| Literature DB >> 35365887 |
Josh M Kennils1, Christina L Maunder1, Marta T Costa1,2.
Abstract
BACKGROUND: Proteinuria can be quantified through the measurement of the urine protein-to-creatinine ratio (UPC). Voided urine samples in cats are often exposed to a non-absorbable litter substrate prior to collection and urinalysis. Little is known about the effect exposure to such substrates has on pre-analytical variability of UPC measurements.Entities:
Keywords: proteinuria; renal; urinalysis
Mesh:
Substances:
Year: 2022 PMID: 35365887 PMCID: PMC9545233 DOI: 10.1111/vcp.13110
Source DB: PubMed Journal: Vet Clin Pathol ISSN: 0275-6382 Impact factor: 1.333
Spearman correlation (r s), slope, and intercept with 95% confidence intervals (CI) of Passing–Bablok and Bland–Altman mean bias with 95% limits of agreement (LOA) for urinary protein (UP), urinary creatinine (UC) and urine protein‐to‐creatinine ratio (UPC) measured at baseline vs UP, UC, and UPC measured after 24‐h exposure with non‐absorbent hydrophobic sand litter
| Spearman correlation | Passing–Bablok | Bland–Altman | ||
|---|---|---|---|---|
|
| Slope (95% CI) | Intercept (95% CI) | Mean Bias (lower LOA to upper LOA) | |
| UP | 0.99 | 1.02 (1.00 to 1.06) | −1.88 (−3.34 to −1.20) | −0.12 (−9.18 to 8.94) |
| UC | 0.99 | 1.02 (1.00 to 1.05) | 0.08 (−2.15 to 3.31) | 13.61 (−29.24 to 56.45) |
| UPC | 0.99 | 0.96 (0.92 to 1.00) | 0.00 (−0.01 to 0.00) | −0.02 (−0.19 to 0.15) |
FIGURE 1Passing–Bablok plot of urine protein‐to‐creatinine (UPC) measurements obtained at baseline (baselineUPC), and those obtained after 24‐h exposure to non‐absorbent hydrophobic sand litter (litterUPC) in 40 samples; The solid line is the regression line (y = −0.0038 + 0.9606), and the dotted lines are the 95%CI. One point (baselineUPC = 5.59, litterUPC = 5.22) was omitted to allow for better visualization of lower UPC values
FIGURE 2Difference vs average Bland–Altman plots of urine protein‐to‐creatinine (UPC) measurements obtained at baseline (baselineUPC) and after 24‐h exposure with non‐absorbent hydrophobic sand litter (litterUPC) in 40 samples. Solid line represents the mean of the difference (bias) = −0.02075. Dashed lines represent the 95% limits of agreement from −0.1896 to 0.1481
Number of samples within each International Renal Interest Society (IRIS) proteinuria category at baseline and at 24 h. Discordant results are highlighted
| IRIS proteinuria category at 24 h | ||||
|---|---|---|---|---|
| Non‐proteinuric | Borderline proteinuric | Proteinuric | ||
| IRIS proteinuria category at baseline | Non‐proteinuric | 22 | 0 | 0 |
| Borderline proteinuric | 1 | 7 | 0 | |
| Proteinuric | 0 | 1 | 9 | |
The mean, standard deviation (SD), bias, coefficient of variation (CV), and observed total error (TEobs) for intra‐ and inter‐assay runs of low‐ and high‐range UPC controls calculated from quality control materials (QCM) for urinary protein (UP) and urinary creatinine (UC) concentrations
| UPC target | QCM | UPC | Intra‐assay | Inter‐assay | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| UP (mg/dL) | UC (mg/dL) | Mean | SD | Bias | CV% | TEobs (%) | Mean | SD | Bias | CV% | TEobs (%) | ||
| Low‐range | 8 | 78.05 | 0.10 | 0.10 | 0.01 | 3.0 | 5.9 | 14.83 | 0.11 | 0.01 | −6.4 | 29.4 | 29.39 |
| High‐range | 70 | 177.60 | 0.39 | 0.39 | 0.01 | 2.3 | 2.0 | 6.36 | 0.37 | 0.02 | 6.1 | 18.9 | 18.87 |