Shahad Mohammmed Altuwaijri1,2, Hanan Nejer Alotaibi1, Talal Mughaileth Alnassar1.
Abstract
PURPOSE: The purpose of this study was to investigate the misfit and screw preload at the implant abutment connection of implant supported fixed dental prosthesis with cantilever (ICFDP) manufactured using different digital manufacturing techniques and to compare the screw preload before and after cyclic loading.Entities:
Keywords: Cyclic loading; Dental implant; Preload; Reverse torque value (RTV); Screw loosening
Year: 2022 PMID: 35284054 PMCID: PMC8891687 DOI: 10.4047/jap.2022.14.1.22
Source DB: PubMed Journal: J Adv Prosthodont ISSN: 2005-7806 Impact factor: 1.904
Fig. 1Mandibular model with temporarily stabilized implants.
Fig. 2Passive fit framework implant assembly.
Fig. 3Indentation on cantilever extension prior to cyclic loading.
Fig. 4Positioning of the model in the cyclic loading chewing simulator (CS-4) with 30 degrees angle.
One-way ANOVA of the effect of independent variables (framework manufacturing technique) on implant abutment misfit
| Group | Mean ± SD (μm) | |
|---|---|---|
| PF | 16.04 ± 2.6 | .000* |
| CAD-cast | 29.2 ± 3.1 | .000* |
| 3DP | 24.5 ± 1.05 | .000* |
PF: passive fit, 3DP: three-dimensional printing.
*statistically significant (P < .05).
Paired sample t-test comparing the applied torque and RTV before and after cyclic loading
| Framework | Torque | RTV | Mean difference | ||
|---|---|---|---|---|---|
| M ± SD | M ± SD | ||||
| PF | Pre cyclic loading | 35.2 ± 0.21 | 33.6 ± 1.44 | 1.6 | .000* |
| Post cyclic loading | 35.2 ± 0.17 | 30.6 ± 4.17 | 4.6 | .000* | |
| CAD-cast | Pre cyclic loading | 35.2 ± 0.19 | 33.3 ± 1.87 | 1.9 | .000* |
| Post cyclic loading | 35.3 ± 0.19 | 30.7 ± 5.51 | 4.6 | .001* | |
| 3DP | Pre cyclic loading | 35.1 ± 0.15 | 34 ± 1.44 | 1.2 | .001* |
| Post cyclic loading | 35.2 ± 0.16 | 29.4 ± 4.9 | 5.8 | .000* |
RTV: reverse torque value, PF: passive fit, 3DP: three-dimensional printing.
*statistically significant (P < .05).
Paired sample t-test comparing pre cyclic loading and post cyclic loading RTV in each group
| Mean difference | df | t | ||
|---|---|---|---|---|
| PF | 3 | 19 | 3.3 | .004* |
| CAD-cast | 2.6 | 19 | 1.87 | .076 |
| 3DP | 4.6 | 19 | 3.83 | .001* |
RTV: reverse torque value, PF: passive fit, 3DP: three-dimensional printing.
*statistically significant (P < .05).
One-way ANOVA of independent variables (framework manufacturing technique) on pre cyclic loading and post cyclic loading RTV
| df | Mean square | F | Sig. | ||
|---|---|---|---|---|---|
| Pre cyclic loading RTV | Between groups | 2 | 2.813 | 1.09 | .340 |
| Within groups | 57 | 2.561 | |||
| Total | 59 | ||||
| Post cyclic loading RTV | Between groups | 2 | 10.840 | 0.448 | .641 |
| Within groups | 57 | 24.195 | |||
| Total | 59 |
RTV: reverse torque value.