| Literature DB >> 35365118 |
Assaf Gershoni1,2, Edward Barayev1,2, Igor Vainer2,3, Raviv Allon4, Roy Yavnieli2,5, Yinon Shapira6, Michael Mimouni6, Noa Geffen1,2, Arie Yehuda Nemet2,5, Ori Segal7,8.
Abstract
BACKGROUND: Optical coherence tomography (OCT) is used worldwide by clinicians to evaluate macular and retinal nerve fiber layer (RNFL) characteristics. It is frequently utilized to assess disease severity, progression and efficacy of treatment, and therefore must be reliable and reproducible.Entities:
Keywords: Macular thickness; Optical Coherence tomography; Retinal nerve fiber layer; Signal strength; Spectralis
Mesh:
Year: 2022 PMID: 35365118 PMCID: PMC8976355 DOI: 10.1186/s12886-022-02356-4
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1Macular Thickness and Volume measurements at different Signal Strength categories
Macular and RNFL measurements by OCT Q score category
| Low | Moderate | Good | Excellent | Differencea (%) | |
|---|---|---|---|---|---|
| Central Macular Thickness, μm | 286.73 ± 21.44 | 284.48 ± 19.38 | †*279.21 ± 19.83 | ‡†*275.59 ± 19.08 | −3.88 |
| Macular Volume, mm3 | 8.79 ± 0.39 | 8.81 ± 0.36 | 8.79 ± 0.34 | ‡†*8.76 ± 0.33 | −0.35 |
| Inner Macular Thickness by quadrant, μm | |||||
| Superior | 349.71 ± 13.84 | 349.83 ± 13.94 | ± 13.47349.77 | ‡†*348.18 ± 13.93 | −0.44 |
| Inferior | 346.98 ± 14.48 | 348.44 ± 13.17 | 348.18 ± 13.03 | 347.31 ± 12.96 | 0.1 |
| Temporal | 336.57 ± 14.10 | 12.77 ± 337.39 | 337.85 ± 12.77 | 335.97 ± 13.06 | −0.18 |
| Nasal | 353.18 ± 13.67 | 352.24 ± 14.40 | *351.07 ± 14.61 | †* 13.69 ± 350.1 | −0.88 |
| Mean RNFL, μm | 101.4 ± 10.99 | 100.85 ± 10.4 | 100.45 ± 10.4 | †*99.68 ± 10.37 | −1.7 |
| RNFL by quadrant, μm | |||||
| Nasal | 77.34 ± 17.31 | 77.2 ± 16.27 | 76.41 ± 16.47 | 76.41 ± 16.77 | −1.2 |
| Inferior-Nasal | 117.89 ± 28.86 | 117.63 ± 28.73 | 117.2 ± 29.42 | 116.2 ± 28.89 | −1.43 |
| Inferior-Temporal | 148.08 ± 19.92 | 147.89 ± 20.36 | 147.99 ± 20.64 | 147.00 ± 21.53 | −0.73 |
| Temporal | 71.60 ± 10.96 | 71.42 ± 10.88 | 70.77 ± 10.91 | 70.08 ± 10.94 | −2.12 |
| Superior-Temporal | 135.40 ± 17.78 | 134.93 ± 17.12 | 133.97 ± 16.44 | 132.38 ± 16.09 | −2.23 |
| Superior-Nasal | 111.60 ± 24.63 | 109.76 ± 24.71 | 109.55 ± 24.19 | 108.15 ± 23.81 | −3.09 |
Data are shown as mean ± SD
Scans were categorized according to signal strength. Low (Q < 15), Moderate (15 ≤ Q < 20), Good (20 ≤ Q ≤ 25), Excellent (Q > 25)
RNFL Retinal Nerve Fiber Layer, OCT Optical Coherence Tomography
aDifference is calculated between excellent and low groups, as a percentage of the low group
*Statistically significant difference, p < 0.05, compared with the low group
† Statistically significant difference, p < 0.05, compared with the moderate group
‡Statistically significant difference, p < 0.05, compared with the good group
Fig. 2Retinal Nerve Fiber Layer Thickness measurements at different Signal Strength categories
Fig. 3Correlation between signal strength and central macular thickness. Thickness measurement increase with decreasing signal strength. r = Pearson correlation coefficient
Correlations between signal strength and macular / RNFL measurements
| r | ||
|---|---|---|
| Central Macular Thickness, μm | −0.204 | |
| Macular Volume, mm3 | −0.009 | 0.873 |
| Macular Thickness by quadrant, μm | ||
| Superior | −0.01 | 0.867 |
| Inferior | 0.034 | 0.572 |
| Temporal | 0.0002 | 0.996 |
| Nasal | −0.05 | 0.407 |
| Mean RNFL, μm | −0.058 | 0.331 |
| RNFL by quadrant, μm | ||
| Nasal | −0.021 | 0.727 |
| Inferior-Nasal | −0.031 | 0.612 |
| Inferior-Temporal | −0.039 | 0.517 |
| Temporal | −0.062 | 0.303 |
| Superior-Temporal | −0.039 | 0.518 |
| Superior-Nasal | −0.033 | 0.588 |
RNFL Retinal Nerve Fiber Layer
r = Pearson correlation coefficient
P values in bold indicate significant associations
Fig. 4Bland-Altman analysis to test for the differences in macular (top) and retinal nerve fiber layer (RNFL) thickness measurements between excellent (>25) and low (<15). In both measurements, dispersion increases with retinal thickness
comparison of the previous studies for other OCT devices
| Current study | Kim et al. [ | Zhang et al. [ | Wu et al. [ | |
|---|---|---|---|---|
| Spectralis | Cirrus | RTVue | Stratus | |
| Acquisition technology | Spectral domain | Spectral domain | Fourier domain | Time domain |
| Scan speed (A-scans per second) | 40,000 | 27,000 | 26,000 | 768 |
| Axial resolution (μm) | 7 | 5 | 5 | 10 |
| Transverse resolution (μm) | 14 | 20 | 15 | 20 |
| Quality score scale | (Q, 0-40) | (SS, 0-10) | (SSI, 0-100) | (SS, 0-10) |
| Minimal suggested quality by manufacturer | 15 | 6 | 30 | 6 |
| Quality score & thickness relationship | Thickness decreases with higher SS (RNFL & macula) | Thickness increases with higher SS (RNFL) | Thickness increases with higher SS (RNFL) | Thickness increases with higher SS (RNFL) |
| Outer retina segmentation border | outer reflective band | outer reflective band | second inner hyperreflective band | inner hyperreflective band |
OCT Optical Coherence Tomography, RNFL Retinal Nerve Fiber Layer, SS Signal strength, SSI Signal Strength Index