Abstract
Purpose: Commercial wearable sleep-tracking devices are growing in popularity and in recent studies have performed well against gold standard sleep measurement techniques. However, most studies were conducted in controlled laboratory conditions. We therefore aimed to test the performance of devices under naturalistic unrestricted home sleep conditions. Participants andEntities:
Keywords: actigraphy; consumer sleep technologies; habitual sleep; mobile EEG; validation
Year: 2022 PMID: 35345630 PMCID: PMC8957400 DOI: 10.2147/NSS.S348795
Source DB: PubMed Journal: Nat Sci Sleep ISSN: 1179-1608
Sleep-Wake Summary Agreement
| Outcome | Device | n | Device Mean (SD) | Reference Mean (SD) |
|---|---|---|---|---|
| TIB | Fatigue Science Readiband | 112 | 459.0 (85.2) | 437.4 (78.9) |
| Fitbit Inspire HR | 106 | 459.8 (78.6) | 443.3 (73.6) | |
| Oura Ring | 111 | 460.0 (90.3) | 439.9 (74.0) | |
| Polar Vantage V Titan | 103 | 452.0 (75.6) | 445.5 (69.4) | |
| TST | Actiwatch | 113 | 406.1 (77.4) | 401.3 (82.6) |
| Fatigue Science Readiband | 112 | 406.0 (89.4) | 401.0 (82.9) | |
| Fitbit Inspire HR | 106 | 406.4 (70.1) | 406.6 (78.8) | |
| Oura Ring | 111 | 412.5 (77.5) | 403.1 (78.8) | |
| Polar Vantage V Titan | 103 | 418.5 (74.0) | 411.6 (72.1) | |
| SE | Actiwatch | 113 | 92.4 (3.2) | 91.2 (6.9) |
| Fatigue Science Readiband | 112 | 88.0 (7.5) | 91.3 (7.0) | |
| Fitbit Inspire HR | 106 | 94.6 (2.9) | 91.3 (7.0) | |
| Oura Ring | 111 | 89.8 (4.9) | 91.2 (7.0) | |
| Polar Vantage V Titan | 103 | 92.5 (4.6) | 92.2 (4.6) | |
| SOL | Actiwatch | 113 | 2.3 (1.7) | 8.5 (5.6) |
| Fatigue Science Readiband | 112 | 18.8 (17.3) | 8.5 (5.6) | |
| Fitbit Inspire HR | 106 | 4.9 (3.4) | 8.6 (5.6) | |
| Oura Ring | 111 | 7.8 (5.5) | 8.5 (5.6) | |
| Polar Vantage V Titan | 103 | 0.6 (1.0) | 8.5 (5.4) | |
| WASO | Actiwatch | 113 | 30.2 (12.7) | 28.1 (22.5) |
| Fatigue Science Readiband | 112 | 21.2 (23.8) | 27.9 (22.5) | |
| Fitbit Inspire HR | 106 | 38.0 (13.1) | 28.1 (21.9) | |
| Oura Ring | 111 | 39.7 (32.0) | 28.4 (22.5) | |
| Polar Vantage V Titan | 103 | 25.8 (8.6) | 25.4 (17.5) |
Notes: Sleep-wake summary agreement between devices and reference. The sleep diary served as reference for TIB, and the Dreem headband served as reference for TST, SE, SOL, and WASO. All nights with available data for both the device and reference were included in analyses, with the total number of nights (n) indicated in each row. Mean and SD are shown for each device and reference.
Abbreviations: SD, standard deviation; SE, sleep efficiency; SOL, sleep onset latency; TIB, time in bed; TST, total sleep time; WASO, wake after sleep onset.
Figure 1Bland-Altman plots: TIB.
Figure 2Bland-Altman plots: TST.
Figure 3Bland-Altman plots: SE.
Figure 4Bland-Altman plots: SOL.
Figure 5Bland-Altman plots: WASO.
Sleep Stage Summary Agreement
| Stage | Device | n | Device Mean (SD) | Reference Mean (SD) |
|---|---|---|---|---|
| Light | Fitbit Inspire HR | 106 | 243.5 (53.0) | 189.4 (54.3) |
| Oura Ring | 111 | 210.8 (66.0) | 189.0 (53.9) | |
| Polar Vantage V Titan | 103 | 226.1 (90.5) | 190.7 (52.9) | |
| Deep | Fitbit Inspire HR | 106 | 70.1 (21.9) | 80.1 (29.3) |
| Oura Ring | 111 | 124.1 (41.7) | 81.4 (29.2) | |
| Polar Vantage V Titan | 103 | 104.5 (92.1) | 82.8 (28.6) | |
| REM | Fitbit Inspire HR | 106 | 92.8 (30.2) | 137.2 (54.8) |
| Oura Ring | 111 | 77.5 (34.8) | 132.7 (52.6) | |
| Polar Vantage V Titan | 103 | 87.8 (40.7) | 138.2 (54.4) |
Notes: Sleep stage summary agreement between devices and reference. The Dreem headband served as reference for all sleep stage outcomes. Results are shown for all devices that output sleep stage classifications. See Table 1 notes for additional table details.
Abbreviations: REM, rapid eye movement; SD, standard deviation.
Figure 6Bland-Altman plots: light sleep.
Figure 7Bland-Altman plots: deep sleep.
Figure 8Bland-Altman plots: REM sleep.
EBE Sleep-Wake Agreement
| Device | Sensitivity | Specificity | Accuracy | PPV | NPV | PABAK | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| All Epochs | Between Nights | All Epochs | Between Nights | All Epochs | Between Nights | All Epochs | Between Nights | All Epochs | Between Nights | All Epochs | Between Nights | |
| Actiwatch | 0.95 | 0.95 (0.02) [0.95, 0.96] | 0.35 | 0.37 (0.15) [0.34, 0.4] | 0.90 | 0.90 (0.06) [0.89, 0.91] | 0.94 | 0.94 (0.06) [0.93, 0.95] | 0.38 | 0.39 (0.16) [0.36, 0.42] | 0.80 | 0.80 (0.11) [0.78, 0.82] |
| Fatigue Science Readiband | 0.94 | 0.94 (0.07) [0.92, 0.95] | 0.40 | 0.43 (0.20) [0.39, 0.46] | 0.90 | 0.89 (0.07) [0.88, 0.90] | 0.95 | 0.94 (0.06) [0.93, 0.95] | 0.37 | 0.45 (0.22) [0.41, 0.49] | 0.79 | 0.78 (0.15) [0.75, 0.81] |
| Fitbit Inspire HR | 0.93 | 0.93 (0.03) [0.93, 0.94] | 0.45 | 0.50 (0.17) [0.47, 0.54] | 0.89 | 0.89 (0.05) [0.88, 0.90] | 0.95 | 0.95 (0.06) [0.94, 0.96] | 0.36 | 0.36 (0.15) [0.33, 0.39] | 0.79 | 0.78 (0.11) [0.76, 0.81] |
| Oura Ring | 0.94 | 0.94 (0.03) [0.94, 0.95] | 0.41 | 0.44 (0.17) [0.41, 0.47] | 0.90 | 0.90 (0.06) [0.89, 0.91] | 0.95 | 0.95 (0.06) [0.94, 0.96] | 0.37 | 0.39 (0.16) [0.36, 0.42] | 0.81 | 0.80 (0.11) [0.78, 0.82] |
| Polar Vantage V Titan | 0.96 | 0.96 (0.02) [0.95, 0.96] | 0.35 | 0.39 (0.16) [0.36, 0.42] | 0.92 | 0.92 (0.04) [0.91, 0.93] | 0.96 | 0.96 (0.04) [0.95, 0.97] | 0.36 | 0.36 (0.15) [0.33, 0.38] | 0.84 | 0.84 (0.08) [0.82, 0.85] |
Notes: Proportions for EBE agreement metrics are shown for all sleep epochs (versus wake epochs) for all combined epochs as well as between all nights (irrespective of participant) for the devices compared with the corresponding epochs from the reference Dreem headband. Values for between-night analyses are presented as mean (SD) [95% confidence interval]. Higher values (closer to 1.0) indicate better performance on that metric.
Abbreviations: EBE, epoch by epoch; NPV, negative predictive value; PABAK, prevalence and bias adjusted kappa; PPV, positive predictive value; SD, standard deviation.
EBE Sleep Stage Agreement
| Stage | Device | Sensitivity | Specificity | Accuracy | PPV | NPV | PABAK | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| All Epochs | Between Nights | All Epochs | Between Nights | All Epochs | Between Nights | All Epochs | Between Nights | All Epochs | Between Nights | All Epochs | Between Nights | ||
| Light | Fitbit Inspire HR | 0.71 | 0.71 (0.09) [0.70, 0.73] | 0.59 | 0.60 (0.10) [0.58, 0.62] | 0.64 | 0.65 (0.07) [0.63, 0.66] | 0.57 | 0.57 (0.13) [0.54, 0.60] | 0.73 | 0.73 (0.10) [0.71, 0.75] | 0.29 | 0.29 (0.13) [0.26, 0.32] |
| Oura Ring | 0.61 | 0.60 (0.12) [0.58, 0.62] | 0.64 | 0.64 (0.12) [0.62, 0.67] | 0.62 | 0.62 (0.07) [0.61, 0.64] | 0.56 | 0.56 (0.13) [0.54, 0.59] | 0.69 | 0.68 (0.10) [0.66, 0.70] | 0.25 | 0.25 (0.15) [0.22, 0.28] | |
| Polar Vantage V Titan | 0.59 | 0.63 (0.10) [0.61, 0.65] | 0.54 | 0.50 (0.10) [0.48, 0.52] | 0.56 | 0.56 (0.07) [0.55, 0.57] | 0.49 | 0.50 (0.13) [0.47, 0.52] | 0.64 | 0.63 (0.11) [0.61, 0.66] | 0.12 | 0.12 (0.14) [0.09, 0.15] | |
| Deep | Fitbit Inspire HR | 0.51 | 0.53 (0.23) [0.49, 0.58] | 0.92 | 0.93 (0.04) [0.92, 0.94] | 0.85 | 0.85 (0.06) [0.84, 0.86] | 0.58 | 0.61 (0.23) [0.57, 0.66] | 0.90 | 0.90 (0.06) [0.88, 0.91] | 0.70 | 0.70 (0.11) [0.68, 0.72] |
| Oura Ring | 0.72 | 0.74 (0.22) [0.70, 0.78] | 0.83 | 0.83 (0.10) [0.81, 0.85] | 0.81 | 0.81 (0.07) [0.79, 0.82] | 0.50 | 0.51 (0.17) [0.47, 0.54] | 0.93 | 0.93 (0.06) [0.92, 0.94] | 0.62 | 0.62 (0.14) [0.59, 0.64] | |
| Polar Vantage V Titan | 0.45 | 0.46 (0.27) [0.41, 0.51] | 0.82 | 0.81 (0.24) [0.76, 0.86] | 0.75 | 0.75 (0.15) [0.72, 0.78] | 0.37 | 0.41 (0.19) [0.37, 0.44] | 0.86 | 0.87 (0.06) [0.86, 0.89] | 0.51 | 0.50 (0.31) [0.44, 0.56] | |
| REM | Fitbit Inspire HR | 0.47 | 0.48 (0.14) [0.45, 0.51] | 0.91 | 0.91 (0.05) [0.90, 0.92] | 0.77 | 0.77 (0.08) [0.76, 0.79] | 0.70 | 0.70 (0.16) [0.66, 0.73] | 0.79 | 0.79 (0.11) [0.77, 0.82] | 0.55 | 0.55 (0.16) [0.52, 0.58] |
| Oura Ring | 0.39 | 0.40 (0.17) [0.37, 0.43] | 0.92 | 0.92 (0.05) [0.91, 0.93] | 0.76 | 0.76 (0.09) [0.74, 0.78] | 0.68 | 0.68 (0.19) [0.64, 0.71] | 0.78 | 0.78 (0.11) [0.76, 0.80] | 0.51 | 0.52 (0.18) [0.48, 0.55] | |
| Polar Vantage V Titan | 0.36 | 0.40 (0.16) [0.37, 0.43] | 0.88 | 0.87 (0.07) [0.86, 0.88] | 0.71 | 0.71 (0.10) [0.69, 0.73] | 0.59 | 0.57 (0.21) [0.53, 0.61] | 0.75 | 0.75 (0.12) [0.73, 0.78] | 0.43 | 0.43 (0.20) [0.39, 0.47] | |
Notes: Proportions for EBE agreement metrics are shown for each sleep stage (versus the combination of all other possible classifications) for all combined epochs as well as between all nights (irrespective of participant) for the devices compared with the corresponding epochs from the reference Dreem headband. Values for between-night analyses are presented as mean (SD) [95% confidence interval]. Higher values (closer to 1.0) indicate better performance on that metric. Results are shown for all devices that output sleep stage classifications.
Abbreviations: EBE, epoch by epoch; NPV, negative predictive value; PABAK, prevalence and bias adjusted kappa; PPV, positive predictive value; REM, rapid eye movement; SD, standard deviation.
EBE Sleep Stage Misclassification Errors
| Fitbit Inspire HR | 0.37 | 0.04 | 0.14 | 0.07 | 0.11 | 0.10 | 0.03 | 0.42 | 0.04 | 0.09 | 0.41 | 0.03 |
| Oura Ring | 0.33 | 0.16 | 0.10 | 0.06 | 0.23 | 0.10 | 0.01 | 0.24 | 0.03 | 0.08 | 0.45 | 0.09 |
| Polar Vantage V Titan | 0.41 | 0.10 | 0.13 | 0.04 | 0.23 | 0.14 | 0.02 | 0.45 | 0.08 | 0.06 | 0.47 | 0.11 |
| Fitbit Inspire HR | 0.32 | 0.06 | 0.28 | 0.05 | 0.14 | 0.24 | 0.02 | 0.31 | 0.07 | 0.05 | 0.22 | 0.03 |
| Oura Ring | 0.31 | 0.03 | 0.28 | 0.06 | 0.09 | 0.29 | 0.04 | 0.37 | 0.10 | 0.04 | 0.25 | 0.03 |
| Polar Vantage V Titan | 0.29 | 0.06 | 0.30 | 0.05 | 0.17 | 0.29 | 0.03 | 0.44 | 0.16 | 0.04 | 0.30 | 0.08 |
Notes: (A) Proportions for device EBE sleep stage misclassification errors versus those scored by the reference Dreem headband. Dreem-scored classifications are the larger column categories, with the three possible device-scored misclassifications under each category. (B) Proportions of sleep stage EBE differences in Dreem-scored classifications from the device-scored epochs. Device-scored classifications are the larger column categories, with the three possible Dreem-scored differences in classification under each category. Results are shown for all devices that output sleep stage classifications.
Abbreviations: EBE, epoch by epoch; REM, rapid eye movement.