant. The implantation of the pin into the femur with hard tissue (bone) and soft tissue (muscle) disturbs the anatomic integrity of the images by introducing artefacts as illustrated in Fig. 7.
Fig. 7

Example of T1-weighted images of rat femur explants with WE43 and Ti pins implanted transcortical to bone.

Fig. 8

Total averaged artefact volume produced by pins implanted into rat femurs imaged in 7 T small animal MRI.

Example of T1-weighted images of rat femur explants with WE43 and Ti pins implanted transcortical to bone. Total averaged artefact volume produced by pins implanted into rat femurs imaged in 7 T small animal MRI.

Clinical results

Fig. 9 shows a decrease in the hypointense area surrounding the implant after one month, three months, and twelve months post-surgery for a single patient. The rest of the patient data follows a comparable trend (Fig. 10). Significant differences were found between healing times and imaging slice positions. An increase in implant degradation and patient healing time resulted in a decrease in artefact production. Imaging the implant at 1.5 months with a T2w TSE sequence in the sagittal direction produced the largest TAV of approximately 650 mm3. Since the screw is implanted into the scaphoid bone, the shape of the implant in Fig. 9 is defined since it is primarily surrounded by one type of material, spongy bone.
Fig. 9

T2-weighted images of scaphoid with Mg-based screw one (left), three months (middle), and twelve months (right) after surgery. White arrows indicate position of implant.

Fig. 10

Total artefact volume produced by Mg-based implant as a function of healing time in different slice planes in 3 T MRI.

T2-weighted images of scaphoid with Mg-based screw one (left), three months (middle), and twelve months (right) after surgery. White arrows indicate position of implant. Total artefact volume produced by Mg-based implant as a function of healing time in different slice planes in 3 T MRI.

Discussion

Postoperative care in the field of orthopaedic surgery relies heavily on precise imaging techniques. Early-stage research of biomaterials concentrates on the assurance of appropriate mechanical and physiological responses between the host and the implant. Further along a product's developmental stage requires insight on product application and translational aspects. Medical imaging compatibility is one of the many focusses. The artefact production of an implant material may sometimes be overlooked but is a key property of the material and must be considered during a product's design process. Previous studies have documented artefact behaviour of magnesium-based implants in MRI [[22], [23], [24]]. Our study is the first which reports on the assessment of artefact behaviour of WE43-based implants in in vitro, ex vivo, and clinical environments. Our work adds to the literature by examining the impact of the magnetic field strength on the artefact behaviour. More importantly, our study provides clinically meaningful insights on how degradation time affects artefact behaviour for the experimental setups used in this work. A well-known and commercially equivalent orthopaedic implant material like Ti is widely used today [25]. With research and popularity of magnesium-based implants increasing, a comparison between the two materials is conceptually appealing. Our findings demonstrate that the Mg-based material outperforms Ti in the in vitro environment by producing lower metallic distortion artefacts. Magnetic field perturbations are present around metallic objects in MRI which leads to a modification of hydrogen nuclei precession rates. The change in precession rates results in signal losses or pixel shifts and may manifest in local hypointensities or hyperintensities in MR images. Magnetic field variations are dependent on the magnetic susceptibility of the implant and the surrounding environment (in our case a CuSO4 solution). Since the difference between the magnetic susceptibility of Ti and CuSO4 is pronounced versus the difference between Mg and CuSO4, larger magnetic field distortions are to be expected around a Ti implant. When introducing corrosion time as a parameter, a decrease in artefact production is noticed as the implant is degraded in vitro. As Mg degrades in the body, the base metallic material corrodes forming non-metallic degradation layers composing mainly of MgO and Mg(OH)2 [26]. As such, the base metallic material begins to shrink in volume resulting in a reduced artefact size. This decrease in artefact size is confirmed by our results obtained for a range of in vitro corrosion times. The question of how the degradation layer affects artefact production can be answered by analysing the magnetic properties of the degradation layer. Mainly composing of Mg(OH)2 [27], the degradation layer consists of diamagnetic material. As in vivo degradation occurs, calcium and magnesium apatite corrosion products additionally form [28]. However, since calcium phosphates are intrinsically diamagnetic, these compounds are trivial with respect to magnetic field interference. Thus, the degradation layer dampens the effects of magnetic field variations caused by the base metallic material. This layer has been reported to be on the micrometer scale [29], which may be assumed to have an almost negligible damping effect. Standards and norms governed by global organisations, such as ASTM F2119, are referred to by regulatory bodies to ensure patient safety and quality control. Our in vitro investigations, based on the ASTM F2119 standard, also considered how parameters such as implant position, encoding direction, and imaging techniques affected artefact production. Of the three, the only parameter that significantly affected artefact production was the MR imaging technique used. In both in vitro investigations, GRE imaging generally produced larger artefact size versus fast spin-echo imaging. Spin echo based imaging techniques are less affected by susceptibility artefacts than gradient echo techniques because of the use of a train of refocusing pulses that helps to partially compensate for dephasing of spins caused by metallic objects. Additionally, imaging the diameter of the screw produced larger artefact size than imaging the length of the screw. Our results add to the literature by providing important insights on how the main magnetic field strength of an MRI system contributes to metallic artefact production. An early study by Ludeke et al. describe show the magnitude of susceptibility artefacts increases with magnetic field strength [30]. This relationship is exemplified in Figs. 4 and 5 where an increase of distortion was observed when moving from 1.5 T to 3.0 T and 7.0 T. It should be noted that conductive implants interact with radio frequency (RF) signals from MRI transmitters. These interferences may cause extra artefacts and may even elevate local RF power deposition and energy absorption in the neighbouring region of the implants which may exceed regulatory limits and RF exposure guidelines [31,32]. Explants analysed in this investigation underwent freezing at −20 °C and thawing before scanned in a small animal 7.0 T MRI system. MRI availability could not be synchronised with the euthanasia of the animals and therefore animals needed to be stored. The freezing and thawing cycle has shown to effect MRI by significantly reducing the effective transversal relaxation time T2* [33]. Such an effect should be considered when interpreting the results. However, the motivation behind our explant analysis is to provide insights into artefact production in hard and soft tissue. Although there were no significant differences obtained for the WE43 samples at different time points, Ti showed an approximately three times larger TAV than the WE43 material. The clinical results and analysis revealed important information on artefact production of Mg-based implants used to treat scaphoid fractures. Averaging measurements of the four patients included in this investigation followed a similar trend of artefact reduction with patient healing time. The decrease in distortion also provides proof of implant degradation or a loss of metallic base material. More importantly, after averaging all patients and artefacts derived for each imaging plane, an extrapolation of the healing time can be made until the artefact size reaches zero. Although this extrapolation has limitations (such as implant size), complete disappearance of artefact is approximately reached at 2.1 years, which is in accordance with a one-to-three-year implant resorption period [4,34,35]. Many methods have been employed to track implant degradation rates in vitro and in vivo, such as computed tomography segmentation and hydrogen gas release [36]. The reduction in metallic artefact size measured in MRI with progressing degradation may offer a viable alternative for screw degradation monitoring and tracking. This investigation utilised two parameters to quantify artefact production caused by the Mg-based and Ti implants. Distortion quantifications were employed in the in vitro measurements while total artefact volume (TAV) was employed in the ex vivo and clinical analysis. The inclusion of two measurement methods was chosen to describe the artefact production behaviour of the implants in more than one way. Analysing the two measured quantities show an agreement in the decline of metallic distortion trend of Mg during the degradation process. Metal artefact reduction techniques are widely used today to negate the effects of metallic implants in MRI [37]. Various specialised software and MRI vendors provide dedicated imaging techniques tailored for reducing metallic artefacts. Since these approaches vary from MRI vendor to vendor, our study excluded their impact as it is out of the scope of this study. However, the literature reports that some metal artefact reduction techniques do not produce statistically significant results when applied to Mg [24]. Examining the efficacy of these metal artefact reduction techniques on Mg-based material warrants further investigation. A limitation of the in vitro aspect of this study is found in the natural behaviour of Mg itself. The inherent corrosion process of the implants made it impractical for samples to be tested more than once due to the interaction with CuSO4 solution. Regarding explant analysis, the movement of blood is avoided in the ex vivo animal studies so that blood motion artefacts are eliminated. An increase in explant samples and clinical patients would enhance the statistical power of our findings. Future studies should include the examination of different animal models to gain insights for other orthopaedic applications or solutions. Alternative in vitro implant degradation methods can be used to mimic specific environments. These methods should also be analysed to investigate how the degradation method affects artefact production. Furthermore, the inclusion of multi-modal data such as clinical computed tomography can be explored in quantifying an implant volume to MRI artefact production ratio. Assessing the artefact production behaviour of Mg-based implants in vitro, ex vivo, and in clinical application provides valuable insights into the material's properties and its implications for MRI image quality. Our results demonstrate that Mg-based implants displayed a decrease in metallic artefact production with implant degradation. As the implant degrades and artefact production decreases, better visualisation of bone healing is achieved. To summarise, the Mg-based implant outperformed the Ti material by inducing lower artefact size.

Conclusion

In this investigation, in vitro, ex vivo, and clinical studies demonstrate a reduction in artefact size with implant degradation time for Mg-based screws. The largest artefact behaviour or “worst-case” scenario caused by the WE43 implants were found when imaging the non-degraded state in the in vitro and ex vivo experiments. In the clinical study, the largest TAV was observed when imaging the implant at 1.5 months in the sagittal plane. When compared against Ti-based implants, the Mg-based material induces significantly lower artefacts. Translational findings from this study suggests that Mg-based implant material should be promoted for future orthopaedic solutions due to lower artefact production. Further clinical studies are required to determine degradation rates based on artefact production as an alternative approach for degradation tracking.

CRediT authorship contribution statement

Jonathan Espiritu: Conceptualization, Methodology, Software, Validation, Formal analysis, Investigation, Resources, Data curation, Writing – original draft, Visualization. Mostafa Berangi: Investigation, Data curation. Christina Yiannakou: Formal analysis, Investigation, Resources, Data curation. Eduarda Silva: Methodology, Resources. Roberto Francischello: Data curation, Resources. Andre Kuehne: Investigation, Data curation, Writing – review & editing, Supervision. Thoralf Niendorf: Resources, Writing – review & editing. Sören Könneker: Methodology, Resources, Data curation. Regine Willumeit-Römer: Conceptualization, Supervision, Project administration, Funding acquisition. Jan-Marten Seitz: Conceptualization, Validation, Resources, Writing – review & editing, Supervision, Project administration, Funding acquisition.

Declaration of competing interest

Authors Kuehne and Berangi are employees of MRI.TOOLS GmbH. Author Niendorf is CEO and founder of MRI.TOOLS GmbH. Syntellix AG is a medical technology manufacturer of metallic and bio-absorbable clinical implants. Authors Espiritu and Seitz are employed as Research Associate and Chief Technical Officer, respectively.
  28 in total

Review 1.  MRI safety: a report of current practice and advancements in patient preparation and screening.

Authors:  Elizabeth K Weidman; Kathryn E Dean; William Rivera; Michael L Loftus; Thomas W Stokes; Robert J Min
Journal:  Clin Imaging       Date:  2015-09-04       Impact factor: 1.605

2.  Composition of corrosion layers on a magnesium rare-earth alloy in simulated body fluids.

Authors:  Ralf Rettig; Sannakaisa Virtanen
Journal:  J Biomed Mater Res A       Date:  2009-02       Impact factor: 4.396

3.  The effect of freeze-thawing on magnetic resonance imaging T2* of freshly harvested bovine patellar tendon.

Authors:  Sarah L Pownder; Parina H Shah; Hollis G Potter; Matthew F Koff
Journal:  Quant Imaging Med Surg       Date:  2015-06

4.  Susceptibility artefacts in NMR imaging.

Authors:  K M Lüdeke; P Röschmann; R Tischler
Journal:  Magn Reson Imaging       Date:  1985       Impact factor: 2.546

Review 5.  Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz). International Commission on Non-Ionizing Radiation Protection.

Authors: 
Journal:  Health Phys       Date:  1998-04       Impact factor: 1.316

Review 6.  New prostate MRI techniques and sequences.

Authors:  Aritrick Chatterjee; Carla Harmath; Aytekin Oto
Journal:  Abdom Radiol (NY)       Date:  2020-12

7.  CONSORT 2010 statement: updated guidelines for reporting parallel group randomised trials.

Authors:  Kenneth F Schulz; Douglas G Altman; David Moher
Journal:  PLoS Med       Date:  2010-03-24       Impact factor: 11.069

8.  Bioabsorbable magnesium screw versus conventional titanium screw fixation for medial malleolar fractures.

Authors:  Hasan May; Yusuf Alper Kati; Gurkan Gumussuyu; Tuluhan Yunus Emre; Melih Unal; Ozkan Kose
Journal:  J Orthop Traumatol       Date:  2020-05-25

9.  A comparison of the effectiveness of functional MRI analysis methods for pain research: The new normal.

Authors:  Patrick W Stroman; Howard J M Warren; Gabriela Ioachim; Jocelyn M Powers; Kaitlin McNeil
Journal:  PLoS One       Date:  2020-12-14       Impact factor: 3.240

10.  Bioabsorbable metal screws in traumatology: A promising innovation.

Authors:  Roland Biber; Johannes Pauser; Matthias Brem; Hermann Josef Bail
Journal:  Trauma Case Rep       Date:  2017-01-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.