Addressing CT metal artifacts using photon-counting detectors and one-step spectral CT image reconstruction.

Taly Gilat Schmidt1, Barbara A Sammut1, Rina Foygel Barber2, Xiaochuan Pan3, Emil Y Sidky3.   

Abstract

PURPOSE: The constrained one-step spectral CT image reconstruction (cOSSCIR) algorithm with a nonconvex alternating direction method of multipliers optimizer is proposed for addressing computed tomography (CT) metal artifacts caused by beam hardening, noise, and photon starvation. The quantitative performance of cOSSCIR is investigated through a series of photon-counting CT simulations.
METHODS: cOSSCIR directly estimates basis material maps from photon-counting data using a physics-based forward model that accounts for beam hardening. The cOSSCIR optimization framework places constraints on the basis maps, which we hypothesize will stabilize the decomposition and reduce streaks caused by noise and photon starvation. Another advantage of cOSSCIR is that the spectral data need not be registered, so that a ray can be used even if some energy window measurements are unavailable. Photon-counting CT acquisitions of a virtual pelvic phantom with low-contrast soft tissue texture and bilateral hip prostheses were simulated. Bone and water basis maps were estimated using the cOSSCIR algorithm and combined to form a virtual monoenergetic image for the evaluation of metal artifacts. The cOSSCIR images were compared to a "two-step" decomposition approach that first estimated basis sinograms using a maximum likelihood algorithm and then reconstructed basis maps using an iterative total variation constrained least-squares optimization (MLE+TV min $_{\text{min}}$ ). Images were also compared to a nonspectral TV min $_{\text{min}}$ reconstruction of the total number of counts detected for each ray with and without normalized metal artifact reduction (NMAR) applied. The simulated metal density was increased to investigate the effects of increasing photon starvation. The quantitative error and standard deviation in regions of the phantom were compared across the investigated algorithms. The ability of cOSSCIR to reproduce the soft-tissue texture, while reducing metal artifacts, was quantitatively evaluated.
RESULTS: Noiseless simulations demonstrated the convergence of the cOSSCIR and MLE+TV min $_{\text{min}}$ algorithms to the correct basis maps in the presence of beam-hardening effects. When noise was simulated, cOSSCIR demonstrated a quantitative error of -1 HU, compared to 2 HU error for the MLE+TV min $_{\text{min}}$ algorithm and -154 HU error for the nonspectral TV min $_{\text{min}}$ +NMAR algorithm. For the cOSSCIR algorithm, the standard deviation in the central iodine region of interest was 20 HU, compared to 299 HU for the MLE+TV min $_{\text{min}}$ algorithm, 41 HU for the MLE+TV min $_{\text{min}}$ +Mask algorithm that excluded rays through metal, and 55 HU for the nonspectral TV min $_{\text{min}}$ +NMAR algorithm. Increasing levels of photon starvation did not impact the bias or standard deviation of the cOSSCIR images. cOSSCIR was able to reproduce the soft-tissue texture when an appropriate regularization constraint value was selected.
CONCLUSIONS: By directly inverting photon-counting CT data into basis maps using an accurate physics-based forward model and a constrained optimization algorithm, cOSSCIR avoids metal artifacts due to beam hardening, noise, and photon starvation. The cOSSCIR algorithm demonstrated improved stability and accuracy compared to a two-step method of decomposition followed by reconstruction.
© 2022 American Association of Physicists in Medicine.

Entities:  

Keywords:  metal artifacts; photon-counting CT; reconstruction

Mesh:

Substances:

Year:  2022        PMID: 35318699      PMCID: PMC9353719          DOI: 10.1002/mp.15621

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.506


  45 in total

1.  Frequency split metal artifact reduction (FSMAR) in computed tomography.

Authors:  Esther Meyer; Rainer Raupach; Michael Lell; Bernhard Schmidt; Marc Kachelrieß
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

2.  Model-Based Iterative Reconstruction for Dual-Energy X-Ray CT Using a Joint Quadratic Likelihood Model.

Authors:  Ruoqiao Zhang; Jean-Baptiste Thibault; Charles A Bouman; Ken D Sauer; Jiang Hsieh
Journal:  IEEE Trans Med Imaging       Date:  2013-09-17       Impact factor: 10.048

3.  Prospects for quantitative computed tomography imaging in the presence of foreign metal bodies using statistical image reconstruction.

Authors:  Jeffrey F Williamson; Bruce R Whiting; Jasenka Benac; Ryan J Murphy; G James Blaine; Joseph A O'Sullivan; David G Politte; Donald L Snyder
Journal:  Med Phys       Date:  2002-10       Impact factor: 4.071

4.  Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility.

Authors:  Seung Chol Han; Yong Eun Chung; Young Han Lee; Kwan Kyu Park; Myeong Jin Kim; Ki Whang Kim
Journal:  AJR Am J Roentgenol       Date:  2014-10       Impact factor: 3.959

5.  Regularization of nonlinear decomposition of spectral x-ray projection images.

Authors:  Nicolas Ducros; Juan Felipe Perez-Juste Abascal; Bruno Sixou; Simon Rit; Françoise Peyrin
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

6.  An algorithm for constrained one-step inversion of spectral CT data.

Authors:  Rina Foygel Barber; Emil Y Sidky; Taly Gilat Schmidt; Xiaochuan Pan
Journal:  Phys Med Biol       Date:  2016-04-15       Impact factor: 3.609

7.  A Spectral CT Method to Directly Estimate Basis Material Maps From Experimental Photon-Counting Data.

Authors:  Taly Gilat Schmidt; Rina Foygel Barber; Emil Y Sidky
Journal:  IEEE Trans Med Imaging       Date:  2017-04-24       Impact factor: 10.048

8.  Validation of a power-law noise model for simulating small-scale breast tissue.

Authors:  I Reiser; A Edwards; R M Nishikawa
Journal:  Phys Med Biol       Date:  2013-08-12       Impact factor: 3.609

Review 9.  Metal artifact reduction techniques in musculoskeletal CT-imaging.

Authors:  R H H Wellenberg; E T Hakvoort; C H Slump; M F Boomsma; M Maas; G J Streekstra
Journal:  Eur J Radiol       Date:  2018-08-12       Impact factor: 3.528

10.  Polychromatic Iterative Statistical Material Image Reconstruction for Photon-Counting Computed Tomography.

Authors:  Thomas Weidinger; Thorsten M Buzug; Thomas Flohr; Steffen Kappler; Karl Stierstorfer
Journal:  Int J Biomed Imaging       Date:  2016-03-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.