Seong Hoon Bae1, Seojin Moon1, Mincheol Jeong1, In Seok Moon1.
Abstract
The nine-step test is a classical method for evaluating Eustachian tube function. It directly assesses the patient's capacity to equilibrate middle ear pressure by swallowing. However, there are insufficient studies to appraise its diagnostic performance. The purpose of this study is to evaluate the sensitivity, specificity, and cut-off value of the nine-step test in patients with obstructive Eustachian tube dysfunction (oETD) and patulous Eustachian tube (PET). Enrolled subjects were divided into three groups. Control (50 ears of healthy volunteers), oETD (19 ears with oETD), and PET (29 ears with PET). Receiver operating characteristics curve analysis was conducted to evaluate the diagnostic performance of maximal peak pressure difference (ETTmd) in the nine-step test. Both the oETD group and the PET group showed decreased ETTmd. The nine-step test showed moderate accuracy when used to diagnose oETD (area under the curve = 0.875) and PET (area under the curve = 0.769). The highest diagnostic performance was observed when the cut-off value was 13 daPa for both the oETD group (sensitivity = 73.7%, specificity = 90.0%) and the PET group (sensitivity = 58.6%, specificity = 90.0%). The nine-step test has moderate diagnostic performance for oETD and PET.Entities:
Keywords: eustachian tube; eustachian tube dysfunction; nine-step test; patulous eustachian tube
Year: 2022 PMID: 35328285 PMCID: PMC8947718 DOI: 10.3390/diagnostics12030732
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Demographic data of enrolled subjects.
| Control | oETD | PET | |||
|---|---|---|---|---|---|
| Age (year) | 29.0 ± 3.9 | 42.7 ± 19.5 | 0.069 | 41.5 ± 17.8 | 0.027 * |
| Male (%) | 28.0 | 36.8 | 0.560 | 34.5 | 0.613 |
| Right side (%) | 50.0 | 42.1 | 0.787 | 55.2 | 0.637 |
| Average ETDQ7 | 1.1 ± 0.2 | 3.5 ± 1.2 | < 0.001 * | 2.8 ± 1.1 | < 0.001 * |
| ETTmd (daPa) | 27.9 ± 15.0 | 10.2 ± 8.4 | <0.001 * | 15.5 ± 12.9 | <0.001 * |
| Middle ear pressure (daPa) | −9.4 ± 11.5 | −146.7 ± 109.0 | <0.001 * | −10.3 ± 21.8 | >0.999 |
| Number of ears | 50 | 19 | 29 |
oETD: obstructive Eustachian tube dysfunction, PET: Patulous Eustachian tube dysfunction, ETTmd: Maximal difference of peak pressure in the nine-step test, p-value: post hoc analysis compared to the control group *: p < 0.05.
Figure 1Eustachian tube function of the three groups. ETTmd: maximal difference of peak pressure in the nine-step test, CTR: control group, oETD: obstructive Eustachian tube dysfunction group, PET: patulous Eustachian tube dysfunction group. ***: p < 0.001. ns: p > 0.05.
Figure 2Receiver operating characteristics curve analysis. (A) For diagnosing oETD, (B) for diagnosing PET. AUC: Area under the curve, CTR: Control group, oETD: obstructive Eustachian tube dysfunction group, PET: patulous Eustachian tube dysfunction group.
Diagnostic values of the modified nine-step test.
| Disease | AUC | Cut-Off | Sensitivity | Specificity | |
|---|---|---|---|---|---|
| 10 daPa | 63.2 | 92.0 | |||
| oETD | 0.875 | <0.001 * | 13 daPa | 73.7 | 90.0 |
| 15 daPa | 79.0 | 82.0 | |||
| 10 daPa | 48.3 | 92.0 | |||
| PET | 0.769 | <0.001 * | 13 daPa | 58.6 | 90.0 |
| 15 daPa | 62.1 | 82.0 |
oETD: obstructive Eustachian tube dysfunction, PET: Patulous Eustachian tube dysfunction, AUC: Area under curve, *: p < 0.05.